8807 - Radiation Detection - Gamma |
TS Flowchart
|
Published
|
06/25/2025 1:06 PM
|
9948 - 9945 - Restart RPM - Decision, Notes: 9946 |
Troubleshooting Element
|
Published
|
06/28/2024 12:22 PM
|
9946 - Restart the RPM by cycling the on/off switch on the LD-260 |
Remark
|
Published
|
10/19/2023 8:50 AM
|
8861 - Did this fix the Issue? |
Remark
|
Published
|
09/25/2023 1:22 PM
|
1544 - 546 - Done - Complete, Notes: 547 |
Troubleshooting Element
|
Published
|
10/01/2020 7:23 PM
|
547 - Troubleshooting is complete. |
Remark
|
Published
|
07/09/2021 7:45 PM
|
546 - Done |
Troubleshoot Element Text
|
Published
|
07/09/2021 7:44 PM
|
9944 - 9941 - Measure discriminator voltages - Decision, Notes: 9942 |
Troubleshooting Element
|
Published
|
10/19/2023 8:47 AM
|
9942 - Use a digital multimeter to verify SCA-774 discriminator voltages are n... |
Remark
|
Published
|
10/19/2023 8:47 AM
|
9943 - Are the ULD/LLD voltages correct? |
Remark
|
Published
|
10/19/2023 8:47 AM
|
9940 - 9937 - Measure gamma high voltage - Decision, Notes: 9938 |
Troubleshooting Element
|
Published
|
10/19/2023 8:25 AM
|
9938 - Measure the HV on the MHV cable at both detectors using a digital multi-meter... |
Remark
|
Published
|
10/19/2023 8:24 AM
|
9939 - Is the measured voltage significantly less than 900 VDC? |
Remark
|
Published
|
10/19/2023 8:25 AM
|
11596 - |
Remark
|
Published
|
06/25/2025 12:15 PM
|
9936 - 9933 - Adjust gamma high voltage - Decision, Notes: 9934 |
Troubleshooting Element
|
Published
|
10/19/2023 8:19 AM
|
9934 - Attempt to adjust gamma high voltage on the gamma HV-448/458 board R1 to +900... |
Remark
|
Published
|
10/19/2023 8:17 AM
|
9935 - Is the gamma high voltage now approximately 900VDC? |
Remark
|
Published
|
10/19/2023 8:18 AM
|
9932 - 9930 - Perform an Electronic Alignment - Formal Procedure, Notes: 9007 |
Troubleshooting Element
|
Published
|
10/19/2023 8:15 AM
|
9007 - Perform the following procedure: |
Remark
|
Published
|
09/26/2023 1:23 PM
|
9931 - DET-RPM-RAP-CM01, RPM Electronic Alignment. |
Remark
|
Published
|
10/19/2023 8:15 AM
|
9928 - 8168 - Did this fix the problem? - Decision, Notes: |
Troubleshooting Element
|
Published
|
10/19/2023 8:12 AM
|
9923 - 9921 - Gently wiggle BNC connectors - Decision, Notes: 9922 |
Troubleshooting Element
|
Published
|
10/19/2023 8:10 AM
|
9922 - Gently wiggle the affected channel’s signal cable BNC connector where it is... |
Remark
|
Published
|
10/19/2023 8:10 AM
|
9913 - Did the displayed count rate change? |
Remark
|
Published
|
10/19/2023 8:00 AM
|
9098 - 9096 - Replace SCA-774 Board - Formal Procedure, Notes: 9007 |
Troubleshooting Element
|
Published
|
10/14/2024 11:52 AM
|
9097 - DET-RPM-RAP-CM07, SCA-774 RPM Single Channel Analyzer Board Replacement. ... |
Remark
|
Published
|
09/27/2023 12:42 PM
|
9096 - Replace SCA-774 Board |
Troubleshoot Element Text
|
Published
|
09/27/2023 12:40 PM
|
9914 - 9910 - Gently wiggle signal cable - Decision, Notes: 9911 |
Troubleshooting Element
|
Published
|
10/19/2023 8:00 AM
|
9911 - Gently wiggle the affected channel’s signal cable above the connector where... |
Remark
|
Published
|
10/19/2023 7:59 AM
|
9912 - Gently wiggle the affected channel’s signal cable above the connector where... |
