7738 - 7646 - Prepare RPM Specific FCT Datasheet; Remark: |
Procedure Step
|
Published
|
07/03/2023 10:36 AM
|
7647 - Note : Step 1 (1 through 1-4) should be performed before traveling to the RPMs. ... |
Annotation
|
Published
|
06/30/2023 1:12 PM
|
7646 - Prepare RPM Specific FCT Datasheet |
Instruction
|
Published
|
06/30/2023 1:11 PM
|
7739 - 7648 - Open most recent baselined FCT datasheet on laptop; Remark: |
Procedure Step
|
Published
|
07/02/2023 12:27 PM
|
7648 - Open most recent baselined FCT datasheet on laptop |
Instruction
|
Published
|
06/30/2023 1:14 PM
|
7740 - 7649 - Save file with unique name (add today's date) and close master copy; Remark:7650 |
Procedure Step
|
Published
|
07/02/2023 12:28 PM
|
7649 - Save file with unique name (add today's date) and close master copy |
Instruction
|
Published
|
06/30/2023 1:15 PM
|
7650 - An example of a unique file name is US_PortofEntry_FCT_Lane1_05132016... |
Remark
|
Published
|
07/02/2023 3:00 AM
|
7741 - 7651 - Enter date of procedure performance in appropriate cell; Remark: |
Procedure Step
|
Published
|
07/02/2023 12:30 PM
|
7652 - Note : Date of procedure in FCT datasheet is extremely important for correct radioac... |
Annotation
|
Published
|
07/02/2023 5:22 PM
|
7651 - Enter date of procedure performance in appropriate cell |
Instruction
|
Published
|
06/30/2023 1:19 PM
|
7742 - 7653 - Enter RPM identification information; Remark:7654 |
Procedure Step
|
Published
|
07/02/2023 12:32 PM
|
7653 - Enter RPM identification information |
Instruction
|
Published
|
06/30/2023 1:21 PM
|
7654 - Complete the General Information section to specifically identify the RPM (se... |
Remark
|
Published
|
07/02/2023 3:01 AM
|
7743 - 7655 - Verify Gamma Radioactive Source is Sufficient for FCT; Remark: |
Procedure Step
|
Published
|
07/02/2023 12:37 PM
|
7655 - Verify Gamma Radioactive Source is Sufficient for FCT |
Instruction
|
Published
|
07/02/2023 5:23 PM
|
7744 - 7656 - Enter Co-57 source identification number; Remark:7657 |
Procedure Step
|
Published
|
07/02/2023 12:40 PM
|
7656 - Enter Co-57 source identification number |
Instruction
|
Published
|
06/30/2023 2:26 PM
|
7657 - Source information is commonly listed directly on the label of radioactive so... |
Remark
|
Published
|
07/02/2023 3:01 AM
|
7745 - 7658 - Enter Co-57 initial activity in kBq (as listed on source label or source cert...; Remark:7659 |
Procedure Step
|
Published
|
07/02/2023 12:40 PM
|
7658 - Enter Co-57 initial activity in kBq (as listed on source label or source cert... |
Instruction
|
Published
|
06/30/2023 2:35 PM
|
7659 - Source Activity Level |
Remark
|
Published
|
07/02/2023 3:02 AM
|
7746 - 7660 - Enter Co-57 source calibration date; Remark:7661 |
Procedure Step
|
Published
|
07/02/2023 12:42 PM
|
7660 - Enter Co-57 source calibration date |
Instruction
|
Published
|
06/30/2023 2:46 PM
|
7661 - Source Calibration Date |
Remark
|
Published
|
07/02/2023 3:02 AM
|
7747 - 7662 - Verify Co-57 calculated current activity is within specified limits; Remark:7664 |
Procedure Step
|
Published
|
04/13/2025 6:20 PM
|
7663 - Standard : Current Co-57 source activity must be between 19 and 400 kBq to perform this ... |
Annotation
|
Published
|
07/02/2023 5:24 PM
|
7662 - Verify Co-57 calculated current activity is within specified limits |
Instruction
|
Published
|
04/13/2025 5:49 PM
|
7664 - Source Current Activity Level |
Remark
|
Published
|
07/02/2023 3:03 AM
|
7748 - 7665 - Verify Neutron Radioactive Source is Sufficient for FCT ...; Remark: |
