7928 - 7927 - Establish Initial Conditions; Remark:
Procedure Step
Published
07/12/2023 12:24 AM
7927 - Establish Initial Conditions
Instruction
Published
07/12/2023 12:23 AM
7969 - 7968 - Close lane; Remark:8224
Procedure Step
Published
07/12/2023 12:35 AM
8226 - Warning : Traffic should temporarily be rerouted or halted during this procedure to ens...
Annotation
Published
09/05/2023 1:04 PM
7968 - Close lane
Instruction
Published
07/12/2023 12:35 AM
8224 - Use orange safety (traffic control) cones to setup barriers. Practice establi...
Remark
Published
09/05/2023 11:26 AM
7926 - 7924 - Open RPM door.; Remark:8223
Procedure Step
Published
07/12/2023 12:25 AM
7924 - Open RPM door.
Instruction
Published
07/12/2023 12:25 AM
8223 - Open all doors needed to access components.
Remark
Published
09/05/2023 11:27 AM
7270 - 7235 - Power off RPM; Remark:7236
Procedure Step
Published
09/01/2023 7:57 PM
7235 - Power off RPM
Instruction
Published
04/24/2023 12:30 PM
7236 - Move LD-260 switch to OFF position. The PWR ON
Remark
Published
04/27/2023 4:25 PM
7964 - 7963 - Inspect Desiccant; Remark:
Procedure Step
Published
07/12/2023 12:44 AM
7963 - Inspect Desiccant
Instruction
Published
10/12/2023 10:01 AM
7959 - 7958 - Inspect existing desiccant bags.; Remark:8233
Procedure Step
Published
07/12/2023 12:47 AM
7958 - Inspect existing desiccant bags.
Instruction
Published
07/12/2023 12:44 AM
8233 - Inspect existing desiccant bags visually and by feel. If the bags are only pa...
Remark
Published
09/05/2023 8:31 PM
7960 - 7956 - Document the current date on the existing bag if no replacement is required.; Remark:
Procedure Step
Published
07/12/2023 12:48 AM
7956 - Document the current date on the existing bag if no replacement is required.
Instruction
Published
07/12/2023 12:48 AM
7961 - 7955 - Inspect interior of RPM cabinets.; Remark:7954
Procedure Step
Published
07/12/2023 12:49 AM
7955 - Inspect interior of RPM cabinets.
Instruction
Published
07/12/2023 12:48 AM
7954 - Inspect cabinets for dirt, debris, and signs of water or rodent/insect infest...
Remark
Published
07/12/2023 12:48 AM
7962 - 7953 - Clean cabinet interior.; Remark:8227
Procedure Step
Published
07/12/2023 12:49 AM
7953 - Clean cabinet interior.
Instruction
Published
07/12/2023 12:49 AM
8227 - Use absorbent cleaning rags if required.
Remark
Published
09/05/2023 1:17 PM
7952 - 7951 - Replace Desiccant; Remark:
Procedure Step
Published
07/12/2023 12:50 AM
7951 - Replace Desiccant
Instruction
Published
10/12/2023 10:02 AM
7947 - 7945 - Remove exhausted desiccant bag(s); Remark:
Procedure Step
Published
07/12/2023 12:51 AM
7945 - Remove exhausted desiccant bag(s)
Instruction
Published
07/12/2023 12:51 AM
7948 - 7944 - Install new desiccant bag(s) in place of removed bag; Remark:
Procedure Step
Published
07/12/2023 12:52 AM
7975 - Caution : Place bags away from electrical connections to avoid damage if desiccant leak...
Annotation
Published
07/21/2023 4:26 PM
7944 - Install new desiccant bag(s) in place of removed bag
Instruction
Published
07/12/2023 12:52 AM
7949 - 7943 - Properly dispose of exhausted desiccant bag; Remark:
Procedure Step
Published
07/12/2023 12:53 AM
7943 - Properly dispose of exhausted desiccant bag
Instruction
Published
07/12/2023 12:52 AM
7279 - 7249 - Return equipment to normal operating condition; Remark:
Procedure Step
Published
04/27/2023 4:13 PM
7249 - Return equipment to normal operating condition
Instruction
Published
04/08/2025 3:22 PM
7280 - 7250 - Power on RPM; Remark:7251
Procedure Step
Published
04/27/2023 4:14 PM
7250 - Power on RPM
Instruction
Published
05/01/2023 8:31 AM
7251 - Move LD-260 switch to ON position. The LED PWR ON
Remark
Published
05/10/2023 3:14 PM
7281 - 7252 - Verify RPM start-up; Remark:7253
Procedure Step
Published
04/20/2025 3:26 PM
7252 - Verify RPM start-up
Instruction
Published
04/27/2023 2:23 PM
7253 - RPM will perform a power on self-test (POST) and collect a background radiati...
