radEAGLET Operational Test

Last Updated: 04/25/2025 1:44 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
ORTEC radEAGLET-R 2SG
Device
Published
10/08/2024 1:48 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
10880 - 10722 - Place instrument next to calibration block; Remark:10720
Procedure Step
Published
07/25/2024 9:39 PM
10722 - Place instrument next to calibration block
Instruction
Published
07/19/2024 2:28 PM
10720 - This is not required for models with an internal Cesium-137 (Cs-137) source. ...
Remark
Published
07/19/2024 2:28 PM
10879 - radEAGLET Next to calibration block
Remark
Published
09/25/2024 8:47 PM
10881 - 220 - Power on instrument; Remark:10718
Procedure Step
Published
07/25/2024 9:44 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
10718 - Hold power button for about 2 seconds.
Remark
Published
07/19/2024 2:35 PM
10878 - radEAGLET Power instrument on
Remark
Published
09/26/2024 6:51 PM
10728 - 612 - Wait for instrument to complete start-up sequence; Remark:10713
Procedure Step
Published
04/13/2025 6:58 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
10713 - This may take 1-3 minutes.
Remark
Published
07/19/2024 2:45 PM
10714 - radEAGLET Start up instrument stabilizing
Remark
Published
09/26/2024 7:15 PM
10715 - radEAGLET Start up sequence
Remark
Published
09/26/2024 7:14 PM
10882 - 10712 - Verify sufficient battery charge; Remark:10710
Procedure Step
Published
07/25/2024 9:50 PM
10712 - Verify sufficient battery charge
Instruction
Published
07/19/2024 2:57 PM
10710 - A fully charged battery will last approximately 8 hours. Confirm remaining ba...
Remark
Published
07/19/2024 3:00 PM
10877 - radEAGLET Verify battery charge
Remark
Published
09/25/2024 9:56 PM
10730 - 10709 - Verify instrument is stabilized; Remark:10707
Procedure Step
Published
07/19/2024 3:04 PM
10709 - Verify instrument is stabilized
Instruction
Published
07/19/2024 3:02 PM
10707 - The instrument is stabilized when the icon is green.
Remark
Published
07/19/2024 3:03 PM
10708 - radEAGLET Verify instrument stablized
Remark
Published
09/25/2024 10:09 PM
2253 - 2208 - Acquire Background Measurement; Remark:
Procedure Step
Published
08/27/2024 4:49 PM
2208 - Acquire Background Measurement
Instruction
Published
04/06/2021 4:47 PM
10900 - 1437 - Verify Response to Gamma Source; Remark:
Procedure Step
Published
07/25/2024 10:12 PM
1437 - Verify Response to Gamma Source
Instruction
Published
09/01/2020 7:45 PM
10918 - 10672 - Verify instrument is in dose rate mode; Remark:10917
Procedure Step
Published
07/28/2024 2:45 PM
10672 - Verify instrument is in dose rate mode
Instruction
Published
07/19/2024 4:25 PM
10917 - radEAGLET verify instrument is in dose rate mode
Remark
Published
09/25/2024 10:07 PM
10896 - 235 - Place instrument next to radioactive gamma source; Remark:10895
Procedure Step
Published
07/25/2024 10:16 PM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
10895 - radEAGLET Next to radioactive gamma source
Remark
Published
09/26/2024 6:59 PM
10897 - 3097 - Verify instrument responds correctly; Remark:10889
Procedure Step
Published
07/25/2024 10:25 PM
3097 - Verify instrument responds correctly
Instruction
Published
07/16/2021 10:50 AM
10889 - Dose rate mode should respond as follows: • Gamma counts per second ...
Remark
Published
07/25/2024 10:18 PM
10890 - radEAGLET Background color will change
Remark
Published
09/26/2024 7:20 PM
10891 - • Gamma LED will flash
Remark
Published
07/25/2024 10:19 PM
10892 - radEAGLET Gamma LED will flash
Remark
Published
09/26/2024 7:22 PM
10893 - • Audible alarm will sound (if enabled) • Instrument will vibrate ...
Remark
Published
07/25/2024 10:23 PM
10894 - radEAGLET Instrument will vibrate
Remark
Published
09/26/2024 7:25 PM
10898 - 10888 - Move instrument away from radioactive source; Remark:10901
Procedure Step
Published
07/25/2024 10:30 PM
10888 - Move instrument away from radioactive source
Instruction
Published
07/25/2024 10:26 PM
10901 - radEAGLET Move instrument away from radioactive source
Remark
Published
09/26/2024 7:33 PM
10916 - 2209 - Verify instrument is in a suitable background radiation area; Remark:10693
Procedure Step
Published
07/28/2024 2:35 PM
2209 - Verify instrument is in a suitable background radiation area
Instruction
Published
06/11/2024 3:31 PM
10693 - The instrument should be located near the secondary inspection area, in simil...
Remark
Published
07/19/2024 3:12 PM
10915 - No source
Remark
Published
09/26/2024 7:36 PM
10899 - 10692 - Enter DETECT mode; Remark:10690
Procedure Step
Published
07/25/2024 10:33 PM
10695 - Note : Left, right, and down arrow Buttons will select screen options. The remainder...
Annotation
Published
07/19/2024 3:14 PM
10692 - Enter DETECT mode
Instruction
Published
07/19/2024 3:16 PM
10690 - Press down arrow to select DETECT.
Remark
Published
07/19/2024 3:16 PM
10885 - radEAGLET Enter DETECT mode
Remark
Published
09/26/2024 7:38 PM
10698 - 2220 - Wait for measurement to complete; Remark:10688
Procedure Step
Published
07/19/2024 3:23 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
10688 - This should take approximately 30 seconds. Instrument is ready when percentage b...
Remark
Published
07/19/2024 3:22 PM
10689 - radEAGLET Wait for measurement to complete
Remark
Published
09/26/2024 7:42 PM
10903 - 235 - Place instrument next to radioactive gamma source; Remark:10938
Procedure Step
Published
07/28/2024 11:53 AM
10938 - radEAGLET next to radioactive gamma source GREEN screen
Remark
Published
09/26/2024 7:49 PM
10912 - 3097 - Verify instrument responds correctly; Remark:10904
Procedure Step
Published
07/28/2024 12:09 PM
10904 - DETECT mode should respond as follows: • Gamma dose...
Remark
Published
07/28/2024 11:56 AM
10905 - radEAGLET Graph values will increase change to red
Remark
Published
09/26/2024 7:50 PM
10906 - radEAGLET Gamma LED will flash
Remark
Published
09/26/2024 7:51 PM
10907 - • Audible alerts will increase in pitch (if enabled) • Instrument ...
Remark
Published
07/28/2024 12:00 PM
10908 - radEAGLET Instrument will vibrate GRAPH screen
Remark
Published
09/26/2024 7:53 PM
10921 - 10888 - Move instrument away from radioactive source; Remark:10919
Procedure Step
Published
07/28/2024 2:56 PM
10919 - radEAGLET move instrument away from source red GRAPH screen
Remark
Published
09/26/2024 7:55 PM
10922 - 10680 - Exit DETECT mode; Remark:10923
Procedure Step
Published
07/28/2024 3:01 PM
10680 - Exit DETECT mode
Instruction
Published
07/19/2024 4:18 PM
10923 - Select EXIT
Remark
Published
07/28/2024 3:11 PM
10920 - radEAGLET Select EXIT
Remark
Published
09/27/2024 7:08 PM
10914 - 10913 - Verify Identification Performance; Remark:
Procedure Step
Published
07/28/2024 12:13 PM
10913 - Verify Identification Performance
Instruction
Published
07/28/2024 12:11 PM
10941 - 235 - Place instrument next to radioactive gamma source; Remark:10940
Procedure Step
Published
07/28/2024 7:51 PM
10940 - radEAGLET Next to radioactive gamma source
Remark
Published
09/27/2024 7:21 PM
10928 - 10672 - Verify instrument is in dose rate mode; Remark:10927
Procedure Step
Published
07/28/2024 3:56 PM
10927 - radEAGLET instrument is in dose rate mode
Remark
Published
09/27/2024 7:22 PM
10926 - 859 - Start identification measurement; Remark:10665
Procedure Step
Published
07/28/2024 3:48 PM
859 - Start identification measurement
Instruction
Published
10/26/2020 7:00 PM
10665 - Select EASY ID
Remark
Published
07/19/2024 4:28 PM
10924 - radeAGLET Select EASY ID
Remark
Published
09/27/2024 7:24 PM
10932 - Highlight Unknown and choose SELECT.
Remark
Published
07/28/2024 4:00 PM
10933 - radEAGLET Highlight Unknown and choose SELECT
Remark
Published
09/27/2024 7:29 PM
10934 - 2220 - Wait for measurement to complete; Remark:10737
Procedure Step
Published
07/28/2024 4:08 PM
10737 - The screen might prompt the operator to move the instrument away or towards t...
Remark
Published
07/20/2024 7:34 PM
10931 - radEAGLET Wait for measurement to complete
Remark
Published
09/27/2024 7:34 PM
10935 - 1442 - Verify identification results are correct; Remark:10929
Procedure Step
Published
07/28/2024 4:14 PM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
10929 - Identification results should match the isotope(s) present.
Remark
Published
07/28/2024 4:10 PM
10930 - radEAGLET results should match isotope present
Remark
Published
09/27/2024 7:38 PM
10937 - 10749 - Exit EASY ID mode; Remark:10678
Procedure Step
Published
07/28/2024 7:41 PM
10749 - Exit EASY ID mode
Instruction
Published
07/20/2024 8:24 PM
10678 - Select EXIT
Remark
Published
07/19/2024 4:18 PM
10936 - radEAGLET Select EXIT
Remark
Published
09/27/2024 7:42 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
radEAGLET Next to calibration block
Image
Published
07/19/2024 4:32 PM
radEAGLET Power instrument on
Image
Published
07/19/2024 4:32 PM
radEAGLET Start up instrument stabilizing
Image
Published
07/19/2024 4:32 PM
radEAGLET start up sequence
Image
Published
07/19/2024 4:32 PM
radEAGLET Verify battery charge
Image
Published
07/19/2024 4:32 PM
radEAGLET Verify instrument stabilized
Image
Published
07/19/2024 4:32 PM
radEAGLET start up sequence
Image
Published
07/21/2024 3:26 PM
radEAGLET Next to radioactive gamma source
Image
Published
07/25/2024 10:49 PM
radEAGLET Background color will change
Image
Published
07/25/2024 10:49 PM
radEAGLET Gamma LED will flash
Image
Published
07/25/2024 10:49 PM
radEAGLET instrument will vibrate
Image
Published
07/28/2024 3:22 PM
radEAGLET Move instrument away from source
Image
Published
07/26/2024 1:25 PM
No Source Found
Image
Published
12/07/2023 2:54 PM
RadEAGLET Enter DETECT mode_V2.png
Image
Published
07/20/2024 1:49 PM
radEAGLET Wait for measurement to complete
Image
Published
07/19/2024 4:32 PM
radEAGLET next to radioactive gamma source GREEN screen
Image
Published
07/28/2024 2:38 PM
radEAGLET Graph values will increase change to red
Image
Published
07/28/2024 12:09 PM
radEAGLET Instrument will vibrate GRAPH.png
Image
Published
07/25/2024 10:49 PM
radEAGLET move instrument away from source red GRAPH screen
Image
Published
07/28/2024 3:01 PM
radEAGLET Select EXIT DETECT mode
Image
Published
07/25/2024 9:20 PM
radEAGLET Verify instrument is in dose rate mode
Image
Published
07/19/2024 4:32 PM
radEAGLET Select EASY ID
Image
Published
07/19/2024 4:32 PM
radEAGLET Highlight Unknown and choose SELECT
Image
Published
07/28/2024 4:14 PM
radEAGLET Wait for measurement to complete
Image
Published
07/20/2024 7:54 PM
radEAGLET results should match isotope present
Image
Published
07/28/2024 4:14 PM
radEAGLET Select EXIT
Image
Published
07/20/2024 8:51 PM