Remark
|
Published
|
10/19/2023 7:59 AM
|
9892 - 9851 - Replace affected signal cable - Decision, Notes: 9852 |
Troubleshooting Element
|
Published
|
10/19/2023 7:51 AM
|
9852 - Replace the signal cable on the affected channel |
Remark
|
Published
|
10/19/2023 7:15 AM
|
9889 - 9884 - Swap signal cables at SCA-774 board - Decision, Notes: 9885 |
Troubleshooting Element
|
Published
|
10/19/2023 7:46 AM
|
9885 - Navigate to SHOW COUNTS on the SC-770 |
Remark
|
Published
|
10/19/2023 7:44 AM
|
9886 - Navigate to SHOW COUNTS on the SC-770 |
Remark
|
Published
|
10/19/2023 7:45 AM
|
9887 - While watching SHOW COUNTS, switch the BNC connectors for gamma CH1 and CH2 a... |
Remark
|
Published
|
10/19/2023 7:46 AM
|
9888 - Did the fault indication move from the original channel to the other? |
Remark
|
Published
|
10/19/2023 7:46 AM
|
9864 - 9861 - Replace VD-580 - Formal Procedure, Notes: 9862 |
Troubleshooting Element
|
Published
|
10/19/2023 7:29 AM
|
9862 - Perform the following procedure on the affected channel: |
Remark
|
Published
|
10/19/2023 7:28 AM
|
9863 - DET-RPM-RAP-CM03, RPM VD-580 Replacement |
Remark
|
Published
|
10/19/2023 7:29 AM
|
9858 - 8168 - Did this fix the problem? - Decision, Notes: |
Troubleshooting Element
|
Published
|
10/19/2023 7:22 AM
|
9853 - 9851 - Replace affected signal cable - Decision, Notes: 9852 |
Troubleshooting Element
|
Published
|
10/19/2023 7:15 AM
|
9848 - 9846 - GH or GL? - Decision, Notes: 9847 |
Troubleshooting Element
|
Published
|
10/19/2023 7:10 AM
|
9847 - Is the fault indication on the affected channel HI or LO? |
Remark
|
Published
|
10/19/2023 7:10 AM
|
9845 - 9843 - Light Leak? - Decision, Notes: 9844 |
Troubleshooting Element
|
Published
|
10/19/2023 7:05 AM
|
9844 - It is unlikely that the gamma detector would need to be replaced. Insp... |
Remark
|
Published
|
10/19/2023 7:04 AM
|
9842 - 9840 - Replace gamma detector - Formal Procedure, Notes: 9007 |
Troubleshooting Element
|
Published
|
10/19/2023 7:00 AM
|
9841 - DET-RPM-RAP-CM22, Gamma Detector Assembly Replacement. |
Remark
|
Published
|
10/19/2023 7:00 AM
|
8823 - 8168 - Did this fix the problem? - Decision, Notes: |
Troubleshooting Element
|
Published
|
09/25/2023 12:14 PM
|
5882 - 545 - Contact Technical Support - Incomplete, Notes: 5881 |
Troubleshooting Element
|
Published
|
11/07/2022 10:02 AM
|
5881 - Troubleshooting incomplete. Contact technical support. Provide the ... |
Remark
|
Published
|
11/07/2022 9:49 AM
|
545 - Contact Technical Support |
Troubleshoot Element Text
|
Published
|
04/05/2021 12:00 PM
|
8168 - Did this fix the problem? |
Troubleshoot Element Text
|
Published
|
08/25/2023 11:59 AM
|
9840 - Replace gamma detector |
Troubleshoot Element Text
|
Published
|
10/19/2023 6:57 AM
|
9843 - Light Leak? |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:02 AM
|
9846 - GH or GL? |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:09 AM
|
9851 - Replace affected signal cable |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:13 AM
|
9861 - Replace VD-580 |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:25 AM
|
9874 - 9869 - Swap SC-770 controller connectors - Decision, Notes: 7236 |
Troubleshooting Element
