Last Updated: 04/25/2025 1:35 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Rapiscan PRM-470 CG
Device
Published
10/28/2024 3:40 PM
Rapiscan PRM-470 CGN
Device
Published
10/28/2024 3:33 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
3092 - 3091 - Disconnect from external power; Remark:
Procedure Step
Published
07/29/2021 11:05 AM
3091 - Disconnect from external power
Instruction
Published
07/23/2021 10:01 AM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
250 - 220 - Power on instrument; Remark:228
Procedure Step
Published
07/29/2021 11:06 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
228 - Press POWER button until the display turns on.
Remark
Published
06/18/2021 6:29 PM
3107 - 229 - Press LIGHT button to brighten display (if needed); Remark:3106
Procedure Step
Published
07/29/2021 11:08 AM
229 - Press LIGHT button to brighten display (if needed)
Instruction
Published
07/29/2021 11:08 AM
3106 - Light may provide better visibility for reading the display...
Remark
Published
07/23/2021 9:59 AM
231 - IO PRM-470 Light Button Red Box
Remark
Published
11/04/2024 9:49 PM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
333 - 233 - Verify instrument is in SEARCH MODE; Remark:234
Procedure Step
Published
06/25/2021 12:41 PM
233 - Verify instrument is in SEARCH MODE
Instruction
Published
06/10/2021 8:46 PM
234 - Use S/F button to change between SEARCH MODE and FIND MODE.
Remark
Published
07/29/2021 11:11 AM
751 - IO PRM-470 Search Mode 65 CPS Gamma
Remark
Published
10/29/2024 9:31 PM
3108 - S/F mode key red box
Remark
Published
11/02/2024 10:09 PM
334 - 235 - Place instrument next to radioactive gamma source; Remark:535
Procedure Step
Published
07/20/2021 11:10 AM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
535 - IO Move PRM-470 approaching source in SEARCH mode
Remark
Published
11/05/2024 9:35 PM
335 - 237 - Verify SEARCH MODE responds correctly; Remark:243
Procedure Step
Published
07/29/2021 4:14 PM
237 - Verify SEARCH MODE responds correctly
Instruction
Published
07/29/2021 3:47 PM
243 - The instrument should respond as follows: Gamma counts per ...
Remark
Published
10/06/2022 12:31 PM
340 - 241 - Move instrument away from radioactive gamma source; Remark:332
Procedure Step
Published
07/20/2021 11:11 AM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
332 - IO Move PRM-470 away from source Alarm SEARCH
Remark
Published
11/05/2024 9:36 PM
342 - 239 - Set instrument to FIND MODE; Remark:234
Procedure Step
Published
07/20/2021 11:12 AM
239 - Set instrument to FIND MODE
Instruction
Published
07/16/2021 10:43 AM
3109 - Find Mode
Remark
Published
11/05/2024 9:39 PM
336 - 235 - Place instrument next to radioactive gamma source; Remark:536
Procedure Step
Published
07/20/2021 11:51 AM
536 - IO Move PRM-470 approaching source in FIND mode
Remark
Published
11/05/2024 9:38 PM
337 - 240 - Verify FIND MODE responds correctly; Remark:245
Procedure Step
Published
07/20/2021 11:51 AM
240 - Verify FIND MODE responds correctly
Instruction
Published
07/16/2021 10:44 AM
245 - The instrument should respond as follows: Gamma counts per ...
Remark
Published
10/06/2022 12:32 PM
338 - 241 - Move instrument away from radioactive gamma source; Remark:339
Procedure Step
Published
07/20/2021 11:53 AM
339 - IO Move PRM-470 away source Alarm FIND
Remark
Published
11/05/2024 9:40 PM
3099 - 3093 - Check instrument neutron response (CGN model only); Remark:
Procedure Step
Published
07/20/2021 11:53 AM
