Last Updated: 04/26/2025 1:36 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Atomtex AT6101C Backpack
Device
Published
06/27/2024 1:47 PM
1469 - 1401 - Establish Initial Conditions; Remark:
Procedure Step
Published
11/14/2020 4:42 PM
1468 - Note : No artificial radioactive sources should be present during the initial startu...
Annotation
Published
07/27/2022 1:16 PM
1401 - Establish Initial Conditions
Instruction
Published
09/30/2022 3:14 PM
1394 - 1393 - Verify no radioactive sources are near instrument; Remark:1392
Procedure Step
Published
11/24/2020 10:54 AM
1393 - Verify no radioactive sources are near instrument
Instruction
Published
05/25/2021 2:05 PM
1392 - Symbol - No radioactive material in immediate area
Remark
Published
09/01/2020 7:07 PM
794 - 793 - Place check sample into side pocket; Remark:791
Procedure Step
Published
05/24/2020 8:17 PM
793 - Place check sample into side pocket
Instruction
Published
10/26/2020 5:47 PM
791 - Atomtex_Place_check_sample_in_side_pocket
Remark
Published
11/08/2024 9:21 PM
1470 - 220 - Power on instrument; Remark:
Procedure Step
Published
11/14/2020 4:42 PM
1467 - Note : The BT-DU3 will power off after 10 minutes if not connected to the handheld c...
Annotation
Published
09/01/2020 7:11 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
1471 - 1377 - Power on BT-DU3 Adapter; Remark:1448
Procedure Step
Published
04/07/2025 12:55 PM
1377 - Power on BT-DU3 Adapter
Instruction
Published
09/30/2020 4:33 PM
1448 - Press and hold the PWR button until the yellow light is on a...
Remark
Published
09/01/2020 7:14 PM
1473 - 1495 - Power on mobile device; Remark:
Procedure Step
Published
04/07/2025 1:34 PM
1495 - Power on mobile device
Instruction
Published
11/12/2020 4:20 PM
1472 - 1427 - Start Atomtex App; Remark:1461
Procedure Step
Published
06/29/2022 11:47 PM
1427 - Start Atomtex App
Instruction
Published
06/29/2022 11:44 PM
1461 - IO RSA App Icon for AT6101C
Remark
Published
09/17/2020 10:05 AM
1474 - 1428 - If prompted, tap OK to perform calibration; Remark:1449
Procedure Step
Published
09/01/2020 7:24 PM
1466 - Note : It is recommended to perform a calibration upon startup, at least once a week...
Annotation
Published
09/01/2020 7:24 PM
1428 - If prompted, tap OK to perform calibration
Instruction
Published
09/01/2020 7:23 PM
1449 - Calibration typically takes about 3 minutes.
Remark
Published
09/01/2020 7:22 PM
1475 - 1429 - Enter scanning file name; Remark:1462
Procedure Step
Published
09/01/2020 7:26 PM
1429 - Enter scanning file name
Instruction
Published
09/01/2020 7:26 PM
1462 - IO RSA Scanning File Name Keypad
Remark
Published
11/08/2024 9:24 PM
1476 - 612 - Wait for instrument to complete start-up sequence; Remark:1450
Procedure Step
Published
04/13/2025 7:30 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
1450 - The start-up sequence may take several minutes to complete.
Remark
Published
09/01/2020 7:28 PM
1477 - 1431 - Verify system is ready; Remark:
Procedure Step
Published
09/01/2020 7:31 PM
1431 - Verify system is ready
Instruction
Published
09/01/2020 7:31 PM
1478 - 1432 - Verify backpack connection to mobile device; Remark:1451
Procedure Step
Published
09/01/2020 7:34 PM
1432 - Verify backpack connection to mobile device
Instruction
Published
09/01/2020 7:34 PM
1451 - Scanning should appear in the status bar. Gamma and neutron ...
Remark
Published
09/01/2020 7:33 PM
1479 - 1433 - Verify GPS is connected; Remark:1452
Procedure Step
Published
09/01/2020 7:36 PM
