DET-RPM-RAP-FC12
10083 - 10082 - What is on the SC-770 controller display? - Decision, Notes:
9948 - 9945 - Restart RPM - Decision, Notes: 9946
9947 - Reminder : Start troubleshooting with the control pillar. If both gamma channels s...
Type
Reminder
Start troubleshooting with the control pillar. If both gamma channels show the same issue, check shared parts. If different, troubleshoot separately
9945 - Restart RPM
Restart RPM
Yes
10000 - 10651 - Restart RPM - Decision, Notes: 9946
9947 - Reminder : Start troubleshooting with the control pillar. If both gamma channels s...
Type
Reminder
Start troubleshooting with the control pillar. If both gamma channels show the same issue, check shared parts. If different, troubleshoot separately
9999 - Note : If there are no or low neutron counts on a channel, place a source on the aff...
Type
Note
If there are no or low neutron counts on a channel, place a source on the affected detector before performing troubleshooting steps
10651 - Restart RPM
Restart RPM
Yes
10081 - 10079 - Alarm and Fault Indications Constantly on? - Decision, Notes: 10080
10079 - Alarm and Fault Indications Constantly on?
Alarm and Fault Indications Constantly on?
9944 - 9941 - Measure discriminator voltages - Decision, Notes: 9942
9941 - Measure discriminator voltages
Measure discriminator voltages
9998 - 9941 - Measure discriminator voltages - Decision, Notes: 9997
9941 - Measure discriminator voltages
Measure discriminator voltages
9203 - 9199 - Push firmly on all cabinet doors - Process, Notes: 9200
9202 - Note : A cabinet tamper fault condition is often caused by a loose cabinet door latc...
Type
Note
A cabinet tamper fault condition is often caused by a loose cabinet door latch.
9199 - Push firmly on all cabinet doors
Push firmly on all cabinet doors
Flowchart Title
Amber Fault LED Illuminated