Fluke 179 Digital Multimeter (DMM)

PackEye Reseat Power Supply Connections

Last Updated: 04/24/2025 1:42 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Thermo FHT1377 PackEye
Device
Published
10/28/2024 3:37 PM
5662 - 5661 - Uninstall Power Supply Module; Remark:
Procedure Step
Published
11/02/2022 5:10 PM
5661 - Uninstall Power Supply Module
Instruction
Published
11/02/2022 5:09 PM
5660 - 313 - Power off instrument; Remark:4086
Procedure Step
Published
04/07/2025 4:34 PM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
4086 - Press and hold the power button for approximately 10 seconds until indicator ...
Remark
Published
05/02/2022 6:04 PM
5666 - 4222 - Place instrument on flat surface; Remark:4221
Procedure Step
Published
11/03/2022 8:17 AM
4222 - Place instrument on flat surface
Instruction
Published
05/10/2022 1:54 PM
4221 - Shoulder straps should be facing down.
Remark
Published
05/10/2022 1:55 PM
5667 - 3876 - Open top of backpack; Remark:5665
Procedure Step
Published
11/03/2022 8:21 AM
3876 - Open top of backpack
Instruction
Published
04/04/2022 3:14 PM
5665 - IO PackEye Open Top for Battery
Remark
Published
11/09/2024 8:06 PM
5668 - 4219 - Remove distribution box; Remark:4218
Procedure Step
Published
11/03/2022 8:22 AM
4219 - Remove distribution box
Instruction
Published
05/10/2022 1:58 PM
4218 - The distribution box is usually located in the top of the backpack.
Remark
Published
05/10/2022 1:58 PM
5669 - 4217 - Disconnect battery cable; Remark:4216
Procedure Step
Published
11/03/2022 8:47 AM
5664 - Note : The other cord connected to the distribution box is the headphone plug.
Annotation
Published
11/03/2022 8:47 AM
4217 - Disconnect battery cable
Instruction
Published
05/10/2022 2:01 PM
4216 - Rotate the collar counterclockwise and pull straight out. The battery cable c...
Remark
Published
05/10/2022 2:03 PM
5673 - 5672 - Disconnect RS232 jumper; Remark:5670
Procedure Step
Published
11/03/2022 9:07 AM
5672 - Disconnect RS232 jumper
Instruction
Published
11/03/2022 9:03 AM
5670 - Use small, slotted screwdriver.
Remark
Published
11/03/2022 9:07 AM
5671 - IO PackEye RS232 Disconnect Cable
Remark
Published
11/09/2024 8:07 PM
5680 - 5679 - Disconnect Bluetooth adapter cable; Remark:5670
Procedure Step
Published
11/03/2022 9:15 AM
5679 - Disconnect Bluetooth adapter cable
Instruction
Published
11/03/2022 9:12 AM
5678 - IO PackEye Cable for BTA Disconnect
Remark
Published
11/09/2024 8:08 PM
5681 - 5677 - Remove wire ties; Remark:5676
Procedure Step
Published
11/03/2022 9:27 AM
5677 - Remove wire ties
Instruction
Published
11/03/2022 9:16 AM
5676 - Note location of wire ties for reassembly. Use wire cutter or scissors.
Remark
Published
11/03/2022 9:27 AM
5770 - IO PackEye Power Cable Ties Location
Remark
Published
11/09/2024 8:08 PM
5682 - 5675 - Remove power supply module; Remark:5674
Procedure Step
Published
11/03/2022 9:35 AM
5675 - Remove power supply module
Instruction
Published
11/03/2022 9:31 AM
5674 - There are Velcro strips securing the power supply module to the FHT 681 ampli...
Remark
Published
11/03/2022 9:35 AM
5783 - 5782 - Inspect Internal Components; Remark:
Procedure Step
Published
11/03/2022 3:19 PM
5782 - Inspect Internal Components
Instruction
Published
11/03/2022 3:19 PM
5821 - 5309 - Remove two (2) screws; Remark:5819
Procedure Step
Published
11/04/2022 7:38 AM
5309 - Remove two (2) screws
Instruction
Published
10/24/2022 9:50 AM
5819 - Only remove the screws on the side with the charger connector. Use T-10 Torx ...
Remark
Published
11/04/2022 7:36 AM
5704 - IO PackEye Power Module Screws
Remark
Published
11/09/2024 8:09 PM
5822 - 5702 - Remove electronics; Remark:5818
Procedure Step
Published
11/04/2022 7:42 AM
5702 - Remove electronics
Instruction
Published
11/03/2022 10:50 AM
5818 - Components will slide out of the housing.
Remark
Published
11/04/2022 7:42 AM
5823 - 5817 - Inspect internal board for damage; Remark:5816
Procedure Step
Published
11/04/2022 7:46 AM
5817 - Inspect internal board for damage
Instruction
Published
11/04/2022 7:44 AM
5816 - Check for loose connections, burn marks, and debris. Check battery for leakag...
Remark
Published
11/04/2022 7:46 AM
5824 - 5815 - Inspect underside of circuit board; Remark:5814
Procedure Step
Published
11/04/2022 7:50 AM
5815 - Inspect underside of circuit board
Instruction
Published
11/04/2022 7:48 AM
5814 - Check for loose connections, burn marks, and debris. Check USB port for damag...
Remark
Published
11/04/2022 7:50 AM
5825 - 5813 - Disconnect spade connectors; Remark:5812
Procedure Step
Published
11/04/2022 7:55 AM
5813 - Disconnect spade connectors
Instruction
Published
11/04/2022 7:52 AM
5812 - If wires are soldered to the board, skip this step.
Remark
Published
11/04/2022 7:55 AM
5826 - 5811 - Reconnect spade connectors from battery to internal board; Remark:5810
Procedure Step
Published
11/04/2022 7:59 AM
5553 - Caution : Correct placement of connectors and wiring is critical to proper operation.
Annotation
Published
10/28/2022 2:19 PM
5811 - Reconnect spade connectors from battery to internal board
Instruction
Published
11/04/2022 7:56 AM
5810 - Ensure the black wire is connected closest to the edge of the board.
