The following table lists all the components that make up the document.
Clean alumina ceramic insulator on GE Reuter-Stokes proportional neutron counter detector tube. This is performed only when detector tubes are removed from moderator box.
If available, use an RPM or desktop RPM to test the neutron detector.
Prior to arrival, notify operators concerning expected work, duration, and anticipated alarms and fault indications.
9584 - 9543 - Neutron Tube Serial Number; Remark:
9543 - Neutron Tube Serial Number
Neutron Tube Serial Number
9585 - 9544 - Record serial number; Remark:9546
9544 - Record serial number
Record serial number
9546 - Only performed on detector tubes removed from the RPM.
Only performed on detector tubes removed from the RPM.
10325 - 10324 - Remove MHV Outer Connector Shell; Remark:
10324 - Remove MHV Outer Connector Shell
Remove MHV Outer Connector Shell
9586 - 9547 - Remove MHV outer connector shell; Remark:9548
9545 - Caution : Oils from skin may contaminate the insulating surfaces. Either lint-free cott...
Oils from skin may contaminate the insulating surfaces. Either lint-free cotton gloves or rubber gloves are used at all times during the cleaning process.
9604 - Caution : Avoid tearing or binding the rubber O-ring at the base of the connector shell...
Avoid tearing or binding the rubber O-ring at the base of the connector shell.
9547 - Remove MHV outer connector shell
Remove MHV outer connector shell
9548 - Use adjustable wrench. Loosen and unscrew connector shell.
Use adjustable wrench. Loosen and unscrew connector shell.
9587 - 9549 - Confirm O-ring is positioned above threaded surface and undamaged; Remark:9550
9604 - Caution : Avoid tearing or binding the rubber O-ring at the base of the connector shell...
Avoid tearing or binding the rubber O-ring at the base of the connector shell.
9549 - Confirm O-ring is positioned above threaded surface and undamaged
Confirm O-ring is positioned above threaded surface and undamaged
9550 - The O-ring size is 0.79 mm (0.031 in.) cross section by 19.69 mm (0.775 in.) ...
The O-ring size is 0.79 mm (0.031 in.) cross section by 19.69 mm (0.775 in.) inner diameter. (Parker size 5-137)
10329 - 10328 - Remove Teflon Insulator Ring; Remark:
10328 - Remove Teflon Insulator Ring
Remove Teflon Insulator Ring
9588 - 9551 - Remove Teflon insulator ring from ceramic core insulator; Remark:9552
9573 - Caution : Do not bend or twist the high voltage wire lead.
Do not bend or twist the high voltage wire lead.
9551 - Remove Teflon insulator ring from ceramic core insulator
Remove Teflon insulator ring from ceramic core insulator
9552 - Carefully pull Teflon insulator ring along the tube axis
Carefully pull Teflon insulator ring along the tube axis

10327 - 10326 - Inspect Neutron Tube Insulator; Remark:
10326 - Inspect Neutron Tube Insulator
Inspect Neutron Tube Insulator
9589 - 9553 - Examine surface condition of neutron tube insulator; Remark:9554
9553 - Examine surface condition of neutron tube insulator
Examine surface condition of neutron tube insulator
9554 - Questionable insulator
Questionable insulator

9555 - Typical insulator
Typical insulator

9590 - 9556 - Clean Insulator Core and Ring; Remark:
9556 - Clean Insulator Core and Ring
Clean Insulator Core and Ring
9591 - 9557 - Rinse ceramic and Teflon insulators; Remark:9558
9559 - Warning : Use nitrile gloves as sensitive skin may react to the alcohol
Use nitrile gloves as sensitive skin may react to the alcohol
9560 - Caution : For quality check, spray alcohol on a piece of clean glass and let evaporate....
For quality check, spray alcohol on a piece of clean glass and let evaporate. If a film is left on the glass, then the alcohol may have contamination and should be avoided.
9557 - Rinse ceramic and Teflon insulators
Rinse ceramic and Teflon insulators
9558 - Using a stream spray of alcohol, coat insulators and allow to evaporate. ...
Using a stream spray of alcohol, coat insulators and allow to evaporate.
9592 - 9561 - Wipe insulator; Remark:9562
9561 - Wipe insulator
Wipe insulator
9562 - Sweep surfaces with cottom swab, or lint-free cloth soaked in alcohol.
Sweep surfaces with cottom swab, or lint-free cloth soaked in alcohol.

10332 - Continue to sweep and rinse until dust particles no longer appear on the swab...
Continue to sweep and rinse until dust particles no longer appear on the swab.
9593 - 9563 - Inspect for and remove cotton fibers; Remark:9564
9563 - Inspect for and remove cotton fibers
Inspect for and remove cotton fibers
9564 - Remove cotton fibers deposited in insulator areas using tweezers.
Remove cotton fibers deposited in insulator areas using tweezers.
9594 - 9566 - Dry insulators; Remark:9567
9565 - Caution : Heating of the detector tube shall not exceed 100°C.
Heating of the detector tube shall not exceed 100°C.
9566 - Dry insulators
Dry insulators
9567 - Allow components to air dry.
Allow components to air dry.
9595 - 9568 - Clean connector shell; Remark:9569
9568 - Clean connector shell
Clean connector shell
9569 - The interior of the narrow-size barrel is the primary contact surface to clea...
The interior of the narrow-size barrel is the primary contact surface to clean to remove oxidation.

7279 - 7249 - Return equipment to normal operating condition; Remark:
7249 - Return equipment to normal operating condition
Return equipment to normal operating condition
9597 - 9571 - Confirm O-ring is in place; Remark:9572
9571 - Confirm O-ring is in place
Confirm O-ring is in place
9572 - No lubrication is necessary.
No lubrication is necessary.
9598 - 9574 - Reinstall ring insulator; Remark:9575
9573 - Caution : Do not bend or twist the high voltage wire lead.
Do not bend or twist the high voltage wire lead.
9574 - Reinstall ring insulator
Reinstall ring insulator
9575 - Slide ring insulator over core insulator and high voltage wire lead.
Slide ring insulator over core insulator and high voltage wire lead.

9599 - 9577 - Screw connector shell in place; Remark:9578
9576 - Note : Carefully place threads to avoid cross-threading of shell with the tube wall....
Carefully place threads to avoid cross-threading of shell with the tube wall.
9577 - Screw connector shell in place
Screw connector shell in place
9578 - Screw connector shell until finger tight.
Screw connector shell until finger tight.
10331 - 10330 - Test detector tube in an RPM or text fixture when possible; Remark:
10330 - Test detector tube in an RPM or text fixture when possible
Test detector tube in an RPM or text fixture when possible
322 - 321 - Документуйте виконання техобслуговування.; Remark:
321 - Документуйте виконання техобслуговування.
Документуйте виконання техобслуговування.
2869 - 2585 - Задокументуйте дії з технічного обслуговування; Remark:2866
2585 - Задокументуйте дії з технічного обслуговування
Задокументуйте дії з технічного обслуговування
2866 - Запишіть спостереження, час і результат у «Звіті про коригувальний ремонт».
Запишіть спостереження, час і результат у «Звіті про коригувальний ремонт».
4706 - 4705 - Надішліть звіт.; Remark:4864
4705 - Надішліть звіт.
Надішліть звіт.
4864 - Як визначено керівництвом або договірними зобов’язаннями.
Як визначено керівництвом або договірними зобов’язаннями.