Replace battery on the Sensor Technology Engineering Handheld Radiation Monitor (HRM).
Prior to arrival, notify operators concerning expected work and duration.
4373 - 313 - Power off instrument; Remark:4372
313 - Power off instrument
Power off instrument
4372 - Move switch to OFF position.
Move switch to OFF position.

4369 - 4368 - Remove battery cover; Remark:11198
4368 - Remove battery cover
Remove battery cover
11198 - Use a coin screwdriver, a medium #2 slotted screwdriver or a coin. Rotate counte...
4366 - 1816 - Remove battery; Remark:4365
1816 - Remove battery
Remove battery
4365 - Properly dispose of faulty battery
Properly dispose of faulty battery
4364 - 1818 - Install replacement battery; Remark:4363
1818 - Install replacement battery
Install replacement battery
4363 - The positive + terminal points to top of instrument.
The positive + terminal points to top of instrument.

4362 - 4360 - Remplacez le joint torique (si nécessaire); Remark:4359
4361- : La taille du joint torique est 2-018.
La taille du joint torique est 2-018.
4360 - Remplacez le joint torique (si nécessaire)
Remplacez le joint torique (si nécessaire)
4359 - Remplacez le joint torique s’il est cassé ou s’il présente des signes de dommage...
Remplacez le joint torique s’il est cassé ou s’il présente des signes de dommages (craquelures ou fissures).

4358 - 4357 - Appliquez de la graisse de silicone (si nécessaire); Remark:4356
4357 - Appliquez de la graisse de silicone (si nécessaire)
Appliquez de la graisse de silicone (si nécessaire)
4356 - De la graisse de silicone doit être appliquée sur les joints toriques secs.
De la graisse de silicone doit être appliquée sur les joints toriques secs.
4355 - 4354 - Reinstall battery cover; Remark:4353
4354 - Reinstall battery cover
Reinstall battery cover
4353 - Use medium #2 slotted screwdriver. A coin may also be used. Rotate clockwise....
Use medium #2 slotted screwdriver. A coin may also be used. Rotate clockwise.

4352 - 220 - Power on instrument; Remark:4351
220 - Power on instrument
Power on instrument
4351 - Move switch to AUDIO or VIBRATE.
Move switch to AUDIO or VIBRATE.

4290 - 612 - Wait for instrument to complete start-up sequence; Remark:4286
612 - Wait for instrument to complete start-up sequence
Wait for instrument to complete start-up sequence
4286 - The unit is ready for operation when the period flashes every 20 seconds....
The unit is ready for operation when the period flashes every 20 seconds.

253 - 242 - Consignez et faites un rapport pour tout problème non résolu; Remark:
242 - Consignez et faites un rapport pour tout problème non résolu
Consignez et faites un rapport pour tout problème non résolu