ID Title Content type Referenced in Step Referenced in Document Published
9547 9547 - Remove MHV outer connector shell Instruction 9586 - 9547 - Remove MHV outer connector shell; Remark:9548 STEP: 9586 - 9547 - Remove MHV outer connector shell; Remark:9548
RPM Neutron Detector Contaminant Removal
9548 9548 - Use adjustable wrench. Loosen and unscrew connector shell. Remark 9586 - 9547 - Remove MHV outer connector shell; Remark:9548 STEP: 9586 - 9547 - Remove MHV outer connector shell; Remark:9548
RPM Neutron Detector Contaminant Removal
9549 9549 - Confirm O-ring is positioned above threaded surface and undamaged Instruction 9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550 STEP: 9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550
RPM Neutron Detector Contaminant Removal
9550 9550 - The O-ring size is 0.79 mm (0.031 in.) cross section by 19.69 mm (0.775 in.) ... Remark 9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550 STEP: 9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550
RPM Neutron Detector Contaminant Removal
9551 9551 - Remove Teflon insulator ring from ceramic core insulator Instruction 9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552 STEP: 9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552
RPM Neutron Detector Contaminant Removal
9552 9552 - Carefully pull Teflon insulator ring along the tube axis Remark 9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552 STEP: 9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552
RPM Neutron Detector Contaminant Removal
9553 9553 - Examine surface condition of neutron tube insulator Instruction 9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554 STEP: 9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554
RPM Neutron Detector Contaminant Removal
9554 9554 - Questionable insulator Remark 9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554 STEP: 9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554
RPM Neutron Detector Contaminant Removal
9555 9555 - Typical insulator Remark 9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554 STEP: 9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554
RPM Neutron Detector Contaminant Removal
9556 9556 - Clean Insulator Core and Ring Instruction 9590 - 9556 - Clean Insulator Core and Ring; Remark: STEP: 9590 - 9556 - Clean Insulator Core and Ring; Remark:
RPM Neutron Detector Contaminant Removal
9557 9557 - Rinse ceramic and Teflon insulators Instruction 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558 STEP: 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558
RPM Neutron Detector Contaminant Removal
9558 9558 - Using a stream spray of alcohol, coat insulators and allow to evaporate.&nbsp... Remark 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558 STEP: 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558
RPM Neutron Detector Contaminant Removal
9559 9559 - Warning : Use nitrile gloves as sensitive skin may react to the alcohol Annotation 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558 STEP: 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558
RPM Neutron Detector Contaminant Removal
9560 9560 - Caution : For quality check, spray alcohol on a piece of clean glass and let evaporate.... Annotation 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558 STEP: 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558
RPM Neutron Detector Contaminant Removal
9561 9561 - Wipe insulator Instruction 9592 - 9561 - Wipe insulator; Remark:9562 STEP: 9592 - 9561 - Wipe insulator; Remark:9562
RPM Neutron Detector Contaminant Removal