Procedure Step
|
Published
|
07/02/2023 12:46 PM
|
7665 - Verify Neutron Radioactive Source is Sufficient for FCT ... |
Instruction
|
Published
|
06/30/2023 2:59 PM
|
7749 - 7666 - Enter Cf-252 source identification number; Remark:7667 |
Procedure Step
|
Published
|
07/02/2023 12:48 PM
|
7666 - Enter Cf-252 source identification number |
Instruction
|
Published
|
06/30/2023 3:00 PM
|
7667 - Source information is commonly listed directly on the label of the radioactiv... |
Remark
|
Published
|
07/03/2023 10:21 AM
|
7750 - 7668 - Enter Cf-252 source activity in kBq (as listed on source label or source cert...; Remark:7669 |
Procedure Step
|
Published
|
07/02/2023 12:49 PM
|
7668 - Enter Cf-252 source activity in kBq (as listed on source label or source cert... |
Instruction
|
Published
|
06/30/2023 3:23 PM
|
7669 - |
Remark
|
Published
|
07/02/2023 12:51 PM
|
7751 - 7670 - Enter Cf-252 source calibration date; Remark:7671 |
Procedure Step
|
Published
|
07/02/2023 12:52 PM
|
7670 - Enter Cf-252 source calibration date |
Instruction
|
Published
|
06/30/2023 3:35 PM
|
7671 - Certificate of Calibration Neutron Date |
Remark
|
Published
|
07/02/2023 3:04 AM
|
7752 - 7673 - Verify Cf-252 calculated current activity is within specified limits; Remark:7674 |
Procedure Step
|
Published
|
07/02/2023 12:54 PM
|
7672 - Standard : Current Cf-252 source activity must be between 37 and 185 kBq to perform this... |
Annotation
|
Published
|
06/30/2023 3:42 PM
|
7673 - Verify Cf-252 calculated current activity is within specified limits |
Instruction
|
Published
|
06/30/2023 3:44 PM
|
7674 - Current Cf-252 source activity must be between 2,000 and 20,000 n/s to perfor... |
Remark
|
Published
|
07/02/2023 3:05 AM
|
7621 - 1401 - Establish Initial Conditions; Remark: |
Procedure Step
|
Published
|
06/29/2023 5:37 PM
|
1401 - Establish Initial Conditions |
Instruction
|
Published
|
09/30/2022 3:14 PM
|
7622 - 7592 - Inform system operators; Remark:11335 |
Procedure Step
|
Published
|
04/20/2025 10:15 AM
|
7592 - Inform system operators |
Instruction
|
Published
|
06/29/2023 3:20 PM
|
11335 - Upon arrival, brief system operators on expected work, duration, anticipated ... |
Remark
|
Published
|
04/20/2025 10:14 AM
|
7753 - 7676 - Brief system operators on expected work, duration, anticipated alarms, and fa...; Remark:7677 |
Procedure Step
|
Published
|
07/02/2023 1:53 PM
|
7676 - Brief system operators on expected work, duration, anticipated alarms, and fa... |
Instruction
|
Published
|
07/01/2023 11:39 PM
|
7677 - Operators should expect numerous fault indications and alarms during procedur... |
Remark
|
Published
|
07/02/2023 3:05 AM
|
7754 - 7594 - Close lane and set up barriers to establish a safety watch; Remark:7595 |
Procedure Step
|
Published
|
07/02/2023 1:55 PM
|
7678 - Warning : Traffic should temporarily be rerouted or halted during conduction of this pr... |
Annotation
|
Published
|
07/03/2023 10:23 AM
|
7594 - Close lane and set up barriers to establish a safety watch |
Instruction
|
Published
|
06/29/2023 3:28 PM
|
7595 - Use orange safety (traffic control) cones or equivalent per site-specific saf... |
Remark
|
Published
|
06/29/2023 3:33 PM
|
7755 - 7679 - Disconnect Ethernet; Remark:11336 |
Procedure Step
|
Published
|
04/20/2025 11:18 AM
|
7680 - Caution : Make sure to press release button on back of Ethernet cable before removing. |
Annotation
|
Published
|
07/01/2023 11:54 PM
|
7679 - Disconnect Ethernet |