TA600 Radiation Pager Operational Test

Last Updated: 04/25/2025 1:42 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech TA600 Radiation Pager
Device
Published
10/28/2024 3:51 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
7555 - 220 - Power on instrument; Remark:7542
Procedure Step
Published
06/05/2023 5:57 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
7542 - Set instrument to audio only mode.
Remark
Published
04/13/2025 5:48 PM
7556 - 612 - Wait for instrument to complete start-up sequence; Remark:7512
Procedure Step
Published
04/13/2025 6:50 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
7512 - Red numerical LED display will flash while collecting background for 30 secon...
Remark
Published
10/19/2023 7:54 AM
7557 - 7543 - Test Audible Alarm; Remark:
Procedure Step
Published
06/05/2023 5:59 PM
7543 - Test Audible Alarm
Instruction
Published
06/05/2023 3:31 PM
7558 - 7544 - Place instrument next to source of radiation; Remark:7545
Procedure Step
Published
06/05/2023 5:59 PM
7544 - Place instrument next to source of radiation
Instruction
Published
06/05/2023 3:32 PM
7545 - TA600 Next to Source
Remark
Published
06/05/2023 3:34 PM
7559 - 2936 - Verify audible alarm functions correctly; Remark:7547
Procedure Step
Published
06/05/2023 6:00 PM
2936 - Verify audible alarm functions correctly
Instruction
Published
06/15/2021 4:09 PM
7547 - The instrument should respond as follows: Emit an audible b...
Remark
Published
06/05/2023 5:00 PM
7560 - 7548 - Acknowledge alarm; Remark:7549
Procedure Step
Published
06/05/2023 6:01 PM
7548 - Acknowledge alarm
Instruction
Published
06/05/2023 5:03 PM
7549 - Press display button to stop alarm. Otherwise, the alarm will automatically s...
Remark
Published
06/05/2023 5:07 PM
7561 - 7550 - Move instrument away from source of radiation; Remark:7551
Procedure Step
Published
06/05/2023 6:02 PM
7550 - Move instrument away from source of radiation
Instruction
Published
06/05/2023 5:09 PM
7551 - Move source away from TA600
Remark
Published
06/05/2023 5:12 PM
2914 - 2912 - Test vibration alarm; Remark:
Procedure Step
Published
06/06/2022 5:28 PM
2912 - Test vibration alarm
Instruction
Published
06/15/2021 4:11 PM
7562 - 7546 - Change to vibrate mode; Remark:7552
Procedure Step
Published
06/05/2023 6:02 PM
7546 - Change to vibrate mode
Instruction
Published
06/05/2023 3:35 PM
7552 - Set instrument to vibrate only mode.
Remark
Published
06/05/2023 5:15 PM
7563 - 612 - Wait for instrument to complete start-up sequence; Remark:7512
Procedure Step
Published
04/13/2025 6:55 PM
7553 - Note : If dial is turned to vibrate mode quickly, the instrument will not repeat sta...
Annotation
Published
06/05/2023 5:17 PM
7564 - 7544 - Place instrument next to source of radiation; Remark:7545
Procedure Step
Published
06/05/2023 6:05 PM
7565 - 2938 - Verify vibration alarm functions correctly; Remark:7554
Procedure Step
Published
06/05/2023 6:06 PM
2938 - Verify vibration alarm functions correctly
Instruction
Published
06/15/2021 4:12 PM
7554 - The instrument should respond as follows: Vibrate
Remark
Published
06/05/2023 5:24 PM
7566 - 7548 - Acknowledge alarm; Remark:7549
Procedure Step
Published
06/05/2023 6:07 PM
7567 - 7550 - Move instrument away from source of radiation; Remark:7551
Procedure Step
Published
06/05/2023 6:08 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
Daily
Frequency
Published
04/01/2020 11:36 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
TA600 Audio Only
Image
Published
06/04/2023 9:55 PM
TA600 Background
Image
Published
06/04/2023 9:59 PM
TA600 Ready
Image
Published
06/04/2023 9:59 PM
TA600 Next to Source
Image
Published
03/29/2024 11:57 AM
TA600 Audible Alarm
Image
Published
06/05/2023 5:00 PM
TA600 Acknowledge Alarm
Image
Published
06/05/2023 5:07 PM
TA600 Away From Source
Image
Published
06/05/2023 5:11 PM
TA600 Vibrate Only
Image
Published
06/04/2023 9:56 PM
TA600 Vibration Alarm
Image
Published
06/05/2023 5:23 PM