Remark
Published
11/04/2022 7:58 AM
5827 - 5720 - Replace electronics; Remark:5719
Procedure Step
Published
11/04/2022 8:03 AM
5721 - Caution : Circuit board must be inserted correctly to prevent damage to components....
Annotation
Published
11/03/2022 1:15 PM
5720 - Replace electronics
Instruction
Published
11/03/2022 1:11 PM
5719 - Align circuit board with third slot from bottom of enclosure and slide into p...
Remark
Published
11/03/2022 1:15 PM
5828 - 5718 - Align gasket; Remark:5717
Procedure Step
Published
11/04/2022 8:06 AM
5718 - Align gasket
Instruction
Published
11/03/2022 1:16 PM
5717 - Verify gasket is aligned between housing and end panel before inserting screw...
Remark
Published
11/03/2022 1:18 PM
5829 - 5716 - Reinstall two (2) enclosure screws; Remark:5714
Procedure Step
Published
11/04/2022 8:09 AM
5716 - Reinstall two (2) enclosure screws
Instruction
Published
11/03/2022 1:19 PM
5714 - Use T-10 Torx screwdriver.
Remark
Published
11/03/2022 1:20 PM
5780 - 5779 - Reinstall Power Supply Module; Remark:
Procedure Step
Published
11/03/2022 1:25 PM
5779 - Reinstall Power Supply Module
Instruction
Published
11/03/2022 1:25 PM
5834 - 5775 - Secure power supply module to FHT 681; Remark:5774
Procedure Step
Published
11/04/2022 9:29 AM
5775 - Secure power supply module to FHT 681
Instruction
Published
11/03/2022 2:46 PM
5774 - The module will snap into place.
Remark
Published
11/03/2022 2:50 PM
5835 - 5773 - Reconnect battery cable; Remark:5772
Procedure Step
Published
11/04/2022 9:30 AM
5773 - Reconnect battery cable
Instruction
Published
11/03/2022 2:51 PM
5772 - Align the connector and push in. Rotate collar clockwise and hand-tighten.
Remark
Published
11/03/2022 2:55 PM
5836 - 5771 - Install wire ties; Remark:5769
Procedure Step
Published
11/04/2022 9:33 AM
5771 - Install wire ties
Instruction
Published
11/03/2022 2:56 PM
5769 - If desired, install new wire ties and cut off excess.
Remark
Published
11/03/2022 3:05 PM
5837 - 4237 - Replace distribution box; Remark:4236
Procedure Step
Published
11/04/2022 9:43 AM
4237 - Replace distribution box
Instruction
Published
05/10/2022 2:15 PM
4236 - IO PackEye Distro Box Install
Remark
Published
10/24/2024 6:53 PM
5838 - 5788 - Reconnect Bluetooth adapter cable; Remark:5787
Procedure Step
Published
11/04/2022 9:45 AM
5788 - Reconnect Bluetooth adapter cable
Instruction
Published
11/03/2022 3:17 PM
5787 - IO PackEye Cable for BTA Connect
Remark
Published
11/09/2024 8:13 PM
5839 - 5786 - Reconnect RS232 jumper; Remark:5670
Procedure Step
Published
11/04/2022 9:49 AM
5786 - Reconnect RS232 jumper
Instruction
Published
11/03/2022 3:39 PM
5830 - IO PackEye RS232 Connect Cable
Remark
Published
11/09/2024 8:14 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
minutes
Duration
Published
05/26/2022 11:41 AM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
Small slotted (flathead) screwdriver
Tools
Published
09/11/2020 11:23 AM
Wire cutter pliers
Tools
Published
10/27/2022 1:20 PM
T10 Torx screwdriver
Tools
Published
11/10/2022 12:57 PM
Medium pliers
Tools
Published
11/03/2022 2:20 PM
Needle nose pliers
Tools
Published
05/06/2020 4:10 PM
Fluke 179 Digital Multimeter (DMM)
Tools
Published
03/20/2021 6:27 PM
Power off Packeye
படம்
Published
02/19/2022 1:48 AM
packeye_di_opentop_redarrow_00758.jpg
படம்
Published
10/20/2023 12:55 PM
packeye_di_distrobox_remove.jpg
படம்
Published
10/20/2023 4:00 PM
packeye_di_batterycable_remove.jpg
படம்
Published
10/20/2023 3:54 PM
packeye_di_batterycable_2prong.jpg
படம்
Published
10/20/2023 3:51 PM
packeye_di_rs232cable_remove.jpg
படம்
Published
10/23/2023 9:49 AM
packeye_di_rs232btadisconnect.jpg
படம்
Published
10/23/2023 1:46 PM
packeye_di_powercable_ties_locate.jpg
படம்
Published
10/20/2023 4:02 PM
packeye_di_powermodule_remove.jpg
படம்
Published
10/20/2023 4:04 PM
packeye_di_powersupp_screws.jpg
படம்
Published
10/23/2023 1:47 PM
packeye_di_powersupp_slideopen.jpg
படம்
Published
10/23/2023 1:48 PM
packeye_di_powersupp_internalview.jpg
படம்
Published
10/23/2023 1:50 PM
packeye_di_powersupp_pcb.jpg
படம்
Published
10/23/2023 1:51 PM
packeye_di_powersupp_spadeoff.jpg
படம்
Published
10/23/2023 1:53 PM
packeye_di_powersupp_spadeon.jpg
படம்
Published
10/23/2023 1:54 PM
packeye_di_powersupp_alignslot3.jpg
படம்
Published
10/23/2023 1:55 PM
packeye_di_powersupp_slideclosed.jpg
படம்
Published
10/23/2023 1:49 PM
packeye_di_powermodule_install.jpg
படம்
Published
10/20/2023 4:03 PM
packeye_di_batterycable_install.jpg
படம்
Published
10/20/2023 3:53 PM
packeye_di_distrobox_install.jpg
படம்
Published
10/20/2023 3:59 PM
packeye_di_rs232btaconnectinsert.jpg
படம்
Published
10/23/2023 1:44 PM
packeye_di_rs232cable_insert.jpg
படம்
Published
10/23/2023 9:48 AM

PRM-470 NHA-472 Board Replacement

Last Updated: 04/24/2025 1:35 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Rapiscan PRM-470 CGN
Device
Published
10/28/2024 3:33 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
2548 - 205 - Disconnect charger; Remark:2547
Procedure Step
Published
08/02/2021 12:46 PM
205 - Disconnect charger
Instruction
Published
07/23/2021 2:42 PM
2547 - IO PRM-470 Remove Charger Cable
Remark
Published
11/02/2024 3:10 PM
397 - 313 - Power off instrument; Remark:398
Procedure Step
Published
06/11/2021 9:31 AM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
398 - Press and hold the POWER button until instrument powers off....