Instruction
|
Published
|
07/01/2023 11:52 PM
|
11336 - Locate Ethernet connector on top of SC-770 controller. Press release button o... |
Remark
|
Published
|
04/20/2025 11:21 AM
|
7756 - 7681 - Disable occupancy sensor; Remark:7682 |
Procedure Step
|
Published
|
07/02/2023 1:59 PM
|
7681 - Disable occupancy sensor |
Instruction
|
Published
|
07/01/2023 11:55 PM
|
7682 - For vehicle monitor, disconnect both break-beam sensor receivers (usually*) l... |
Remark
|
Published
|
07/02/2023 3:06 AM
|
7757 - 344 - Verify parameter settings are correct; Remark: |
Procedure Step
|
Published
|
07/02/2023 2:00 PM
|
344 - Verify parameter settings are correct |
Instruction
|
Published
|
07/03/2023 10:24 AM
|
7280 - 7250 - Power on RPM; Remark:7251 |
Procedure Step
|
Published
|
04/27/2023 4:14 PM
|
7250 - Power on RPM |
Instruction
|
Published
|
05/01/2023 8:31 AM
|
7251 - Move LD-260 switch to ON position. The LED PWR ON |
Remark
|
Published
|
05/10/2023 3:14 PM
|
7281 - 7252 - Verify RPM start-up; Remark:7253 |
Procedure Step
|
Published
|
04/20/2025 3:26 PM
|
7252 - Verify RPM start-up |
Instruction
|
Published
|
04/27/2023 2:23 PM
|
7253 - RPM will perform a power on self-test (POST) and collect a background radiati... |
Remark
|
Published
|
05/10/2023 3:14 PM
|
7760 - 7683 - Record SC-770 parameter values; Remark:7684 |
Procedure Step
|
Published
|
07/02/2023 2:09 PM
|
7685 - Caution : Each SC-770 parameter must match the values obtained from the LANL configurat... |
Annotation
|
Published
|
07/02/2023 12:20 AM
|
7683 - Record SC-770 parameter values |
Instruction
|
Published
|
07/02/2023 12:09 AM
|
7684 - Enter Program Mode by pressing "0." Enter the password 1234... |
Remark
|
Published
|
07/02/2023 3:06 AM
|
7761 - 7686 - Manually enter variable values; Remark:7687 |
Procedure Step
|
Published
|
07/02/2023 2:12 PM
|
7686 - Manually enter variable values |
Instruction
|
Published
|
07/02/2023 12:22 AM
|
7687 - Variable values include parameters such as N*Sigma and Alpha value. |
Remark
|
Published
|
04/13/2025 5:49 PM
|
7762 - 7688 - Click the check box to confirm fixed values; Remark: |
Procedure Step
|
Published
|
04/13/2025 6:20 PM
|
7688 - Click the check box to confirm fixed values |
Instruction
|
Published
|
04/13/2025 5:49 PM
|
7763 - 7689 - Return to SC-770 main menu; Remark:7690 |
Procedure Step
|
Published
|
07/02/2023 2:14 PM
|
7689 - Return to SC-770 main menu |
Instruction
|
Published
|
07/02/2023 12:28 AM
|
7690 - Press "#" button until the SC-770 displays the main menu. |
Remark
|
Published
|
04/13/2025 5:50 PM
|
7764 - 7691 - Perform Gamma Detector Efficiency Measurements; Remark: |
Procedure Step
|
Published
|
04/13/2025 6:20 PM
|
7691 - Perform Gamma Detector Efficiency Measurements |
Instruction
|
Published
|
04/13/2025 5:50 PM
|
7765 - 7694 - Navigate SC-770 to SHOW COUNTS for gamma detectors; Remark:7695 |
Procedure Step
|
Published
|
07/02/2023 2:18 PM
|
7692 - Caution : Keep gamma detectors protected from direct sunlight. |
Annotation
|
Published
|
04/13/2025 5:50 PM
|
7693 - Note : Verify there are no radiation sources sources within 20 meters of the RPM whi... |
Annotation
|
Published
|
07/02/2023 12:40 AM
|
7694 - Navigate SC-770 to SHOW COUNTS for gamma detectors |
Instruction
|
Published
|
07/02/2023 12:41 AM
|
7695 - From SC-770 main programming menu, press GAMMA |
Remark
|
Published
|
04/13/2025 5:52 PM
|
7766 - 7696 - Record measured gamma background count rate for each gamma detector; Remark:7697 |
