6412 - 6328 - Vraťte konektory gama na místo; Remark:6326

Step 00

6328 - Reseat gamma connectors

Reseat gamma connectors

6326 - The gamma detector is connected at the FHT 681 amplifier at the connection la...

The gamma detector is connected at the FHT 681 amplifier at the connection labeled PMT1. Disconnect and reconnect both connections to the 90-degree elbow several times to remove oxidation.

6409 -

Image
PackEye FHT681 Install PMT1 Connector Gamma