Handheld Radiation Monitor Operational Test

Last Updated: 04/25/2025 1:39 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech Handheld Radiation Monitor
Device
Published
10/28/2024 3:52 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
4316 - 220 - Power on instrument; Remark:4315
Procedure Step
Published
06/06/2022 5:58 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
4315 - Set instrument to AUDIO mode.
Remark
Published
06/06/2022 5:58 PM
4290 - 612 - Wait for instrument to complete start-up sequence; Remark:4286
Procedure Step
Published
04/13/2025 6:50 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
4286 - The unit is ready for operation when the period flashes every 20 seconds....
Remark
Published
06/06/2022 4:11 PM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
4312 - 235 - Place instrument next to radioactive gamma source; Remark:4884
Procedure Step
Published
06/06/2022 6:00 PM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
4884 - IO HRM Move Closer to Rad Source
Remark
Published
08/04/2022 1:44 PM
4313 - 2936 - Verify audible alarm functions correctly; Remark:4310
Procedure Step
Published
06/06/2022 6:02 PM
2936 - Verify audible alarm functions correctly
Instruction
Published
06/15/2021 4:09 PM
4310 - The instrument should respond as follows: Instrument will e...
Remark
Published
06/06/2022 6:01 PM
4311 - Numeric value will appear in the gamma LED [light-emitting diode] dis...
Remark
Published
06/06/2022 6:02 PM
4314 - 241 - Move instrument away from radioactive gamma source; Remark:4883
Procedure Step
Published
06/06/2022 6:03 PM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
4883 - IO HRM Move Away from Rad Source
Remark
Published
08/04/2022 1:47 PM
2914 - 2912 - Test vibration alarm; Remark:
Procedure Step
Published
06/06/2022 5:28 PM
2912 - Test vibration alarm
Instruction
Published
06/15/2021 4:11 PM
4320 - 4319 - Change to VIBRATE mode; Remark:4318
Procedure Step
Published
06/06/2022 6:08 PM
4319 - Change to VIBRATE mode
Instruction
Published
06/06/2022 6:07 PM
4318 - Quickly move switch to VIBRATE.
Remark
Published
06/06/2022 6:08 PM
4323 - 2938 - Verify vibration alarm functions correctly; Remark:4321
Procedure Step
Published
06/06/2022 6:12 PM
2938 - Verify vibration alarm functions correctly
Instruction
Published
06/15/2021 4:12 PM
4321 - The instrument should respond as follows: Instrument will v...
Remark
Published
06/06/2022 6:11 PM
4322 - Numeric value will appear in the gamma LED display.
Remark
Published
06/06/2022 6:11 PM
4329 - 4328 - Check response with neutron radioactive source; Remark:
Procedure Step
Published
06/06/2022 6:21 PM
4332 - Note : The neutron response can be performed using a neutron radioactive source or b...
Annotation
Published
06/06/2022 6:37 PM
4328 - Check response with neutron radioactive source
Instruction
Published
06/06/2022 6:21 PM
4326 - 3095 - Place instrument next to radioactive neutron source; Remark:4884
Procedure Step
Published
06/06/2022 6:31 PM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
4325 - 3097 - Verify instrument responds correctly; Remark:4310
Procedure Step
Published
06/06/2022 6:24 PM
3096 - Note : Most radioactive neutron sources also have a gamma component, which will likely ...
Annotation
Published
07/23/2021 9:57 AM
3097 - Verify instrument responds correctly
Instruction
Published
07/16/2021 10:50 AM
4324 - Numeric value will appear in the neutron LED display. Blue...
Remark
Published
06/06/2022 6:24 PM
4327 - 3098 - Move instrument away from radioactive neutron source; Remark:4883
Procedure Step
Published
06/06/2022 6:32 PM
3098 - Move instrument away from radioactive neutron source
Instruction
Published
07/16/2021 10:54 AM
4338 - 3095 - Place instrument next to radioactive neutron source; Remark:4884
Procedure Step
Published
06/06/2022 6:31 PM
4331 - 3097 - Verify instrument responds correctly; Remark:4321
Procedure Step
Published
06/06/2022 6:51 PM
4330 - Numeric value will appear in the neutron LED display. Blue...
Remark
Published
06/06/2022 6:51 PM
4339 - 3098 - Move instrument away from radioactive neutron source; Remark:4883
Procedure Step
Published
06/06/2022 6:32 PM
4337 - 4335 - Check response with background neutron radiation; Remark:
Procedure Step
Published
06/06/2022 7:01 PM
4336 - Note : If no neutron radioactive source is available, the following step can be used...
Annotation
Published
06/06/2022 7:01 PM
4335 - Check response with background neutron radiation
Instruction
Published
06/06/2022 7:01 PM
4344 - 4343 - Monitor neutron LED; Remark:4342
Procedure Step
Published
06/06/2022 7:11 PM
4343 - Monitor neutron LED
Instruction
Published
06/06/2022 7:10 PM
4342 - Watch neutron LED carefully in a low-light environment for at least 5 minutes...
Remark
Published
06/06/2022 7:11 PM
4345 - 4341 - Verify instrument responds to background neutron radiation; Remark:4340
Procedure Step
Published
06/06/2022 7:13 PM
4341 - Verify instrument responds to background neutron radiation
Instruction
Published
06/06/2022 7:11 PM
4340 - In natural background, the instrument should respond about one (1) to three (...
Remark
Published
06/06/2022 7:13 PM
4349 - 4343 - Monitor neutron LED; Remark:4342
Procedure Step
Published
06/06/2022 7:11 PM
4350 - 4341 - Verify instrument responds to background neutron radiation; Remark:4348
Procedure Step
Published
06/06/2022 7:17 PM
4348 - In natural background, the instrument should respond about one (1) to three (...
Remark
Published
06/06/2022 7:16 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
HRM Switch in AUDIO position
Image
Published
08/02/2022 5:03 PM
HRM Red LED Dot Period
Image
Published
08/02/2022 4:53 PM
HRM Move to Rad Source Trefoil
Image
Published
08/02/2022 7:57 PM
HRM Alert Audio
Image
Published
08/02/2022 7:51 PM
HRM Red LED Display 3
Image
Published
08/02/2022 7:47 PM
HRM Move away from Rad Source Trefoil
Image
Published
08/02/2022 7:59 PM
HRM Switch in VIBRATE position
Image
Published
08/02/2022 5:06 PM
HRM Alert Vibrate
Image
Published
08/02/2022 7:52 PM
HRM Alert Audio Neutron Full Size
Image
Published
08/04/2022 4:58 PM
HRM Alert Blue LED Neutron Level 3
Image
Published
08/03/2022 10:29 AM
HRM Alert Vibrate Neutron Full Size
Image
Published
08/04/2022 5:00 PM
HRM Alert Audio Neutron
Image
Published
08/03/2022 10:32 AM
HRM Alert Vibrate Neutron
Image
Published
08/03/2022 10:34 AM