Procedure Step
|
Published
|
07/02/2023 2:19 PM
|
7696 - Record measured gamma background count rate for each gamma detector |
Instruction
|
Published
|
07/02/2023 12:57 AM
|
7697 - Stand clear of all gamma detectors, watch displayed count rate for each chann... |
Remark
|
Published
|
07/02/2023 3:07 AM
|
7767 - 7700 - Place label side of Co-57 gamma source against yellow dot at center of detect...; Remark:7701 |
Procedure Step
|
Published
|
07/02/2023 2:23 PM
|
7698 - Standard : Due to low energy of Co-57 gammas, the count rate on either side of the sourc... |
Annotation
|
Published
|
07/02/2023 1:07 AM
|
7699 - Note : Verify distance between source and detector is minimized to warrant accurate ... |
Annotation
|
Published
|
07/02/2023 1:08 AM
|
7700 - Place label side of Co-57 gamma source against yellow dot at center of detect... |
Instruction
|
Published
|
07/02/2023 1:09 AM
|
7701 - The source must be held firmly against the detector with a low-density tool s... |
Remark
|
Published
|
07/02/2023 3:08 AM
|
7768 - 7702 - Record gamma count rate in appropriate "Source Count Rate (Co-57)" cell; Remark:7704 |
Procedure Step
|
Published
|
07/02/2023 2:25 PM
|
7703 - Note : Verify that there are no additional radiation sources within 20 meters of the... |
Annotation
|
Published
|
07/02/2023 1:17 AM
|
7702 - Record gamma count rate in appropriate "Source Count Rate (Co-57)" cell |
Instruction
|
Published
|
07/02/2023 1:15 AM
|
7704 - Stand clear of all gamma detectors, watch displayed count rate for each chann... |
Remark
|
Published
|
07/02/2023 1:23 AM
|
7769 - 7705 - Repeat for each gamma detector in RPM; Remark: |
Procedure Step
|
Published
|
07/02/2023 2:27 PM
|
7705 - Repeat for each gamma detector in RPM |
Instruction
|
Published
|
07/02/2023 1:25 AM
|
7770 - 7707 - Verify gamma detector efficiency; Remark:7708 |
Procedure Step
|
Published
|
04/13/2025 6:20 PM
|
7706 - Standard : Gamma detector efficiency should be greater than 20% |
Annotation
|
Published
|
07/02/2023 1:31 AM
|
7707 - Verify gamma detector efficiency |
Instruction
|
Published
|
04/13/2025 5:52 PM
|
7708 - Gamma Detector Efficiency |
Remark
|
Published
|
07/02/2023 3:08 AM
|
7771 - 7709 - Go to SC-770 main menu; Remark:7710 |
Procedure Step
|
Published
|
07/02/2023 2:31 PM
|
7709 - Go to SC-770 main menu |
Instruction
|
Published
|
07/02/2023 1:41 AM
|
7710 - Press "#" button until the SC-770 displays the main menu. |
Remark
|
Published
|
07/02/2023 3:09 AM
|
7772 - 7711 - Perform Neutron Detector Efficiency Measurements; Remark: |
Procedure Step
|
Published
|
07/02/2023 2:33 PM
|
7711 - Perform Neutron Detector Efficiency Measurements |
Instruction
|
Published
|
07/02/2023 1:49 AM
|
7773 - 7712 - Enter "fixed" neutron background count rate for each neutron detector; Remark:7713 |
Procedure Step
|
Published
|
07/02/2023 2:34 PM
|
7712 - Enter "fixed" neutron background count rate for each neutron detector |
Instruction
|
Published
|
07/02/2023 1:52 AM
|
7713 - Record 1 counts per second (cps) for each neutron tube installed on each chan... |
Remark
|
Published
|
07/02/2023 2:00 AM
|
7774 - 7714 - Navigate SC-770 to SHOW COUNTS for neutron detectors; Remark:7715 |
Procedure Step
|
Published
|
07/02/2023 2:36 PM
|
7714 - Navigate SC-770 to SHOW COUNTS for neutron detectors |
Instruction
|
Published
|
07/02/2023 2:02 AM
|
7715 - From SC-770 main programming menu, press NEUTRON ... |
Remark
|
Published
|
04/13/2025 6:19 PM
|