Remark
Published
06/18/2021 6:02 PM
261 - 65 - Loosen four (4) cover screws; Remark:264
Procedure Step
Published
08/12/2021 7:25 AM
65 - Loosen four (4) cover screws
Instruction
Published
07/22/2021 12:38 PM
264 - Use a medium Phillips (PH2) screwdriver. Screws do not need to be completely ...
Remark
Published
07/20/2021 3:59 PM
263 - 262 - Remove instrument cover; Remark:413
Procedure Step
Published
08/11/2021 1:41 PM
2771 - Note : There are high-voltage detector circuit boards inside the instrument; however...
Annotation
Published
10/04/2022 11:06 AM
262 - Remove instrument cover
Instruction
Published
07/20/2021 3:35 PM
413 - PRM-470 Remove Instrument Cover
Remark
Published
11/01/2024 8:32 PM
2734 - 266 - Locate battery connection; Remark:2733
Procedure Step
Published
08/18/2021 10:14 PM
3033 - Note : PRM-470 CGN models have an additional board and connectors not shown in remar...
Annotation
Published
10/04/2022 11:07 AM
266 - Locate battery connection
Instruction
Published
07/20/2021 3:36 PM
2733 - Battery connection is labeled 1.
Remark
Published
08/17/2021 3:23 PM
271 - 270 - Disconnect battery; Remark:414
Procedure Step
Published
08/11/2021 1:42 PM
1345 - Note : It is good practice to ensure all cables and connectors are labeled prior to ...
Annotation
Published
10/04/2022 11:08 AM
270 - Disconnect battery
Instruction
Published
07/20/2021 3:37 PM
414 - PRM-470_Disconnect_battery
Remark
Published
11/02/2024 1:21 PM
2737 - 2736 - Disconnect NHA-472 board wiring; Remark:2735
Procedure Step
Published
08/18/2021 10:15 PM
2736 - Disconnect NHA-472 board wiring
Instruction
Published
08/17/2021 3:39 PM
2735 - Remove connectors at locations 5, 6, and 7.
Remark
Published
08/17/2021 3:24 PM
2739 - 2738 - Remove faulty NHA-472; Remark:
Procedure Step
Published
08/18/2021 10:15 PM
2738 - Remove faulty NHA-472
Instruction
Published
08/12/2021 10:10 AM
2742 - 2741 - Locate NHA-472 board; Remark:2740
Procedure Step
Published
08/18/2021 10:16 PM
2741 - Locate NHA-472 board
Instruction
Published
08/12/2021 10:25 AM
2740 - PRM-470CGN Neutron high voltage board
Remark
Published
11/02/2024 6:01 PM
2745 - 2744 - Remove NHA-472 mounting nut; Remark:2743
Procedure Step
Published
08/19/2021 3:12 PM
2744 - Remove NHA-472 mounting nut
Instruction
Published
08/12/2021 10:27 AM
2743 - Removing the NHA-472 mounting nut
Remark
Published
11/02/2024 6:02 PM
2748 - 2747 - Remove NHA-472 board; Remark:2746
Procedure Step
Published
08/18/2021 10:16 PM
2747 - Remove NHA-472 board
Instruction
Published
08/12/2021 10:29 AM
2746 - Gently pry up each side and lift.
Remark
Published
08/17/2021 3:25 PM
2750 - 2749 - Install new NHA-472; Remark:
Procedure Step
Published
08/18/2021 10:17 PM
2749 - Install new NHA-472
Instruction
Published
08/12/2021 10:31 AM
2753 - 2752 - Mount new NHA-472; Remark:2751
Procedure Step
Published
08/18/2021 10:18 PM
2752 - Mount new NHA-472
Instruction
Published
08/12/2021 10:32 AM
2751 - Verify connectors are correctly aligned then press down.
Remark
Published
08/17/2021 3:26 PM
2756 - 2755 - Reinstall mounting nut; Remark:2754
Procedure Step
Published
08/18/2021 10:18 PM
504 - Caution : Do not overtighten the mounting nut.
Annotation
Published
10/04/2022 11:11 AM
2755 - Reinstall mounting nut
Instruction
Published
08/12/2021 10:34 AM
2754 - Reinstall NHA-472 mounting nut
Remark
Published
11/02/2024 6:03 PM
2758 - 2757 - Reconnect wiring; Remark:
Procedure Step
Published
08/18/2021 10:18 PM
2757 - Reconnect wiring
Instruction
Published
08/12/2021 10:35 AM
2761 - 2760 - Reconnect all connectors except battery; Remark:2759
Procedure Step
Published
08/18/2021 10:19 PM
2760 - Reconnect all connectors except battery
Instruction
Published
08/12/2021 10:37 AM
2759 - Do not connect location 1.
Remark
Published
08/17/2021 3:27 PM
2763 - 305 - Reconnect battery; Remark:2762
Procedure Step
Published
08/18/2021 10:19 PM
305 - Reconnect battery
Instruction
Published
07/29/2021 3:48 PM
2762 - Reconnect PRM-470 battery
Remark
Published
11/02/2024 6:05 PM
2766 - 2764 - Adjust neutron high voltage; Remark:
Procedure Step
Published
08/18/2021 10:19 PM
2765 - Note : The same circuit boards are used for both gamma and neutron detectors; howeve...
Annotation
Published
10/04/2022 11:13 AM
2764 - Adjust neutron high voltage
Instruction
Published
08/12/2021 10:38 AM
2770 - 2769 - Determine manufacturer of the neutron detector; Remark:2767
Procedure Step
Published
08/18/2021 10:20 PM
2769 - Determine manufacturer of the neutron detector
Instruction
Published
08/12/2021 10:39 AM
2767 - To access the neutron detector information, remove the moderator from the cas...
