Last Updated: 04/26/2025 1:34 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Rapiscan PRM-470 CG
Device
Published
10/28/2024 3:40 PM
Rapiscan PRM-470 CGN
Device
Published
10/28/2024 3:33 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
607 - 205 - Disconnect charger; Remark:
Procedure Step
Published
07/23/2021 2:42 PM
205 - Disconnect charger
Instruction
Published
07/23/2021 2:42 PM
260 - 313 - Power off instrument; Remark:
Procedure Step
Published
07/23/2021 2:43 PM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
261 - 65 - Loosen four (4) cover screws; Remark:264
Procedure Step
Published
08/12/2021 7:25 AM
65 - Loosen four (4) cover screws
Instruction
Published
07/22/2021 12:38 PM
264 - Use a medium Phillips (PH2) screwdriver. Screws do not need to be completely ...
Remark
Published
07/20/2021 3:59 PM
263 - 262 - Remove instrument cover; Remark:413
Procedure Step
Published
08/11/2021 1:41 PM
2771 - Note : There are high-voltage detector circuit boards inside the instrument; however...
Annotation
Published
10/04/2022 11:06 AM
262 - Remove instrument cover
Instruction
Published
07/20/2021 3:35 PM
413 - PRM-470 Remove Instrument Cover
Remark
Published
11/01/2024 8:32 PM
267 - 266 - Locate battery connection; Remark:272
Procedure Step
Published
08/11/2021 1:42 PM
3033 - Note : PRM-470 CGN models have an additional board and connectors not shown in remar...
Annotation
Published
10/04/2022 11:07 AM
266 - Locate battery connection
Instruction
Published
07/20/2021 3:36 PM
272 - Battery connection is labeled 1.
Remark
Published
07/21/2021 12:11 PM
271 - 270 - Disconnect battery; Remark:414
Procedure Step
Published
08/11/2021 1:42 PM
1345 - Note : It is good practice to ensure all cables and connectors are labeled prior to ...
Annotation
Published
10/04/2022 11:08 AM
270 - Disconnect battery
Instruction
Published
07/20/2021 3:37 PM
414 - PRM-470_Disconnect_battery
Remark
Published
11/02/2024 1:21 PM
443 - 442 - Disconnect remaining connectors; Remark:
Procedure Step
Published
08/11/2021 1:43 PM
442 - Disconnect remaining connectors
Instruction
Published
07/22/2021 12:39 PM
618 - 615 - Uninstall PRCB-472 board; Remark:
Procedure Step
Published
08/12/2021 7:16 AM
615 - Uninstall PRCB-472 board
Instruction
Published
08/07/2021 11:00 AM
446 - 445 - Remove four (4) mounting screws; Remark:444
Procedure Step
Published
08/11/2021 1:43 PM
445 - Remove four (4) mounting screws
Instruction
Published
07/22/2021 12:41 PM
444 - Use a small Phillips (PH1) screwdriver. Retain screws.
Remark
Published
07/22/2021 1:19 PM
450 - 449 - Raise edge of PRCB-472 board to access underside; Remark:448
Procedure Step
Published
08/11/2021 1:48 PM
447 - Caution : Do not damage ribbon cable attached to the underside of PRCB-472.
Annotation
Published
10/04/2022 11:11 AM
449 - Raise edge of PRCB-472 board to access underside
Instruction
Published
07/22/2021 12:42 PM
448 - PRM-470 Lifting PRCB-472 circuit board
Remark
Published
11/02/2024 4:30 PM
453 - 452 - Disconnect overlay button connector; Remark:451
Procedure Step
Published
08/11/2021 1:48 PM
452 - Disconnect overlay button connector
Instruction
Published
07/22/2021 12:43 PM
451 - PRM-470 Display circuit board
Remark
Published
11/02/2024 3:35 PM
595 - 572 - Remove four (4) cover screws; Remark:594
Procedure Step
Published
08/12/2021 7:26 AM
572 - Remove four (4) cover screws
Instruction
Published
08/09/2021 1:57 PM
594 - Completely remove screws by continuing to thread them out of the front cover....
Remark
Published
08/12/2021 7:34 AM
596 - IO PRM-470 Cover Held by 4 screws exposed
Remark
Published
11/02/2024 4:33 PM
597 - 573 - Remove two (2) handle mounting screws; Remark:593
Procedure Step
Published
08/12/2021 7:26 AM
573 - Remove two (2) handle mounting screws
Instruction
Published
08/09/2021 1:58 PM
593 - Use a medium Phillips screwdriver (PH2). Retain screws.
Remark
Published
08/11/2021 11:19 AM
598 - 574 - Uninstall faulty overlay; Remark:
Procedure Step
Published
08/12/2021 7:27 AM
574 - Uninstall faulty overlay
Instruction
Published
08/09/2021 1:59 PM
599 - 575 - Peel off faulty overlay; Remark:592
Procedure Step
Published
08/12/2021 7:27 AM
575 - Peel off faulty overlay
Instruction
Published
08/09/2021 2:00 PM
592 - Use a scraping tool or blade to begin. The overlay has two (2) layers. Verify...
