  1. Home

9940 - 9937 - Measure gamma high voltage - Decision, Notes: 9938

9937 - Measure gamma high voltage

Measure gamma high voltage
Yes
Go to
9936 - 9933 - Adjust gamma high voltage - Decision, Notes: 9934
No
Go to
9923 - 9921 - Gently wiggle BNC connectors - Decision, Notes: 9922
Note
9938 - Measure the HV on the MHV cable at both detectors using a digital multi-meter...
9939 - Is the measured voltage significantly less than 900 VDC?