Remark
Published
08/17/2021 3:27 PM
2768 - Detector information is printed on the detector. There are two (2) common neu...
Remark
Published
06/18/2021 6:47 PM
2772 - 220 - Power on instrument; Remark:228
Procedure Step
Published
08/18/2021 10:20 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
228 - Press POWER button until the display turns on.
Remark
Published
06/18/2021 6:29 PM
2777 - 2776 - Connect HV probe to DMM; Remark:2774
Procedure Step
Published
08/18/2021 10:21 PM
2776 - Connect HV probe to DMM
Instruction
Published
08/17/2021 3:40 PM
2774 - Use Fluke 80K-40 High Voltage Probe and Fluke 179 DMM. The GND tab should lin...
Remark
Published
08/17/2021 3:28 PM
2658 - 2566 - Switch DMM to DC voltage; Remark:2613
Procedure Step
Published
08/18/2021 10:21 PM
2566 - Switch DMM to DC voltage
Instruction
Published
08/11/2021 3:41 PM
2613 - IO Fluke DMM Dial DC Volt
Remark
Published
11/02/2024 3:15 PM
2785 - 2567 - Locate ground test point; Remark:2782
Procedure Step
Published
08/18/2021 10:21 PM
2567 - Locate ground test point
Instruction
Published
08/11/2021 3:42 PM
2782 - The neutron ground TP38 is located on the NHA-472 neutron board and might not...
Remark
Published
08/17/2021 3:29 PM
2790 - 2789 - Connect HV probe to ground; Remark:2787
Procedure Step
Published
08/19/2021 11:33 AM
2789 - Connect HV probe to ground
Instruction
Published
08/19/2021 11:28 AM
2787 - The ground connector for the HV probe may be directly connected to the ground...
Remark
Published
08/17/2021 3:30 PM
2788 - It may be easier to use a separate test lead (jumper wire) and connect the HV...
Remark
Published
08/17/2021 3:31 PM
2794 - 2792 - Locate high voltage test point; Remark:2791
Procedure Step
Published
08/18/2021 10:22 PM
2793 - Warning : The test point has high voltage. Only contact the test point with the high vo...
Annotation
Published
10/04/2022 11:20 AM
2792 - Locate high voltage test point
Instruction
Published
08/17/2021 3:41 PM
2791 - The neutron high voltage TP34 is located on the NHA-472 neutron board and mig...
Remark
Published
08/17/2021 3:32 PM
2797 - 2796 - Contact HV probe to test point; Remark:2795
Procedure Step
Published
08/19/2021 11:32 AM
2796 - Contact HV probe to test point
Instruction
Published
08/19/2021 11:30 AM
2795 - Contact TP34 (not labeled) with tip of red HV probe.
Remark
Published
08/17/2021 3:32 PM
2874 - 2873 - Locate neutron high voltage adjustment; Remark:2872
Procedure Step
Published
08/18/2021 10:23 PM
2873 - Locate neutron high voltage adjustment
Instruction
Published
08/17/2021 3:42 PM
2872 - R10 on the NHA-472 neutron board is used to adjust the high voltage for the n...
Remark
Published
08/17/2021 3:33 PM
2885 - 2881 - Adjust neutron high voltage; Remark:2875
Procedure Step
Published
08/18/2021 10:23 PM
2884 - Standard : High voltage for a GE Reuter-Stokes neutron detection is 990 Volts. High volt...
Annotation
Published
08/19/2021 11:25 AM
2881 - Adjust neutron high voltage
Instruction
Published
08/17/2021 3:43 PM
2875 - Use precision slotted (tweaker) screwdriver to adjust R10. The voltage should...
Remark
Published
08/17/2021 3:35 PM
2878 - The number indicated on the DMM must be multiplied by 1000 if the Fluke 80K-4...
Remark
Published
08/17/2021 3:35 PM
2889 - 2888 - Remove HV probe leads; Remark:
Procedure Step
Published
08/19/2021 11:31 AM
2888 - Remove HV probe leads
Instruction
Published
08/19/2021 11:31 AM
2893 - 2892 - Locate neutron stage 2 gain adjustment; Remark:2891
Procedure Step
Published
08/18/2021 10:24 PM
2892 - Locate neutron stage 2 gain adjustment
Instruction
Published
08/17/2021 4:05 PM
2891 - R17 on the NHA-472 neutron board is used to adjust the stage 2 gain.
Remark
Published
08/17/2021 3:36 PM
2896 - 2895 - Set stage 2 gain at maximum; Remark:2894
Procedure Step
Published
08/18/2021 10:24 PM
2895 - Set stage 2 gain at maximum
Instruction
Published
08/17/2021 4:05 PM
2894 - Use precision slotted (tweaker) screwdriver to rotate R17 10 full turns clock...
Remark
Published
08/17/2021 3:37 PM
303 - 302 - Reassemble instrument; Remark:
Procedure Step
Published
10/07/2022 1:49 PM
302 - Reassemble instrument
Instruction
Published
06/16/2021 11:28 PM
2898 - 307 - Replace front cover; Remark:2897
Procedure Step
Published
08/18/2021 10:24 PM
304 - Caution : Ensure all wiring is inside the instrument while replacing cover.
Annotation
Published
10/04/2022 11:10 AM
307 - Replace front cover
Instruction
Published
07/20/2021 3:57 PM
2897 - Replace front cover - PRM470
Remark
Published
11/02/2024 6:06 PM
310 - 309 - Tighten four (4) cover screws; Remark:417
Procedure Step
Published
08/11/2021 2:04 PM
569 - Caution : Do not overtighten screws.
Annotation
Published
06/16/2021 2:31 PM
309 - Tighten four (4) cover screws
Instruction
Published
07/22/2021 1:02 PM
417 - Use a medium Phillips (PH2) screwdriver.
Remark
Published
07/22/2021 1:18 PM
320 - 319 - Perform operational test; Remark:347
Procedure Step
Published
08/12/2021 7:12 AM
319 - Perform operational test
Instruction
Published
06/10/2021 8:31 PM
347 - Perform DET-HHD-RAP-RM01 PRM-470 Operational Test.
