Title | Type | Status | Last Updated |
---|---|---|---|
Thermo FHT1377 PackEye | Device
|
Published
|
10/28/2024 3:37 PM
|
6445 - 6444 - Connect to PackEye; Remark: | Procedure Step
|
Published
|
11/30/2022 1:47 PM
|
6444 - Connect to PackEye | Instruction
|
Published
|
11/30/2022 1:47 PM
|
6451 - 6353 - Connect PackEye to Computer; Remark:6450 | Procedure Step
|
Published
|
11/30/2022 2:06 PM
|
6353 - Connect PackEye to Computer | Instruction
|
Published
|
11/30/2022 10:56 AM
|
6450 - Refer to DET-HHD-THR-PAK-RM03, PackEye Computer Connection. | Remark
|
Published
|
11/30/2022 2:01 PM
|
6452 - 6449 - Click Settings...; Remark:6448 | Procedure Step
|
Published
|
11/30/2022 2:34 PM
|
6449 - Click Settings... | Instruction
|
Published
|
11/30/2022 2:07 PM
|
6448 - IO PackEye Settings Button | Remark
|
Published
|
11/09/2024 9:07 PM
|
6453 - 6447 - Enter Administrator mode; Remark:6446 | Procedure Step
|
Published
|
11/30/2022 2:50 PM
|
6447 - Enter Administrator mode | Instruction
|
Published
|
11/30/2022 2:48 PM
|
6446 - Select Options, then Configurations, then | Remark
|
Published
|
11/30/2022 2:50 PM
|
6822 - 6821 - Adjust High Voltage; Remark: | Procedure Step
|
Published
|
12/19/2022 2:03 PM
|
6821 - Adjust High Voltage | Instruction
|
Published
|
12/19/2022 2:03 PM
|
6832 - 6831 - Select HV Adjustment tab; Remark:6830 | Procedure Step
|
Published
|
12/19/2022 3:35 PM
|
6831 - Select HV Adjustment tab | Instruction
|
Published
|
12/19/2022 3:32 PM
|
6830 - IO PackEye Interface HV Adjustment | Remark
|
Published
|
11/09/2024 9:08 PM
|
6833 - 6829 - Document current gamma high voltage; Remark:6828 | Procedure Step
|
Published
|
12/19/2022 3:36 PM
|
6829 - Document current gamma high voltage | Instruction
|
Published
|
12/19/2022 3:36 PM
|
6828 - IO PackEye Interface Current HV | Remark
|
Published
|
11/09/2024 9:08 PM
|
6834 - 6827 - Select Cs-137; Remark:6826 | Procedure Step
|
Published
|
12/19/2022 3:38 PM
|
6827 - Select Cs-137 | Instruction
|
Published
|
12/19/2022 3:37 PM
|
6826 - IO PackEye Interface Select Cesium CS-137 | Remark
|
Published
|
11/16/2024 8:03 PM
|
6835 - 3828 - Place gamma source on instrument; Remark:6825 | Procedure Step
|
Published
|
12/19/2022 3:39 PM
|
3828 - Place gamma source on instrument | Instruction
|
Published
|
03/28/2022 6:18 PM
|
6825 - Gamma source should be located in center of PackEye for best results. | Remark
|
Published
|
12/19/2022 3:39 PM
|
6836 - 6824 - Click Start HV Setup; Remark:6823 | Procedure Step
|
Published
|
12/19/2022 3:41 PM
|
6824 - Click Start HV Setup | Instruction
|
Published
|
12/19/2022 3:40 PM
|
6823 - IO PackEye Interface Start HV Setup | Remark
|
Published
|
11/16/2024 8:04 PM
|
6844 - 6843 - Wait for adjustment to complete; Remark:6842 | Procedure Step
|
Published
|
12/19/2022 4:37 PM
|
6843 - Wait for adjustment to complete | Instruction
|
Published
|
12/19/2022 4:37 PM
|
6842 - The screen will update and the Status window will be shaded blue. | Remark
|
Published
|
12/19/2022 4:37 PM
|
6845 - 6841 - Click Update HV; Remark:6840 | Procedure Step
|
Published
|
12/19/2022 4:39 PM
|
6841 - Click Update HV | Instruction
|
Published
|
12/19/2022 4:38 PM
|
6840 - The Status window will be shaded green when the process is done. | Remark
|
Published
|
12/19/2022 4:39 PM
|
6846 - 6492 - Click Send; Remark:6839 | Procedure Step
|
Published
|
12/19/2022 4:41 PM
|
6492 - Click Send | Instruction
|
Published
|
12/01/2022 7:46 PM
|
6839 - This will update the high voltage. | Remark
|
Published
|
12/19/2022 4:40 PM
|
6847 - 6838 - Click OK to close Settings window; Remark:6837 | Procedure Step
|
Published
|
12/19/2022 5:10 PM
|
6838 - Click OK to close Settings window | Instruction
|
Published
|
12/19/2022 4:42 PM
|
6837 - IO PackEye Interface OK Close HV | Remark
|
Published
|
11/16/2024 8:10 PM
|
6475 - 6465 - Close FHT 1377 software; Remark:6632 | Procedure Step
|
Published
|
12/01/2022 12:49 PM
|
6465 - Close FHT 1377 software | Instruction
|
Published
|
12/01/2022 12:46 PM
|
6632 - Click File, then Exit. Click Yes | Remark
|
Published
|
12/08/2022 9:55 AM
|
6464 - IO PackEye Exit Menu Interface | Remark
|
Published
|
11/16/2024 8:10 PM
|
6441 - 6440 - Restore Instrument Operation; Remark: | Procedure Step
|
Published
|
11/30/2022 1:48 PM
|
6440 - Restore Instrument Operation | Instruction
|
Published
|
11/30/2022 1:48 PM
|
3869 - 313 - Power off instrument; Remark:3850 | Procedure Step
|
Published
|
03/29/2022 2:23 AM
|
313 - Power off instrument | Instruction
|
Published
|
07/23/2021 2:41 PM
|
3850 - Press and hold the power button for approximately 10 seconds until indicator ... | Remark
