Last Updated: 04/27/2025 1:43 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Viken Detection Nighthawk HBI XP WAD
Device
Published
06/04/2024 9:41 AM
10489 - 257 - Prepare instrument; Remark:
Procedure Step
Published
12/06/2023 3:03 PM
10487 - Warning : Prior to use, the operator should read and understand the complete user’s g...
Annotation
Published
12/06/2023 3:02 PM
10488 - Caution : Do not block the air intake vents or outlet vents when the instrument is in u...
Annotation
Published
12/06/2023 3:03 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
10485 - 10484 - Insert battery pack; Remark:10483
Procedure Step
Published
12/06/2023 3:07 PM
10484 - Insert battery pack
Instruction
Published
12/06/2023 3:04 PM
10483 -
Remark
Published
12/06/2023 3:07 PM
10486 - 220 - Power on instrument; Remark:10482
Procedure Step
Published
12/06/2023 4:50 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
10482 - Press and hold the Power button.
Remark
Published
12/06/2023 4:48 PM
10498 - 10497 - Select HBI application; Remark:10496
Procedure Step
Published
12/06/2023 4:55 PM
10497 - Select HBI application
Instruction
Published
12/06/2023 4:55 PM
10496 - On the home screen, tap on the Viken icon.
Remark
Published
12/06/2023 4:55 PM
10499 - 10494 - Log into the application; Remark:10493
Procedure Step
Published
12/06/2023 5:00 PM
10495 - Note : See the user’s guide to set up the first administrator. Keep a secure recor...
Annotation
Published
12/06/2023 5:00 PM
10494 - Log into the application
Instruction
Published
12/06/2023 5:00 PM
10493 - Type in your username and password.
Remark
Published
12/06/2023 5:00 PM
10500 - 10492 - Acknowledge warnings; Remark:10491
Procedure Step
Published
12/06/2023 5:04 PM
10492 - Acknowledge warnings
Instruction
Published
12/06/2023 5:04 PM
10491 -
Remark
Published
12/06/2023 5:04 PM
10505 - 10503 - Perform a system check; Remark:
Procedure Step
Published
12/07/2023 9:17 AM
10504 - Note : It is recommended to perform a system check if you have not used the instrume...
Annotation
Published
12/07/2023 9:17 AM
10503 - Perform a system check
Instruction
Published
12/07/2023 9:17 AM
10502 - 10501 - Tap Menu icon, select Settings, if needed; Remark:10518
Procedure Step
Published
12/07/2023 9:24 AM
10501 - Tap Menu icon, select Settings, if needed
Instruction
Published
12/07/2023 9:24 AM
10518 - The system check screen will automatically pop-up if it has been more than 7 ...
Remark
Published
12/07/2023 1:47 PM
10519 - 10517 - Place instrument on sample radiation shield; Remark:10516
Procedure Step
Published
12/07/2023 1:51 PM
10517 - Place instrument on sample radiation shield
Instruction
Published
12/07/2023 1:51 PM
10516 - Place face down on flat surface.
Remark
Published
12/07/2023 1:51 PM
10520 - 10515 - Tap YES on screen; Remark:10514
Procedure Step
Published
12/07/2023 1:54 PM
10515 - Tap YES on screen
Instruction
Published
12/07/2023 1:54 PM
10514 - Viken Nighthawk system check screen YES
Remark
Published
07/24/2024 12:02 PM
10521 - 10513 - Wait for system check to complete; Remark:10512
Procedure Step
Published
12/07/2023 1:59 PM
10513 - Wait for system check to complete
Instruction
Published
12/07/2023 1:59 PM
10512 - This will take approximately 10 seconds.
Remark
Published
12/07/2023 1:59 PM
10522 - 10511 - Tap OK on screen; Remark:10510
Procedure Step
Published
12/07/2023 2:01 PM
10511 - Tap OK on screen
Instruction
Published
12/07/2023 2:01 PM
10510 -
Remark
Published
12/07/2023 2:01 PM
10509 - 10506 - Conduct scan; Remark:
Procedure Step
Published
12/07/2023 2:06 PM
10507 - Warning : The HBI should not be used to scan people directly. Never point the instrumen...
Annotation
Published
12/07/2023 2:05 PM
10508 - Warning : Do not stare directly into the red laser beams.
Annotation
Published
12/07/2023 2:06 PM
10506 - Conduct scan
Instruction
Published
12/07/2023 2:06 PM
10539 - 10538 - Position instrument over item to be scanned; Remark:
Procedure Step
Published
12/07/2023 2:10 PM
10538 - Position instrument over item to be scanned
Instruction
Published
12/07/2023 2:10 PM
