Last Updated: 04/25/2025 1:40 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Smiths Detection RadSeeker CS
Device
Published
06/27/2024 12:54 PM
5925 - 5924 - Prepare Calibration Source; Remark:
Procedure Step
Published
11/07/2022 1:09 PM
5924 - Prepare Calibration Source
Instruction
Published
11/07/2022 1:09 PM
5923 - 5922 - Position Eu-152 calibration source; Remark:5921
Procedure Step
Published
11/07/2022 1:21 PM
5922 - Position Eu-152 calibration source
Instruction
Published
11/07/2022 1:10 PM
5921 - Position Eu-152 source in front of the detector.
Remark
Published
11/07/2022 1:21 PM
5920 - 5919 - Calibrate Instrument; Remark:
Procedure Step
Published
11/07/2022 1:22 PM
5919 - Calibrate Instrument
Instruction
Published
11/07/2022 1:22 PM
5914 - 5913 - Enter Advanced Mode; Remark:5912
Procedure Step
Published
11/07/2022 1:24 PM
5913 - Enter Advanced Mode
Instruction
Published
11/07/2022 1:23 PM
5912 - Select Menu from the Detect screen. Scroll down to E...
Remark
Published
11/07/2022 1:24 PM
5915 - 5911 - Enter password; Remark:5910
Procedure Step
Published
11/07/2022 1:26 PM
5911 - Enter password
Instruction
Published
11/07/2022 1:25 PM
5910 - Enter default Nuclear Smuggling Detection and Deterrence (NSDD) password
Remark
Published
11/07/2022 1:26 PM
5916 - 5592 - Enter Service Mode; Remark:5909
Procedure Step
Published
11/07/2022 1:28 PM
5592 - Enter Service Mode
Instruction
Published
10/31/2022 11:37 AM
5909 - Select Enter Service Mode from the Advanced Mode menu.
Remark
Published
11/07/2022 1:28 PM
5917 - 5908 - Start calibration; Remark:5907
Procedure Step
Published
11/07/2022 1:31 PM
5908 - Start calibration
Instruction
Published
11/07/2022 1:30 PM
5907 - Select Calibrate Detector. Wait several minutes for the cali...
Remark
Published
11/07/2022 1:31 PM
5918 - 5906 - Confirm completion; Remark:5905
Procedure Step
Published
11/07/2022 1:32 PM
5906 - Confirm completion
Instruction
Published
11/07/2022 1:32 PM
5905 - The Calibrated message will appear indicating calibration is...
Remark
Published
11/07/2022 1:32 PM
5904 - 5903 - Prepare Instrument for Use; Remark:
Procedure Step
Published
11/07/2022 1:33 PM
5903 - Prepare Instrument for Use
Instruction
Published
11/07/2022 1:33 PM
5900 - 5899 - Exit Service Mode; Remark:5898
Procedure Step
Published
11/07/2022 1:35 PM
5899 - Exit Service Mode
Instruction
Published
11/07/2022 1:34 PM
5898 - Return Back to Service Mode options.
Remark
Published
11/07/2022 1:35 PM
5901 - 5897 - Select Exit Advanced Mode; Remark:5957
Procedure Step
Published
11/07/2022 1:36 PM
5897 - Select Exit Advanced Mode
Instruction
Published
11/07/2022 1:36 PM
5957 - IO RadSeeker Detect Screen Exit Advanced Mode
Remark
Published
11/15/2022 11:43 AM
5902 - 5896 - Restart Instrument; Remark:
Procedure Step
Published
11/07/2022 1:37 PM
5896 - Restart Instrument
Instruction
Published
11/07/2022 1:37 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
minutes
Duration
Published
05/26/2022 11:41 AM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Technician 2
Skill Level
Published
04/09/2020 1:17 AM
None
Tools
Published
05/06/2020 9:50 PM
radseeker_di_front_rad_symbol.jpg
Image
Published
10/10/2023 12:32 AM
radseeker_ui_detect_screen_menu.jpg
Image
Published
10/10/2023 12:20 AM
radseeker_ui_detect_screen_enter_arrow.jpg
Image
Published
10/10/2023 12:22 AM
radseeker_ui_detect_screen_password.jpg
Image
Published
10/10/2023 12:14 AM
radseeker_ui_detect_screen_enter_service_mode.jpg
Image
Published
10/10/2023 12:26 AM
radseeker_ui_detect_screen_calibrate_detector.jpg
Image
Published
10/10/2023 12:27 AM
radseeker_ui_detect_screen_calibrated.jpg
Image
Published
10/10/2023 12:26 AM
radseeker_ui_detect_screen_back_left_arrow.jpg
Image
Published
10/10/2023 12:27 AM
radseeker_ui_detect_screen_back.jpg
Image
Published
10/10/2023 12:28 AM
radseeker_ui_detect_screen_exit_advanced_mode.jpg
Image
Published
10/10/2023 12:21 AM
DET-HHD-SMD-RSK-CM03

This procedure describes how to perform an energy calibration on the RadSeeker. This is only required following some corrective maintenance procedures. It is not required as a routine maintenance procedure as with other radiation detection instruments.

Equipment Hierarchy
Smiths Detection RadSeeker CS
Frequency
As Needed
Technician Level
Technician Level
Tehnik 2
x 1
Duration
Time increment
minutes
Length of time
30
Tools Required
None
Materials Required
Radioactive gamma source of Eu-152 activity 37-370 kBq (1-10 µCi)
Originators
Dylan Toneff
Brian Tucker
Originating Organization
PNNL
Approver
Cody Hostick
Safety Concerns
Radioactive Source
Final Documentation
Corrective Maintenance Report
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration.

Date of Origin
Version
0.00
Step 2
Document Type
Corrective Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media