Last Updated: 04/25/2025 1:34 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Rapiscan PRM-470 CG
Device
Published
10/28/2024 3:40 PM
Rapiscan PRM-470 CGN
Device
Published
10/28/2024 3:33 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
607 - 205 - Disconnect charger; Remark:
Procedure Step
Published
07/23/2021 2:42 PM
205 - Disconnect charger
Instruction
Published
07/23/2021 2:42 PM
314 - 313 - Power off instrument; Remark:
Procedure Step
Published
08/12/2021 6:54 AM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
275 - 65 - Loosen four (4) cover screws; Remark:264
Procedure Step
Published
07/26/2021 12:42 PM
65 - Loosen four (4) cover screws
Instruction
Published
07/22/2021 12:38 PM
264 - Use a medium Phillips (PH2) screwdriver. Screws do not need to be completely ...
Remark
Published
07/20/2021 3:59 PM
263 - 262 - Remove instrument cover; Remark:413
Procedure Step
Published
08/11/2021 1:41 PM
2771 - Note : There are high-voltage detector circuit boards inside the instrument; however...
Annotation
Published
10/04/2022 11:06 AM
262 - Remove instrument cover
Instruction
Published
07/20/2021 3:35 PM
413 - PRM-470 Remove Instrument Cover
Remark
Published
11/01/2024 8:32 PM
267 - 266 - Locate battery connection; Remark:272
Procedure Step
Published
08/11/2021 1:42 PM
3033 - Note : PRM-470 CGN models have an additional board and connectors not shown in remar...
Annotation
Published
10/04/2022 11:07 AM
266 - Locate battery connection
Instruction
Published
07/20/2021 3:36 PM
272 - Battery connection is labeled 1.
Remark
Published
07/21/2021 12:11 PM
271 - 270 - Disconnect battery; Remark:414
Procedure Step
Published
08/11/2021 1:42 PM
1345 - Note : It is good practice to ensure all cables and connectors are labeled prior to ...
Annotation
Published
10/04/2022 11:08 AM
270 - Disconnect battery
Instruction
Published
07/20/2021 3:37 PM
414 - PRM-470_Disconnect_battery
Remark
Published
11/02/2024 1:21 PM
443 - 442 - Disconnect remaining connectors; Remark:
Procedure Step
Published
08/11/2021 1:43 PM
442 - Disconnect remaining connectors
Instruction
Published
07/22/2021 12:39 PM
571 - 570 - Uninstall faulty GHA-472; Remark:
Procedure Step
Published
08/12/2021 6:55 AM
570 - Uninstall faulty GHA-472
Instruction
Published
08/11/2021 4:34 PM
478 - 476 - Remove mounting nut; Remark:477
Procedure Step
Published
08/11/2021 1:54 PM
476 - Remove mounting nut
Instruction
Published
08/07/2021 10:57 AM
477 - PRM-470 - Removing GHA-472 mounting nut
Remark
Published
11/02/2024 2:00 PM
481 - 479 - Remove GHA-472 board; Remark:480
Procedure Step
Published
08/11/2021 1:55 PM
479 - Remove GHA-472 board
Instruction
Published
07/22/2021 12:56 PM
480 - Gently pry up the GHA-472 board from the PRCB-472 board.
Remark
Published
07/22/2021 1:13 PM
741 - 740 - Install new GHA-472 board; Remark:
Procedure Step
Published
08/12/2021 6:56 AM
740 - Install new GHA-472 board
Instruction
Published
08/11/2021 4:35 PM
484 - 482 - Replace GHA-472 board; Remark:483
Procedure Step
Published
08/11/2021 1:55 PM
482 - Replace GHA-472 board
Instruction
Published
07/22/2021 12:56 PM
483 - Verify connectors are correctly aligned then press down.
Remark
Published
07/22/2021 1:14 PM
744 - 743 - Reinstall mounting nut; Remark:489
Procedure Step
Published
08/12/2021 6:56 AM
504 - Caution : Do not overtighten the mounting nut.
Annotation
Published
10/04/2022 11:11 AM
743 - Reinstall mounting nut
Instruction
Published
08/11/2021 4:36 PM
489 - PRM-470 - Location of GHA-472 mounting nut
Remark
Published
11/02/2024 2:01 PM
303 - 302 - Reassemble instrument; Remark:
Procedure Step
Published
10/07/2022 1:49 PM
302 - Reassemble instrument
Instruction
Published
06/16/2021 11:28 PM