9562 9562 - Sweep surfaces with cottom swab, or lint-free cloth soaked in alcohol. Remark 9592 - 9561 - Wipe insulator; Remark:9562 STEP: 9592 - 9561 - Wipe insulator; Remark:9562
RPM Neutron Detector Contaminant Removal
9563 9563 - Inspect for and remove cotton fibers Instruction 9593 - 9563 - Inspect for and remove cotton fibers; Remark:9564 STEP: 9593 - 9563 - Inspect for and remove cotton fibers; Remark:9564
RPM Neutron Detector Contaminant Removal
9564 9564 - Remove cotton fibers deposited in insulator areas using tweezers. Remark 9593 - 9563 - Inspect for and remove cotton fibers; Remark:9564 STEP: 9593 - 9563 - Inspect for and remove cotton fibers; Remark:9564
RPM Neutron Detector Contaminant Removal
9565 9565 - Caution : Heating of the detector tube shall not exceed 100°C. Annotation 9594 - 9566 - Dry insulators; Remark:9567 STEP: 9594 - 9566 - Dry insulators; Remark:9567
RPM Neutron Detector Contaminant Removal
9566 9566 - Dry insulators Instruction 9594 - 9566 - Dry insulators; Remark:9567 STEP: 9594 - 9566 - Dry insulators; Remark:9567
RPM Neutron Detector Contaminant Removal
9567 9567 - Allow components to air dry. Remark 9594 - 9566 - Dry insulators; Remark:9567 STEP: 9594 - 9566 - Dry insulators; Remark:9567
RPM Neutron Detector Contaminant Removal
9568 9568 - Clean connector shell Instruction 9595 - 9568 - Clean connector shell; Remark:9569 STEP: 9595 - 9568 - Clean connector shell; Remark:9569
RPM Neutron Detector Contaminant Removal
9569 9569 - The interior of the narrow-size barrel is the primary contact surface to clea... Remark 9595 - 9568 - Clean connector shell; Remark:9569 STEP: 9595 - 9568 - Clean connector shell; Remark:9569
RPM Neutron Detector Contaminant Removal
9570 9570 - Reinstall Cleaned Neutron Tube Instruction
9571 9571 - Confirm O-ring is in place Instruction 9597 - 9571 - Confirm O-ring is in place; Remark:9572 STEP: 9597 - 9571 - Confirm O-ring is in place; Remark:9572
RPM Neutron Detector Contaminant Removal
9572 9572 - No lubrication is necessary. Remark 9597 - 9571 - Confirm O-ring is in place; Remark:9572 STEP: 9597 - 9571 - Confirm O-ring is in place; Remark:9572
RPM Neutron Detector Contaminant Removal
9573 9573 - Caution : Do not bend or twist the high voltage wire lead. Annotation 9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552
9598 - 9574 - Reinstall ring insulator; Remark:9575
STEP: 9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552
RPM Neutron Detector Contaminant Removal

STEP: 9598 - 9574 - Reinstall ring insulator; Remark:9575
RPM Neutron Detector Contaminant Removal
9574 9574 - Reinstall ring insulator Instruction 9598 - 9574 - Reinstall ring insulator; Remark:9575 STEP: 9598 - 9574 - Reinstall ring insulator; Remark:9575
RPM Neutron Detector Contaminant Removal
9575 9575 - Slide ring insulator over core insulator and high voltage wire lead. Remark 9598 - 9574 - Reinstall ring insulator; Remark:9575 STEP: 9598 - 9574 - Reinstall ring insulator; Remark:9575
RPM Neutron Detector Contaminant Removal
9576 9576 - Note : Carefully place threads to avoid cross-threading of shell with the tube wall.... Annotation 9599 - 9577 - Screw connector shell in place; Remark:9578 STEP: 9599 - 9577 - Screw connector shell in place; Remark:9578
RPM Neutron Detector Contaminant Removal
9577 9577 - Screw connector shell in place Instruction 9599 - 9577 - Screw connector shell in place; Remark:9578 STEP: 9599 - 9577 - Screw connector shell in place; Remark:9578
RPM Neutron Detector Contaminant Removal