Remark
Published
08/11/2021 11:21 AM
600 - 576 - Clean case cover; Remark:591
Procedure Step
Published
08/12/2021 7:28 AM
576 - Clean case cover
Instruction
Published
08/11/2021 11:24 AM
591 - Use scraper, adhesive remover, and rags to remove excess adhesive.
Remark
Published
08/11/2021 11:28 AM
601 - 580 - Install new button overlay; Remark:
Procedure Step
Published
08/12/2021 7:28 AM
580 - Install new button overlay
Instruction
Published
08/09/2021 2:03 PM
602 - 577 - Expose adhesive; Remark:590
Procedure Step
Published
08/12/2021 7:29 AM
577 - Expose adhesive
Instruction
Published
08/09/2021 2:04 PM
590 - Peel protective layer from new overlay.
Remark
Published
08/11/2021 11:29 AM
603 - 578 - Feed overlay button connector through slot; Remark:589
Procedure Step
Published
08/12/2021 7:29 AM
578 - Feed overlay button connector through slot
Instruction
Published
08/09/2021 2:05 PM
589 - IO Overlay of PRM-470 Wire Ribbon Slot
Remark
Published
11/02/2024 4:35 PM
708 - 579 - Apply overlay to cover; Remark:707
Procedure Step
Published
08/12/2021 7:31 AM
579 - Apply overlay to cover
Instruction
Published
08/09/2021 2:05 PM
707 - Align holes for screws and LEDs.
Remark
Published
08/11/2021 11:30 AM
604 - 581 - Pull connector through slot; Remark:587
Procedure Step
Published
08/12/2021 7:30 AM
581 - Pull connector through slot
Instruction
Published
08/09/2021 2:06 PM
587 - Overlay should be flat against the case.
Remark
Published
08/11/2021 11:32 AM
605 - 582 - Verify overlay is securely attached; Remark:586
Procedure Step
Published
08/12/2021 7:31 AM
582 - Verify overlay is securely attached
Instruction
Published
08/09/2021 2:07 PM
586 - Apply pressure over entire overlay to ensure contact with cover.
Remark
Published
08/11/2021 11:34 AM
303 - 302 - Reassemble instrument; Remark:
Procedure Step
Published
10/07/2022 1:49 PM
302 - Reassemble instrument
Instruction
Published
06/16/2021 11:28 PM
709 - 583 - Reattach handle; Remark:585
Procedure Step
Published
08/12/2021 7:32 AM
583 - Reattach handle
Instruction
Published
08/09/2021 2:08 PM
585 - Use a medium Phillips screwdriver (PH2).
Remark
Published
08/11/2021 11:37 AM
710 - 584 - Reinsert four (4) assembly screws; Remark:596
Procedure Step
Published
08/12/2021 7:32 AM
584 - Reinsert four (4) assembly screws
Instruction
Published
08/09/2021 2:09 PM
465 - 463 - Reconnect overlay button connector; Remark:464
Procedure Step
Published
08/11/2021 1:51 PM
462 - Caution : Correct placement of connectors and wiring is critical to proper operation.
Annotation
Published
09/28/2023 2:13 PM
463 - Reconnect overlay button connector
Instruction
Published
07/22/2021 12:50 PM
464 - PRM-470 - Connecting ribbon cable to overlay button connector
Remark
Published
11/02/2024 3:35 PM
470 - 468 - Set PRCB-472 board over mounting posts; Remark:469
Procedure Step
Published
08/11/2021 1:52 PM
468 - Set PRCB-472 board over mounting posts
Instruction
Published
07/22/2021 12:52 PM
469 - Position ribbon cable to avoid obstructing PRCB-472 mounting hole.
Remark
Published
07/22/2021 1:12 PM
473 - 471 - Reinstall four (4) mounting screws; Remark:472
Procedure Step
Published
08/11/2021 1:53 PM
471 - Reinstall four (4) mounting screws
Instruction
Published
07/22/2021 12:53 PM
472 - Use a small Phillips (PH1) screwdriver.
Remark
Published
07/22/2021 1:13 PM
500 - 498 - Reconnect all connectors except battery; Remark:499
Procedure Step
Published
08/11/2021 2:02 PM
498 - Reconnect all connectors except battery
Instruction
Published
07/22/2021 1:02 PM
499 - Do not connect location 1.
Remark
Published
07/22/2021 1:17 PM
306 - 305 - Reconnect battery; Remark:416
Procedure Step
Published
08/11/2021 3:11 PM
305 - Reconnect battery
Instruction
Published
07/29/2021 3:48 PM
416 - PRM-470 Battery Replacement - Reconnect battery
Remark
Published
11/02/2024 2:02 PM
308 - 307 - Replace front cover; Remark:
Procedure Step
Published
08/11/2021 2:03 PM
304 - Caution : Ensure all wiring is inside the instrument while replacing cover.