|
Published
|
10/19/2023 7:40 AM
|
7236 - Move LD-260 switch to OFF position. The PWR ON |
Remark
|
Published
|
04/27/2023 4:25 PM
|
9870 - Switch the positions of the large 19-pin connectors on the bottom of the SC-7... |
Remark
|
Published
|
10/19/2023 7:37 AM
|
7367 - Move LD-260 switch to ON position. The PWR ON |
Remark
|
Published
|
05/11/2023 12:42 PM
|
9871 - Wait for controller to complete POST and obtain initial background |
Remark
|
Published
|
10/19/2023 7:39 AM
|
9872 - Did the fault indication move from the original pillar (for example: DET 1/2 ... |
Remark
|
Published
|
10/19/2023 7:40 AM
|
9095 - 9093 - Replace SC-771 Controller Board - Formal Procedure, Notes: 9007 |
Troubleshooting Element
|
Published
|
09/27/2023 12:36 PM
|
9094 - DET-RPM-RAP-CM09, RPM SC-771 Controller Board Replacement. |
Remark
|
Published
|
09/27/2023 12:36 PM
|
9093 - Replace SC-771 Controller Board |
Troubleshoot Element Text
|
Published
|
09/27/2023 12:34 PM
|
9869 - Swap SC-770 controller connectors |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:32 AM
|
9884 - Swap signal cables at SCA-774 board |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:42 AM
|
9910 - Gently wiggle signal cable |
Troubleshoot Element Text
|
Published
|
10/19/2023 8:00 AM
|
9921 - Gently wiggle BNC connectors |
Troubleshoot Element Text
|
Published
|
10/19/2023 8:09 AM
|
9930 - Perform an Electronic Alignment |
Troubleshoot Element Text
|
Published
|
10/19/2023 8:14 AM
|
9857 - 9855 - Replace High Voltage Board - Formal Procedure, Notes: 9007 |
Troubleshooting Element
|
Published
|
10/19/2023 7:18 AM
|
9856 - DET-RPM-RAP-CM25, RPM HHV-448/458 High Voltage Board Replacement. |
Remark
|
Published
|
10/19/2023 7:18 AM
|
9855 - Replace High Voltage Board |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:16 AM
|
9933 - Adjust gamma high voltage |
Troubleshoot Element Text
|
Published
|
10/19/2023 8:16 AM
|
9937 - Measure gamma high voltage |
Troubleshoot Element Text
|
Published
|
10/19/2023 8:23 AM
|
9905 - 9898 - Reset ULD/LLD - Decision, Notes: 9899 |
Troubleshooting Element
|
Published
|
10/19/2023 7:57 AM
|
9899 - Navigate to DSCR. ADJUST on the SC-770 |
Remark
|
Published
|
10/19/2023 7:54 AM
|
9900 - Press #Pre... |
Remark
|
Published
|
10/19/2023 7:54 AM
|
9898 - Reset ULD/LLD |
Troubleshoot Element Text
|
Published
|
10/19/2023 7:53 AM
|
9941 - Measure discriminator voltages |
Troubleshoot Element Text
|
Published
|
10/19/2023 8:27 AM
|
9945 - Restart RPM |
Troubleshoot Element Text
|
Published
|
06/28/2024 11:59 AM
|
11285 - Local gamma alarm indication on every occupancy |
Troubleshoot Symptom
|
Published
|
10/14/2024 10:53 AM
|
11284 - Gamma alarm rate higher than normal |
Troubleshoot Symptom
|
Published
|
10/14/2024 10:54 AM
|
Complete |
Troubleshoot Step Type
|
Published
|
04/10/2020 12:12 AM
|
Formal Procedure |
Troubleshoot Step Type
|
Published
|
04/10/2020 12:12 AM
|
Incomplete |
Troubleshoot Step Type
|
Published
|
04/10/2020 12:13 AM
|
Decision |
Troubleshoot Step Type
|
Published
|
04/10/2020 12:12 AM
|
LD-260 Disconnect Switch Power Off |
படம்
|
Published
|
04/24/2023 12:49 PM
|
LD-260 Disconnect Switch Power On |
படம்
|
Published
|
06/04/2025 1:28 PM
|