Remark
Published
07/21/2021 12:22 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
30 minutes
Duration
Published
04/01/2020 11:35 AM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Warning
Annotation Type
Published
03/02/2020 3:25 PM
Standard
Annotation Type
Published
03/02/2020 3:25 PM
Technician 2
Skill Level
Published
04/09/2020 1:17 AM
Medium Phillips (PH2) screwdriver
Tools
Published
04/24/2020 4:40 PM
5-mm nut (1/4 inch) nut driver or socket (optional)
Tools
Published
03/21/2021 5:00 PM
Needle nose pliers
Tools
Published
05/06/2020 4:10 PM
Precision slotted (tweaker) screwdriver
Tools
Published
03/20/2021 6:23 PM
Fluke 179 Digital Multimeter (DMM)
Tools
Published
03/20/2021 6:27 PM
Fluke 80K-40 high voltage (HV) probe
Tools
Published
03/20/2021 6:23 PM
Mini Hook Test Leads (optional)
Tools
Published
03/21/2021 5:02 PM
PRM-470 Disconnect Charger Cable
படம்
Published
03/16/2021 1:39 AM
PRM-470_Partial_Power_Button_Red_Highlight.png
படம்
Published
06/18/2021 6:28 PM
PRM470_Remove_Screws_Top_Case
படம்
Published
05/08/2020 6:05 PM
PRM-470_di_CGN_open_00536
படம்
Published
02/19/2023 11:26 AM
PRM-470 connectors shows 1-7
படம்
Published
02/19/2023 11:22 AM
PRM-470_Disconnect_battery
படம்
Published
05/09/2020 8:40 PM
PRM-470 connectors shows 5-7
படம்
Published
03/16/2021 1:39 AM
PRM-470 locate neutron CB
படம்
Published
03/16/2021 1:39 AM
PRM-470 NHA-470 CB mounting nut
படம்
Published
03/16/2021 1:39 AM
PRM-470 remove NHA-472 1 of 3
படம்
Published
03/16/2021 1:39 AM
PRM-470 Remove NHA-472 2 of 3
படம்
Published
03/16/2021 1:39 AM
PRM-470 Remove NHA-472 3 of 3
படம்
Published
03/16/2021 1:39 AM
PRM-470 align NHA-472 connectors 1 of 2
படம்
Published
03/16/2021 1:39 AM
PRM-470 press NHA-472 connectors 2 of 2
படம்
Published
03/16/2021 1:39 AM
PRM-470 mount NHA-472 nut
படம்
Published
03/16/2021 1:39 AM
PRM-470 connectors shows 2-7
படம்
Published
02/19/2023 11:35 AM
PRM-470 connector shows 1 for battery
படம்
Published
03/16/2021 1:39 AM
PRM-470_Reconnect_battery
படம்
Published
05/10/2020 8:16 PM
PRM-470 Neutron sleeve remove 1
படம்
Published
03/16/2021 1:39 AM
PRM-470_di_ntube2_000532.png
படம்
Published
03/16/2021 1:39 AM
PRM-470 Neutron locate manufacturer
படம்
Published
03/16/2021 1:39 AM
Fluke DMM HV Probe connector
படம்
Published
03/16/2021 1:39 AM
Fluke DMM on DC Voltage Dial
படம்
Published
03/16/2021 1:39 AM
PRM-470 locate TP38 ground 1 of 4
படம்
Published
03/16/2021 1:39 AM
PRM-470 locate TP38 ground 2 of 4
படம்
Published
03/16/2021 1:39 AM
PRM-470 locate TP38 ground 3 of 4
படம்
Published
03/16/2021 1:39 AM
PRM-470 locate TP38 ground 4 of 4
படம்
Published
03/16/2021 1:39 AM
PRM-470 locate TP34 high voltage
படம்
Published
03/16/2021 1:39 AM
PRM-470 TP34 high voltage measure
படம்
Published
02/19/2023 11:45 AM
PRM-470 adjust R10 location
படம்
Published
02/19/2023 11:32 AM
PRM-470 adjust R10 twist
படம்
Published
03/16/2021 1:39 AM
Fluke DMM Display in HV Probe Multiplier
படம்
Published
03/16/2021 1:39 AM
PRM-470 adjust R17 location
படம்
Published
02/19/2023 11:40 AM
PRM-470 adjust R17 twist
படம்
Published
03/16/2021 1:39 AM
PRM-470_di_close_000508.png
படம்
Published
06/18/2021 6:08 PM
PRM-470_Tighten_4_cover_screws
படம்
Published
05/10/2020 8:22 PM

PRM-470 Gamma Alignment

Last Updated: 04/24/2025 1:34 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
2548 - 205 - Disconnect charger; Remark:2547
Procedure Step
Published
08/02/2021 12:46 PM
205 - Disconnect charger
Instruction
Published
07/23/2021 2:42 PM
2547 - IO PRM-470 Remove Charger Cable
Remark
Published
11/02/2024 3:10 PM
2633 - 220 - Power on instrument; Remark:2545
Procedure Step
Published
08/12/2021 12:19 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
2545 - IO PRM-470 Power Button Red Box
Remark
Published
10/29/2024 9:30 PM
261 - 65 - Loosen four (4) cover screws; Remark:264
Procedure Step
Published
08/12/2021 7:25 AM
65 - Loosen four (4) cover screws
Instruction
Published
07/22/2021 12:38 PM
264 - Use a medium Phillips (PH2) screwdriver. Screws do not need to be completely ...
Remark
Published
07/20/2021 3:59 PM
2642 - 262 - Remove instrument cover; Remark:2631
Procedure Step
Published
08/23/2021 11:44 AM
2634 - Warning : Do not touch circuit board components as there are high voltage energized tes...
