10000 - 10651 - Restart RPM - Decision, Notes: 9946
9947 - Reminder : Start troubleshooting with the control pillar. If both gamma channels s...
Start troubleshooting with the control pillar. If both gamma channels show the same issue, check shared parts. If different, troubleshoot separately
9999 - Note : If there are no or low neutron counts on a channel, place a source on the aff...
If there are no or low neutron counts on a channel, place a source on the affected detector before performing troubleshooting steps
10651 - Restart RPM
9998 - 9941 - Measure discriminator voltages - Decision, Notes: 9997
9996 - 9921 - Gently wiggle BNC connectors - Decision, Notes: 9922
9921 - Gently wiggle BNC connectors
9995 - 9910 - Gently wiggle signal cable - Decision, Notes: 9911
9910 - Gently wiggle signal cable
9992 - 9990 - Measure PMFX Box input high voltage - Decision, Notes: 9991
9990 - Measure PMFX Box input high voltage
9976 - 9973 - Measure PMFX Box output high voltages - Decision, Notes: 9974
9973 - Measure PMFX Box output high voltages
9970 - 8168 - Did this fix the problem? - Decision, Notes:
9969 - 9884 - Swap signal cables at SCA-774 board - Decision, Notes: 9965
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9884 - Swap signal cables at SCA-774 board
9958 - 9869 - Swap SC-770 controller connectors - Decision, Notes: 7236
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9869 - Swap SC-770 controller connectors
9989 - 9988 - Adjust neutron high voltage - Decision, Notes: 9985
9988 - Adjust neutron high voltage
9987 - 9983 - Replace high voltage cable - Decision, Notes: 9984
9983 - Replace high voltage cable
9982 - 9978 - Perform Functional Compliance Test - Formal Procedure, Notes: 9007
9980 - Note : Only the neutron portion of the FCT is required
Only the neutron portion of the FCT is required
9978 - Perform Functional Compliance Test
9981 - 8168 - Did this fix the problem? - Decision, Notes:
9967 - 9964 - Swap cables at bottom of PMFX Box - Decision, Notes: 9885
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9964 - Swap cables at bottom of PMFX Box
9963 - 9898 - Reset ULD/LLD - Decision, Notes: 9899
9961 - 9959 - Replace MHV cable - Decision, Notes: 9960
9955 - Note : Helium-3 tube failures are extremely rare.
Helium-3 tube failures are extremely rare.
9959 - Replace MHV cable
9098 - 9096 - Replace SCA-774 Board - Formal Procedure, Notes: 9007
9095 - 9093 - Replace SC-771 Controller Board - Formal Procedure, Notes: 9007
9952 - 9949 - Replace Helium-3 tube - Formal Procedure, Notes: 9950
9955 - Note : Helium-3 tube failures are extremely rare.
Helium-3 tube failures are extremely rare.