9578 9578 - Screw connector shell until finger tight. Remark 9599 - 9577 - Screw connector shell in place; Remark:9578 STEP: 9599 - 9577 - Screw connector shell in place; Remark:9578
RPM Neutron Detector Contaminant Removal
9579 9579 - Insert new He3 tube into moderator Instruction
9580 9580 - Isolate cleaned He3 tube Instruction 9603 - 9580 - Isolate cleaned He3 tube; Remark:9581
9581 9581 - Disconnect cables from all He3 tubes in moderator (white box) except ... Remark 9603 - 9580 - Isolate cleaned He3 tube; Remark:9581
9582 9582 - Caution : Do not disconnect cable from cleaned tube. Annotation 9603 - 9580 - Isolate cleaned He3 tube; Remark:9581
9583 9583 - Insert cleaned He3 tube into moderator Instruction 9600 - 9583 - Insert cleaned He3 tube into moderator; Remark:
9584 9584 - 9543 - Neutron Tube Serial Number; Remark: Procedure Step
9585 9585 - 9544 - Record serial number; Remark:9546 Procedure Step
9586 9586 - 9547 - Remove MHV outer connector shell; Remark:9548 Procedure Step
9587 9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550 Procedure Step
9588 9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552 Procedure Step
9589 9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554 Procedure Step
9590 9590 - 9556 - Clean Insulator Core and Ring; Remark: Procedure Step
9591 9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558 Procedure Step
9592 9592 - 9561 - Wipe insulator; Remark:9562 Procedure Step
9593 9593 - 9563 - Inspect for and remove cotton fibers; Remark:9564 Procedure Step
9594 9594 - 9566 - Dry insulators; Remark:9567 Procedure Step
9595 9595 - 9568 - Clean connector shell; Remark:9569 Procedure Step
9596 9596 - 7249 - Return equipment to normal operating condition; Remark: Procedure Step
9597 9597 - 9571 - Confirm O-ring is in place; Remark:9572 Procedure Step
9598 9598 - 9574 - Reinstall ring insulator; Remark:9575 Procedure Step
9599 9599 - 9577 - Screw connector shell in place; Remark:9578 Procedure Step
9600 9600 - 9583 - Insert cleaned He3 tube into moderator; Remark: Procedure Step
9602 9602 - 8520 - Remove stabilizing brace screws; Remark:8521 Procedure Step
9603 9603 - 9580 - Isolate cleaned He3 tube; Remark:9581 Procedure Step
9604 9604 - Caution : Avoid tearing or binding the rubber O-ring at the base of the connector shell... Annotation 9586 - 9547 - Remove MHV outer connector shell; Remark:9548
9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550
STEP: 9586 - 9547 - Remove MHV outer connector shell; Remark:9548
RPM Neutron Detector Contaminant Removal

STEP: 9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550
RPM Neutron Detector Contaminant Removal
9605 9605 - Note : This step is only needed if cleaning He3 tube from upper detector. Annotation 8566 - 8523 - Loosen bracket screw (if applicable); Remark:8525 STEP: 8566 - 8523 - Loosen bracket screw (if applicable); Remark:8525
RPM Neutron Detector Helium-3 (He3) Tube Replacement
9606 9606 - Verify counts from cleaned He3 tube Instruction 8582 - 9606 - Verify counts from cleaned He3 tube; Remark:8553 STEP: 8582 - 9606 - Verify counts from cleaned He3 tube; Remark:8553
RPM Neutron Detector Helium-3 (He3) Tube Replacement
9607 9607 - Warning : 110/240 VAC is present at circuit breaker and at the AC line filter. Annotation 9632 - 9609 - Power off AC circuit breaker; Remark:9610
9642 - 9628 - Power on AC circuit breaker; Remark:9629
STEP: 9632 - 9609 - Power off AC circuit breaker; Remark:9610
RPM AC Line Filter Replacement

STEP: 9642 - 9628 - Power on AC circuit breaker; Remark:9629
RPM AC Line Filter Replacement
