DET-RPM-RAP-FC05
5882 - 545 - Contact Technical Support - Incomplete, Notes: 5881
8823 - 8168 - Did this fix the problem? - Decision, Notes:
9095 - 9093 - Replace SC-771 Controller Board - Formal Procedure, Notes: 9007
9093 - Replace SC-771 Controller Board
Replace SC-771 Controller Board
9853 - 9851 - Replace affected signal cable - Decision, Notes: 9852
9851 - Replace affected signal cable
Replace affected signal cable
Yes
9858 - 8168 - Did this fix the problem? - Decision, Notes:
9874 - 9869 - Swap SC-770 controller connectors - Decision, Notes: 7236
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Type
Note
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9869 - Swap SC-770 controller connectors
Swap SC-770 controller connectors
Note
9889 - 9884 - Swap signal cables at SCA-774 board - Decision, Notes: 9885
9873 - Note : Be sure to return all cables and connectors to their correct positions at the...
Type
Note
Be sure to return all cables and connectors to their correct positions at the end of troubleshooting and repair effort
9884 - Swap signal cables at SCA-774 board
Swap signal cables at SCA-774 board
9905 - 9898 - Reset ULD/LLD - Decision, Notes: 9899
9914 - 9910 - Gently wiggle signal cable - Decision, Notes: 9911
9910 - Gently wiggle signal cable
Gently wiggle signal cable
9923 - 9921 - Gently wiggle BNC connectors - Decision, Notes: 9922
9921 - Gently wiggle BNC connectors
Gently wiggle BNC connectors
9928 - 8168 - Did this fix the problem? - Decision, Notes:
9936 - 9933 - Adjust gamma high voltage - Decision, Notes: 9934
9933 - Adjust gamma high voltage
Adjust gamma high voltage
9940 - 9937 - Measure gamma high voltage - Decision, Notes: 9938
9937 - Measure gamma high voltage
Measure gamma high voltage
9944 - 9941 - Measure discriminator voltages - Decision, Notes: 9041
9941 - Measure discriminator voltages
Measure discriminator voltages
9948 - 9945 - Restart RPM - Decision, Notes: 9946
9947 - Reminder : There are a limited number of faults that can affect both pillars; troublesho...
Type
Reminder
There are a limited number of faults that can affect both pillars; troubleshoot each pillar independently starting with the control pillar. Failure of components that are common to both gamma detector channels should affect both channels similarly. If the problems in the channels are different (GH in one and GL in the other), troubleshoot the gamma channels separately.
9945 - Restart RPM
Restart RPM
Yes
Flowchart Title
Radiation Detection - Gamma