500 - 498 - Reconnect all connectors except battery; Remark:499
Procedure Step
Published
08/11/2021 2:02 PM
498 - Reconnect all connectors except battery
Instruction
Published
07/22/2021 1:02 PM
499 - Do not connect location 1.
Remark
Published
07/22/2021 1:17 PM
306 - 305 - Reconnect battery; Remark:416
Procedure Step
Published
08/11/2021 3:11 PM
305 - Reconnect battery
Instruction
Published
07/29/2021 3:48 PM
416 - PRM-470 Battery Replacement - Reconnect battery
Remark
Published
11/02/2024 2:02 PM
308 - 307 - Replace front cover; Remark:
Procedure Step
Published
08/11/2021 2:03 PM
304 - Caution : Ensure all wiring is inside the instrument while replacing cover.
Annotation
Published
10/04/2022 11:10 AM
307 - Replace front cover
Instruction
Published
07/20/2021 3:57 PM
310 - 309 - Tighten four (4) cover screws; Remark:417
Procedure Step
Published
08/11/2021 2:04 PM
569 - Caution : Do not overtighten screws.
Annotation
Published
06/16/2021 2:31 PM
309 - Tighten four (4) cover screws
Instruction
Published
07/22/2021 1:02 PM
417 - Use a medium Phillips (PH2) screwdriver.
Remark
Published
07/22/2021 1:18 PM
747 - 746 - Perform gamma alignment; Remark:745
Procedure Step
Published
08/12/2021 6:56 AM
746 - Perform gamma alignment
Instruction
Published
08/11/2021 4:37 PM
745 - An alignment is required following replacement of the gamma board or the gamm...
Remark
Published
08/12/2021 7:00 AM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
20 minutes
Duration
Published
05/18/2020 5:44 PM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Technician 2
Skill Level
Published
04/09/2020 1:17 AM
Medium Phillips (PH2) screwdriver
Tools
Published
04/24/2020 4:40 PM
Small Phillips (PH1) screwdriver
Tools
Published
05/14/2020 2:48 PM
5-mm (1/4 inch) nut driver or socket
Tools
Published
05/14/2020 3:21 PM
Needle nose pliers
Tools
Published
05/06/2020 4:10 PM
PRM470_Remove_Screws_Top_Case
Image
Published
05/08/2020 6:05 PM
PRM-470_di_CGN_open_00536
Image
Published
02/19/2023 11:26 AM
Battery_connection_is_labeled_1
Image
Published
02/19/2023 10:46 AM
PRM-470_Disconnect_battery
Image
Published
05/09/2020 8:40 PM
PRM-470 - GHA-472 mounting nut location
Image
Published
02/19/2023 11:37 AM
PRM-470 - GHA-472 mounting nut removal
Image
Published
05/12/2020 1:29 PM
PRM-470 - Removing GHA-472 board
Image
Published
05/12/2020 1:40 PM
PRM-470 - Re-seating GHA-472 board
Image
Published
05/12/2020 1:47 PM
PRM-470 - Re-seating the GHA-472 board
Image
Published
05/12/2020 1:53 PM
PRM-470 - GHA-472 mounting nut location
Image
Published
05/12/2020 2:00 PM
PRM470_Battery_connection_is_labeled_1
Image
Published
02/19/2023 10:40 AM
PRM-470_Reconnect_battery
Image
Published
05/10/2020 8:16 PM
PRM-470_Tighten_4_cover_screws
Image
Published
05/10/2020 8:22 PM
DET-HHD-RAP-CM04

Replace GHA-472 (gamma high voltage board) on PRM-470.

Equipment Hierarchy
Rapiscan PRM-470 CG
Rapiscan PRM-470 CGN
Frequency
As Needed
Technician Level
Technician Level
Tehnician 2
x 1
Duration
Time increment
20 minutes
Length of time
20
Tools Required
Medium Phillips (PH2) screwdriver
Small Phillips (PH1) screwdriver
5-mm (1/4 inch) nut driver or socket
Needle nose pliers
Materials Required
GHA-472 Board (Rapiscan part number 23111634)
PRM-470 Smart battery charger
Originators
Tucker, Brian
Originating Organization
PNNL
Approver
Cody Hostick
Safety Concerns
Electrical
Retest Requirements
Final Documentation
Corrective Maintenance Report
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration.

Date of Origin
Version
0.00
Step 1

263 - 262 - Îndepărtați capacul instrumentului; Remark:413

Step 00
Step 5
Main Step
Document Type
Corrective Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00