PackEye Operational Test

Last Updated: 04/25/2025 1:38 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Thermo FHT1377 PackEye
Device
Published
10/28/2024 3:37 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
3852 - 220 - Power on instrument; Remark:3836
Procedure Step
Published
03/29/2022 1:59 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
3836 - Press power button on the indicator unit for approximately 1 second.
Remark
Published
03/29/2022 1:05 AM
3853 - 612 - Wait for instrument to complete start-up sequence; Remark:3837
Procedure Step
Published
04/13/2025 6:49 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
3837 - Initialization and background measurement may take about 20 seconds. During t...
Remark
Published
03/29/2022 1:08 AM
3854 - 3825 - Verify instrument is ready; Remark:3838
Procedure Step
Published
03/29/2022 2:01 AM
3825 - Verify instrument is ready
Instruction
Published
03/28/2022 6:15 PM
3838 - The flashing green nat light indicates PackEye is ready for ...
Remark
Published
03/29/2022 1:10 AM
3855 - 3826 - Verify fail light is off; Remark:3839
Procedure Step
Published
03/29/2022 2:02 AM
3826 - Verify fail light is off
Instruction
Published
03/28/2022 6:16 PM
3839 - Solid yellow light: neutron detector or electronic failure.
Remark
Published
03/29/2022 1:12 AM
3840 - Flashing yellow light: high gamma radiation level or gamma detector failure.
Remark
Published
03/29/2022 1:13 AM
3856 - 1437 - Verify Response to Gamma Source; Remark:
Procedure Step
Published
03/29/2022 2:03 AM
1437 - Verify Response to Gamma Source
Instruction
Published
09/01/2020 7:45 PM
3857 - 3827 - Select Alarm mode; Remark:3841
Procedure Step
Published
03/29/2022 2:04 AM
3827 - Select Alarm mode
Instruction
Published
03/28/2022 6:17 PM
3841 - The green LED is off. Briefly press power button to change modes.
Remark
Published
03/29/2022 1:15 AM
3858 - 3828 - Place gamma source on instrument; Remark:3842
Procedure Step
Published
03/29/2022 2:05 AM
3828 - Place gamma source on instrument
Instruction
Published
03/28/2022 6:18 PM
3842 - IO PackEye Gamma Source Placed
Remark
Published
03/29/2022 1:23 AM
3859 - 1439 - Verify instrument responds to radioactive source; Remark:3843
Procedure Step
Published
03/29/2022 2:06 AM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
3843 - Additional gamma indicator lights will illuminate on indicator unit and an au...
Remark
Published
03/29/2022 1:22 AM
3860 - 3829 - Acknowledge gamma alarm; Remark:3844
Procedure Step
Published
03/29/2022 2:07 AM
3829 - Acknowledge gamma alarm
Instruction
Published
03/28/2022 6:20 PM
3844 - Briefly press gamma button to acknowledge alarm.
Remark
Published
03/29/2022 1:25 AM
3835 - 3830 - Move source away from instrument; Remark:246
Procedure Step
Published
03/28/2022 6:31 PM
3830 - Move source away from instrument
Instruction
Published
03/28/2022 6:20 PM
246 - IO No source trefoil
Remark
Published
10/07/2024 11:39 AM
3861 - 3831 - Select Chirper mode; Remark:3845
Procedure Step
Published
03/29/2022 2:08 AM
3831 - Select Chirper mode
Instruction
Published
03/28/2022 6:21 PM
3845 - The green LED is on. Briefly press power button to change modes.
Remark
Published
03/29/2022 1:26 AM
3871 - 3828 - Place gamma source on instrument; Remark:3842
Procedure Step
Published
04/04/2022 11:00 AM
3862 - 1439 - Verify instrument responds to radioactive source; Remark:3846
Procedure Step
Published
03/29/2022 2:09 AM
3846 - Additional gamma indicator lights will illuminate. Audible chirper pitch will...
Remark
Published
03/29/2022 1:28 AM
3872 - 3830 - Move source away from instrument; Remark:246
Procedure Step
Published
04/04/2022 11:00 AM
3863 - 3832 - Verify Response to Neutron Source; Remark:
Procedure Step
Published
03/29/2022 2:10 AM
3832 - Verify Response to Neutron Source
Instruction
Published
03/29/2022 2:56 AM
3864 - 3827 - Select Alarm mode; Remark:3841
Procedure Step
Published
03/29/2022 2:11 AM
3865 - 3833 - Place neutron source on instrument; Remark:3847
Procedure Step
Published
03/29/2022 2:17 AM
3833 - Place neutron source on instrument
Instruction
Published
03/28/2022 6:24 PM
3847 - IO PackEye Neutron Source Placed
Remark
Published
03/29/2022 1:32 AM
3866 - 1439 - Verify instrument responds to radioactive source; Remark:3848
Procedure Step
Published
03/29/2022 2:19 AM
3848 - Additional neutron indicator lights will illuminate and an alarm will sound. ...
Remark
Published
03/29/2022 1:34 AM
3867 - 3834 - Acknowledge neutron alarm; Remark:3849
Procedure Step
Published
03/29/2022 2:20 AM
3834 - Acknowledge neutron alarm
Instruction
Published
03/28/2022 6:25 PM
3849 - Briefly press neutron button to acknowledge alarm.
Remark
Published
03/29/2022 1:35 AM
3873 - 3833 - Place neutron source on instrument; Remark:3847
Procedure Step
Published
04/04/2022 11:15 AM
3868 - 1439 - Verify instrument responds to radioactive source; Remark:3851
Procedure Step
Published
04/04/2022 11:03 AM
3851 - Additional neutron indicator lights will illuminate and an alarm will sound. ...
Remark
Published
04/01/2022 1:26 PM
3869 - 313 - Power off instrument; Remark:3850
Procedure Step
Published
03/29/2022 2:23 AM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
3850 - Press and hold the power button for approximately 10 seconds until indicator ...
Remark
Published
03/29/2022 1:38 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
10 minutes
Duration
Published
04/01/2020 11:35 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Daily
Frequency
Published
04/01/2020 11:36 AM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
packeye_ui_power_on_00719
Image
Published
02/07/2022 1:46 AM
PackEye art light highlighted
Image
Published
02/19/2022 1:48 AM
PackEye Nat light flashing
Image
Published
02/19/2022 1:48 AM
packeye_fail_solid_yellow.jpg
Image
Published
10/20/2023 10:36 AM
packeye_fail_flashing_yellow.jpg
Image
Published
10/20/2023 10:36 AM
Alarm mode
Image
Published
02/19/2022 1:48 AM
packeye_radsource_placeright.jpg
Image
Published
10/20/2023 10:46 AM
packeye_gamma_radsource_indicator.jpg
Image
Published
10/20/2023 10:39 AM
Packeye acknowledge gamma alarm
Image
Published
02/19/2022 1:48 AM
Chirper mode
Image
Published
02/19/2022 1:48 AM
packeye_gamma_radsource_chirp.jpg
Image
Published
10/20/2023 10:38 AM
packeye_neutronradsource_placeright.jpg
Image
Published
10/20/2023 10:45 AM
packeye_neutron_radsource_indicator.jpg
Image
Published
10/20/2023 10:44 AM
Acknowledge_neutron.png
Image
Published
02/19/2022 1:48 AM
packeye_neutron_radsource_chirp.jpg
Image
Published
10/20/2023 10:43 AM
Power off Packeye
Image
Published
02/19/2022 1:48 AM

