6981 - 6977 - Reseat detector connections - Process, Notes: 6980

6978 - Warning : Power off instrument prior to disconnecting neutron detector.

Type
Warning

Power off instrument prior to disconnecting neutron detector.

6979 - Note : Access to the connectors is possible from the top of the backpack without rem...

Type
Note

Access to the connectors is possible from the top of the backpack without removing the foam assembly. Picture is shown for clarification only.

6977 - Reseat detector connections

Reseat detector connections
Go to
6976 - 6966 - False Alarms? - Decision, Notes: 6965
Step Reference Number
12
Note
6980 - Disconnect and reconnect the neutron tubes several times. ...