Remark
Published
05/10/2023 3:14 PM
7282 - 7254 - Confirm network connection; Remark:7255
Procedure Step
Published
04/27/2023 4:15 PM
7254 - Confirm network connection
Instruction
Published
04/27/2023 2:22 PM
7255 - Contact operators to confirm communications have been reestablished between RPM ...
Remark
Published
05/10/2023 3:15 PM
7283 - 7256 - Close and lock all RPM doors; Remark:
Procedure Step
Published
09/01/2023 11:45 PM
7256 - Close and lock all RPM doors
Instruction
Published
04/27/2023 2:21 PM
7842 - 7820 - Inform system operators upon completion of maintenance action; Remark:
Procedure Step
Published
07/07/2023 11:15 PM
7820 - Inform system operators upon completion of maintenance action
Instruction
Published
07/07/2023 10:02 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Warning
Annotation Type
Published
03/02/2020 3:25 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
Keys to access RPM
Replacement Parts
Published
04/27/2023 6:56 PM
Reflective safety vest if required
Replacement Parts
Published
06/20/2023 4:47 PM
Ladder; 2 meters, 6 feet tall; non-conducting if required
Replacement Parts
Published
07/02/2023 4:48 PM
Traffic control cones or equivalent if required
Replacement Parts
Published
06/20/2023 4:47 PM
Safety glasses if required
Replacement Parts
Published
06/20/2023 4:47 PM
Medium Phillips (PH2) screwdriver
Tools
Published
04/24/2020 4:40 PM
Small Phillips (PH1) screwdriver
Tools
Published
05/14/2020 2:48 PM
Marker for marking the date on the desiccant bag
Replacement Parts
Published
07/12/2023 12:31 AM
LD-260 Disconnect Switch Power Off
படம்
Published
04/24/2023 12:49 PM
DET-RPM-RAP-RM01 Desiccant Inspection and Replacement step 2-1
படம்
Published
07/12/2023 9:19 AM
LD-260 Disconnect Switch Power On
படம்
Published
05/04/2023 3:37 PM

TA600 Radiation Pager Battery Replacement

Last Updated: 04/24/2025 1:43 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech TA600 Radiation Pager
Device
Published
10/28/2024 3:51 PM
7500 - 7489 - Remove existing battery; Remark:
Procedure Step
Published
04/08/2025 4:30 PM
7489 - Remove existing battery
Instruction
Published
06/03/2023 12:16 AM
7501 - 313 - Power off instrument; Remark:7490
Procedure Step
Published
06/03/2023 1:50 AM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
7490 - Move mode switch to 0
Remark
Published
06/03/2023 12:42 AM
7502 - 4370 - Locate battery cover; Remark:7491
Procedure Step
Published
06/03/2023 1:51 AM
4370 - Locate battery cover
Instruction
Published
06/06/2022 9:11 PM
7491 - TA600 Battery Cover Location
Remark
Published
06/03/2023 12:44 AM
7503 - 4368 - Remove battery cover; Remark:7492
Procedure Step
Published
06/03/2023 1:53 AM
4368 - Remove battery cover
Instruction
Published
06/06/2022 9:13 PM
7492 - Use medium slotted screwdriver. A coin may also be used. Rotate counterclockw...
Remark
Published
10/19/2023 8:36 AM
7504 - 7493 - Remove old battery; Remark:7494
Procedure Step
Published
06/03/2023 1:54 AM
7493 - Remove old battery
Instruction
Published
06/03/2023 12:58 AM
7494 - Properly dispose of the old battery.
Remark
Published
06/03/2023 1:03 AM
7505 - 4378 - Replace battery; Remark:
Procedure Step
Published
06/03/2023 1:57 AM
4378 - Replace battery
Instruction
Published
06/06/2022 9:41 PM
7506 - 4792 - Insert new battery; Remark:7495
Procedure Step
Published
06/03/2023 1:59 AM
7496 - Caution : Do NOT use alkaline batteries.
Annotation
Published
04/13/2025 5:46 PM
7497 - Note : It is recommended to use a single AA Lithium battery.
Annotation
Published
04/13/2025 5:46 PM
4792 - Insert new battery
Instruction
Published
06/30/2022 6:05 PM
7495 - Insert positive + terminal first.
Remark
Published
06/03/2023 1:06 AM
4362 - 4360 - Replace O-ring (if needed); Remark:4359
Procedure Step
Published
06/06/2022 9:22 PM
4361 - Standard : O-ring size is 2-018.
Annotation
Published
06/06/2022 9:21 PM
4360 - Replace O-ring (if needed)
Instruction
Published
06/06/2022 9:21 PM
4359 - Replace the O-ring if it is broken or shows signs of damage (cracks or splits...
Remark
Published
06/06/2022 9:21 PM
4358 - 4357 - Apply silicone grease (if needed); Remark:4356
Procedure Step
Published
06/06/2022 9:23 PM
4357 - Apply silicone grease (if needed)
Instruction
Published
06/06/2022 9:22 PM
4356 - Silicone grease should be applied to dry O-rings.