7775 - 7718 - Place Cf-252 neutron source against yellow dot on center of detector; Remark:7719 |
Procedure Step
|
Published
|
07/02/2023 2:37 PM
|
7716 - Warning : Do not hold the neutron source on the detector face with your hand or stand n... |
Annotation
|
Published
|
07/02/2023 2:07 AM
|
7717 - Note : During the neutron detector efficiency tests, verify no one is near t... |
Annotation
|
Published
|
07/02/2023 2:10 AM
|
7718 - Place Cf-252 neutron source against yellow dot on center of detector |
Instruction
|
Published
|
07/02/2023 2:12 AM
|
7719 - It is acceptable to use adhesive tape for this step |
Remark
|
Published
|
04/13/2025 5:54 PM
|
7776 - 7720 - Record neutron count rate in appropriate "Source Count Rate (Cf-252)" cell; Remark:7722 |
Procedure Step
|
Published
|
07/02/2023 2:38 PM
|
7721 - Note : Ensure that threre are no additional radiation sources within 20 meters of th... |
Annotation
|
Published
|
07/02/2023 2:18 AM
|
7720 - Record neutron count rate in appropriate "Source Count Rate (Cf-252)" cell |
Instruction
|
Published
|
07/02/2023 2:15 AM
|
7722 - Stand clear of all neutron detectors, watch displayed count rate for each cha... |
Remark
|
Published
|
07/02/2023 2:22 AM
|
7777 - 7723 - Repeat for all neutron detectors in RPM; Remark: |
Procedure Step
|
Published
|
07/02/2023 2:40 PM
|
7723 - Repeat for all neutron detectors in RPM |
Instruction
|
Published
|
07/02/2023 2:24 AM
|
7778 - 7724 - Verify neutron detector efficiency; Remark:7726 |
Procedure Step
|
Published
|
07/02/2023 2:42 PM
|
7725 - Standard : Neutron detector efficiency should be greater than 5% (VM with 2 helium tubes... |
Annotation
|
Published
|
07/02/2023 2:27 AM
|
7724 - Verify neutron detector efficiency |
Instruction
|
Published
|
07/02/2023 2:25 AM
|
7726 - Neutron Source Efficiency |
Remark
|
Published
|
07/02/2023 2:29 AM
|
7279 - 7249 - Return equipment to normal operating condition; Remark: |
Procedure Step
|
Published
|
04/27/2023 4:13 PM
|
7249 - Return equipment to normal operating condition |
Instruction
|
Published
|
04/08/2025 3:22 PM
|
7780 - 7728 - Exit SC-770 programming mode; Remark:7729 |
Procedure Step
|
Published
|
07/02/2023 2:44 PM
|
7728 - Exit SC-770 programming mode |
Instruction
|
Published
|
07/02/2023 2:33 AM
|
7729 - Press "#" until SC-770 exits programming mode and return to normal operating ... |
Remark
|
Published
|
07/02/2023 2:34 AM
|
7781 - 7730 - Reconnect Ethernet; Remark:7731 |
Procedure Step
|
Published
|
04/20/2025 11:00 AM
|
7730 - Reconnect Ethernet |
Instruction
|
Published
|
07/02/2023 2:34 AM
|
7731 - Insert Ethernet connector at top of SC-770 module. Connector will audibly "cl... |
Remark
|
Published
|
07/02/2023 2:36 AM
|
7782 - 319 - Perform operational test; Remark:7734 |
Procedure Step
|
Published
|
07/02/2023 2:47 PM
|
7732 - Caution : It is critical that the RPM be returned to the proper operational configurati... |
Annotation
|
Published
|
07/02/2023 2:39 AM
|
319 - Perform operational test |
Instruction
|
Published
|
06/10/2021 8:31 PM
|
7734 - Perform procedure DET-RPM-RAP‐RM03 RPM Operational Testing. |
Remark
|
Published
|
07/02/2023 2:43 AM
|
7842 - 7820 - Inform system operators upon completion of maintenance action; Remark: |
Procedure Step
|
Published
|
07/07/2023 11:15 PM
|
7820 - Inform system operators upon completion of maintenance action |
Instruction
|
Published
|
07/07/2023 10:02 PM
|
322 - 321 - Document maintenance actions; Remark: |
Procedure Step
|
Published
|
06/17/2021 9:49 AM
|
321 - Document maintenance actions |
