Last Updated: 05/05/2025 1:40 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Thermo FHT1377 PackEye
Device
Published
10/28/2024 3:37 PM
5690 - 5689 - Dobíjecí cyklus napájení; Remark:
Procedure Step
Published
11/03/2022 10:35 AM
5689 - Power Cycle Power Supply
Instruction
Published
11/03/2022 10:35 AM
5758 - 205 - Odpojte nabíječku; Remark:3946
Procedure Step
Published
11/03/2022 2:01 PM
205 - Disconnect charger
Instruction
Published
07/23/2021 2:42 PM
3946 - Rotate the collar counterclockwise and pull.
Remark
Published
04/12/2022 1:37 PM
3852 - 220 - Zapněte přístroj; Remark:3836
Procedure Step
Published
03/29/2022 1:59 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
3836 - Press power button on the indicator unit for approximately 1 second.
Remark
Published
03/29/2022 1:05 AM
5757 - 5756 - Počkejte, dokud se přístroj nevypne; Remark:5755
Procedure Step
Published
11/03/2022 2:00 PM
5756 - Wait for instrument to power off
Instruction
Published
11/03/2022 2:00 PM
5755 - This will discharge the battery. Time for this step is between 12 and 48 hour...
Remark
Published
11/03/2022 2:00 PM
5754 - 5753 - Zapojte nabíječku; Remark:5751
Procedure Step
Published
11/03/2022 1:56 PM
5753 - Reconnect charger
Instruction
Published
11/03/2022 1:49 PM
5751 - IO PackEye Reconnect Charger Cable to Power Supply
Remark
Published
10/19/2024 4:44 PM
5752 - CHARGE (yellow) and READY (green) LEDs will...
Remark
Published
11/10/2022 9:17 AM
5750 - 5749 - Počkejte, dokud se přístroj zcela nenabije; Remark:5748
Procedure Step
Published
11/03/2022 1:48 PM
5749 - Wait for instrument to completely charge
Instruction
Published
11/03/2022 1:47 PM
5748 - Time for this step is between 2 and 4 hours. Once fully charged, only the
Remark
Published
11/03/2022 1:48 PM
5688 - 5687 - Dobíjecí cyklus záložní baterie; Remark:
Procedure Step
Published
11/03/2022 10:35 AM
5687 - Power Cycle Backup Battery
Instruction
Published
11/03/2022 10:35 AM
5747 - 5746 - Připojte záložní baterii k nabíječce; Remark:5745
Procedure Step
Published
11/03/2022 1:46 PM
5746 - Connect backup battery to charger
Instruction
Published
11/03/2022 1:45 PM
5745 - The backup battery uses a different charger than the PackEye.
Remark
Published
12/21/2022 12:52 PM
5696 - 5695 - Zahajte cyklus nabíjení; Remark:5694
Procedure Step
Published
11/03/2022 10:41 AM
5695 - Start charging cycle
Instruction
Published
11/03/2022 10:40 AM
5694 - The yellow button will start the discharge cycle. There are different charger...
Remark
Published
11/03/2022 10:41 AM
5693 - 5692 - Vyčkejte na dokončení cyklu nabíjení; Remark:5691
Procedure Step
Published
11/03/2022 10:39 AM
5692 - Wait for charging cycle to complete
Instruction
Published
11/03/2022 10:38 AM
5691 - The charger will fully discharge the battery and then fully recharge it. Time...
Remark
Published
11/03/2022 10:39 AM
253 - 242 - Zdokumentujte a nahlaste veškeré nevyřešené problémy; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
hours
Duration
Published
05/26/2022 11:45 AM
Semi-annually
Frequency
Published
11/03/2022 2:10 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
packeye_di_charger_disconnect_redarrow_00782.jpg
Image
Published
10/20/2023 12:54 PM
packeye_ui_power_on_00719
Image
Published
02/07/2022 1:46 AM
packeye_di_charger_connect_redarrow_00783.jpg
Image
Published
10/20/2023 12:53 PM
packeye_di_batteryCHARGING_0.jpg
Image
Published
10/23/2023 1:43 PM
packeye_di_batteryREADY.jpg
Image
Published
10/23/2023 9:34 AM
packeye_di_backupbattery_oncharger.jpg
Image
Published
10/20/2023 3:50 PM
packeye_di_battery_charger_discharging.jpg
Image
Published
10/23/2023 9:11 AM
packeye_di_battery_charger_status_LED.jpg
Image
Published
10/23/2023 9:12 AM
DET-HHD-THR-PAK-RM02

Routine battery power cycle for the power supply and the backup battery for the Thermo Scientific FHT 1377 PackEye. The labor required for this procedure is minimal; however, the time required for discharging and charging the batteries can be up to 48 hours.

Equipment Hierarchy
Thermo FHT1377 PackEye
Frequency
Semi-annually
Technician Level
Technician Level
Τεχνικός 1
x 1
Duration
Time increment
hours
Length of time
48
Tools Required
None
Materials Required
None
Originators
Tucker, Brian
Originating Organization
PNNL
Approver
Cody Hostick
Safety Concerns
None
Additional Retest Requirements Information

None

Departmental Coordination

Prior to arrival, notify operators concerning expected work, duration, and anticipated alarms and fault indications.

Date of Origin
Version
0.00
Step 1
Step 2
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00