Radiation Pager Operational Test

Last Updated: 04/25/2025 1:37 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech Radiation Pager
Device
Published
10/28/2024 3:52 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
2933 - 220 - Power on instrument; Remark:2921
Procedure Step
Published
06/22/2021 10:03 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
2921 - Set instrument to audio-only mode.
Remark
Published
06/15/2021 4:14 PM
2934 - 2922 - Verify instrument initializes successfully; Remark:2923
Procedure Step
Published
06/22/2021 10:03 AM
2922 - Verify instrument initializes successfully
Instruction
Published
06/15/2021 4:06 PM
2923 - Yellow light will illuminate first.
Remark
Published
06/15/2021 4:15 PM
2924 - Green light will illuminate briefly when pager is ready.
Remark
Published
06/15/2021 4:16 PM
2913 - 2911 - Test audio alarm; Remark:
Procedure Step
Published
06/22/2021 10:04 AM
2911 - Test audio alarm
Instruction
Published
06/15/2021 4:06 PM
2935 - 235 - Place instrument next to radioactive gamma source; Remark:2951
Procedure Step
Published
06/22/2021 10:06 AM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
2951 - IO RadPager Move To Source No Alarm Audible
Remark
Published
10/26/2024 9:16 AM
2939 - 2936 - Verify audible alarm functions correctly; Remark:2926
Procedure Step
Published
06/22/2021 10:07 AM
2936 - Verify audible alarm functions correctly
Instruction
Published
06/15/2021 4:09 PM
2926 - Alarm will be heard and the yellow light will illuminate in response to the s...
Remark
Published
07/06/2021 10:44 AM
2940 - 2937 - Press acknowledge button; Remark:2927
Procedure Step
Published
06/22/2021 10:14 AM
2937 - Press acknowledge button
Instruction
Published
06/15/2021 4:09 PM
2927 - Alarm level will display.
Remark
Published
07/09/2021 12:21 PM
2941 - 241 - Move instrument away from radioactive gamma source; Remark:2964
Procedure Step
Published
07/09/2021 2:18 PM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
2964 - IO RadPager Move Source Audible Mode
Remark
Published
10/26/2024 9:24 AM
2942 - 313 - Power off instrument; Remark:2929
Procedure Step
Published
06/22/2021 10:19 AM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
2929 - IO RadPager Power Off Button
Remark
Published
10/26/2024 10:45 AM
2914 - 2912 - Test vibration alarm; Remark:
Procedure Step
Published
06/06/2022 5:28 PM
2912 - Test vibration alarm
Instruction
Published
06/15/2021 4:11 PM
2943 - 220 - Power on instrument; Remark:2947
Procedure Step
Published
06/22/2021 10:20 AM
2947 - Set instrument to vibration-only mode.
Remark
Published
07/06/2021 10:45 AM
2944 - 235 - Place instrument next to radioactive gamma source; Remark:2925
Procedure Step
Published
06/22/2021 10:21 AM
2925 - IO RadPager Move To Source No Alarm
Remark
Published
10/26/2024 10:46 AM
2945 - 2938 - Verify vibration alarm functions correctly; Remark:2930
Procedure Step
Published
06/22/2021 10:22 AM
2938 - Verify vibration alarm functions correctly
Instruction
Published
06/15/2021 4:12 PM
2930 - Pager will vibrate and the yellow light will illuminate in response to the so...
Remark
Published
07/07/2021 9:54 AM
2946 - 2937 - Press acknowledge button; Remark:2931
Procedure Step
Published
06/22/2021 10:22 AM
2931 - Alarm level will display.
Remark
Published
07/06/2021 10:46 AM
2967 - 241 - Move instrument away from radioactive gamma source; Remark:2928
Procedure Step
Published
07/09/2021 2:17 PM
2928 - IO RadPager Move Away Source No Alarm Vib
Remark
Published
10/26/2024 10:50 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
2 minutes
Duration
Published
03/26/2021 3:33 PM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Daily
Frequency
Published
04/01/2020 11:36 AM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
RadPager Power On Audio On Mode
Image
Published
03/27/2021 1:42 AM
Radiation Pager evaluate alarm
Image
Published
03/27/2021 1:42 AM
Radiation Pager green light
Image
Published
03/27/2021 1:42 AM
Radiation Pager by source
Image
Published
03/27/2021 1:42 AM
Pager_Alarm_Noise_000581.png
Image
Published
03/27/2021 1:42 AM
Radiation Pager acknowledge button
Image
Published
03/27/2021 1:42 AM
Radiation Pager acknowledge alarm
Image
Published
03/27/2021 1:42 AM
Radiation Pager move away from source
Image
Published
03/27/2021 1:42 AM
power_off_000564.png
Image
Published
03/26/2021 1:40 AM
Pager_Alarm_Mode_000575.png
Image
Published
03/27/2021 1:42 AM
RadPager Vibrate Mode Toward Source
Image
Published
03/27/2021 1:42 AM
Radiation Pager vibration
Image
Published
03/27/2021 1:42 AM
Radiation Pager Away Source Vibrate Mode
Image
Published
03/27/2021 1:42 AM