Annotation
Published
08/23/2021 11:27 AM
262 - Remove instrument cover
Instruction
Published
07/20/2021 3:35 PM
2631 - IO PRM-470 Cover Open Action
Remark
Published
11/02/2024 3:12 PM
2643 - 2551 - Connect Multichannel Analyzer (MCA); Remark:
Procedure Step
Published
08/23/2021 11:44 AM
2551 - Connect Multichannel Analyzer (MCA)
Instruction
Published
08/03/2021 2:13 PM
2644 - 2552 - Connect MCA to computer; Remark:2630
Procedure Step
Published
08/23/2021 11:44 AM
2552 - Connect MCA to computer
Instruction
Published
08/03/2021 2:15 PM
2630 - Use provided Mini-USB [universal serial bus] to USB cable.
Remark
Published
08/23/2021 11:29 AM
2645 - 2553 - Attach test leads to MCA; Remark:2629
Procedure Step
Published
08/23/2021 11:37 AM
2553 - Attach test leads to MCA
Instruction
Published
08/03/2021 2:15 PM
2629 - Connect to the SIGNAL input.
Remark
Published
08/23/2021 11:38 AM
2646 - 2554 - Collect pre-alignment Spectra; Remark:
Procedure Step
Published
08/23/2021 11:45 AM
2554 - Collect pre-alignment Spectra
Instruction
Published
08/03/2021 2:16 PM
2647 - 2555 - Start Amptek software; Remark:2627
Procedure Step
Published
08/23/2021 12:04 PM
2555 - Start Amptek software
Instruction
Published
08/03/2021 2:17 PM
2627 - IO Amptek Icon DppMCA
Remark
Published
11/02/2024 3:13 PM
2648 - 2556 - Verify MCA settings; Remark:2626
Procedure Step
Published
08/23/2021 11:40 AM
2556 - Verify MCA settings
Instruction
Published
08/23/2021 11:39 AM
2626 - Settings should match the image below.
Remark
Published
08/23/2021 11:39 AM
2649 - 2557 - Place Cs-137 source on gamma detector; Remark:2625
Procedure Step
Published
08/23/2021 11:46 AM
2557 - Place Cs-137 source on gamma detector
Instruction
Published
08/11/2021 3:25 PM
2625 - Center the source on the detector.
Remark
Published
08/11/2021 3:27 PM
2650 - 2558 - Connect leads to stage 1 test points; Remark:2624
Procedure Step
Published
08/23/2021 11:46 AM
2558 - Connect leads to stage 1 test points
Instruction
Published
08/11/2021 3:30 PM
2624 - Black test lead connects to TP20. Red test lead connects to TP1.
Remark
Published
08/11/2021 3:59 PM
2651 - 2559 - Collect stage 1 spectrum; Remark:2622
Procedure Step
Published
08/23/2021 11:47 AM
2635 - Standard : At least 2,000 counts should be recorded on the peak channel.
Annotation
Published
10/04/2022 11:15 AM
2636 - Note : Hot keys for DppMCA software: “A” will clear spectrum and restart acquisi...
Annotation
Published
10/04/2022 11:13 AM
2559 - Collect stage 1 spectrum
Instruction
Published
08/11/2021 3:30 PM
2622 - The traffic light icon will start/stop data acquisition.
Remark
Published
08/11/2021 4:00 PM
2623 - Placing the cursor at the peak location will aid in monitoring the counts in ...
Remark
Published
08/11/2021 4:01 PM
2652 - 2560 - Save stage 1 spectrum; Remark:2620
Procedure Step
Published
08/23/2021 11:47 AM
2637 - Standard : Use the following file naming convention: CO_SITE_MO_SN_STG_PP CO =...
Annotation
Published
08/23/2021 11:50 AM
2560 - Save stage 1 spectrum
Instruction
Published
08/11/2021 3:31 PM
2620 - IO Amptek Save File in DppMCA
Remark
Published
11/02/2024 3:14 PM
2621 - Use the correct file naming convention. Example: US_Richland_CG_102...
Remark
Published
08/11/2021 4:03 PM
2653 - 2561 - Connect leads to stage 2 test point; Remark:2619
Procedure Step
Published
08/23/2021 11:51 AM
2561 - Connect leads to stage 2 test point
Instruction
Published
08/11/2021 3:32 PM
2619 - Red test lead connects to TP2.
Remark
Published
08/11/2021 4:03 PM
2654 - 2562 - Collect stage 2 spectrum; Remark:2617
Procedure Step
Published
08/23/2021 11:52 AM
2562 - Collect stage 2 spectrum
Instruction
Published
08/11/2021 3:33 PM
2617 - Start data collection.
Remark
Published
08/11/2021 4:04 PM
2618 - Placing the cursor at the peak location will aid in monitoring the counts in ...
Remark
Published
08/11/2021 4:05 PM
2655 - 2563 - Save stage 2 spectrum; Remark:2616
Procedure Step
Published
08/23/2021 11:52 AM
2563 - Save stage 2 spectrum
Instruction
Published
08/11/2021 3:38 PM
2616 - Use the correct file naming convention. Example: US_Richland_CG_102...
Remark
Published
08/11/2021 4:06 PM
2656 - 2564 - Measure pre-alignment high voltage; Remark:
Procedure Step
Published
05/03/2022 12:00 PM
2564 - Measure pre-alignment high voltage
Instruction
Published
05/03/2022 12:00 PM
2657 - 2565 - Connect HV probe to DMM; Remark:2614
Procedure Step
Published
08/23/2021 11:53 AM
2565 - Connect HV probe to DMM
Instruction
Published
08/11/2021 3:41 PM
2614 - Use Fluke 80K-40 high voltage (HV) probe and Fluke 179 DMM.
Remark
Published
08/11/2021 4:06 PM
2658 - 2566 - Switch DMM to DC voltage; Remark:2613
Procedure Step
Published
08/18/2021 10:21 PM
2566 - Switch DMM to DC voltage
Instruction
Published
08/11/2021 3:41 PM
2613 - IO Fluke DMM Dial DC Volt
Remark
Published
11/02/2024 3:15 PM
2659 - 2567 - Locate ground test point; Remark:2610
Procedure Step
Published
08/23/2021 11:53 AM
2567 - Locate ground test point
Instruction
Published
08/11/2021 3:42 PM
2610 - The ground TP38 is located on the GHA-472 gamma board and might not be clearl...
