The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech TA600 Radiation Pager
Device
Published
10/28/2024 3:51 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
7555 - 220 - Power on instrument; Remark:7542
Procedure Step
Published
06/05/2023 5:57 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
7542 - Set instrument to audio only mode.
Remark
Published
04/13/2025 5:48 PM
7556 - 612 - Wait for instrument to complete start-up sequence; Remark:7512
Procedure Step
Published
04/13/2025 6:50 PM
612 - Wait for instrument to complete start-up sequence
Instruction
Published
04/13/2025 6:38 PM
7512 - Red numerical LED display will flash while collecting background for 30 secon...
Remark
Published
10/19/2023 7:54 AM
7557 - 7543 - Test Audible Alarm; Remark:
Procedure Step
Published
06/05/2023 5:59 PM
7543 - Test Audible Alarm
Instruction
Published
06/05/2023 3:31 PM
7558 - 7544 - Place instrument next to source of radiation; Remark:7545
Procedure Step
Published
06/05/2023 5:59 PM
7544 - Place instrument next to source of radiation
Instruction
Published
06/05/2023 3:32 PM
7545 - TA600 Next to Source
Remark
Published
06/05/2023 3:34 PM
7559 - 2936 - Verify audible alarm functions correctly; Remark:7547
Procedure Step
Published
06/05/2023 6:00 PM
2936 - Verify audible alarm functions correctly
Instruction
Published
06/15/2021 4:09 PM
7547 - The instrument should respond as follows: Emit an audible b...
Remark
Published
06/05/2023 5:00 PM
7560 - 7548 - Acknowledge alarm; Remark:7549
Procedure Step
Published
06/05/2023 6:01 PM
7548 - Acknowledge alarm
Instruction
Published
06/05/2023 5:03 PM
7549 - Press display button to stop alarm. Otherwise, the alarm will automatically s...
Remark
Published
06/05/2023 5:07 PM
7561 - 7550 - Move instrument away from source of radiation; Remark:7551
Procedure Step
Published
06/05/2023 6:02 PM
7550 - Move instrument away from source of radiation
Instruction
Published
06/05/2023 5:09 PM
7551 - Move source away from TA600
Remark
Published
06/05/2023 5:12 PM
2914 - 2912 - Test vibration alarm; Remark:
Procedure Step
Published
06/06/2022 5:28 PM
2912 - Test vibration alarm
Instruction
Published
06/15/2021 4:11 PM
7562 - 7546 - Change to vibrate mode; Remark:7552
Procedure Step
Published
06/05/2023 6:02 PM
7546 - Change to vibrate mode
Instruction
Published
06/05/2023 3:35 PM
7552 - Set instrument to vibrate only mode.
Remark
Published
06/05/2023 5:15 PM
7563 - 612 - Wait for instrument to complete start-up sequence; Remark:7512
Procedure Step
Published
04/13/2025 6:55 PM
7553 - Note : If dial is turned to vibrate mode quickly, the instrument will not repeat sta...
Annotation
Published
06/05/2023 5:17 PM
7564 - 7544 - Place instrument next to source of radiation; Remark:7545
Procedure Step
Published
06/05/2023 6:05 PM
7565 - 2938 - Verify vibration alarm functions correctly; Remark:7554
Procedure Step
Published
06/05/2023 6:06 PM
2938 - Verify vibration alarm functions correctly
Instruction
Published
06/15/2021 4:12 PM
7554 - The instrument should respond as follows: Vibrate
Remark
Published
06/05/2023 5:24 PM
7566 - 7548 - Acknowledge alarm; Remark:7549
Procedure Step
Published
06/05/2023 6:07 PM
7567 - 7550 - Move instrument away from source of radiation; Remark:7551
Procedure Step
Published
06/05/2023 6:08 PM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
Daily
Frequency
Published
04/01/2020 11:36 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
TA600 Audio Only
Image
Published
06/04/2023 9:55 PM
TA600 Background
Image
Published
06/04/2023 9:59 PM
TA600 Ready
Image
Published
06/04/2023 9:59 PM
TA600 Next to Source
Image
Published
03/29/2024 11:57 AM
TA600 Audible Alarm
Image
Published
06/05/2023 5:00 PM
TA600 Acknowledge Alarm
Image
Published
06/05/2023 5:07 PM
TA600 Away From Source
Image
Published
06/05/2023 5:11 PM
TA600 Vibrate Only
Image
Published
06/04/2023 9:56 PM
TA600 Vibration Alarm
Image
Published
06/05/2023 5:23 PM
DET-HHD-STE-TA6-RM01

Check the Sensor Technology Engineering TA600 Radiation Pager operation and response to a source of radiation. The source of radiation may be any material that causes an alarm, including consumer products or naturally occuring radioactive materials. 

Equipment Hierarchy
Sensor Tech TA600 Radiation Pager
Frequency
Daily
Before each use
Technician Level
Technician Level
Technik 1
x 1
Duration
Length of time
5
Tools Required
None
Materials Required
Cs-137 activity 37 kBq (1-10 μCi) or gamma radioactive source with activity high enough to create alarm
Originators
Dylan Toneff
Brian Tucker
Originating Organization
PNNL
Approver
James Kitchen
Safety Concerns
Radioactive Source
Final Documentation
None
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration. 

Date of Origin
Príprava
Step 1
Prevádzka
Step 3

7563 - 612 - Wait for instrument to complete start-up sequence; Remark:7512

Step 00
Prevádzka
Step 4
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00