AT6101C Operational Test Windows

Last Updated: 04/25/2025 1:36 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Atomtex AT6101C Backpack
Device
Published
06/27/2024 1:47 PM
1469 - 1401 - Establish Initial Conditions; Remark:
Procedure Step
Published
11/14/2020 4:42 PM
1468 - Note : No artificial radioactive sources should be present during the initial startu...
Annotation
Published
07/27/2022 1:16 PM
1401 - Establish Initial Conditions
Instruction
Published
09/30/2022 3:14 PM
1394 - 1393 - Verify no radioactive sources are near instrument; Remark:1392
Procedure Step
Published
11/24/2020 10:54 AM
1393 - Verify no radioactive sources are near instrument
Instruction
Published
05/25/2021 2:05 PM
1392 - Symbol - No radioactive material in immediate area
Remark
Published
09/01/2020 7:07 PM
794 - 793 - Place check sample into side pocket; Remark:791
Procedure Step
Published
05/24/2020 8:17 PM
793 - Place check sample into side pocket
Instruction
Published
10/26/2020 5:47 PM
791 - Atomtex_Place_check_sample_in_side_pocket
Remark
Published
11/08/2024 9:21 PM
1470 - 220 - Power on instrument; Remark:
Procedure Step
Published
11/14/2020 4:42 PM
1467 - Note : The BT-DU3 will power off after 10 minutes if not connected to the handheld c...
Annotation
Published
09/01/2020 7:11 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
1471 - 1377 - Power on BT-DU3 Adapter; Remark:1448
Procedure Step
Published
04/07/2025 12:55 PM
1377 - Power on BT-DU3 Adapter
Instruction
Published
09/30/2020 4:33 PM
1448 - Press and hold the PWR button until the yellow light is on a...
Remark
Published
09/01/2020 7:14 PM
1589 - 1495 - Power on mobile device; Remark:1569
Procedure Step
Published
04/07/2025 1:33 PM
1495 - Power on mobile device
Instruction
Published
11/12/2020 4:20 PM
1569 - Press the Power button and wait for the screen to activate. The ATAS Scanner ...
Remark
Published
11/14/2020 4:32 PM
1570 - If not ATAS does not automatically start, press F1.
Remark
Published
11/14/2020 4:33 PM
1590 - 1429 - Enter scanning file name; Remark:1571
Procedure Step
Published
07/28/2020 12:41 PM
1429 - Enter scanning file name
Instruction
Published
09/01/2020 7:26 PM
1571 - IO NautizX8 ATAS Scanning file name
Remark
Published
11/14/2020 6:35 PM
1591 - 1565 - If prompted, tap Yes to perform stabilization; Remark:1572
Procedure Step
Published
11/14/2020 4:10 PM
1466 - Note : It is recommended to perform a calibration upon startup, at least once a week...
Annotation
Published
09/01/2020 7:24 PM
1565 - If prompted, tap Yes to perform stabilization
Instruction
Published
11/14/2020 3:04 PM
1572 - Stabilization typically takes about 3 minutes.
Remark
Published
11/14/2020 3:43 PM
1592 - 612 - Wait for instrument to complete start-up sequence; Remark:1573
Procedure Step
Published
04/13/2025 6:48 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
1573 - The start-up sequence may take several minutes to complete.
Remark
Published
11/14/2020 6:32 PM
1477 - 1431 - Verify system is ready; Remark:
Procedure Step
Published
09/01/2020 7:31 PM
1431 - Verify system is ready
Instruction
Published
09/01/2020 7:31 PM
1593 - 1566 - Verify backpack connection to handheld computer; Remark:1574
Procedure Step
Published
11/14/2020 4:11 PM
1566 - Verify backpack connection to handheld computer
Instruction
Published
11/14/2020 3:05 PM
1574 - SCANNING should appear in the status bar. Gamma and neutron ...
Remark
Published
11/14/2020 3:41 PM
1594 - 1433 - Verify GPS is connected; Remark:1575
Procedure Step
Published
07/28/2020 12:45 PM
1433 - Verify GPS is connected
Instruction
Published
09/01/2020 7:36 PM
1575 - The handheld computer may need to be outside for GPS to function correctly.
Remark
Published
11/14/2020 6:30 PM
1595 - 1434 - Verify the correct background mode is selected; Remark:1576
Procedure Step
Published
07/28/2020 12:46 PM
1434 - Verify the correct background mode is selected
Instruction
Published
09/01/2020 7:39 PM
1576 - Permanent background is used in most situations.
Remark
Published
11/14/2020 6:47 PM
1596 - 1435 - Check BT-DU3 battery; Remark:1577
Procedure Step
Published
07/28/2020 12:47 PM
1435 - Check BT-DU3 battery
Instruction
Published
09/01/2020 7:41 PM
1577 - Momentarily press the POWER button on the BT-DU3 to view the battery charge; ...
Remark
Published
11/14/2020 6:16 PM
1597 - 1567 - Check handheld computer battery; Remark:1578
Procedure Step
Published
11/14/2020 4:11 PM
1567 - Check handheld computer battery
Instruction
Published
11/14/2020 3:35 PM
1578 - IO NautizX8 ATAS Battery Level of PDA
Remark
Published
11/14/2020 4:08 PM
1598 - 1568 - Verify Handheld is in Function mode; Remark:1579
Procedure Step
Published
11/14/2020 4:12 PM
1568 - Verify Handheld is in Function mode
Instruction
Published
11/14/2020 3:36 PM
1579 - Fn will appear at the top of the screen.
Remark
Published
11/14/2020 4:10 PM
1580 - Press the blue button to change between function “Fn” and number “1” ...
Remark
Published
11/14/2020 3:39 PM
1483 - 1437 - Verify Response to Gamma Source; Remark:
Procedure Step
Published
09/01/2020 7:45 PM
1437 - Verify Response to Gamma Source
Instruction
Published
09/01/2020 7:45 PM
1484 - 1438 - Place gamma source next to backpack; Remark:1463
Procedure Step
Published
09/01/2020 7:47 PM
1438 - Place gamma source next to backpack
Instruction
Published
09/01/2020 7:46 PM
1463 - IO AT6101C Trefoil Move Toward Backpack
Remark
Published
09/02/2020 3:11 PM
1599 - 1439 - Verify instrument responds to radioactive source; Remark:1581
Procedure Step
Published
07/28/2020 12:50 PM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
1581 - If radioactive source activity is of high enough to trigger an alarm, ...
Remark
Published
11/14/2020 5:50 PM
1600 - 1440 - Perform identification; Remark:1582
Procedure Step
Published
07/28/2020 12:51 PM
1440 - Perform identification
Instruction
Published
09/01/2020 8:16 PM
1582 - The identification mode may start automatically after the gamma alarm. If not...
Remark
Published
11/14/2020 5:41 PM
1583 - Alternately, Tap Menu -> Scanner -> Identification to ...
Remark
Published
11/14/2020 5:40 PM
1601 - 2220 - Wait for measurement to complete; Remark:1584
Procedure Step
Published
04/20/2025 12:54 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
1584 - IO NautizX8 ATAS Identifying 30 Wait
Remark
Published
11/14/2020 5:38 PM
1602 - 1442 - Verify identification results are correct; Remark:1585
Procedure Step
Published
07/28/2020 12:53 PM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
1585 - Identification results should match the isotopes associated with the radioact...
Remark
Published
11/14/2020 5:36 PM
1603 - 1443 - Move instrument away from gamma source of radiation when status changes to; Remark:1586
Procedure Step
Published
07/28/2020 12:54 PM
1443 - Move instrument away from gamma source of radiation when status changes to
Instruction
Published
11/14/2020 6:50 PM
1586 - IO NautizX8 ATAS Wait 15
Remark
Published
11/14/2020 5:29 PM
246 - IO No source trefoil
Remark
Published
10/07/2024 11:39 AM
1604 - 1444 - Wait for Scanning mode to resume; Remark:1587
Procedure Step
Published
07/28/2020 12:55 PM
1444 - Wait for Scanning mode to resume
Instruction
Published
09/02/2020 3:08 PM
1587 - IO NautizX8 ATAS Scanning Mode .048
Remark
Published
11/14/2020 5:27 PM
1491 - 1445 - Verify Response to Neutron Source; Remark:
Procedure Step
Published
09/02/2020 3:09 PM
1445 - Verify Response to Neutron Source
Instruction
Published
09/02/2020 3:09 PM
1492 - 1446 - Place neutron source next to backpack; Remark:1463
Procedure Step
Published
09/02/2020 3:11 PM
1446 - Place neutron source next to backpack
Instruction
Published
09/02/2020 3:11 PM
1605 - 1439 - Verify instrument responds to radioactive source; Remark:1588
Procedure Step
Published
11/14/2020 5:21 PM
1588 - If radioactive source activity is of high enough to trigger an alarm, ...
Remark
Published
11/14/2020 5:20 PM
1494 - 1447 - Move instrument away from neutron source; Remark:246
Procedure Step
Published
09/02/2020 3:19 PM
1447 - Move instrument away from neutron source
Instruction
Published
09/02/2020 3:18 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
10 minutes
Duration
Published
04/01/2020 11:35 AM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
Atomtex_Backpack_Side_Pocket_Red_Arrow_0.png
Image
Published
09/04/2020 10:44 PM
AT6101C_Partial_pocket_side_KCL Potassium check sample_Puck_red_arrow_1.png
Image
Published
09/04/2020 10:26 PM
AT6101C Full BTDU3 Location Open Front Pocket
Image
Published
09/04/2020 10:40 PM
AT6101C Partial BTDU3 Face Power Button On LED
Image
Published
09/04/2020 10:39 PM
NautizX8 Power Button on Keypad
Image
Published
09/04/2020 10:26 PM
ATAS Scanner NautizX8 Screen on start up
Image
Published
09/04/2020 10:26 PM
NautizX8_keypad_F1
Image
Published
09/04/2020 10:29 PM
NautizX8_ATAS_Win_File_Name Enter
Image
Published
03/30/2022 2:22 AM
NautizX8_ATAS_Scanner_Win_Stabilization_Request
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_screen_background_timer_red_highlight.png
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_screen_scanning_51nsv_red_highlight.png
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_GPS_screen_display.png
Image
Published
03/30/2022 2:22 AM
NautizX8_ATAS_Win_Permanent_Background_Setting
Image
Published
03/30/2022 2:22 AM
AT6101C BTDU3 Checking Battery Level Red Arrow
Image
Published
09/04/2020 10:40 PM
NautizX8_ATAS_Scanner_Battery_Level_of_BTDU_Segments
Image
Published
03/30/2022 2:22 AM
NautizX8_ATAS_Scanner_Battery_Level_of_PDA_Segments
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_screen_FN_indication.png
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_Keypad_blue_bar
Image
Published
09/04/2020 10:29 PM
AT6101C Full Move Source Toward Backpack
Image
Published
09/04/2020 10:40 PM
NautizX8_AT_screen_alarm_gamma_red_highlight.png
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_keypad_2_button
Image
Published
09/04/2020 10:34 PM
NautizX8_AT_Screen_Time_Remaining.png
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_screen_isotope_list.png
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_Screen_Waiting_15_Sec.png
Image
Published
03/30/2022 2:22 AM
NautizX8_AT_screen_scanning_mode_red_highlight.png
Image
Published
03/30/2022 2:22 AM
NautizX8 Neutron Alarm Indication Screen
Image
Published
03/30/2022 2:22 AM
NautizX8 Neutron Alarm and Gamma Alarm indications
Image
Published
03/30/2022 2:22 AM