10540 - 10537 - Swipe slider to the right or touch the right side of the line; Remark:10536
Procedure Step
Published
12/07/2023 2:13 PM
10537 - Swipe slider to the right or touch the right side of the line
Instruction
Published
12/07/2023 2:13 PM
10536 - The blue slider at the top of the screen must be swiped to the right edge of ...
Remark
Published
12/07/2023 2:13 PM
10541 - 10535 - Press and hold the triggers; Remark:10534
Procedure Step
Published
12/07/2023 2:16 PM
10535 - Press and hold the triggers
Instruction
Published
12/07/2023 2:16 PM
10534 - The triggers may be red or blue, depending on the model. The HBI will start t...
Remark
Published
12/07/2023 2:16 PM
10542 - 10530 - Scan item of interest; Remark:10528
Procedure Step
Published
12/07/2023 2:19 PM
10531 - Note : Longer scan times produce larger image files, shorter scans are preferred.
Annotation
Published
12/07/2023 2:20 PM
10532 - Note : There is a 30 second maximum scan time.
Annotation
Published
12/07/2023 2:20 PM
10533 - Note : Multiple scans may be required to fully image an item.
Annotation
Published
12/07/2023 2:20 PM
10530 - Scan item of interest
Instruction
Published
12/07/2023 2:19 PM
10528 - The instrument must be moved to capture a two (2)-dimensional image. Moving t...
Remark
Published
12/07/2023 2:19 PM
10529 -
Remark
Published
12/07/2023 2:20 PM
10543 - 10527 - Start new scan, if needed; Remark:10526
Procedure Step
Published
12/07/2023 2:21 PM
10527 - Start new scan, if needed
Instruction
Published
12/07/2023 2:21 PM
10526 - The instrument will need to be re-armed. Repeat steps 3-2 to 3-4.
Remark
Published
12/07/2023 2:21 PM
10525 - 313 - Power off instrument; Remark:10523
Procedure Step
Published
04/07/2025 4:41 PM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
10523 - Press the Power button for 1 second.
Remark
Published
12/07/2023 2:23 PM
Operator Procedure
Document Type
Published
03/02/2020 3:31 PM
Warning
Annotation Type
Published
03/02/2020 3:25 PM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Viken Nighthawk-HBI Insert Battery
Image
Published
12/06/2023 4:50 PM
Viken Nighthawk-HBI Power Button
Image
Published
12/06/2023 4:50 PM
Viken Nighthawk home screen tap Viken icon
Image
Published
12/06/2023 5:04 PM
Viken Nighthawk type username password
Image
Published
12/06/2023 5:04 PM
Viken Nighthawk acknowledge warnings
Image
Published
12/06/2023 5:04 PM
Viken Nighthawk menu icon settings
Image
Published
12/07/2023 2:06 PM
Viken Nighthawk on sample radiation shield
Image
Published
12/07/2023 2:06 PM
Viken Nighthawk face down on sample radiation shield
Image
Published
12/07/2023 2:06 PM
Viken Nighthawk system check screen YES
Image
Published
12/07/2023 2:06 PM
Viken Nighthawk system check running
Image
Published
12/07/2023 2:06 PM
Viken Nighthawk system check successful OK
Image
Published
12/07/2023 2:06 PM
Viken Nighthawk conduct scan swipe slider
Image
Published
12/07/2023 2:23 PM
Viken Nighthawk press hold triggers
Image
Published
12/07/2023 2:23 PM
Viken Nighthawk scan item
Image
Published
12/07/2023 2:23 PM
OTH-OTH-VIK-NHK-OP01

Scan for contraband using the Nighthawk handheld backscatter x-ray device.

Equipment Hierarchy
Viken Detection Nighthawk HBI XP WAD
Technician Level
Duration
Materials Required
Battery pack
Sample radiation shield
Originators
Demboski, Erica
Patton, Greg
Originating Organization
PNNL
Approver
Simons, Brian
Date of Origin
Version
0.00
Operation
Step 1
Main Step
Step 2
Main Step
Step 3
Main Step

10542 - 10530 - Scan item of interest; Remark:10528

Step 00

10528 - The instrument must be moved to capture a two (2)-dimensional image. Moving t...

The instrument must be moved to capture a two (2)-dimensional image. Moving the instrument vertically from high to low is usually preferred.
Scan slowly and methodically. The blue arrow will slowly move to the left as the remaining time counts down to zero. Once the blue arrow reaches the left side of the screen, the instrument will no longer be armed.

Image
Viken Nighthawk scan item
Step 4
Document Type
Operator Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00