Last Updated: 04/25/2025 1:38 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
FLIR identiFINDER 2 NG
Device
Published
10/24/2024 10:14 AM
3292 - 220 - Power on instrument; Remark:3246
Procedure Step
Published
04/07/2025 2:24 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
3246 - IDF2 Dose Rate Screen
Remark
Published
05/25/2021 7:26 PM
3293 - 3259 - Perform factory test on instrument and report any issues; Remark:
Procedure Step
Published
05/25/2021 7:29 PM
3259 - Perform factory test on instrument and report any issues
Instruction
Published
05/25/2021 5:38 PM
3294 - 3260 - Press “R” button to select Options; Remark:
Procedure Step
Published
05/25/2021 7:55 PM
3260 - Press “R” button to select Options
Instruction
Published
05/25/2021 7:54 PM
3295 - 3261 - Press “M” button to select Advanced Menu; Remark:3277
Procedure Step
Published
06/03/2021 2:28 PM
3261 - Press “M” button to select Advanced Menu
Instruction
Published
05/25/2021 5:49 PM
3277 - If prompted, enter the password (normally “LLLLLLL”) and press “
Remark
Published
10/05/2022 4:41 PM
3296 - 3262 - Press “M” button to select More Options; Remark:
Procedure Step
Published
05/25/2021 7:33 PM
3262 - Press “M” button to select More Options
Instruction
Published
05/25/2021 5:51 PM
3297 - 3263 - Press “L” button to scroll down and press “M” button to select Advanc...; Remark:
Procedure Step
Published
05/25/2021 6:53 PM
3263 - Press “L” button to scroll down and press “M” button to select Advanc...
Instruction
Published
05/25/2021 5:54 PM
3298 - 3264 - Press “L” button to scroll down and press “M” button to select Mainte...; Remark:
Procedure Step
Published
05/25/2021 7:04 PM
3264 - Press “L” button to scroll down and press “M” button to select Mainte...
Instruction
Published
05/25/2021 6:04 PM
3299 - 3265 - Press “L” button to scroll down and press “M” button to select Self-T...; Remark:
Procedure Step
Published
05/25/2021 7:05 PM
3265 - Press “L” button to scroll down and press “M” button to select Self-T...
Instruction
Published
05/25/2021 6:07 PM
3300 - 3266 - Press “M” button to start Self-Test; Remark:3279
Procedure Step
Published
05/25/2021 7:06 PM
3266 - Press “M” button to start Self-Test
Instruction
Published
05/25/2021 6:09 PM
3279 - The instrument will perform a series of self-checks (various displays, alarm ...
Remark
Published
05/25/2021 6:36 PM
3301 - 3278 - Press “R” button to exit back to Dose Rate screen; Remark:
Procedure Step
Published
05/25/2021 7:07 PM
3278 - Press “R” button to exit back to Dose Rate screen
Instruction
Published
05/25/2021 6:29 PM
3302 - 3267 - Verify Finder mode functionality; Remark:3291
Procedure Step
Published
05/25/2021 7:08 PM
3280 - Note : Before entering the Finder mode, ensure that the instrument is not near any r...
Annotation
Published
05/25/2021 6:38 PM
3267 - Verify Finder mode functionality
Instruction
Published
10/07/2022 3:47 PM
3291 - IO IDF Exclude Radiation Source
Remark
Published
05/11/2021 6:58 PM
3303 - 3268 - Enter Finder mode by pressing “L” button; Remark:3282
Procedure Step
Published
05/25/2021 7:09 PM
3268 - Enter Finder mode by pressing “L” button
Instruction
Published
05/25/2021 6:13 PM
3282 - IO IDF Dose Mode to Finder
Remark
Published
05/12/2021 11:10 AM
3281 - Upon entering Finder mode, the instrument will collect a background for 10 se...
Remark
Published
05/25/2021 6:41 PM
3304 - 3269 - Move instrument close to radioactive source and then farther away.; Remark:3283
Procedure Step
Published
05/25/2021 7:10 PM
3269 - Move instrument close to radioactive source and then farther away.
Instruction
Published
05/25/2021 6:15 PM
3283 - The identiFINDER will indicate the presence of elevated radiation levels with...
Remark
Published
05/25/2021 6:43 PM
3290 - IO IDF Vibrate Near Source Crate
Remark
Published
05/11/2021 6:55 PM
3305 - 3270 - Verify Identify mode functionality; Remark:
Procedure Step
Published
05/25/2021 7:12 PM
3270 - Verify Identify mode functionality
Instruction
Published
05/25/2021 6:17 PM
