5882 - 545 - Contact Technical Support - Incomplete, Notes: 5881
8823 - 8168 - Did this fix the problem? - Decision, Notes:
9952 - 9949 - Replace Helium-3 tube - Formal Procedure, Notes: 9950
9955 - Note : Helium-3 tube failures are extremely rare.
Helium-3 tube failures are extremely rare.
9949 - Replace Helium-3 tube
9095 - 9093 - Replace SC-771 Controller Board - Formal Procedure, Notes: 9007
9093 - Replace SC-771 Controller Board
9958 - 9869 - Swap SC-770 controller connectors - Decision, Notes: 7236
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9869 - Swap SC-770 controller connectors
9961 - 9959 - Replace MHV cable - Decision, Notes: 9960
9955 - Note : Helium-3 tube failures are extremely rare.
Helium-3 tube failures are extremely rare.
9959 - Replace MHV cable
9963 - 9898 - Reset ULD/LLD - Decision, Notes: 9899
9967 - 9964 - Swap cables at bottom of PMFX Box - Decision, Notes: 9885
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9964 - Swap cables at bottom of PMFX Box
9969 - 9884 - Swap signal cables at SCA-774 board - Decision, Notes: 9885
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9884 - Swap signal cables at SCA-774 board
9970 - 8168 - Did this fix the problem? - Decision, Notes:
9976 - 9973 - Measure PMFX Box output high voltages - Decision, Notes: 9974
9973 - Measure PMFX Box output high voltages
9981 - 8168 - Did this fix the problem? - Decision, Notes:
9982 - 9978 - Perform Functional Compliance Test - Formal Procedure, Notes: 9007
9980 - Note : Only the neutron portion of the FCT is required
Only the neutron portion of the FCT is required
9978 - Perform Functional Compliance Test
9987 - 9983 - Replace high voltage cable - Decision, Notes: 9984
9983 - Replace high voltage cable
9989 - 9988 - Adjust neutron high voltage - Decision, Notes: 9985
9988 - Adjust neutron high voltage
9992 - 9990 - Measure PMFX Box input high voltage - Decision, Notes: 9991
9990 - Measure PMFX Box input high voltage
9993 - 9851 - Replace affected signal cable - Decision, Notes: 9852
9851 - Replace affected signal cable
9995 - 9910 - Gently wiggle signal cable - Decision, Notes: 9911
9910 - Gently wiggle signal cable
9996 - 9921 - Gently wiggle BNC connectors - Decision, Notes: 9922
9921 - Gently wiggle BNC connectors
9998 - 9941 - Measure discriminator voltages - Decision, Notes: 9997
9941 - Measure discriminator voltages
10000 - 10651 - Restart RPM - Decision, Notes: 9946
9947 - Reminder : There are a limited number of faults that can affect both pillars; troublesho...
There are a limited number of faults that can affect both pillars; troubleshoot each pillar independently starting with the control pillar. Failure of components that are common to both gamma detector channels should affect both channels similarly. If the problems in the channels are different (GH in one and GL in the other), troubleshoot the gamma channels separately.
9999 - Note : If there are no or low neutron counts on a channel, place a source on the aff...
If there are no or low neutron counts on a channel, place a source on the affected detector before performing troubleshooting steps