1433 - Verify GPS is connected
Instruction
Published
09/01/2020 7:36 PM
1452 - The mobile device may need to be outside for GPS to function correctly.
Remark
Published
07/29/2021 4:18 PM
1480 - 1434 - Verify the correct background mode is selected; Remark:1453
Procedure Step
Published
09/01/2020 7:39 PM
1434 - Verify the correct background mode is selected
Instruction
Published
09/01/2020 7:39 PM
1453 - Permanent background is used in most situations.
Remark
Published
09/01/2020 7:38 PM
1481 - 1435 - Check BT-DU3 battery; Remark:1454
Procedure Step
Published
09/01/2020 7:41 PM
1435 - Check BT-DU3 battery
Instruction
Published
09/01/2020 7:41 PM
1454 - Momentarily press the POWER button on the BT-DU3 to view the battery charge; ...
Remark
Published
09/01/2020 7:41 PM
1482 - 1436 - Check mobile device battery; Remark:1455
Procedure Step
Published
09/01/2020 7:43 PM
1436 - Check mobile device battery
Instruction
Published
09/01/2020 7:43 PM
1455 - The Atomtex App also displays the mobile device battery life.
Remark
Published
09/01/2020 7:43 PM
1483 - 1437 - Verify Response to Gamma Source; Remark:
Procedure Step
Published
09/01/2020 7:45 PM
1437 - Verify Response to Gamma Source
Instruction
Published
09/01/2020 7:45 PM
1484 - 1438 - Place gamma source next to backpack; Remark:1463
Procedure Step
Published
09/01/2020 7:47 PM
1438 - Place gamma source next to backpack
Instruction
Published
09/01/2020 7:46 PM
1463 - IO AT6101C Trefoil Move Toward Backpack
Remark
Published
09/02/2020 3:11 PM
1485 - 1439 - Verify instrument responds to radioactive source; Remark:1456
Procedure Step
Published
06/29/2022 11:50 PM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
1456 - If radioactive source activity is high enough to trigger an alarm,
Remark
Published
09/22/2020 10:12 AM
1486 - 1440 - Perform identification; Remark:1457
Procedure Step
Published
09/01/2020 8:16 PM
1440 - Perform identification
Instruction
Published
09/01/2020 8:16 PM
1457 - Tap Start Identification.
Remark
Published
09/11/2020 11:55 PM
1487 - 2220 - Wait for measurement to complete; Remark:1989
Procedure Step
Published
04/20/2025 12:54 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
1989 - Identification time might automatically be extended for low activity sources....
Remark
Published
11/16/2020 6:58 PM
1488 - 1442 - Verify identification results are correct; Remark:1459
Procedure Step
Published
09/02/2020 3:04 PM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
1459 - Identification results should match the isotopes associated with the radioact...
Remark
Published
09/02/2020 3:03 PM
1489 - 1443 - Move instrument away from gamma source of radiation when status changes to; Remark:1464
Procedure Step
Published
09/02/2020 3:06 PM
1443 - Move instrument away from gamma source of radiation when status changes to
Instruction
Published
11/14/2020 6:50 PM
1464 - IO RSA App Waiting No Trefoil
Remark
Published
11/16/2020 7:07 PM
246 - IO No source trefoil
Remark
Published
10/07/2024 11:39 AM
1490 - 1444 - Wait for Scanning mode to resume; Remark:1465
Procedure Step
Published
09/02/2020 3:08 PM
1444 - Wait for Scanning mode to resume
Instruction
Published
09/02/2020 3:08 PM
1465 - IO RSA App Scanning No Alarm
Remark
Published
11/12/2020 5:32 PM
1491 - 1445 - Verify Response to Neutron Source; Remark:
Procedure Step
Published
09/02/2020 3:09 PM
1445 - Verify Response to Neutron Source
Instruction
Published
09/02/2020 3:09 PM
1492 - 1446 - Place neutron source next to backpack; Remark:1463