AT6101C Operational Test Android

Last Updated: 04/25/2025 1:36 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Atomtex AT6101C Backpack
Device
Published
06/27/2024 1:47 PM
1469 - 1401 - Establish Initial Conditions; Remark:
Procedure Step
Published
11/14/2020 4:42 PM
1468 - Note : No artificial radioactive sources should be present during the initial startu...
Annotation
Published
07/27/2022 1:16 PM
1401 - Establish Initial Conditions
Instruction
Published
09/30/2022 3:14 PM
1394 - 1393 - Verify no radioactive sources are near instrument; Remark:1392
Procedure Step
Published
11/24/2020 10:54 AM
1393 - Verify no radioactive sources are near instrument
Instruction
Published
05/25/2021 2:05 PM
1392 - Symbol - No radioactive material in immediate area
Remark
Published
09/01/2020 7:07 PM
794 - 793 - Place check sample into side pocket; Remark:791
Procedure Step
Published
05/24/2020 8:17 PM
793 - Place check sample into side pocket
Instruction
Published
10/26/2020 5:47 PM
791 - Atomtex_Place_check_sample_in_side_pocket
Remark
Published
11/08/2024 9:21 PM
1470 - 220 - Power on instrument; Remark:
Procedure Step
Published
11/14/2020 4:42 PM
1467 - Note : The BT-DU3 will power off after 10 minutes if not connected to the handheld c...
Annotation
Published
09/01/2020 7:11 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
1471 - 1377 - Power on BT-DU3 Adapter; Remark:1448
Procedure Step
Published
04/07/2025 12:55 PM
1377 - Power on BT-DU3 Adapter
Instruction
Published
09/30/2020 4:33 PM
1448 - Press and hold the PWR button until the yellow light is on a...
Remark
Published
09/01/2020 7:14 PM
1473 - 1495 - Power on mobile device; Remark:
Procedure Step
Published
04/07/2025 1:34 PM
1495 - Power on mobile device
Instruction
Published
11/12/2020 4:20 PM
1472 - 1427 - Start Atomtex App; Remark:1461
Procedure Step
Published
06/29/2022 11:47 PM
1427 - Start Atomtex App
Instruction
Published
06/29/2022 11:44 PM
1461 - IO RSA App Icon for AT6101C
Remark
Published
09/17/2020 10:05 AM
1474 - 1428 - If prompted, tap OK to perform calibration; Remark:1449
Procedure Step
Published
09/01/2020 7:24 PM
1466 - Note : It is recommended to perform a calibration upon startup, at least once a week...
Annotation
Published
09/01/2020 7:24 PM
1428 - If prompted, tap OK to perform calibration
Instruction
Published
09/01/2020 7:23 PM
1449 - Calibration typically takes about 3 minutes.
Remark
Published
09/01/2020 7:22 PM
1475 - 1429 - Enter scanning file name; Remark:1462
Procedure Step
Published
09/01/2020 7:26 PM
1429 - Enter scanning file name
Instruction
Published
09/01/2020 7:26 PM
1462 - IO RSA Scanning File Name Keypad
Remark
Published
11/08/2024 9:24 PM
1476 - 612 - Wait for instrument to complete start-up sequence; Remark:1450
Procedure Step
Published
04/13/2025 7:30 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
1450 - The start-up sequence may take several minutes to complete.
Remark
Published
09/01/2020 7:28 PM
1477 - 1431 - Verify system is ready; Remark:
Procedure Step
Published
09/01/2020 7:31 PM
1431 - Verify system is ready
Instruction
Published
09/01/2020 7:31 PM
1478 - 1432 - Verify backpack connection to mobile device; Remark:1451
Procedure Step
Published
09/01/2020 7:34 PM
1432 - Verify backpack connection to mobile device
Instruction
Published
09/01/2020 7:34 PM
1451 - Scanning should appear in the status bar. Gamma and neutron ...
Remark
Published
09/01/2020 7:33 PM
1479 - 1433 - Verify GPS is connected; Remark:1452
Procedure Step
Published
09/01/2020 7:36 PM
1433 - Verify GPS is connected
Instruction
Published
09/01/2020 7:36 PM
1452 - The mobile device may need to be outside for GPS to function correctly.
Remark
Published
07/29/2021 4:18 PM
1480 - 1434 - Verify the correct background mode is selected; Remark:1453
Procedure Step
Published
09/01/2020 7:39 PM
1434 - Verify the correct background mode is selected
Instruction
Published
09/01/2020 7:39 PM
1453 - Permanent background is used in most situations.
Remark
Published
09/01/2020 7:38 PM
1481 - 1435 - Check BT-DU3 battery; Remark:1454
Procedure Step
Published
09/01/2020 7:41 PM
1435 - Check BT-DU3 battery
Instruction
Published
09/01/2020 7:41 PM
1454 - Momentarily press the POWER button on the BT-DU3 to view the battery charge; ...
Remark
Published
09/01/2020 7:41 PM
1482 - 1436 - Check mobile device battery; Remark:1455
Procedure Step
Published
09/01/2020 7:43 PM
1436 - Check mobile device battery
Instruction
Published
09/01/2020 7:43 PM
1455 - The Atomtex App also displays the mobile device battery life.
Remark
Published
09/01/2020 7:43 PM
1483 - 1437 - Verify Response to Gamma Source; Remark:
Procedure Step
Published
09/01/2020 7:45 PM
1437 - Verify Response to Gamma Source
Instruction
Published
09/01/2020 7:45 PM
1484 - 1438 - Place gamma source next to backpack; Remark:1463
Procedure Step
Published
09/01/2020 7:47 PM
1438 - Place gamma source next to backpack
Instruction
Published
09/01/2020 7:46 PM
1463 - IO AT6101C Trefoil Move Toward Backpack
Remark
Published
09/02/2020 3:11 PM
1485 - 1439 - Verify instrument responds to radioactive source; Remark:1456
Procedure Step
Published
06/29/2022 11:50 PM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
1456 - If radioactive source activity is high enough to trigger an alarm,
Remark
Published
09/22/2020 10:12 AM
1486 - 1440 - Perform identification; Remark:1457
Procedure Step
Published
09/01/2020 8:16 PM
1440 - Perform identification
Instruction
Published
09/01/2020 8:16 PM
1457 - Tap Start Identification.
Remark
Published
09/11/2020 11:55 PM
1487 - 2220 - Wait for measurement to complete; Remark:1989
Procedure Step
Published
04/20/2025 12:54 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
1989 - Identification time might automatically be extended for low activity sources....
Remark
Published
11/16/2020 6:58 PM
1488 - 1442 - Verify identification results are correct; Remark:1459
Procedure Step
Published
09/02/2020 3:04 PM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
1459 - Identification results should match the isotopes associated with the radioact...
Remark
Published
09/02/2020 3:03 PM
1489 - 1443 - Move instrument away from gamma source of radiation when status changes to; Remark:1464
Procedure Step
Published
09/02/2020 3:06 PM
1443 - Move instrument away from gamma source of radiation when status changes to
Instruction
Published
11/14/2020 6:50 PM
1464 - IO RSA App Waiting No Trefoil
Remark
Published
11/16/2020 7:07 PM
246 - IO No source trefoil
Remark
Published
10/07/2024 11:39 AM
1490 - 1444 - Wait for Scanning mode to resume; Remark:1465
Procedure Step
Published
09/02/2020 3:08 PM
1444 - Wait for Scanning mode to resume
Instruction
Published
09/02/2020 3:08 PM
1465 - IO RSA App Scanning No Alarm
Remark
Published
11/12/2020 5:32 PM
1491 - 1445 - Verify Response to Neutron Source; Remark:
Procedure Step
Published
09/02/2020 3:09 PM
1445 - Verify Response to Neutron Source
Instruction
Published
09/02/2020 3:09 PM
1492 - 1446 - Place neutron source next to backpack; Remark:1463
Procedure Step
Published
09/02/2020 3:11 PM
1446 - Place neutron source next to backpack
Instruction
Published
09/02/2020 3:11 PM
1493 - 1439 - Verify instrument responds to radioactive source; Remark:1460
Procedure Step
Published
09/02/2020 3:17 PM
1460 - If radioactive source activity is of high enough to trigger an alarm, ...
Remark
Published
09/02/2020 3:15 PM
1494 - 1447 - Move instrument away from neutron source; Remark:246
Procedure Step
Published
09/02/2020 3:19 PM
1447 - Move instrument away from neutron source
Instruction
Published
09/02/2020 3:18 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
10 minutes
Duration
Published
04/01/2020 11:35 AM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
Atomtex_Backpack_Side_Pocket_Red_Arrow_0.png
Image
Published
09/04/2020 10:44 PM
AT6101C_Partial_pocket_side_KCL Potassium check sample_Puck_red_arrow_1.png
Image
Published
09/04/2020 10:26 PM
AT6101C Full BTDU3 Location Open Front Pocket
Image
Published
09/04/2020 10:40 PM
AT6101C Partial BTDU3 Face Power Button On LED
Image
Published
09/04/2020 10:39 PM
AT6101C_RSA_Icon04.png
Image
Published
09/04/2020 4:11 PM
AT6101C RSA Screen Request Backpack Stabilization
Image
Published
03/30/2022 2:22 AM
AT6101C RSA Screen Scanning File Name Enter
Image
Published
03/30/2022 2:22 AM
ATAS_Collecting_Background.png
Image
Published
03/30/2022 2:22 AM
ATAS_Scanning.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Scanning_GPS_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C RSA Permanent Background Mode
Image
Published
03/30/2022 2:22 AM
AT6101C BTDU3 Checking Battery Level Red Arrow
Image
Published
09/04/2020 10:40 PM
AT6101C RSA Battery Level of BTDU3 Segments
Image
Published
03/30/2022 2:22 AM
AT6101C RSA Battery Level of mobile device Segments
Image
Published
03/30/2022 2:22 AM
AT6101C Full Move Source Toward Backpack
Image
Published
09/04/2020 10:40 PM
AT6101C_RSA_Scanning_Alarm_Gamma311_Graph_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Scanning_StartIdentify_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Identifying_Alarm_Gamma352_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Identifying_Results_K40_Th_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Waiting5_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Scanning_NoAlarm_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_Android_English_Neutron_Alarm_Blue-LED_Red-Arrow.png
Image
Published
03/30/2022 2:22 AM
AT6101C_Android_English_Gamma-and-Neutron_Alarm_Blue-Red-LED_Red_Arrow
Image
Published
03/30/2022 2:22 AM