3094 - Note : The next step is only applicable to CGN models with neutron detection capability...
Annotation
Published
10/04/2022 11:14 AM
3093 - Check instrument neutron response (CGN model only)
Instruction
Published
07/16/2021 10:47 AM
3100 - 3095 - Place instrument next to radioactive neutron source; Remark:536
Procedure Step
Published
07/20/2021 11:54 AM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
3102 - 3097 - Verify instrument responds correctly; Remark:3103
Procedure Step
Published
07/20/2021 11:54 AM
3096 - Note : Most radioactive neutron sources also have a gamma component, which will likely ...
Annotation
Published
07/23/2021 9:57 AM
3097 - Verify instrument responds correctly
Instruction
Published
07/16/2021 10:50 AM
3103 - The instrument should respond as follows: Neutron CPS chang...
Remark
Published
10/06/2022 12:33 PM
3101 - PRM-470 Search Mode Neutron
Remark
Published
11/05/2024 9:40 PM
3104 - 3098 - Move instrument away from radioactive neutron source; Remark:339
Procedure Step
Published
07/20/2021 11:54 AM
3098 - Move instrument away from radioactive neutron source
Instruction
Published
07/16/2021 10:54 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
5 minutes
Duration
Published
04/01/2020 11:35 AM
Daily
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
PRM-470_Partial_Power_Button_Red_Highlight.png
Image
Published
06/18/2021 6:28 PM
PRM-470_Partial-View_Light-Button_RED-Highlight.png
Image
Published
05/05/2020 9:17 PM
PRM-470 Display SEARCH
Image
Published
04/01/2021 11:22 PM
PRM-470_Partial-View_S-F-Button_RED-Highlight.png
Image
Published
04/01/2021 11:22 PM
PRM-470 Full Approaching Source No Alarm SEARCH Red Arrow
Image
Published
04/03/2021 7:06 PM
PRM-470 Display Gamma Alarm SEARCH Red Highlight
Image
Published
05/07/2020 4:22 PM
prm-470_di_galarm_SEARCH_00055.png
Image
Published
08/01/2021 2:40 AM
PRM-470 Move Away from Source Alarm SEARCH Red Arrow
Image
Published
05/13/2020 1:38 PM
PRM-470 Display FIND
Image
Published
05/07/2020 5:17 PM
PRM-470 Full Approaching Source No Alarm FIND Red Arrow
Image
Published
04/03/2021 7:06 PM
prm-470_di_galarm_FIND_00058.png
Image
Published
08/01/2021 2:40 AM
PRM-470 Display Gamma Alarm FIND Red Highlight
Image
Published
05/07/2020 5:19 PM
PRM-470_Full_Away_GAlarm_Red_Arrow FIND
Image
Published
05/13/2020 1:35 PM
PRM-470 Alarm Reaction
Image
Published
04/03/2021 1:42 AM
Search Mode Neutron
Image
Published
04/03/2021 1:42 AM
DET-HHD-RAP-RM01

Verify instrument is operating correctly

Equipment Hierarchy
Rapiscan PRM-470 CG
Rapiscan PRM-470 CGN
Frequency
Daily
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
5 minutes
Tools Required
None
Materials Required
Radioactive gamma source of Cs-137 activity 37-370 kBq (1-10 μCi), or equivalent
Radioactive neutron source of Cf-252 activity 37-185 kBq (1-5 μCi), or equivalent
Originators
Tucker, Brian
Originating Organization
Pacific Northwest National Laboratory
Approver
Cody Hostick
Safety Concerns
Radioactive Source
Additional Retest Requirements Information

None

Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration.

Version
2.00
Version #
Revision #
1.00
Revision Date
Revision Author
Revision Description

Entire document: Changed title, number, added new pictures, refined text on instructions, added neutron response check

Revision Date
Revision Author
Step 2
Step 3
Main Step

3102 - 3097 - Verify instrument responds correctly; Remark:3103

Step 00
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00