Procedure Step
Published
09/02/2020 3:11 PM
1446 - Place neutron source next to backpack
Instruction
Published
09/02/2020 3:11 PM
1493 - 1439 - Verify instrument responds to radioactive source; Remark:1460
Procedure Step
Published
09/02/2020 3:17 PM
1460 - If radioactive source activity is of high enough to trigger an alarm, ...
Remark
Published
09/02/2020 3:15 PM
1494 - 1447 - Move instrument away from neutron source; Remark:246
Procedure Step
Published
09/02/2020 3:19 PM
1447 - Move instrument away from neutron source
Instruction
Published
09/02/2020 3:18 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
10 minutes
Duration
Published
04/01/2020 11:35 AM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
Atomtex_Backpack_Side_Pocket_Red_Arrow_0.png
Image
Published
09/04/2020 10:44 PM
AT6101C_Partial_pocket_side_KCL Potassium check sample_Puck_red_arrow_1.png
Image
Published
09/04/2020 10:26 PM
AT6101C Full BTDU3 Location Open Front Pocket
Image
Published
09/04/2020 10:40 PM
AT6101C Partial BTDU3 Face Power Button On LED
Image
Published
09/04/2020 10:39 PM
AT6101C_RSA_Icon04.png
Image
Published
09/04/2020 4:11 PM
AT6101C RSA Screen Request Backpack Stabilization
Image
Published
03/30/2022 2:22 AM
AT6101C RSA Screen Scanning File Name Enter
Image
Published
03/30/2022 2:22 AM
ATAS_Collecting_Background.png
Image
Published
03/30/2022 2:22 AM
ATAS_Scanning.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Scanning_GPS_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C RSA Permanent Background Mode
Image
Published
03/30/2022 2:22 AM
AT6101C BTDU3 Checking Battery Level Red Arrow
Image
Published
09/04/2020 10:40 PM
AT6101C RSA Battery Level of BTDU3 Segments
Image
Published
03/30/2022 2:22 AM
AT6101C RSA Battery Level of mobile device Segments
Image
Published
03/30/2022 2:22 AM
AT6101C Full Move Source Toward Backpack
Image
Published
09/04/2020 10:40 PM
AT6101C_RSA_Scanning_Alarm_Gamma311_Graph_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Scanning_StartIdentify_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Identifying_Alarm_Gamma352_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Identifying_Results_K40_Th_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Waiting5_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_RSA_Scanning_NoAlarm_Red_Box.png
Image
Published
03/30/2022 2:22 AM
AT6101C_Android_English_Neutron_Alarm_Blue-LED_Red-Arrow.png
Image
Published
03/30/2022 2:22 AM
AT6101C_Android_English_Gamma-and-Neutron_Alarm_Blue-Red-LED_Red_Arrow
Image
Published
03/30/2022 2:22 AM
DET-HHD-ATM-RM01

Conduct operational test of Atomtex AT6101C Backpack. This procedure is only applicable to the Android Mobile version of the Atomtex application (App).

Equipment Hierarchy
Atomtex AT6101C Backpack
Frequency
Daily
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
10 minutes
Tools Required
None
Materials Required
Atomtex check sample
Radioactive gamma source of Cs-137 activity 37-370 kBq (1-10 μCi), or equivalent
Radioactive neutron source of Cf-252 activity 37-185 kBq (1-5 μCi), or equivalent
Originators
Tucker, Brian
Originating Organization
PNNL
Additional Retest Requirements Information

None

Departmental Coordination

As necessary

Date of Origin
Version
0.00
Step 2
Main Step

1474 - 1428 - If prompted, tap OK to perform calibration; Remark:1449

Step 00
Step 3
Step 4
Step 5

1493 - 1439 - Verify instrument responds to radioactive source; Remark:1460

Step 00
Percent Complete
100.00
Percent Complete W/Media
100.00