DET-HHD-THR-PAK-FC06
7036 - 537 - Start - Begin, Notes: 558
537 - Start
Start
Step Reference Number
01
Note
7033 - 6870 - Fault resolved? - Decision, Notes: 7032
7031 - 7030 - Reseat Neutron tube connections - Process, Notes: 6980
6978 - Warning : Power off instrument prior to disconnecting neutron detector.
Type
Warning
Power off instrument prior to disconnecting neutron detector.
6979 - Note : Access to the connectors is possible from the top of the backpack without rem...
Type
Note
Access to the connectors is possible from the top of the backpack without removing the foam assembly. Picture is shown for clarification only.
7030 - Reseat Neutron tube connections
Reseat Neutron tube connections
Step Reference Number
04
7029 - 6870 - Fault resolved? - Decision, Notes: 7028
7022 - 7021 - Both detectors? - Decision, Notes: 7020
7019 - 7012 - Swap tubes - Process, Notes: 7018
6978 - Warning : Power off instrument prior to disconnecting neutron detector.
Type
Warning
Power off instrument prior to disconnecting neutron detector.
7012 - Swap tubes
Swap tubes
Step Reference Number
09
7017 - 7015 - Changed channels? - Decision, Notes: 7016
7014 - 7012 - Swap tubes - Process, Notes: 7011
6978 - Warning : Power off instrument prior to disconnecting neutron detector.
Type
Warning
Power off instrument prior to disconnecting neutron detector.
7012 - Swap tubes
Swap tubes
Step Reference Number
11
7013 - 7012 - Swap tubes - Process, Notes: 7011
6978 - Warning : Power off instrument prior to disconnecting neutron detector.
Type
Warning
Power off instrument prior to disconnecting neutron detector.
7012 - Swap tubes
Swap tubes
Step Reference Number
12
7010 - 6948 - Check tubes - Process, Notes: 6947
6948 - Check tubes
Check tubes
Step Reference Number
13
7009 - 545 - Contact Technical Support - Incomplete, Notes: 7008
7007 - 545 - Contact Technical Support - Incomplete, Notes: 7006
Flowchart Title
PackEye Low Neutron