The following table lists all the components that make up the document.

Title Type Status Last Updated
Radiation Solutions Inc. RS-350
Device
Published
10/20/2024 8:53 AM
4819 - 257 - Prepare instrument; Remark:
Procedure Step
Published
07/27/2022 1:18 PM
1468 - Note : No artificial radioactive sources should be present during the initial startu...
Annotation
Published
07/27/2022 1:16 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
4424 - 220 - Power on instrument; Remark:4423
Procedure Step
Published
06/10/2022 6:25 AM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
4423 - Open backpack and press power button for approximately 1 second.
Remark
Published
06/10/2022 5:14 AM
1496 - 1495 - Power on mobile device; Remark:
Procedure Step
Published
11/12/2020 4:40 PM
1495 - Power on mobile device
Instruction
Published
11/12/2020 4:20 PM
4425 - 4421 - Open RadMobile app; Remark:4811
Procedure Step
Published
06/10/2022 5:16 AM
4422 - Note : Refer to DET-HHD-RSI-RS3-OP02, RS-350 RadMobile App Pairing for app ...
Annotation
Published
06/10/2022 5:16 AM
4421 - Open RadMobile app
Instruction
Published
06/10/2022 5:15 AM
4811 - IO RS350 RadMobile App Logo No Words
Remark
Published
07/07/2022 12:54 PM
4426 - 4420 - Select name of RS-350; Remark:4419
Procedure Step
Published
06/10/2022 5:17 AM
4420 - Select name of RS-350
Instruction
Published
06/10/2022 5:17 AM
4419 - The last connected device will appear as the first option on the screen. ...
Remark
Published
06/10/2022 5:17 AM
4427 - 1432 - Verify backpack connection to mobile device; Remark:4418
Procedure Step
Published
06/10/2022 5:20 AM
1432 - Verify backpack connection to mobile device
Instruction
Published
09/01/2020 7:34 PM
4418 - Gamma and neutron values will be changing.
Remark
Published
06/10/2022 5:20 AM
4428 - 2295 - Verify sufficient battery charge; Remark:4416
Procedure Step
Published
06/10/2022 5:23 AM
2295 - Verify sufficient battery charge
Instruction
Published
04/06/2021 4:43 PM
4416 - Backpack battery life is indicated by the battery icon in the app.
Remark
Published
06/10/2022 5:22 AM
4417 - Mobile device battery life is typically indicated at the top of the display.
Remark
Published
06/10/2022 5:23 AM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
4497 - 235 - Place instrument next to radioactive gamma source; Remark:4817
Procedure Step
Published
06/10/2022 7:59 AM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
4817 - IO RS350 Move Closer to Rad Source Trefoil
Remark
Published
07/07/2022 7:25 PM
4498 - 1439 - Verify instrument responds to radioactive source; Remark:4495
Procedure Step
Published
06/10/2022 8:00 AM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
4495 - Gamma dose rate and count rate values will increase. Movin...
Remark
Published
07/07/2022 7:21 PM
4496 - Automatic identification of the radiation source may occur as the RS-350 is b...
Remark
Published
07/07/2022 7:23 PM
2251 - 2217 - Perform Identification Measurement; Remark:
Procedure Step
Published
07/28/2021 9:55 PM
2217 - Perform Identification Measurement
Instruction
Published
04/06/2021 5:13 PM
4452 - 859 - Start identification measurement; Remark:4430
Procedure Step
Published
06/10/2022 6:07 AM
859 - Start identification measurement
Instruction
Published
10/26/2020 7:00 PM
4430 - Tap menu icon.
Remark
Published
06/10/2022 5:38 AM
4451 - Tap Start dwell.
Remark
Published
06/10/2022 6:07 AM
4453 - 2220 - Wait for measurement to complete; Remark:4809
Procedure Step
Published
04/20/2025 12:56 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
4809 - IO RS350 Dwell Mode Timer Running 3/60