PRM-470 Operational Test

Last Updated: 04/25/2025 1:35 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Rapiscan PRM-470 CG
Device
Published
10/28/2024 3:40 PM
Rapiscan PRM-470 CGN
Device
Published
10/28/2024 3:33 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
3092 - 3091 - Disconnect from external power; Remark:
Procedure Step
Published
07/29/2021 11:05 AM
3091 - Disconnect from external power
Instruction
Published
07/23/2021 10:01 AM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
250 - 220 - Power on instrument; Remark:228
Procedure Step
Published
07/29/2021 11:06 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
228 - Press POWER button until the display turns on.
Remark
Published
06/18/2021 6:29 PM
3107 - 229 - Press LIGHT button to brighten display (if needed); Remark:3106
Procedure Step
Published
07/29/2021 11:08 AM
229 - Press LIGHT button to brighten display (if needed)
Instruction
Published
07/29/2021 11:08 AM
3106 - Light may provide better visibility for reading the display...
Remark
Published
07/23/2021 9:59 AM
231 - IO PRM-470 Light Button Red Box
Remark
Published
11/04/2024 9:49 PM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
333 - 233 - Verify instrument is in SEARCH MODE; Remark:234
Procedure Step
Published
06/25/2021 12:41 PM
233 - Verify instrument is in SEARCH MODE
Instruction
Published
06/10/2021 8:46 PM
234 - Use S/F button to change between SEARCH MODE and FIND MODE.
Remark
Published
07/29/2021 11:11 AM
751 - IO PRM-470 Search Mode 65 CPS Gamma
Remark
Published
10/29/2024 9:31 PM
3108 - S/F mode key red box
Remark
Published
11/02/2024 10:09 PM
334 - 235 - Place instrument next to radioactive gamma source; Remark:535
Procedure Step
Published
07/20/2021 11:10 AM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
535 - IO Move PRM-470 approaching source in SEARCH mode
Remark
Published
11/05/2024 9:35 PM
335 - 237 - Verify SEARCH MODE responds correctly; Remark:243
Procedure Step
Published
07/29/2021 4:14 PM
237 - Verify SEARCH MODE responds correctly
Instruction
Published
07/29/2021 3:47 PM
243 - The instrument should respond as follows: Gamma counts per ...
Remark
Published
10/06/2022 12:31 PM
340 - 241 - Move instrument away from radioactive gamma source; Remark:332
Procedure Step
Published
07/20/2021 11:11 AM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
332 - IO Move PRM-470 away from source Alarm SEARCH
Remark
Published
11/05/2024 9:36 PM
342 - 239 - Set instrument to FIND MODE; Remark:234
Procedure Step
Published
07/20/2021 11:12 AM
239 - Set instrument to FIND MODE
Instruction
Published
07/16/2021 10:43 AM
3109 - Find Mode
Remark
Published
11/05/2024 9:39 PM
336 - 235 - Place instrument next to radioactive gamma source; Remark:536
Procedure Step
Published
07/20/2021 11:51 AM
536 - IO Move PRM-470 approaching source in FIND mode
Remark
Published
11/05/2024 9:38 PM
337 - 240 - Verify FIND MODE responds correctly; Remark:245
Procedure Step
Published
07/20/2021 11:51 AM
240 - Verify FIND MODE responds correctly
Instruction
Published
07/16/2021 10:44 AM
245 - The instrument should respond as follows: Gamma counts per ...
Remark
Published
10/06/2022 12:32 PM
338 - 241 - Move instrument away from radioactive gamma source; Remark:339
Procedure Step
Published
07/20/2021 11:53 AM
339 - IO Move PRM-470 away source Alarm FIND
Remark
Published
11/05/2024 9:40 PM
3099 - 3093 - Check instrument neutron response (CGN model only); Remark:
Procedure Step
Published
07/20/2021 11:53 AM
3094 - Note : The next step is only applicable to CGN models with neutron detection capability...
Annotation
Published
10/04/2022 11:14 AM
3093 - Check instrument neutron response (CGN model only)
Instruction
Published
07/16/2021 10:47 AM
3100 - 3095 - Place instrument next to radioactive neutron source; Remark:536
Procedure Step
Published
07/20/2021 11:54 AM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
3102 - 3097 - Verify instrument responds correctly; Remark:3103
Procedure Step
Published
07/20/2021 11:54 AM
3096 - Note : Most radioactive neutron sources also have a gamma component, which will likely ...
Annotation
Published
07/23/2021 9:57 AM
3097 - Verify instrument responds correctly
Instruction
Published
07/16/2021 10:50 AM
3103 - The instrument should respond as follows: Neutron CPS chang...
Remark
Published
10/06/2022 12:33 PM
3101 - PRM-470 Search Mode Neutron
Remark
Published
11/05/2024 9:40 PM
3104 - 3098 - Move instrument away from radioactive neutron source; Remark:339
Procedure Step
Published
07/20/2021 11:54 AM
3098 - Move instrument away from radioactive neutron source
Instruction
Published
07/16/2021 10:54 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
5 minutes
Duration
Published
04/01/2020 11:35 AM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
PRM-470_Partial_Power_Button_Red_Highlight.png
Image
Published
06/18/2021 6:28 PM
PRM-470_Partial-View_Light-Button_RED-Highlight.png
Image
Published
05/05/2020 9:17 PM
PRM-470 Display SEARCH
Image
Published
04/01/2021 11:22 PM
PRM-470_Partial-View_S-F-Button_RED-Highlight.png
Image
Published
04/01/2021 11:22 PM
PRM-470 Full Approaching Source No Alarm SEARCH Red Arrow
Image
Published
04/03/2021 7:06 PM
PRM-470 Display Gamma Alarm SEARCH Red Highlight
Image
Published
05/07/2020 4:22 PM
prm-470_di_galarm_SEARCH_00055.png
Image
Published
08/01/2021 2:40 AM
PRM-470 Move Away from Source Alarm SEARCH Red Arrow
Image
Published
05/13/2020 1:38 PM
PRM-470 Display FIND
Image
Published
05/07/2020 5:17 PM
PRM-470 Full Approaching Source No Alarm FIND Red Arrow
Image
Published
04/03/2021 7:06 PM
prm-470_di_galarm_FIND_00058.png
Image
Published
08/01/2021 2:40 AM
PRM-470 Display Gamma Alarm FIND Red Highlight
Image
Published
05/07/2020 5:19 PM
PRM-470_Full_Away_GAlarm_Red_Arrow FIND
Image
Published
05/13/2020 1:35 PM
PRM-470 Alarm Reaction
Image
Published
04/03/2021 1:42 AM
Search Mode Neutron
Image
Published
04/03/2021 1:42 AM