4756 - Note : Failures during the self-test are most likely a result of cables not properly...
Type
Note
Failures during the self-test are most likely a result of cables not properly connected to the processor printed circuit board (PCB). If any components fail during the self-test, disassemble the RadSeeker and ensure all the cabling to the processor PCB is properly in place.
Step 00
4753 - Respond to self-test
Respond to self-test
4752 - Answer the question prompts on the RadSeeker regarding instrument feedback su...
Answer the question prompts on the RadSeeker regarding instrument feedback such as vibrations and alarms. After feedback questions, the RadSeeker will test its gamma detection parameters. This will take several minutes.