Remark
Published
07/06/2022 4:29 PM
4492 - 1442 - Verify identification results are correct; Remark:4491
Procedure Step
Published
06/10/2022 7:28 AM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
4491 - Identification results should match the isotopes associated with the radioact...
Remark
Published
07/07/2022 7:19 PM
4493 - 4490 - Tap CLOSE; Remark:4489
Procedure Step
Published
06/10/2022 7:29 AM
4490 - Tap CLOSE
Instruction
Published
06/10/2022 7:29 AM
4489 - This will return the instrument to Automatic mode.
Remark
Published
07/07/2022 7:18 PM
4494 - 4488 - Move instrument away from radioactive gamma source; Remark:4818
Procedure Step
Published
06/10/2022 7:30 AM
4488 - Move instrument away from radioactive gamma source
Instruction
Published
06/10/2022 7:30 AM
4818 - IO RS350 Move Away from Rad Source Trefoil
Remark
Published
07/07/2022 7:26 PM
4481 - 4480 - Check instrument neutron response (if available); Remark:
Procedure Step
Published
06/10/2022 7:19 AM
4480 - Check instrument neutron response (if available)
Instruction
Published
06/10/2022 7:19 AM
4485 - 3095 - Place instrument next to radioactive neutron source; Remark:4817
Procedure Step
Published
06/10/2022 7:31 AM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
4486 - 1439 - Verify instrument responds to radioactive source; Remark:4484
Procedure Step
Published
06/10/2022 7:32 AM
4484 - Neutron count rate values will increase. The mobile device...
Remark
Published
07/07/2022 7:17 PM
4487 - 4483 - Move instrument away from radioactive neutron source; Remark:4818
Procedure Step
Published
06/10/2022 7:33 AM
4483 - Move instrument away from radioactive neutron source
Instruction
Published
06/10/2022 7:33 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
RS350 Power Button Green Light
Image
Published
07/05/2022 3:50 PM
RS-350 RadMobile App Logo
Image
Published
07/05/2022 4:34 PM
RS-350 App Connecting Serial Number Full Screen
Image
Published
07/06/2022 1:18 PM
RS-350 Graph 70.5 nSv Boxes Dose, Counts, Graph
Image
Published
07/05/2022 4:18 PM
RS-350 App Battery Icon Red Box
Image
Published
07/06/2022 1:09 PM
RS-350 App Mobile Battery Percent
Image
Published
07/06/2022 1:37 PM
RS-350 Move to Rad Source
Image
Published
07/05/2022 3:56 PM
RS-350 Graph Alarm Response Red Arrow
Image
Published
07/05/2022 4:22 PM
RS-350 Identification Dwell Result Co-60
Image
Published
07/05/2022 5:35 PM
RS-350 App Menu Icon Red Box
Image
Published
07/06/2022 1:27 PM
RS-350 Identification Dwell Start
Image
Published
07/05/2022 5:37 PM
RS-350 Wait Dwell 3 of 60 sec
Image
Published
07/05/2022 4:05 PM
RS-350 Close Dwell Mode
Image
Published
07/05/2022 5:32 PM
RS-350 Move away from Rad Source
Image
Published
07/05/2022 3:58 PM
RS-350 Alarm Neutron Red Box 10 CPS
Image
Published
07/05/2022 5:20 PM
DET-HHD-RSI-RS3-RM01

Conduct operational test of RS-350 backpack.

Equipment Hierarchy
Radiation Solutions Inc. RS-350
Frequency
Annually
Before each use
Monthly
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
minutes
Length of time
10
Tools Required
None
Materials Required
Cs-137 activity 37 kBq (1-10 μCi) or gamma radioactive source with activity high enough to create alarm
Cf-252 activity 37 kBq (1-10 μCi)
Originators
Demboski, Erica
Johns, Paul
Originating Organization
PNNL
Approver
Simons, Brian
Safety Concerns
Radioactive Source
Final Documentation
None
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration.

Date of Origin
Version
0.00
Step 1
Main Step
Step 2

4498 - 1439 - Verify instrument responds to radioactive source; Remark:4495

Step 00
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00