7031 - 7030 - Reseat Neutron tube connections - Process, Notes: 6980

6978 - Warning : Power off instrument prior to disconnecting neutron detector.

Type
Warning

Power off instrument prior to disconnecting neutron detector.

6979 - Note : Access to the connectors is possible from the top of the backpack without rem...

Type
Note

Access to the connectors is possible from the top of the backpack without removing the foam assembly. Picture is shown for clarification only.

7030 - Reseat Neutron tube connections

Reseat Neutron tube connections
Go to
7029 - 6870 - Fault resolved? - Decision, Notes: 7028
Step Reference Number
04
Note
6980 - ●Несколько раз отсоедините и снова подсоедините нейтронные трубки.●Проверьте нал...