7003 - 537 - Start - Begin, Notes: 558
537 - Start
7002 - 5953 - Power off instrument - Process, Notes: 5939
5953 - Power off instrument
7001 - 6937 - Disconnect cables - Process, Notes: 7000
6937 - Disconnect cables
6999 - 6934 - Power on instrument - Process, Notes: 6935
6934 - Power on instrument
6998 - 6966 - False Alarms? - Decision, Notes: 6996
6997 - 6994 - Black connector? - Decision, Notes: 6995
6993 - 6990 - Disconnect headphone connector - Process, Notes: 6992
6991 - Note : The black plastic headphone connectors can cause false alarms. It is highly r...
The black plastic headphone connectors can cause false alarms. It is highly recommended to order a new replacement headphone connector (Manufacturer part # 425506102 Headphone cable EMV PackEye FHT1377). Contact technical support for assistance.
6990 - Disconnect headphone connector
6989 - 6966 - False Alarms? - Decision, Notes: 6988
6987 - 6985 - Shield cable - Process, Notes: 6986
6985 - Shield cable
6984 - 6883 - Tighten screws - Process, Notes: 6983
6883 - Tighten screws
6982 - 6966 - False Alarms? - Decision, Notes: 6965
6981 - 6977 - Reseat detector connections - Process, Notes: 6980
6978 - Warning : Power off instrument prior to disconnecting neutron detector.
Power off instrument prior to disconnecting neutron detector.
6979 - Note : Access to the connectors is possible from the top of the backpack without rem...
Access to the connectors is possible from the top of the backpack without removing the foam assembly. Picture is shown for clarification only.
6977 - Reseat detector connections
6976 - 6966 - False Alarms? - Decision, Notes: 6965
6975 - 6974 - Connectors Loose? - Decision, Notes: 6972
6973 - Warning : Power off instrument prior to checking relief connections.
Power off instrument prior to checking relief connections.
6974 - Connectors Loose?
6971 - 6968 - FHT 681 Connections - Formal Procedure, Notes: 6970
6969 - Caution : Checking connections inside the FHT 681 amplifier and tightening the strain r...
Checking connections inside the FHT 681 amplifier and tightening the strain relief connectors is an advanced procedure and should only be performed by an experienced electronics technician with the necessary tools. Contact expert support if needed.