|
Published
|
03/29/2022 1:38 AM
|
6457 - 6456 - Disconnect USB/Serial adapter; Remark:6455 | Procedure Step
|
Published
|
11/30/2022 3:05 PM
|
6456 - Disconnect USB/Serial adapter | Instruction
|
Published
|
11/30/2022 3:03 PM
|
6455 - IO PackEye Serial Cable Disconnect | Remark
|
Published
|
11/16/2024 8:11 PM
|
5792 - 5786 - Reconnect RS232 jumper; Remark:5785 | Procedure Step
|
Published
|
11/03/2022 3:42 PM
|
5786 - Reconnect RS232 jumper | Instruction
|
Published
|
11/03/2022 3:39 PM
|
5785 - Use small, slotted screwdriver to secure connector. | Remark
|
Published
|
11/16/2022 9:35 AM
|
5830 - IO PackEye RS232 Connect Cable | Remark
|
Published
|
11/09/2024 8:14 PM
|
6458 - 319 - Perform operational test; Remark:6454 | Procedure Step
|
Published
|
11/30/2022 3:06 PM
|
319 - Perform operational test | Instruction
|
Published
|
06/10/2021 8:31 PM
|
6454 - DET-HHD-THR-PAK-RM01, PackEye Operational Test. | Remark
|
Published
|
11/30/2022 3:06 PM
|
322 - 321 - Document maintenance actions; Remark: | Procedure Step
|
Published
|
06/17/2021 9:49 AM
|
321 - Document maintenance actions | Instruction
|
Published
|
04/21/2025 11:19 AM
|
2869 - 2585 - Document maintenance performed; Remark:2866 | Procedure Step
|
Published
|
06/18/2021 5:58 PM
|
2585 - Document maintenance performed | Instruction
|
Published
|
07/09/2021 12:15 PM
|
2866 - Record observations, times, and results for the maintenance report. | Remark
|
Published
|
04/21/2025 11:39 AM
|
4706 - 4705 - Submit report; Remark:4864 | Procedure Step
|
Published
|
06/29/2022 9:50 AM
|
4705 - Submit report | Instruction
|
Published
|
06/29/2022 9:50 AM
|
4864 - As specified by management or contractual obligations. | Remark
|
Published
|
07/28/2022 1:36 PM
|
Routine Maintenance Procedure | Document Type
|
Published
|
03/02/2020 3:31 PM
|
minutes | Duration
|
Published
|
05/26/2022 11:41 AM
|
Annually | Frequency
|
Published
|
04/01/2020 11:37 AM
|
Directed to do so by maintenance support | Frequency
|
Published
|
12/19/2022 5:55 PM
|
FHT 681 replacement | Frequency
|
Published
|
12/19/2022 5:54 PM
|
Gamma detector replacement | Frequency
|
Published
|
12/19/2022 5:53 PM
|
Instrument is not functioning correctly (nuisance alarms) | Frequency
|
Published
|
12/19/2022 5:53 PM
|
Technician 2 | Skill Level
|
Published
|
04/09/2020 1:17 AM
|
Computer | Tools
|
Published
|
05/09/2020 3:05 PM
|
FHT 1377 software | Tools
|
Published
|
11/30/2022 1:25 PM
|
USB-A male to RS-232 (DB9) male serial adapter (USB/Serial adapter) | Tools
|
Published
|
11/30/2022 1:26 PM
|
packeye_ui_fhtapp_settings.jpg | Image
|
Published
|
10/23/2023 10:14 AM
|
packeye_ui_fhtapp_administrator.jpg | Image
|
Published
|
10/23/2023 1:24 PM
|
packeye_ui_fhtapp_hv_selecttab.jpg | Image
|
Published
|
10/23/2023 1:26 PM
|
packeye_ui_fhtapp_hv_currentvalue.jpg | Image
|
Published
|
10/23/2023 1:27 PM
|
packeye_ui_fhtapp_hv_cs137.jpg | Image
|
Published
|
10/23/2023 1:29 PM
|
packeye_diradsource_placecenter.jpg | Image
|
Published
|
10/23/2023 9:46 AM
|
packeye_ui_config_hv_startsetup.jpg | Image
|
Published
|
10/23/2023 1:30 PM
|
packeye_ui_fhtapp_hv_bluescreen.jpg | Image
|
Published
|
10/23/2023 1:31 PM
|
packeye_ui_fhtapp_hv_greenoptimized.jpg | Image
|
Published
|
10/23/2023 1:33 PM
|
packeye_ui_fhtapp_sendt1031.jpg | Image
|
Published
|
10/23/2023 1:40 PM
|
packeye_ui_fhtapp_hv_okclose.jpg | Image
|
Published
|
10/23/2023 1:34 PM
|
packeye_ui_fhtapp_exit.jpg | Image
|
Published
|
10/23/2023 1:35 PM
|
Power off Packeye | Image
|
Published
|
02/19/2022 1:48 AM
|
packeye_di_serialcable_remove.jpg | Image
|
Published
|
10/23/2023 9:51 AM
|
packeye_di_rs232cable_insert.jpg | Image
|
Published
|
10/23/2023 9:48 AM
|
Conduct high voltage adjustment of the PackEye gamma detector using FHT 1377 software. This procedure can also be performed using the Handheld Assessment and Troubleshooting Software (HATS).
Prior to arrival, notify operators concerning expected work, duration, and anticipated alarms and fault indications.
6445 - 6444 - Connect to PackEye; Remark:
6444 - Connect to PackEye
Connect to PackEye
6451 - 6353 - Connect PackEye to Computer; Remark:6450
6353 - Connect PackEye to Computer
Connect PackEye to Computer
6450 - Refer to DET-HHD-THR-PAK-RM03, PackEye Computer Connection.
Refer to DET-HHD-THR-PAK-RM03, PackEye Computer Connection.
6453 - 6447 - Enter Administrator mode; Remark:6446
6447 - Enter Administrator mode
Enter Administrator mode
6446 - Select Options, then Configurations, then
Select Options, then Configurations, then Administrator.