Remark
Published
06/06/2022 9:23 PM
7507 - 4354 - Reinstall battery cover; Remark:7498
Procedure Step
Published
06/03/2023 2:03 AM
4354 - Reinstall battery cover
Instruction
Published
06/06/2022 9:24 PM
7498 - Use medium #2 slotted screwdriver. A coin may also be used. Rotate clockwise....
Remark
Published
06/03/2023 1:29 AM
7508 - 319 - Perform operational test; Remark:7499
Procedure Step
Published
04/08/2025 2:18 PM
319 - Perform operational test
Instruction
Published
06/10/2021 8:31 PM
7499 - DET-HHD-STE-TA6-RM01, TA600 Radiation Pager Operational Test.
Remark
Published
04/13/2025 5:47 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Gamma LED displays L every 20 seconds
Frequency
Published
06/06/2022 3:45 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Standard
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
Medium slotted screwdriver or coin
Replacement Parts
Published
06/03/2023 2:35 AM
Power off TA600
படம்
Published
06/03/2023 12:22 AM
Locate Battery Cover on TA600
படம்
Published
06/03/2023 12:25 AM
Remove Battery Cover
படம்
Published
06/03/2023 12:28 AM
Remove Battery Cover Using Quarter
படம்
Published
06/03/2023 12:34 AM
Remove Old TA600 Battery
படம்
Published
06/03/2023 12:35 AM
Insert TA600 New Battery
படம்
Published
06/03/2023 12:37 AM
HRM Battery Cover Sealing Ring
படம்
Published
08/03/2022 10:45 AM
Reinstall Battery Cover
படம்
Published
06/03/2023 12:40 AM

D3S ID Operational Test

Last Updated: 04/24/2025 1:41 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Kromek D3S-ID
Device
Published
10/28/2024 4:01 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
4679 - 220 - Power on instrument; Remark:4572
Procedure Step
Published
06/23/2022 1:26 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
4572 - Press and hold power button for about 1 second. The D3S indicator light will ...
Remark
Published
12/22/2022 12:14 PM
1496 - 1495 - Power on mobile device; Remark:
Procedure Step
Published
11/12/2020 4:40 PM
1495 - Power on mobile device
Instruction
Published
11/12/2020 4:20 PM
4577 - 4571 - Open D3S ID app; Remark:4825
Procedure Step
Published
06/15/2022 1:56 PM
4571 - Open D3S ID app
Instruction
Published
06/15/2022 1:55 PM
4825 - IO D3S App Icon Dark Background
Remark
Published
08/01/2022 12:33 PM
4702 - 4569 - Verify mobile device is communicating with D3S; Remark:4567
Procedure Step
Published
06/29/2022 11:56 PM
4569 - Verify mobile device is communicating with D3S
Instruction
Published
06/15/2022 1:58 PM
4567 - The indicator light will flash blue every 5 seconds when connected via Blueto...
Remark
Published
06/23/2022 12:03 PM
4568 - Additionally, the status icon will display a check mark when connected.
Remark
Published
06/15/2022 1:58 PM
4680 - 2295 - Verify sufficient battery charge; Remark:4587
Procedure Step
Published
06/23/2022 1:34 PM
2295 - Verify sufficient battery charge
Instruction
Published
04/06/2021 4:43 PM
4587 - Check the battery charge on the mobile device. Tap the status icon to verify ...
Remark
Published
06/23/2022 12:04 PM
4690 - 4662 - Verify instrument is in SEARCH mode; Remark:4586
Procedure Step
Published
06/23/2022 2:17 PM
4662 - Verify instrument is in SEARCH mode
Instruction
Published
06/30/2022 12:00 AM
4586 - SEARCH will be highlighted on the screen. If not, tap
Remark
Published
06/15/2022 2:03 PM
4691 - 4585 - Select alarm setting; Remark:4870
Procedure Step
Published
06/23/2022 2:32 PM
4585 - Select alarm setting
Instruction
Published
06/15/2022 2:04 PM
4870 - Tap menu icon.
Remark
Published
08/01/2022 12:34 PM
4689 - Tap speaker icon to enable both Sound and Vibrate.
Remark
Published
06/23/2022 2:20 PM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
4687 - 235 - Place instrument next to radioactive gamma source; Remark:4872
Procedure Step
Published
06/23/2022 2:23 PM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
4872 - IO D3S Move To Rad Source Trefoil
Remark
Published
07/29/2022 10:20 AM
4688 - 1439 - Verify instrument responds to radioactive source; Remark:4686
Procedure Step
Published
06/23/2022 2:24 PM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
4686 - The gamma radiation level will increase. The mobile device will vibrate and e...
