Last Updated: 04/25/2025 1:37 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech Radiation Pager
Device
Published
10/28/2024 3:52 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
2933 - 220 - Power on instrument; Remark:2921
Procedure Step
Published
06/22/2021 10:03 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
2921 - Set instrument to audio-only mode.
Remark
Published
06/15/2021 4:14 PM
2934 - 2922 - Verify instrument initializes successfully; Remark:2923
Procedure Step
Published
06/22/2021 10:03 AM
2922 - Verify instrument initializes successfully
Instruction
Published
06/15/2021 4:06 PM
2923 - Yellow light will illuminate first.
Remark
Published
06/15/2021 4:15 PM
2924 - Green light will illuminate briefly when pager is ready.
Remark
Published
06/15/2021 4:16 PM
2913 - 2911 - Test audio alarm; Remark:
Procedure Step
Published
06/22/2021 10:04 AM
2911 - Test audio alarm
Instruction
Published
06/15/2021 4:06 PM
2935 - 235 - Place instrument next to radioactive gamma source; Remark:2951
Procedure Step
Published
06/22/2021 10:06 AM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
2951 - IO RadPager Move To Source No Alarm Audible
Remark
Published
10/26/2024 9:16 AM
2939 - 2936 - Verify audible alarm functions correctly; Remark:2926
Procedure Step
Published
06/22/2021 10:07 AM
2936 - Verify audible alarm functions correctly
Instruction
Published
06/15/2021 4:09 PM
2926 - Alarm will be heard and the yellow light will illuminate in response to the s...
Remark
Published
07/06/2021 10:44 AM
2940 - 2937 - Press acknowledge button; Remark:2927
Procedure Step
Published
06/22/2021 10:14 AM
2937 - Press acknowledge button
Instruction
Published
06/15/2021 4:09 PM
2927 - Alarm level will display.
Remark
Published
07/09/2021 12:21 PM
2941 - 241 - Move instrument away from radioactive gamma source; Remark:2964
Procedure Step
Published
07/09/2021 2:18 PM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
2964 - IO RadPager Move Source Audible Mode
Remark
Published
10/26/2024 9:24 AM
2942 - 313 - Power off instrument; Remark:2929
Procedure Step
Published
06/22/2021 10:19 AM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
2929 - IO RadPager Power Off Button
Remark
Published
10/26/2024 10:45 AM
2914 - 2912 - Test vibration alarm; Remark:
Procedure Step
Published
06/06/2022 5:28 PM
2912 - Test vibration alarm
Instruction
Published
06/15/2021 4:11 PM
2943 - 220 - Power on instrument; Remark:2947
Procedure Step
Published
06/22/2021 10:20 AM
2947 - Set instrument to vibration-only mode.
Remark
Published
07/06/2021 10:45 AM
2944 - 235 - Place instrument next to radioactive gamma source; Remark:2925
Procedure Step
Published
06/22/2021 10:21 AM
2925 - IO RadPager Move To Source No Alarm
Remark
Published
10/26/2024 10:46 AM
2945 - 2938 - Verify vibration alarm functions correctly; Remark:2930
Procedure Step
Published
06/22/2021 10:22 AM
2938 - Verify vibration alarm functions correctly
Instruction
Published
06/15/2021 4:12 PM
2930 - Pager will vibrate and the yellow light will illuminate in response to the so...
Remark
Published
07/07/2021 9:54 AM
2946 - 2937 - Press acknowledge button; Remark:2931
Procedure Step
Published
06/22/2021 10:22 AM
2931 - Alarm level will display.
Remark
Published
07/06/2021 10:46 AM
2967 - 241 - Move instrument away from radioactive gamma source; Remark:2928
Procedure Step
Published
07/09/2021 2:17 PM
2928 - IO RadPager Move Away Source No Alarm Vib
Remark
Published
10/26/2024 10:50 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
2 minutes
Duration
Published
03/26/2021 3:33 PM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Daily
Frequency
Published
04/01/2020 11:36 AM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
RadPager Power On Audio On Mode
Image
Published
03/27/2021 1:42 AM
Radiation Pager evaluate alarm
Image
Published
03/27/2021 1:42 AM
Radiation Pager green light
Image
Published
03/27/2021 1:42 AM
Radiation Pager by source
Image
Published
03/27/2021 1:42 AM
Pager_Alarm_Noise_000581.png
Image
Published
03/27/2021 1:42 AM
Radiation Pager acknowledge button
Image
Published
03/27/2021 1:42 AM
Radiation Pager acknowledge alarm
Image
Published
03/27/2021 1:42 AM
Radiation Pager move away from source
Image
Published
03/27/2021 1:42 AM
power_off_000564.png
Image
Published
03/26/2021 1:40 AM
Pager_Alarm_Mode_000575.png
Image
Published
03/27/2021 1:42 AM
RadPager Vibrate Mode Toward Source
Image
Published
03/27/2021 1:42 AM
Radiation Pager vibration
Image
Published
03/27/2021 1:42 AM
Radiation Pager Away Source Vibrate Mode
Image
Published
03/27/2021 1:42 AM
DET-HHD-STE-RPG-RM01

Check the radiation pager operation and response to a source of radiation. The source of radiation may be any material that causes an alarm, including consumer products or naturally occurring radioactive materials

Equipment Hierarchy
Sensor Tech Radiation Pager
Frequency
Before each use
Daily
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
2 minutes
Length of time
2
Tools Required
None
Materials Required
Known source of gamma radiation
Originators
Tucker, Brian
Originating Organization
PNNL
Approver
Cody Hostick
Safety Concerns
Radioactive Source
Final Documentation
None
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration.

Date of Origin
Version
1.00
Version #
Revision #
1.00
Revision Description

Updated text to match new FLO curriculum, updated pictures, optimized content for mobile delivery

Revision Date
Revision Author
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00