6822 - 6821 - Adjust High Voltage; Remark:
6821 - Adjust High Voltage
Adjust High Voltage
6832 - 6831 - Select HV Adjustment tab; Remark:6830
6831 - Select HV Adjustment tab
Select HV Adjustment tab
6833 - 6829 - Document current gamma high voltage; Remark:6828
6829 - Document current gamma high voltage
Document current gamma high voltage
6835 - 3828 - Place gamma source on instrument; Remark:6825
3828 - Place gamma source on instrument
Place gamma source on instrument
6825 - Gamma source should be located in center of PackEye for best results.
Gamma source should be located in center of PackEye for best results.

6836 - 6824 - Click Start HV Setup; Remark:6823
6824 - Click Start HV Setup
Click Start HV Setup
6844 - 6843 - Wait for adjustment to complete; Remark:6842
6843 - Wait for adjustment to complete
Wait for adjustment to complete
6842 - The screen will update and the Status window will be shaded blue.
The screen will update and the Status window will be shaded blue.

6845 - 6841 - Click Update HV; Remark:6840
6841 - Click Update HV
Click Update HV
6840 - The Status window will be shaded green when the process is done.
The Status window will be shaded green when the process is done.

6846 - 6492 - Click Send; Remark:6839
6492 - Click Send
Click Send
6839 - This will update the high voltage.
This will update the high voltage.

6847 - 6838 - Click OK to close Settings window; Remark:6837
6838 - Click OK to close Settings window
Click OK to close Settings window
6475 - 6465 - Close FHT 1377 software; Remark:6632
6465 - Close FHT 1377 software
Close FHT 1377 software
6632 - Click File, then Exit. Click Yes
Click File, then Exit. Click Yes on the next window.
6441 - 6440 - Restore Instrument Operation; Remark:
6440 - Restore Instrument Operation
Restore Instrument Operation
3869 - 313 - Power off instrument; Remark:3850
313 - Power off instrument
Power off instrument
3850 - Press and hold the power button for approximately 10 seconds until indicator ...
Press and hold the power button for approximately 10 seconds until indicator lights are off.

6457 - 6456 - Disconnect USB/Serial adapter; Remark:6455
6456 - Disconnect USB/Serial adapter
Disconnect USB/Serial adapter
5792 - 5786 - Reconnect RS232 jumper; Remark:5785
5786 - Reconnect RS232 jumper
Reconnect RS232 jumper
5785 - Use small, slotted screwdriver to secure connector.
Use small, slotted screwdriver to secure connector.
6458 - 319 - Perform operational test; Remark:6454
319 - Perform operational test
Perform operational test
6454 - DET-HHD-THR-PAK-RM01, PackEye Operational Test.
DET-HHD-THR-PAK-RM01, PackEye Operational Test.
322 - 321 - Document maintenance actions; Remark:
321 - Document maintenance actions
Document maintenance actions
2869 - 2585 - Document maintenance performed; Remark:2866
2585 - Document maintenance performed
Document maintenance performed
2866 - Record observations, times, and results for the maintenance report.
Record observations, times, and results for the maintenance report.
4706 - 4705 - Submit report; Remark:4864
4705 - Submit report
Submit report
4864 - As specified by management or contractual obligations.
As specified by management or contractual obligations.