Remark
Published
06/23/2022 2:24 PM
4658 - 4657 - Perform identification; Remark:
Procedure Step
Published
06/16/2022 4:40 PM
4657 - Perform identification
Instruction
Published
06/16/2022 4:40 PM
4683 - 4621 - Select confirmation mode; Remark:4620
Procedure Step
Published
06/23/2022 2:26 PM
4621 - Select confirmation mode
Instruction
Published
06/24/2022 9:01 AM
4620 - Tap CONFIRMATION or Move to Confirmation Mode
Remark
Published
06/15/2022 5:11 PM
4684 - 2218 - Place instrument close to measurement location; Remark:4619
Procedure Step
Published
06/23/2022 2:27 PM
2218 - Place instrument close to measurement location
Instruction
Published
04/06/2021 5:15 PM
4619 - Position the D3S close to RN material. For optimal measurement, the pointer w...
Remark
Published
06/15/2022 5:12 PM
4685 - 859 - Start identification measurement; Remark:4618
Procedure Step
Published
06/23/2022 2:29 PM
859 - Start identification measurement
Instruction
Published
10/26/2020 7:00 PM
4618 - Tap Start Scan. Stay in the same location for duration of me...
Remark
Published
06/15/2022 5:13 PM
4698 - 2220 - Wait for measurement to complete; Remark:4616
Procedure Step
Published
04/20/2025 1:01 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
4616 - Wait for at least 120 seconds. Tap STOP ID.
Remark
Published
06/23/2022 3:37 PM
4699 - 1442 - Verify identification results are correct; Remark:4826
Procedure Step
Published
06/23/2022 3:38 PM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
4826 - IO D3S Gamma Identify Cs-137 4.43 uSv
Remark
Published
07/29/2022 9:40 AM
4700 - 241 - Move instrument away from radioactive gamma source; Remark:4871
Procedure Step
Published
06/23/2022 3:39 PM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
4871 - IO D3S Move Away Rad Source Trefoil
Remark
Published
07/29/2022 10:19 AM
4656 - 4480 - Check instrument neutron response (if available); Remark:
Procedure Step
Published
06/16/2022 4:42 PM
4480 - Check instrument neutron response (if available)
Instruction
Published
06/10/2022 7:19 AM
4693 - 3095 - Place instrument next to radioactive neutron source; Remark:4872
Procedure Step
Published
06/23/2022 3:40 PM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
4694 - 1439 - Verify instrument responds to radioactive source; Remark:4692
Procedure Step
Published
06/23/2022 3:43 PM
4692 - The neutron radiation level will increase. The mobile device will vibrate and...
Remark
Published
06/23/2022 3:42 PM
4695 - 3098 - Move instrument away from radioactive neutron source; Remark:4871
Procedure Step
Published
06/23/2022 3:43 PM
3098 - Move instrument away from radioactive neutron source
Instruction
Published
07/16/2021 10:54 AM
4565 - 4564 - Tap SNOOZE; Remark:4566
Procedure Step
Published
06/15/2022 10:43 AM
4564 - Tap SNOOZE
Instruction
Published
06/15/2022 10:11 AM
4566 - IO D3S Snooze Neutron Red
Remark
Published
06/15/2022 10:43 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
No Radiation Symbol
படம்
Published
08/13/2024 2:56 PM
Power on/off D3S
படம்
Published
07/26/2022 12:40 PM
D3S App Icon
படம்
Published
07/26/2022 2:28 PM
KRO D3S Blue Indicator Light
படம்
Published
07/11/2022 3:45 PM
KRO D3S Gamma 02uSv/h
படம்
Published
07/12/2022 10:53 AM
KRO D3S Battery Status Icon
படம்
Published
07/12/2022 10:56 AM
D3S Battery Status
படம்
Published
07/26/2022 1:44 PM
D3S Battery Status Data
படம்
Published
07/26/2022 1:44 PM
D3S Search mode
படம்
Published
07/26/2022 1:44 PM
KRO D3S Menu Icon
படம்
Published
07/12/2022 1:34 PM
D3S Sound Vibrate Icon
படம்
Published
07/26/2022 2:24 PM
D3S Rad Source Closer
படம்
Published
07/29/2022 10:12 AM
D3S Gamma Alarm
படம்
Published
07/29/2022 9:07 AM
D3S Confirmation Mode
படம்
Published
07/27/2022 11:23 AM
D3S Device distance
படம்
Published
07/27/2022 11:23 AM
D3S Begin Scan
படம்
Published
07/27/2022 11:23 AM
D3S Stop Scan ID
படம்
Published
07/27/2022 11:23 AM
D3S Observe Results
படம்
Published
07/27/2022 11:23 AM
D3S Rad Source Away
படம்
Published
07/29/2022 10:14 AM
D3S Neutron Alarm
படம்
Published
07/26/2022 2:24 PM
D3S Snooze High Neutron
படம்
Published
06/15/2022 10:49 AM

RS-350 Operational Test

Last Updated: 04/24/2025 1:41 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Radiation Solutions Inc. RS-350
Device
Published
10/20/2024 8:53 AM
4819 - 257 - Prepare instrument; Remark:
Procedure Step
Published
07/27/2022 1:18 PM
1468 - Note : No artificial radioactive sources should be present during the initial startu...