Annotation
Published
10/04/2022 11:10 AM
307 - Replace front cover
Instruction
Published
07/20/2021 3:57 PM
310 - 309 - Tighten four (4) cover screws; Remark:417
Procedure Step
Published
08/11/2021 2:04 PM
569 - Caution : Do not overtighten screws.
Annotation
Published
06/16/2021 2:31 PM
309 - Tighten four (4) cover screws
Instruction
Published
07/22/2021 1:02 PM
417 - Use a medium Phillips (PH2) screwdriver.
Remark
Published
07/22/2021 1:18 PM
359 - 319 - Perform operational test; Remark:347
Procedure Step
Published
07/20/2021 10:55 AM
319 - Perform operational test
Instruction
Published
06/10/2021 8:31 PM
347 - Perform DET-HHD-RAP-RM01 PRM-470 Operational Test.
Remark
Published
07/21/2021 12:22 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
45 minutes
Duration
Published
05/18/2020 12:58 PM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Technician 2
Skill Level
Published
04/09/2020 1:17 AM
Medium Phillips (PH2) screwdriver
Tools
Published
04/24/2020 4:40 PM
Small Phillips (PH1) screwdriver
Tools
Published
05/14/2020 2:48 PM
Needle nose pliers
Tools
Published
05/06/2020 4:10 PM
Scraping tool or blade
Tools
Published
05/14/2020 2:49 PM
PRM470_Remove_Screws_Top_Case
Image
Published
05/08/2020 6:05 PM
PRM-470_di_CGN_open_00536
Image
Published
02/19/2023 11:26 AM
Battery_connection_is_labeled_1
Image
Published
02/19/2023 10:46 AM
PRM-470_Disconnect_battery
Image
Published
05/09/2020 8:40 PM
PRM-470 Remove four mounting screws on PRCB-472
Image
Published
05/11/2020 7:57 PM
PRM-470_Raise_edge_of _PRCB-472_board_gamma_on
Image
Published
05/20/2020 2:04 PM
PRM-470 Remove ribbon cable from PRCB-472 LCD Functional
Image
Published
05/20/2020 10:03 AM
PRM-470 Cover Screw Total Extraction
Image
Published
05/14/2020 5:50 PM
PRM-470 Handle Screws Interior
Image
Published
05/14/2020 6:59 PM
PRM-470 Handle Removal Exterior
Image
Published
05/14/2020 5:55 PM
PRM-470 Overlay Remove 1 with Scraper
Image
Published
05/14/2020 5:57 PM
PRM-470 Overlay Remove 2
Image
Published
05/14/2020 5:59 PM
PRM-470 Overlay Remove 3
Image
Published
05/14/2020 6:01 PM
PRM-470 Overlay Install Wire Ribbon Slot
Image
Published
05/14/2020 6:04 PM
PRM-470 Overlay Install Align to Cover
Image
Published
05/14/2020 6:07 PM
PRM-470 Overlay Flatten to Cover
Image
Published
05/14/2020 6:09 PM
PRM-470 - Connecting ribbon cable to overlay button connector
Image
Published
05/12/2020 12:56 PM
PRM-470_Set_PRCB-472_board_gamma_on
Image
Published
05/20/2020 2:13 PM
Remove metal brackets
Image
Published
09/08/2023 2:50 PM
Remove metal brackets 2
Image
Published
09/08/2023 2:51 PM
PRM470_Battery_connection_is_labeled_1
Image
Published
02/19/2023 10:40 AM
PRM-470_Reconnect_battery
Image
Published
05/10/2020 8:16 PM
PRM-470_Tighten_4_cover_screws
Image
Published
05/10/2020 8:22 PM
DET-HHD-RAP-CM06

Replace button overlay on PRM-470

Equipment Hierarchy
Rapiscan PRM-470 CG
Rapiscan PRM-470 CGN
Frequency
As Needed
Technician Level
Technician Level
Technician 2
x 1
Duration
Time increment
45 minutes
Tools Required
Medium Phillips (PH2) screwdriver
Small Phillips (PH1) screwdriver
Needle nose pliers
Scraping tool or blade
Materials Required
Disposable rags
Adhesive remover
Overlay PRM-470 CG (Rapiscan Part #3653)
Overlay PRM-470 CGN (Rapiscan Part #3653B)
Originators
Tucker, Brian
Originating Organization
PNNL
Safety Concerns
Electrical
Retest Requirements
Final Documentation
Corrective Maintenance Report
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration.

Version
0.00
Step 1
Step 3
Main Step
Step 4
Step 6
Step 7
Main Step
Document Type
Corrective Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00