移至主內容
首頁
FRONTLINE Maintenance Tool

導航連結

  1. 首頁

7948 - 7944 - Install new desiccant bag(s) in place of removed bag; Remark:

7975 - Caution : Place bags away from electrical connections to avoid damage if desiccant leak...

Type
Caution

Place bags away from electrical connections to avoid damage if desiccant leaks. Do not place desiccant bag on top of the neutron detector due to leaking CaCl desiccant bags causing issues in the field. A different location needs to be determined

Step 00

7944 - Install new desiccant bag(s) in place of removed bag

Install new desiccant bag(s) in place of removed bag