Sensor Tech Handheld Radiation Monitor
Device
Published
10/28/2024 3:52 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
4316 - 220 - Power on instrument; Remark:4315
Procedure Step
Published
06/06/2022 5:58 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
4315 - Set instrument to AUDIO mode.
Remark
Published
06/06/2022 5:58 PM
4290 - 612 - Wait for instrument to complete start-up sequence; Remark:4286
Procedure Step
Published
04/13/2025 6:50 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
4286 - The unit is ready for operation when the period flashes every 20 seconds....
Remark
Published
06/06/2022 4:11 PM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
4312 - 235 - Place instrument next to radioactive gamma source; Remark:4884
Procedure Step
Published
06/06/2022 6:00 PM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
4884 - IO HRM Move Closer to Rad Source
Remark
Published
08/04/2022 1:44 PM
4313 - 2936 - Verify audible alarm functions correctly; Remark:4310
Procedure Step
Published
06/06/2022 6:02 PM
2936 - Verify audible alarm functions correctly
Instruction
Published
06/15/2021 4:09 PM
4310 - The instrument should respond as follows: Instrument will e...
Remark
Published
06/06/2022 6:01 PM
4311 - Numeric value will appear in the gamma LED [light-emitting diode] dis...
Remark
Published
06/06/2022 6:02 PM
4314 - 241 - Move instrument away from radioactive gamma source; Remark:4883
Procedure Step
Published
06/06/2022 6:03 PM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
4883 - IO HRM Move Away from Rad Source
Remark
Published
08/04/2022 1:47 PM
2914 - 2912 - Test vibration alarm; Remark:
Procedure Step
Published
06/06/2022 5:28 PM
2912 - Test vibration alarm
Instruction
Published
06/15/2021 4:11 PM
4320 - 4319 - Change to VIBRATE mode; Remark:4318
Procedure Step
Published
06/06/2022 6:08 PM
4319 - Change to VIBRATE mode
Instruction
Published
06/06/2022 6:07 PM
4318 - Quickly move switch to VIBRATE.
Remark
Published
06/06/2022 6:08 PM
4323 - 2938 - Verify vibration alarm functions correctly; Remark:4321
Procedure Step
Published
06/06/2022 6:12 PM
2938 - Verify vibration alarm functions correctly
Instruction
Published
06/15/2021 4:12 PM
4321 - The instrument should respond as follows: Instrument will v...
Remark
Published
06/06/2022 6:11 PM
4322 - Numeric value will appear in the gamma LED display.
Remark
Published
06/06/2022 6:11 PM
4329 - 4328 - Check response with neutron radioactive source; Remark:
Procedure Step
Published
06/06/2022 6:21 PM
4332 - Note : The neutron response can be performed using a neutron radioactive source or b...
Annotation
Published
06/06/2022 6:37 PM
4328 - Check response with neutron radioactive source
Instruction
Published
06/06/2022 6:21 PM
4326 - 3095 - Place instrument next to radioactive neutron source; Remark:4884
Procedure Step
Published
06/06/2022 6:31 PM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
4325 - 3097 - Verify instrument responds correctly; Remark:4310
Procedure Step
Published
06/06/2022 6:24 PM
3096 - Note : Most radioactive neutron sources also have a gamma component, which will likely ...
Annotation
Published
07/23/2021 9:57 AM
3097 - Verify instrument responds correctly
Instruction
Published
07/16/2021 10:50 AM
4324 - Numeric value will appear in the neutron LED display. Blue...
Remark
Published
06/06/2022 6:24 PM
4327 - 3098 - Move instrument away from radioactive neutron source; Remark:4883
Procedure Step
Published
06/06/2022 6:32 PM
3098 - Move instrument away from radioactive neutron source
Instruction
Published
07/16/2021 10:54 AM
4338 - 3095 - Place instrument next to radioactive neutron source; Remark:4884
Procedure Step
Published
06/06/2022 6:31 PM
4331 - 3097 - Verify instrument responds correctly; Remark:4321
