The following table lists all the components that make up the document.
Verify and update parameter settings on the PRM-470.
This procedure may be performed using the Handheld Assessment and Troubleshooting Software (HATS).
Prior to arrival, notify operators concerning expected work and duration.
Changed procedure number, Updated front matter, updated Step 2-4 Remarks.
420 - 418 - Connect to PRM-470 with PRMCOM software; Remark:419
418 - Connect to PRM-470 with PRMCOM software
Connect to PRM-470 with PRMCOM software
419 - Refer to DET-HHD-RAP-RM02 PRM-470 Computer Connection
Refer to DET-HHD-RAP-RM02 PRM-470 Computer Connection
421 - 344 - Verify parameter settings are correct; Remark:
344 - Verify parameter settings are correct
Verify parameter settings are correct
435 - 422 - Select Configuration tab; Remark:434
422 - Select Configuration tab
Select Configuration tab
436 - 423 - Click Connect; Remark:433
423 - Click Connect
Click Connect
433 - Values will appear in the Parameters section.
Values will appear in the Parameters section.


437 - 424 - Compare values; Remark:432
424 - Compare values
Compare values
432 - Note differences in parameter values with approved Nuclear Smuggling Detectio...
Note differences in parameter values with approved Nuclear Smuggling Detection and Deterrence (NSDD) parameter values in the table below.

438 - 425 - Update parameters in PRMCOM; Remark:430
431 - Note : It is important that the “Restrict Setup” parameter be set to ON. This se...
It is important that the “Restrict Setup” parameter be set to ON. This setting requires a computer connection to change parameters.
425 - Update parameters in PRMCOM
Update parameters in PRMCOM
430 - Do not change Sigma parameter if between 4 and 7. For remaining parameters, u...
Do not change Sigma parameter if between 4 and 7. For remaining parameters, update to approved NSDD parameter values.
439 - 426 - Update parameters on PRM-470; Remark:429
426 - Update parameters on PRM-470
Update parameters on PRM-470
429 - Click on Send Parameters to update settings on the PRM-470.
Click on Send Parameters to update settings on the PRM-470.

440 - 427 - Verify parameters were successfully updated; Remark:428
427 - Verify parameters were successfully updated
Verify parameters were successfully updated
428 - Click on Get Parameters to ensure values were successfully u...
Click on Get Parameters to ensure values were successfully updated.

253 - 242 - Documentați și raportați problemele nerezolvate; Remark:
242 - Documentați și raportați problemele nerezolvate
Documentați și raportați problemele nerezolvate