Annotation
Published
07/27/2022 1:16 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
4424 - 220 - Power on instrument; Remark:4423
Procedure Step
Published
06/10/2022 6:25 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
4423 - Open backpack and press power button for approximately 1 second.
Remark
Published
06/10/2022 5:14 AM
1496 - 1495 - Power on mobile device; Remark:
Procedure Step
Published
11/12/2020 4:40 PM
1495 - Power on mobile device
Instruction
Published
11/12/2020 4:20 PM
4425 - 4421 - Open RadMobile app; Remark:4811
Procedure Step
Published
06/10/2022 5:16 AM
4422 - Note : Refer to DET-HHD-RSI-RS3-OP02, RS-350 RadMobile App Pairing for app ...
Annotation
Published
06/10/2022 5:16 AM
4421 - Open RadMobile app
Instruction
Published
06/10/2022 5:15 AM
4811 - IO RS350 RadMobile App Logo No Words
Remark
Published
07/07/2022 12:54 PM
4426 - 4420 - Select name of RS-350; Remark:4419
Procedure Step
Published
06/10/2022 5:17 AM
4420 - Select name of RS-350
Instruction
Published
06/10/2022 5:17 AM
4419 - The last connected device will appear as the first option on the screen. ...
Remark
Published
06/10/2022 5:17 AM
4427 - 1432 - Verify backpack connection to mobile device; Remark:4418
Procedure Step
Published
06/10/2022 5:20 AM
1432 - Verify backpack connection to mobile device
Instruction
Published
09/01/2020 7:34 PM
4418 - Gamma and neutron values will be changing.
Remark
Published
06/10/2022 5:20 AM
4428 - 2295 - Verify sufficient battery charge; Remark:4416
Procedure Step
Published
06/10/2022 5:23 AM
2295 - Verify sufficient battery charge
Instruction
Published
04/06/2021 4:43 PM
4416 - Backpack battery life is indicated by the battery icon in the app.
Remark
Published
06/10/2022 5:22 AM
4417 - Mobile device battery life is typically indicated at the top of the display.
Remark
Published
06/10/2022 5:23 AM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
4497 - 235 - Place instrument next to radioactive gamma source; Remark:4817
Procedure Step
Published
06/10/2022 7:59 AM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
4817 - IO RS350 Move Closer to Rad Source Trefoil
Remark
Published
07/07/2022 7:25 PM
4498 - 1439 - Verify instrument responds to radioactive source; Remark:4495
Procedure Step
Published
06/10/2022 8:00 AM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
4495 - Gamma dose rate and count rate values will increase. Movin...
Remark
Published
07/07/2022 7:21 PM
4496 - Automatic identification of the radiation source may occur as the RS-350 is b...
Remark
Published
07/07/2022 7:23 PM
2251 - 2217 - Perform Identification Measurement; Remark:
Procedure Step
Published
07/28/2021 9:55 PM
2217 - Perform Identification Measurement
Instruction
Published
04/06/2021 5:13 PM
4452 - 859 - Start identification measurement; Remark:4430
Procedure Step
Published
06/10/2022 6:07 AM
859 - Start identification measurement
Instruction
Published
10/26/2020 7:00 PM
4430 - Tap menu icon.
Remark
Published
06/10/2022 5:38 AM
4451 - Tap Start dwell.
Remark
Published
06/10/2022 6:07 AM
4453 - 2220 - Wait for measurement to complete; Remark:4809
Procedure Step
Published
04/20/2025 12:56 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
4809 - IO RS350 Dwell Mode Timer Running 3/60
Remark
Published
07/06/2022 4:29 PM
4492 - 1442 - Verify identification results are correct; Remark:4491
Procedure Step
Published
06/10/2022 7:28 AM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
4491 - Identification results should match the isotopes associated with the radioact...
Remark
Published
07/07/2022 7:19 PM
4493 - 4490 - Tap CLOSE; Remark:4489
Procedure Step
Published
06/10/2022 7:29 AM
4490 - Tap CLOSE
Instruction
Published
06/10/2022 7:29 AM
4489 - This will return the instrument to Automatic mode.