Remark
Published
08/23/2021 11:41 AM
2660 - 2568 - Connect HV probe to ground; Remark:2609
Procedure Step
Published
08/23/2021 11:54 AM
2568 - Connect HV probe to ground
Instruction
Published
08/11/2021 3:42 PM
2609 - Connect black test lead of HV probe to TP38. Verify good connection.
Remark
Published
08/11/2021 4:08 PM
2661 - 2569 - Locate HV test point; Remark:2607
Procedure Step
Published
08/23/2021 11:42 AM
2569 - Locate HV test point
Instruction
Published
08/23/2021 11:41 AM
2607 - The HV TP34 is located on the GHA-472 gamma board and might not be clearly la...
Remark
Published
08/11/2021 4:09 PM
2662 - 2570 - Connect HV probe to test point; Remark:2605
Procedure Step
Published
08/23/2021 11:54 AM
2638 - Warning : The test point has high voltage. Only contact the test point with the high vo...
Annotation
Published
10/04/2022 11:21 AM
2570 - Connect HV probe to test point
Instruction
Published
08/23/2021 11:30 AM
2605 - Contact TP34 with tip of red HV probe.
Remark
Published
08/11/2021 4:10 PM
2663 - 2571 - Document voltage; Remark:2589
Procedure Step
Published
08/23/2021 11:55 AM
2639 - Standard : Typical values are between 650 and 850 Volts DC. Older units might have highe...
Annotation
Published
10/04/2022 11:17 AM
2571 - Document voltage
Instruction
Published
08/11/2021 3:45 PM
2589 - The number indicated on the DMM must be multiplied by 1000 if the 80K-40 prob...
Remark
Published
08/11/2021 4:11 PM
2664 - 2572 - Remove HV probe from PRM-470; Remark:
Procedure Step
Published
08/23/2021 11:55 AM
2572 - Remove HV probe from PRM-470
Instruction
Published
08/11/2021 3:46 PM
2665 - 2573 - Adjust gains; Remark:
Procedure Step
Published
08/23/2021 11:55 AM
2573 - Adjust gains
Instruction
Published
08/11/2021 3:46 PM
2666 - 2574 - Connect leads to stage 1 test points; Remark:2602
Procedure Step
Published
08/23/2021 11:56 AM
2574 - Connect leads to stage 1 test points
Instruction
Published
08/11/2021 3:47 PM
2602 - Black test lead connects to TP20. Red test lead connects to TP1.
Remark
Published
05/03/2022 4:24 PM
2667 - 2575 - Verify Cs-137 source is on center of gamma detector; Remark:2601
Procedure Step
Published
08/23/2021 11:56 AM
2575 - Verify Cs-137 source is on center of gamma detector
Instruction
Published
08/11/2021 3:47 PM
2601 - IO PRM-470 Source on Gamma Detector
Remark
Published
11/02/2024 3:02 PM
2668 - 2576 - Adjust stage 1 gain; Remark:2599
Procedure Step
Published
08/23/2021 11:56 AM
2640 - Standard : 1st stage gain = Peak location should be at channel 115 +/- 3 channels.
Annotation
Published
10/04/2022 11:18 AM
2576 - Adjust stage 1 gain
Instruction
Published
08/11/2021 3:48 PM
2599 - Use precision slotted (tweaker) screwdriver to adjust R10 until peak is cente...
Remark
Published
08/11/2021 4:14 PM
2669 - 2559 - Collect stage 1 spectrum; Remark:2598
Procedure Step
Published
08/23/2021 11:57 AM
2598 - Place the cursor at channel 115 to monitor the counts.
Remark
Published
05/03/2022 12:01 PM
2670 - 2560 - Save stage 1 spectrum; Remark:2597
Procedure Step
Published
08/23/2021 11:57 AM
2597 - Use the correct file naming convention. Example: US_Richland_CG_102...
Remark
Published
08/11/2021 4:15 PM
2671 - 2577 - Connect to stage 2 test point; Remark:2596
Procedure Step
Published
08/23/2021 11:58 AM
2577 - Connect to stage 2 test point
Instruction
Published
08/11/2021 3:49 PM
2596 - Move red test lead to TP2.
Remark
Published
08/11/2021 4:16 PM
2672 - 2578 - Adjust stage 2 gain; Remark:2593
Procedure Step
Published
08/23/2021 11:58 AM
2641 - Standard : 2nd stage gain = Peak location should be at channel 310 +/- 5 channels.
Annotation
Published
10/04/2022 11:17 AM
2578 - Adjust stage 2 gain
Instruction
Published
08/11/2021 3:50 PM
2593 - Use Tweaker to adjust R17 (may not be labeled on board) until peak is centere...
Remark
Published
08/11/2021 4:16 PM
2673 - 2562 - Collect stage 2 spectrum; Remark:2592
Procedure Step
Published
08/23/2021 11:59 AM
2592 - Place the cursor at channel 310 to monitor the counts.
Remark
Published
08/11/2021 4:17 PM
2674 - 2563 - Save stage 2 spectrum; Remark:2591
Procedure Step
Published
08/23/2021 11:59 AM
2591 - Use the correct file naming convention. Example: US_Richland_CG_102...
Remark
Published
08/11/2021 4:18 PM
2675 - 2579 - Remove test leads; Remark:
Procedure Step
Published
08/23/2021 11:59 AM
2579 - Remove test leads
Instruction
Published
08/11/2021 3:51 PM
2676 - 2580 - Disconnect MCA; Remark:
Procedure Step
Published
08/23/2021 12:00 PM
2580 - Disconnect MCA
Instruction
Published
08/11/2021 3:51 PM
2677 - 2581 - Remove radioactive source; Remark:
Procedure Step
Published
08/23/2021 12:00 PM
2581 - Remove radioactive source
Instruction
Published
08/11/2021 3:52 PM
2678 - 2582 - Measure post-alignment high voltage; Remark:
Procedure Step
Published
08/23/2021 12:00 PM
2582 - Measure post-alignment high voltage
Instruction
Published
08/11/2021 3:53 PM
2679 - 2583 - Measure high voltage; Remark:2590
Procedure Step
Published
08/23/2021 12:01 PM
2583 - Measure high voltage
Instruction
Published
08/11/2021 3:54 PM
2590 - Repeat high voltage measurement steps provided earlier in this procedure....
