2884 - Standard : High voltage for a GE Reuter-Stokes neutron detection is 990 Volts. High volt...
Type
Standard
High voltage for a GE Reuter-Stokes neutron detection is 990 Volts. High voltage for a LND neutron detector is 1075 Volts.
Step 00
2881 - Adjust neutron high voltage
Adjust neutron high voltage
2875 - Use precision slotted (tweaker) screwdriver to adjust R10. The voltage should...
Use precision slotted (tweaker) screwdriver to adjust R10. The voltage should match the manufacturer’s recommendation.
GE Reuter-Stokes: 990 Volts
LND: 1075 Volts
படம்

2878 - The number indicated on the DMM must be multiplied by 1000 if the Fluke 80K-4...
The number indicated on the DMM must be multiplied by 1000 if the Fluke 80K-40 probe is being used.
படம்