Remark
Published
07/07/2022 7:18 PM
4494 - 4488 - Move instrument away from radioactive gamma source; Remark:4818
Procedure Step
Published
06/10/2022 7:30 AM
4488 - Move instrument away from radioactive gamma source
Instruction
Published
06/10/2022 7:30 AM
4818 - IO RS350 Move Away from Rad Source Trefoil
Remark
Published
07/07/2022 7:26 PM
4481 - 4480 - Check instrument neutron response (if available); Remark:
Procedure Step
Published
06/10/2022 7:19 AM
4480 - Check instrument neutron response (if available)
Instruction
Published
06/10/2022 7:19 AM
4485 - 3095 - Place instrument next to radioactive neutron source; Remark:4817
Procedure Step
Published
06/10/2022 7:31 AM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
4486 - 1439 - Verify instrument responds to radioactive source; Remark:4484
Procedure Step
Published
06/10/2022 7:32 AM
4484 - Neutron count rate values will increase. The mobile device...
Remark
Published
07/07/2022 7:17 PM
4487 - 4483 - Move instrument away from radioactive neutron source; Remark:4818
Procedure Step
Published
06/10/2022 7:33 AM
4483 - Move instrument away from radioactive neutron source
Instruction
Published
06/10/2022 7:33 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
படம்
Published
08/13/2024 2:56 PM
RS350 Power Button Green Light
படம்
Published
07/05/2022 3:50 PM
RS-350 RadMobile App Logo
படம்
Published
07/05/2022 4:34 PM
RS-350 App Connecting Serial Number Full Screen
படம்
Published
07/06/2022 1:18 PM
RS-350 Graph 70.5 nSv Boxes Dose, Counts, Graph
படம்
Published
07/05/2022 4:18 PM
RS-350 App Battery Icon Red Box
படம்
Published
07/06/2022 1:09 PM
RS-350 App Mobile Battery Percent
படம்
Published
07/06/2022 1:37 PM
RS-350 Move to Rad Source
படம்
Published
07/05/2022 3:56 PM
RS-350 Graph Alarm Response Red Arrow
படம்
Published
07/05/2022 4:22 PM
RS-350 Identification Dwell Result Co-60
படம்
Published
07/05/2022 5:35 PM
RS-350 App Menu Icon Red Box
படம்
Published
07/06/2022 1:27 PM
RS-350 Identification Dwell Start
படம்
Published
07/05/2022 5:37 PM
RS-350 Wait Dwell 3 of 60 sec
படம்
Published
07/05/2022 4:05 PM
RS-350 Close Dwell Mode
படம்
Published
07/05/2022 5:32 PM
RS-350 Move away from Rad Source
படம்
Published
07/05/2022 3:58 PM
RS-350 Alarm Neutron Red Box 10 CPS
படம்
Published
07/05/2022 5:20 PM

Handheld Radiation Monitor Battery Replacement

Last Updated: 04/24/2025 1:40 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech Handheld Radiation Monitor
Device
Published
10/28/2024 3:52 PM
4373 - 313 - Power off instrument; Remark:4372
Procedure Step
Published
06/06/2022 9:29 PM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
4372 - Move switch to OFF position.
Remark
Published
06/06/2022 9:11 PM
4371 - 4370 - Locate battery cover; Remark:4885
Procedure Step
Published
06/06/2022 9:12 PM
4370 - Locate battery cover
Instruction
Published
06/06/2022 9:11 PM
4885 - IO HRM Battery Location Cover
Remark
Published
08/17/2022 2:24 PM
4369 - 4368 - Remove battery cover; Remark:11198
Procedure Step
Published
06/06/2022 9:13 PM
4368 - Remove battery cover
Instruction
Published
06/06/2022 9:13 PM
11198 - Use a coin screwdriver, a medium #2 slotted screwdriver or a coin. Rotate counte...
Remark
Published
10/01/2024 10:19 PM
11199 - TA600 Use a coin screwdriver, rotate counterclockwise
Remark
Published
10/01/2024 10:22 PM
11200 - TA600 Remove battery cover, rotate counterclockwise
Remark
Published
10/01/2024 10:23 PM
11201 - TA600 Use coin, rotate counterclockwise
Remark
Published
10/01/2024 10:25 PM
4366 - 1816 - Remove battery; Remark:4365
Procedure Step
Published
06/06/2022 9:17 PM
1816 - Remove battery
Instruction
Published
11/20/2020 9:29 AM
4365 - Properly dispose of faulty battery
Remark
Published
06/06/2022 9:17 PM
4364 - 1818 - Install replacement battery; Remark:4363
Procedure Step
Published
06/06/2022 9:20 PM
1818 - Install replacement battery
Instruction
Published
11/20/2020 9:28 AM
4363 - The positive + terminal points to top of instrument.
Remark
Published
06/06/2022 9:20 PM
4362 - 4360 - Replace O-ring (if needed); Remark:4359
Procedure Step
Published
06/06/2022 9:22 PM
4361 - Standard : O-ring size is 2-018.
Annotation
Published
06/06/2022 9:21 PM
4360 - Replace O-ring (if needed)
Instruction
Published
06/06/2022 9:21 PM
4359 - Replace the O-ring if it is broken or shows signs of damage (cracks or splits...