Remark
Published
08/11/2021 4:19 PM
2680 - 2571 - Document voltage; Remark:2589
Procedure Step
Published
08/23/2021 12:01 PM
2681 - 2572 - Remove HV probe from PRM-470; Remark:2588
Procedure Step
Published
08/23/2021 12:01 PM
2588 - Power off DMM. Carefully remove test leads.
Remark
Published
08/11/2021 4:20 PM
303 - 302 - Reassemble instrument; Remark:
Procedure Step
Published
10/07/2022 1:49 PM
302 - Reassemble instrument
Instruction
Published
06/16/2021 11:28 PM
2684 - 307 - Replace front cover; Remark:
Procedure Step
Published
08/23/2021 12:02 PM
304 - Caution : Ensure all wiring is inside the instrument while replacing cover.
Annotation
Published
10/04/2022 11:10 AM
307 - Replace front cover
Instruction
Published
07/20/2021 3:57 PM
310 - 309 - Tighten four (4) cover screws; Remark:417
Procedure Step
Published
08/11/2021 2:04 PM
569 - Caution : Do not overtighten screws.
Annotation
Published
06/16/2021 2:31 PM
309 - Tighten four (4) cover screws
Instruction
Published
07/22/2021 1:02 PM
417 - Use a medium Phillips (PH2) screwdriver.
Remark
Published
07/22/2021 1:18 PM
320 - 319 - Perform operational test; Remark:347
Procedure Step
Published
08/12/2021 7:12 AM
319 - Perform operational test
Instruction
Published
06/10/2021 8:31 PM
347 - Perform DET-HHD-RAP-RM01 PRM-470 Operational Test.
Remark
Published
07/21/2021 12:22 PM
2682 - 2584 - Perform efficiency test; Remark:2587
Procedure Step
Published
08/23/2021 12:02 PM
2584 - Perform efficiency test
Instruction
Published
08/11/2021 3:55 PM
2587 - DET-HHD-RAP-RM05, PRM-470 Efficiency Test
Remark
Published
08/11/2021 4:22 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
30 minutes
Duration
Published
04/01/2020 11:35 AM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Warning
Annotation Type
Published
03/02/2020 3:25 PM
Standard
Annotation Type
Published
03/02/2020 3:25 PM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Technician 2
Skill Level
Published
04/09/2020 1:17 AM
Precision slotted (tweaker) screwdriver
Tools
Published
03/20/2021 6:23 PM
Amptek MCA8000D multichannel analyzer (MCA)
Tools
Published
03/20/2021 6:22 PM
Computer with MCA software installed (Amptek DppMCA)
Tools
Published
03/20/2021 6:22 PM
Mini-USB male to USB-A male cable
Tools
Published
03/19/2021 3:37 PM
Test leads for MCA
Tools
Published
03/19/2021 3:37 PM
Fluke 179 Digital Multimeter (DMM)
Tools
Published
03/20/2021 6:27 PM
Test leads for DMM
Tools
Published
03/20/2021 6:28 PM
Fluke 80K-40 high voltage (HV) probe
Tools
Published
03/20/2021 6:23 PM
Medium Phillips (PH2) screwdriver
Tools
Published
04/24/2020 4:40 PM
PRM-470 Disconnect Charger Cable
படம்
Published
03/16/2021 1:39 AM
PRM-470_Partial_Power_Button_Red_Highlight.png
படம்
Published
06/18/2021 6:28 PM
PRM470_Remove_Screws_Top_Case
படம்
Published
05/08/2020 6:05 PM
PRM-470_di_CGN_open_00536
படம்
Published
02/19/2023 11:26 AM
Amptek MCA miniUSB connection
படம்
Published
03/17/2021 2:46 AM
Amptek MCA connection to computer
படம்
Published
03/17/2021 2:46 AM
Amptek MCA connection to signal input
படம்
Published
03/17/2021 2:46 AM
Amptek Software Logo DppMCA
படம்
Published
03/17/2021 2:46 AM
Amptek Software DppMCA Load Settings
படம்
Published
03/17/2021 2:46 AM
Amptek Software DppMCA Settings
படம்
Published
03/17/2021 2:46 AM
PRM-470 Gamma Source on Detector Interior
படம்
Published
03/17/2021 2:46 AM
PRM-470 TP20 Ground on main CB
படம்
Published
03/17/2021 2:46 AM
Amptek Software DppMCA Traffic Light
படம்
Published
03/17/2021 2:46 AM
Amptek Software DppMCA Spectra Channel 120
படம்
Published
03/17/2021 2:46 AM
Amptek Software DppMCA Save File
படம்
Published
03/17/2021 2:46 AM
PRM-470 GHA-472 TP2
படம்
Published
03/17/2021 2:46 AM
Amptek Software DppMCA Spectra Channel 291
படம்
Published
03/17/2021 2:46 AM
Fluke DMM on DC Voltage Dial
படம்
Published
03/16/2021 1:39 AM
PRM-470 GHA-472 Locate TP38
படம்
Published
03/17/2021 2:46 AM
PRM-470 GHA-472 Connect TP38
படம்
Published
03/17/2021 2:46 AM
PRM-470 GHA-472 Locate TP34
படம்
Published
03/17/2021 2:46 AM
PRM-470 GHA-472 Connect TP34
படம்
Published
03/17/2021 2:46 AM
Fluke DMM Display 0.745 in HV Probe Multiplier
படம்
Published
03/17/2021 2:46 AM
PRM-470 GHA-472 Connect TP1 Stage 1
படம்
Published
03/17/2021 2:46 AM
PRM-470 GHA-472 Adjust R10
படம்
Published
02/19/2023 11:19 AM
Amptek Software DppMCA Spectra Channel 115
படம்
Published
03/17/2021 2:46 AM
PRM-470 GHA-472 Adjust R17
படம்
Published
03/17/2021 2:46 AM
Amptek Software DppMCA Spectra Channel 310
படம்
Published
03/17/2021 2:46 AM
PRM-470_Tighten_4_cover_screws
படம்
Published
05/10/2020 8:22 PM