Procedure Step
Published
06/06/2022 6:51 PM
4330 - Numeric value will appear in the neutron LED display. Blue...
Remark
Published
06/06/2022 6:51 PM
4339 - 3098 - Move instrument away from radioactive neutron source; Remark:4883
Procedure Step
Published
06/06/2022 6:32 PM
4337 - 4335 - Check response with background neutron radiation; Remark:
Procedure Step
Published
06/06/2022 7:01 PM
4336 - Note : If no neutron radioactive source is available, the following step can be used...
Annotation
Published
06/06/2022 7:01 PM
4335 - Check response with background neutron radiation
Instruction
Published
06/06/2022 7:01 PM
4344 - 4343 - Monitor neutron LED; Remark:4342
Procedure Step
Published
06/06/2022 7:11 PM
4343 - Monitor neutron LED
Instruction
Published
06/06/2022 7:10 PM
4342 - Watch neutron LED carefully in a low-light environment for at least 5 minutes...
Remark
Published
06/06/2022 7:11 PM
4345 - 4341 - Verify instrument responds to background neutron radiation; Remark:4340
Procedure Step
Published
06/06/2022 7:13 PM
4341 - Verify instrument responds to background neutron radiation
Instruction
Published
06/06/2022 7:11 PM
4340 - In natural background, the instrument should respond about one (1) to three (...
Remark
Published
06/06/2022 7:13 PM
4349 - 4343 - Monitor neutron LED; Remark:4342
Procedure Step
Published
06/06/2022 7:11 PM
4350 - 4341 - Verify instrument responds to background neutron radiation; Remark:4348
Procedure Step
Published
06/06/2022 7:17 PM
4348 - In natural background, the instrument should respond about one (1) to three (...
Remark
Published
06/06/2022 7:16 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
HRM Switch in AUDIO position
Image
Published
08/02/2022 5:03 PM
HRM Red LED Dot Period
Image
Published
08/02/2022 4:53 PM
HRM Move to Rad Source Trefoil
Image
Published
08/02/2022 7:57 PM
HRM Alert Audio
Image
Published
08/02/2022 7:51 PM
HRM Red LED Display 3
Image
Published
08/02/2022 7:47 PM
HRM Move away from Rad Source Trefoil
Image
Published
08/02/2022 7:59 PM
HRM Switch in VIBRATE position
Image
Published
08/02/2022 5:06 PM
HRM Alert Vibrate
Image
Published
08/02/2022 7:52 PM
HRM Alert Audio Neutron Full Size
Image
Published
08/04/2022 4:58 PM
HRM Alert Blue LED Neutron Level 3
Image
Published
08/03/2022 10:29 AM
HRM Alert Vibrate Neutron Full Size
Image
Published
08/04/2022 5:00 PM
HRM Alert Audio Neutron
Image
Published
08/03/2022 10:32 AM
HRM Alert Vibrate Neutron
Image
Published
08/03/2022 10:34 AM
DET-HHD-STE-HRM-RM01

Check the handheld radiation monitor (HRM) operation and response to a source of radiation. The source of radiation may be any material that causes an alarm, including consumer products or naturally occurring radioactive materials.

Equipment Hierarchy
Sensor Tech Handheld Radiation Monitor
Frequency
Before each use
Daily
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
minutes
Length of time
5
Tools Required
None
Materials Required
Cs-137 activity 37 kBq (1-10 μCi) or gamma radioactive source with activity high enough to create alarm
Cf-252 activity 37 kBq (1-10 μCi)
Originators
Demboski, Erica
Tucker, Brian
Originating Organization
PNNL
Approver
Simons, Brian
Safety Concerns
Radioactive Source
Final Documentation
None
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration.

Date of Origin
Version
1.00
Step 4
Main Step

4325 - 3097 - Verify instrument responds correctly; Remark:4310

Step 00

4331 - 3097 - Verify instrument responds correctly; Remark:4321

Step 00
Step 5
Main Step

4345 - 4341 - Verify instrument responds to background neutron radiation; Remark:4340

Step 00

4350 - 4341 - Verify instrument responds to background neutron radiation; Remark:4348

Step 00
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00