Remark
Published
06/06/2022 9:21 PM
4358 - 4357 - Apply silicone grease (if needed); Remark:4356
Procedure Step
Published
06/06/2022 9:23 PM
4357 - Apply silicone grease (if needed)
Instruction
Published
06/06/2022 9:22 PM
4356 - Silicone grease should be applied to dry O-rings.
Remark
Published
06/06/2022 9:23 PM
4355 - 4354 - Reinstall battery cover; Remark:4353
Procedure Step
Published
06/06/2022 9:25 PM
4354 - Reinstall battery cover
Instruction
Published
06/06/2022 9:24 PM
4353 - Use medium #2 slotted screwdriver. A coin may also be used. Rotate clockwise....
Remark
Published
06/06/2022 9:25 PM
4352 - 220 - Power on instrument; Remark:4351
Procedure Step
Published
06/06/2022 9:29 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
4351 - Move switch to AUDIO or VIBRATE.
Remark
Published
06/06/2022 9:27 PM
4290 - 612 - Wait for instrument to complete start-up sequence; Remark:4286
Procedure Step
Published
04/13/2025 6:50 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
4286 - The unit is ready for operation when the period flashes every 20 seconds....
Remark
Published
06/06/2022 4:11 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Gamma LED displays L every 20 seconds
Frequency
Published
06/06/2022 3:45 PM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Standard
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
Medium #2 slotted screwdriver
Tools
Published
06/06/2022 3:41 PM
(or a coin)
Tools
Published
06/06/2022 3:41 PM
HRM Switch in OFF position
படம்
Published
08/02/2022 5:05 PM
HRM Battery Cover Location
படம்
Published
08/03/2022 10:36 AM
TA600 Use a coin screwdriver, rotate counterclockwise
படம்
Published
10/01/2024 11:06 PM
TA600 Remove battery cover, rotate counterclockwise
படம்
Published
10/01/2024 11:06 PM
TA600 Use coin, rotate counterclockwise
படம்
Published
10/01/2024 11:06 PM
HRM Battery Positive Orientation
படம்
Published
08/04/2022 8:35 PM
HRM Battery Cover Sealing Ring
படம்
Published
08/03/2022 10:45 AM
HRM Battery Cover Screwdriver Install
படம்
Published
08/03/2022 10:43 AM
HRM Switch in AUDIO position
படம்
Published
08/02/2022 5:03 PM
HRM Red LED Dot Period
படம்
Published
08/02/2022 4:53 PM

RadSeeker Operational Test

Last Updated: 04/24/2025 1:39 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Smiths Detection RadSeeker CS
Device
Published
06/27/2024 12:54 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
3512 - 2295 - Verify sufficient battery charge; Remark:2297
Procedure Step
Published
07/28/2021 9:32 PM
2295 - Verify sufficient battery charge
Instruction
Published
04/06/2021 4:43 PM
2297 - IO RadSeeker 3 bar Battery
Remark
Published
10/19/2024 10:28 PM
2460 - 220 - Power on instrument; Remark:2303
Procedure Step
Published
11/04/2020 5:34 PM
2360 - Caution : Temperature must be above -15°C for instrument to start. Do not operate the ...
Annotation
Published
11/04/2020 5:27 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
2303 - Press and hold ENTER button until display powers on (3 to 5 ...
Remark
Published
07/23/2021 8:41 PM
2298 - IO RadSeeker Button Pad Select Button Yellow Box
Remark
Published
10/21/2024 7:26 PM
2462 - 612 - Wait for instrument to complete start-up sequence; Remark:2306
Procedure Step
Published
04/13/2025 6:49 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
2306 - The instrument will automatically enter Detect mode after sy...
Remark
Published
07/23/2021 8:45 PM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
3513 - 2352 - Verify Audible Dose Rate Indicator (ADI) is on; Remark:3502
Procedure Step
Published
07/28/2021 9:34 PM
2352 - Verify Audible Dose Rate Indicator (ADI) is on
Instruction
Published
07/29/2021 12:52 PM
3502 - Use ADI button to activate an auditory equivalent to the visual historical gr...
Remark
Published
07/28/2021 9:18 PM
2324 - IO RadSeeker ADI Bar
Remark
Published
10/21/2024 7:33 PM
3514 - 235 - Place instrument next to radioactive gamma source; Remark:3507
Procedure Step
Published
06/06/2022 5:25 PM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
3507 - IO RadSeeker Closer to Trefoil
Remark
Published
10/21/2024 8:54 PM
3515 - 3509 - Verify DETECT mode responds correctly; Remark:3510
Procedure Step
Published
07/28/2021 9:36 PM
3509 - Verify DETECT mode responds correctly
Instruction
Published
07/28/2021 9:37 PM
3510 - The instrument should respond as follows:
Remark
Published
07/28/2021 9:28 PM
2310 - Gamma Dose Rate (µSv/hr) values will increase. Gamma coun...