Instruction
|
Published
|
04/21/2025 11:19 AM
|
2869 - 2585 - Document maintenance performed; Remark:2866 |
Procedure Step
|
Published
|
06/18/2021 5:58 PM
|
2585 - Document maintenance performed |
Instruction
|
Published
|
07/09/2021 12:15 PM
|
2866 - Record observations, times, and results for the maintenance report. |
Remark
|
Published
|
04/21/2025 11:39 AM
|
4706 - 4705 - Submit report; Remark:4864 |
Procedure Step
|
Published
|
06/29/2022 9:50 AM
|
4705 - Submit report |
Instruction
|
Published
|
06/29/2022 9:50 AM
|
4864 - As specified by management or contractual obligations. |
Remark
|
Published
|
07/28/2022 1:36 PM
|
Routine Maintenance Procedure |
Document Type
|
Published
|
03/02/2020 3:31 PM
|
Annually |
Frequency
|
Published
|
04/01/2020 11:37 AM
|
As specified by corrective maintenance retest requirements following guidance from the Nuclear Smuggling Detection and Deterrence (NSDD) Program |
Frequency
|
Published
|
07/02/2023 3:28 PM
|
Note |
Annotation Type
|
Published
|
03/02/2020 3:25 PM
|
Standard |
Annotation Type
|
Published
|
03/02/2020 3:25 PM
|
Warning |
Annotation Type
|
Published
|
03/02/2020 3:25 PM
|
Caution |
Annotation Type
|
Published
|
03/02/2020 3:24 PM
|
Technician 1 |
Skill Level
|
Published
|
05/14/2020 2:56 PM
|
Technician 2 |
Skill Level
|
Published
|
04/09/2020 1:17 AM
|
Keys to access RPM |
Replacement Parts
|
Published
|
04/27/2023 6:56 PM
|
Reflective safety vest if required |
Replacement Parts
|
Published
|
06/20/2023 4:47 PM
|
safety glasses |
Replacement Parts
|
Published
|
07/02/2023 4:48 PM
|
Ladder; 2 meters, 6 feet tall; non-conducting if required |
Replacement Parts
|
Published
|
07/02/2023 4:48 PM
|
Traffic control cones or equivalent if required |
Replacement Parts
|
Published
|
06/20/2023 4:47 PM
|
Source-holding device ~ 1.5 m long extending rod or pole |
Replacement Parts
|
Published
|
07/02/2023 4:48 PM
|
Flashlight/headlight is recommended |
Replacement Parts
|
Published
|
07/02/2023 4:48 PM
|
Laptop computer with current baselined functional compliance test, FCT, datasheet |
Replacement Parts
|
Published
|
07/02/2023 4:48 PM
|
Functional Compliance Testting Data |
படம்
|
Published
|
06/30/2023 1:26 PM
|
Source Activity Label |
படம்
|
Published
|
06/30/2023 2:37 PM
|
Certificate of Calibration Source Activity |
படம்
|
Published
|
06/30/2023 2:38 PM
|
Source Calibration Date |
படம்
|
Published
|
06/30/2023 2:43 PM
|
Certificate of Calibration Source Date |
படம்
|
Published
|
06/30/2023 2:44 PM
|
Certificate of Calibration Current Source Activity |
படம்
|
Published
|
06/30/2023 2:51 PM
|
Certificate of Calibration Current Neutron Date |
படம்
|
Published
|
06/30/2023 3:37 PM
|
Certificate of Calibration Current Neutron Current Level |
படம்
|
Published
|
06/30/2023 3:45 PM
|
Disconnect Ethernet cable on top of SC-770 box |
படம்
|
Published
|
10/04/2023 2:59 PM
|
LD-260 Disconnect Switch Power On |
படம்
|
Published
|
05/04/2023 3:37 PM
|
SC-770 Main Menu |
படம்
|
Published
|
07/02/2023 12:29 AM
|
Record Measured Gamma Background |
படம்
|
Published
|
07/02/2023 12:56 AM
|
Gamma Count in Co-57 Source |
படம்
|
Published
|
07/02/2023 1:20 AM
|
Gamma Detector Efficiency |
படம்
|
Published
|
07/02/2023 1:39 AM
|
Neutron Background Count Rate |
படம்
|
Published
|
07/02/2023 1:54 AM
|
Neutron Count Rate |
படம்
|
Published
|
07/02/2023 2:21 AM
|
Neutron Source Efficiency |
படம்
|
Published
|
07/02/2023 2:28 AM
|
Functional Compliance Test Results |
படம்
|
Published
|
07/02/2023 2:43 AM
|