3306 - 235 - Place instrument next to radioactive gamma source; Remark:3288
Procedure Step
Published
05/25/2021 7:13 PM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
3288 - IO IDF Near Radiation Source
Remark
Published
05/11/2021 6:47 PM
3307 - 3272 - From Dose Rate or Finder mode, press “M” button to enter Identify mode; Remark:3289
Procedure Step
Published
05/19/2022 5:53 PM
3272 - From Dose Rate or Finder mode, press “M” button to enter Identify mode
Instruction
Published
05/25/2021 6:20 PM
3289 - IO IDF Menu Identify Button M
Remark
Published
05/11/2021 6:53 PM
3287 - Ensure the instrument is as close to the source as possible, trying to mainta...
Remark
Published
05/25/2021 6:51 PM
3167 - IdentiFINDER2-NG Point unit to Source
Remark
Published
05/19/2022 5:52 PM
3308 - 3273 - Leave instrument next to the source(s); Remark:3284
Procedure Step
Published
05/25/2021 7:17 PM
3273 - Leave instrument next to the source(s)
Instruction
Published
05/25/2021 6:22 PM
3284 - Usually 120 seconds, until the instrument completes the data collection and a...
Remark
Published
05/19/2022 5:57 PM
3309 - 3274 - Verify results agree with known radioactive source(s) (e.g. Co-57, Na-22, Ba-...; Remark:3201
Procedure Step
Published
05/25/2021 7:18 PM
3274 - Verify results agree with known radioactive source(s) (e.g. Co-57, Na-22, Ba-...
Instruction
Published
05/25/2021 6:24 PM
3201 - IDF2 - List of Identified Nuclides
Remark
Published
05/05/2021 2:20 PM
3310 - 3275 - Select Exit (may need to select Skip first) to return to Dose Rate or Finder ...; Remark:
Procedure Step
Published
05/25/2021 7:20 PM
3275 - Select Exit (may need to select Skip first) to return to Dose Rate or Finder ...
Instruction
Published
05/25/2021 6:26 PM
3311 - 3276 - Record name, maintenance performed, problems observed, and date and time in o...; Remark:3285
Procedure Step
Published
05/25/2021 7:21 PM
3276 - Record name, maintenance performed, problems observed, and date and time in o...
Instruction
Published
05/25/2021 6:28 PM
3285 - Report any unresolved problems to the local maintenance provider or sustainab...
Remark
Published
05/25/2021 6:47 PM
3286 - If identiFINDER is not functioning correctly, perform troubleshooting by refe...
Remark
Published
05/25/2021 6:49 PM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
5 minutes
Duration
Published
04/01/2020 11:35 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
IDF2_ui_doseratemode2_609.png
Image
Published
05/02/2021 2:52 AM
IDF Middle Button
Image
Published
05/25/2021 7:45 PM
IDF No Radiation Source Near
Image
Published
05/11/2021 2:03 AM
IDF Select Finder Button L
Image
Published
05/12/2021 3:05 AM
IDF Finder Mode Partial chart 4 sec
Image
Published
05/01/2021 2:01 AM
IDF Finder Mode Rolling Graph Peaks
Image
Published
05/01/2021 2:01 AM
IDF Gamma Alarm Vibrate
Image
Published
05/11/2021 2:03 AM
IDF_di_Trefoil_640_0.png
Image
Published
05/13/2021 2:05 AM
IDF Select Identify Button M Dose Screen
Image
Published
05/12/2021 3:05 AM
IDF Select Identify Button M
Image
Published
05/12/2021 3:05 AM
identiFINDER Keep Distance RedArrow
Image
Published
05/19/2022 5:46 PM
IDF Isotope List
Image
Published
05/01/2021 2:01 AM
DET-HHD-FLR-IDF-RM08

Perform a functional test on the identiFINDER 2

Equipment Hierarchy
FLIR identiFINDER 2 NG
Frequency
Annually
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
5 minutes
Length of time
5
Tools Required
None
Materials Required
Known source of gamma radiation
Safety Concerns
None
Final Documentation
None
Step 2
Step 3
Main Step

3303 - 3268 - Enter Finder mode by pressing “L” button; Remark:3282

Step 00

3304 - 3269 - Move instrument close to radioactive source and then farther away.; Remark:3283

Step 00
Step 4

3307 - 3272 - From Dose Rate or Finder mode, press “M” button to enter Identify mode; Remark:3289

Step 00
Step 5
Main Step

3311 - 3276 - Record name, maintenance performed, problems observed, and date and time in o...; Remark:3285

Step 00
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00