Remark
Published
11/04/2020 7:38 AM
3516 - 241 - Move instrument away from radioactive gamma source; Remark:3508
Procedure Step
Published
06/06/2022 5:27 PM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
3508 - IO RadSeeker Away from Trefoil
Remark
Published
10/21/2024 8:54 PM
3517 - 3511 - Check Instrument Neutron Response; Remark:
Procedure Step
Published
07/28/2021 9:39 PM
3511 - Check Instrument Neutron Response
Instruction
Published
07/28/2021 9:29 PM
3519 - 3095 - Place instrument next to radioactive neutron source; Remark:3507
Procedure Step
Published
06/06/2022 6:30 PM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
3518 - 3501 - Verify DETECT mode responds correctly; Remark:3500
Procedure Step
Published
07/28/2021 9:40 PM
3096 - Note : Most radioactive neutron sources also have a gamma component, which will likely ...
Annotation
Published
07/23/2021 9:57 AM
3501 - Verify DETECT mode responds correctly
Instruction
Published
07/28/2021 9:12 PM
3500 - The instrument should respond as follows: Vibrate and alarm...
Remark
Published
07/28/2021 9:42 PM
3499 - N-cps values will increase and have a red color. Moving gr...
Remark
Published
07/28/2021 9:10 PM
3520 - 3098 - Move instrument away from radioactive neutron source; Remark:3508
Procedure Step
Published
06/06/2022 6:33 PM
3098 - Move instrument away from radioactive neutron source
Instruction
Published
07/16/2021 10:54 AM
2251 - 2217 - Perform Identification Measurement; Remark:
Procedure Step
Published
07/28/2021 9:55 PM
2217 - Perform Identification Measurement
Instruction
Published
04/06/2021 5:13 PM
3521 - 859 - Start identification measurement; Remark:3498
Procedure Step
Published
07/28/2021 9:46 PM
859 - Start identification measurement
Instruction
Published
10/26/2020 7:00 PM
3498 - Press ENTER under Start ID
Remark
Published
07/28/2021 9:07 PM
3486 - IO RadSeeker StartID Small Menu
Remark
Published
10/21/2024 8:54 PM
2491 - 2220 - Wait for measurement to complete; Remark:2321
Procedure Step
Published
04/20/2025 12:55 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
2321 - IO RadSeeker Collecting Spectrum Wait
Remark
Published
10/21/2024 8:36 PM
2492 - 2357 - Acknowledge alarms; Remark:2326
Procedure Step
Published
11/04/2020 5:49 PM
2357 - Acknowledge alarms
Instruction
Published
07/23/2021 9:29 PM
2326 - Press ENTER to acknowledge innocent or threat alarms.
Remark
Published
07/23/2021 9:35 PM
2320 - IO RadSeeker Alarm Acknowledge Threat
Remark
Published
10/21/2024 8:40 PM
3522 - 1442 - Verify identification results are correct; Remark:3497
Procedure Step
Published
07/28/2021 9:47 PM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
3497 - Identification results should match the isotopes associated with the radioact...
Remark
Published
07/28/2021 9:04 PM
2319 - IO RadSeeker Identify Isotope Co60
Remark
Published
10/21/2024 8:42 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Operator Procedure
Document Type
Published
03/02/2020 3:31 PM
5 minutes
Duration
Published
04/01/2020 11:35 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
படம்
Published
08/13/2024 2:56 PM
radseeker_di_battbar_000477.jpg
படம்
Published
10/18/2023 4:32 PM
radseeker_di_selectmiddle_00473.jpg
படம்
Published
10/20/2023 9:35 AM
radseeker_ui_modedetect_217cps_00467.jpg
படம்
Published
10/20/2023 10:15 AM
radseeker_ui_adiselect_00471.jpg
படம்
Published
10/20/2023 9:42 AM
radseeker_di_MoveToward_rad_00668.jpg
படம்
Published
10/20/2023 9:33 AM
radseeker_ui_increasgr_11164cps_00463.jpg
படம்
Published
10/20/2023 10:07 AM
radseeker_di_MoveAway_rad_00669.jpg
படம்
Published
10/20/2023 9:32 AM
radseeker_ui_neutronalarmack_nobutton_252cps_00654.jpg
படம்
Published
10/20/2023 10:18 AM
radseeker_ui_alarm_neutron_1783cps_00656.jpg
படம்
Published
10/20/2023 9:43 AM
radseeker_ui_menustartid_button_00651.jpg
படம்
Published
10/20/2023 10:11 AM
radseeker_ui_collectspec_00468.jpg
படம்
Published
10/20/2023 9:46 AM
radseeker_ui_threatack_1025cps_00457.jpg
படம்
Published
10/20/2023 10:22 AM
radseeker_ui_idco60_902cps_00465.jpg
படம்
Published
10/20/2023 10:06 AM