Skip to main content
Home
FRONTLINE Maintenance Tool
  • Structures
  • Management
    • Content Management
  • Exports
  • Add Content
  • Content Views

Breadcrumb

  1. Home

9987 - 9983 - Replace high voltage cable - Decision, Notes: 9984

9983 - Replace high voltage cable

Replace high voltage cable
Yes
Go to
9982 - 9978 - Perform Functional Compliance Test - Formal Procedure, Notes: 9007
No
Go to
9971 - 9855 - Replace High Voltage Board - Formal Procedure, Notes: 9007
Note
9984 - Replace the high voltage cable between the SCA-775 module and the PMFX Box
9985 - Attempt to adjust neutron high voltage on the neutron HV-448/458 board R1 to ...
9986 - Is the neutron high voltage between +1365 VDC and +1385 VDC?