9608 9608 - Note : Normally, the OFF position will be when the circuit breaker ... Annotation 9632 - 9609 - Power off AC circuit breaker; Remark:9610 STEP: 9632 - 9609 - Power off AC circuit breaker; Remark:9610
RPM AC Line Filter Replacement
9609 9609 - Power off AC circuit breaker Instruction 9632 - 9609 - Power off AC circuit breaker; Remark:9610 STEP: 9632 - 9609 - Power off AC circuit breaker; Remark:9610
RPM AC Line Filter Replacement
9610 9610 - The AC circuit breaker is located at the bottom of the control pillar. ... Remark 9632 - 9609 - Power off AC circuit breaker; Remark:9610 STEP: 9632 - 9609 - Power off AC circuit breaker; Remark:9610
RPM AC Line Filter Replacement
9611 9611 - Uninstall AC Line Filter Instruction 9633 - 9611 - Uninstall AC Line Filter; Remark: STEP: 9633 - 9611 - Uninstall AC Line Filter; Remark:
RPM AC Line Filter Replacement
9612 9612 - Confirm no AC voltage is present Instruction 9634 - 9612 - Confirm no AC voltage is present; Remark:9613 STEP: 9634 - 9612 - Confirm no AC voltage is present; Remark:9613
RPM AC Line Filter Replacement
9613 9613 - Using a digital multimeter, place the negative (-) lead on the mounting screw... Remark 9634 - 9612 - Confirm no AC voltage is present; Remark:9613 STEP: 9634 - 9612 - Confirm no AC voltage is present; Remark:9613
RPM AC Line Filter Replacement
9614 9614 - Warning : DO NOT proceed if voltage is present at either wire location... Annotation 9635 - 9615 - Measure each black wire connection; Remark:9616
9636 - 9617 - Measure each white wire connection; Remark:9618
STEP: 9635 - 9615 - Measure each black wire connection; Remark:9616
RPM AC Line Filter Replacement

STEP: 9636 - 9617 - Measure each white wire connection; Remark:9618
RPM AC Line Filter Replacement
9615 9615 - Measure each black wire connection Instruction 9635 - 9615 - Measure each black wire connection; Remark:9616 STEP: 9635 - 9615 - Measure each black wire connection; Remark:9616
RPM AC Line Filter Replacement
9616 9616 - Using a digital multimeter, place the positive (+) lead on each of the mounti... Remark 9635 - 9615 - Measure each black wire connection; Remark:9616 STEP: 9635 - 9615 - Measure each black wire connection; Remark:9616
RPM AC Line Filter Replacement
9617 9617 - Measure each white wire connection Instruction 9636 - 9617 - Measure each white wire connection; Remark:9618 STEP: 9636 - 9617 - Measure each white wire connection; Remark:9618
RPM AC Line Filter Replacement
9618 9618 - Using a digital multimeter, place the positive (+) lead on each of the mounti... Remark 9636 - 9617 - Measure each white wire connection; Remark:9618 STEP: 9636 - 9617 - Measure each white wire connection; Remark:9618
RPM AC Line Filter Replacement
9619 9619 - Disconnect all wiring Instruction 9637 - 9619 - Disconnect all wiring; Remark:9620 STEP: 9637 - 9619 - Disconnect all wiring; Remark:9620
RPM AC Line Filter Replacement
9620 9620 - Using a small Phillips screwdriver, disconnect all the wires attached to the ... Remark 9637 - 9619 - Disconnect all wiring; Remark:9620 STEP: 9637 - 9619 - Disconnect all wiring; Remark:9620
RPM AC Line Filter Replacement
9621 9621 - Remove AC line filter Instruction 9638 - 9621 - Remove AC line filter; Remark:9622 STEP: 9638 - 9621 - Remove AC line filter; Remark:9622
RPM AC Line Filter Replacement
9622 9622 - Remove two M4 mounting screws and washers. Retain hardware. Remark 9638 - 9621 - Remove AC line filter; Remark:9622 STEP: 9638 - 9621 - Remove AC line filter; Remark:9622
RPM AC Line Filter Replacement
9623 9623 - Install New AC Line Filter Instruction 9639 - 9623 - Install New AC Line Filter; Remark: STEP: 9639 - 9623 - Install New AC Line Filter; Remark:
RPM AC Line Filter Replacement
9624 9624 - Mount new AC line filter Instruction 9640 - 9624 - Mount new AC line filter; Remark:9625 STEP: 9640 - 9624 - Mount new AC line filter; Remark:9625
RPM AC Line Filter Replacement
9625 9625 - Reuse retained M4 screws and internal washers. Remember to mount the ground w... Remark 9640 - 9624 - Mount new AC line filter; Remark:9625 STEP: 9640 - 9624 - Mount new AC line filter; Remark:9625
RPM AC Line Filter Replacement
9626 9626 - Reconnect AC line filter wiring Instruction 9641 - 9626 - Reconnect AC line filter wiring; Remark:9627 STEP: 9641 - 9626 - Reconnect AC line filter wiring; Remark:9627
RPM AC Line Filter Replacement
9627 9627 - Make sure the load wire (+) is accross from the line wire (+). Remark 9641 - 9626 - Reconnect AC line filter wiring; Remark:9627 STEP: 9641 - 9626 - Reconnect AC line filter wiring; Remark:9627
RPM AC Line Filter Replacement
9628 9628 - Power on AC circuit breaker Instruction 9642 - 9628 - Power on AC circuit breaker; Remark:9629 STEP: 9642 - 9628 - Power on AC circuit breaker; Remark:9629
RPM AC Line Filter Replacement
9629 9629 - AC Circuit Breaker On Remark 9642 - 9628 - Power on AC circuit breaker; Remark:9629 STEP: 9642 - 9628 - Power on AC circuit breaker; Remark:9629
RPM AC Line Filter Replacement
9630 9630 - Verify AC line filter voltage Instruction 9643 - 9630 - Verify AC line filter voltage; Remark:9631 STEP: 9643 - 9630 - Verify AC line filter voltage; Remark:9631
RPM AC Line Filter Replacement
9631 9631 - Voltage should be present at both the input and output. 110-240 VAC will be p... Remark 9643 - 9630 - Verify AC line filter voltage; Remark:9631 STEP: 9643 - 9630 - Verify AC line filter voltage; Remark:9631
RPM AC Line Filter Replacement
9632 9632 - 9609 - Power off AC circuit breaker; Remark:9610 Procedure Step
9633 9633 - 9611 - Uninstall AC Line Filter; Remark: Procedure Step
9634 9634 - 9612 - Confirm no AC voltage is present; Remark:9613 Procedure Step
9635 9635 - 9615 - Measure each black wire connection; Remark:9616 Procedure Step
9636 9636 - 9617 - Measure each white wire connection; Remark:9618 Procedure Step
9637 9637 - 9619 - Disconnect all wiring; Remark:9620 Procedure Step
9638 9638 - 9621 - Remove AC line filter; Remark:9622 Procedure Step
9639 9639 - 9623 - Install New AC Line Filter; Remark: Procedure Step
9640 9640 - 9624 - Mount new AC line filter; Remark:9625 Procedure Step
9641 9641 - 9626 - Reconnect AC line filter wiring; Remark:9627 Procedure Step
9642 9642 - 9628 - Power on AC circuit breaker; Remark:9629 Procedure Step
9643 9643 - 9630 - Verify AC line filter voltage; Remark:9631 Procedure Step
9645 9645 - Prepare for Faulty High Voltage Board Removal Instruction 9646 - 9645 - Prepare for Faulty High Voltage Board Removal; Remark: STEP: 9646 - 9645 - Prepare for Faulty High Voltage Board Removal; Remark:
RPM HHV-448/458 High Voltage Board Replacement
9646 9646 - 9645 - Prepare for Faulty High Voltage Board Removal; Remark: Procedure Step
9647 9647 - Remove SCA-774 board from SCA-775 module Instruction 9658 - 9647 - Remove SCA-774 board from SCA-775 module; Remark:9648 STEP: 9658 - 9647 - Remove SCA-774 board from SCA-775 module; Remark:9648
RPM HHV-448/458 High Voltage Board Replacement
9648 9648 - BNC connectors are attached to the SCA-774 board and easily slide down then o... Remark 9658 - 9647 - Remove SCA-774 board from SCA-775 module; Remark:9648 STEP: 9658 - 9647 - Remove SCA-774 board from SCA-775 module; Remark:9648