10000 - 10651 - Restart RPM - Decision, Notes: 9946

9947 - Reminder : There are a limited number of faults that can affect both pillars; troublesho...

Type
Reminder

There are a limited number of faults that can affect both pillars; troubleshoot each pillar independently starting with the control pillar. Failure of components that are common to both gamma detector channels should affect both channels similarly. If the problems in the channels are different (GH in one and GL in the other), troubleshoot the gamma channels separately.

9999 - Note : If there are no or low neutron counts on a channel, place a source on the aff...

Type
Note

If there are no or low neutron counts on a channel, place a source on the affected detector before performing troubleshooting steps

10651 - Restart RPM

Restart RPM
Yes
Go to
1544 - 546 - Done - Complete, Notes: 547
No
Go to
9998 - 9941 - Measure discriminator voltages - Decision, Notes: 9997
Note
9946 - Restart the RPM by cycling the on/off switch on the LD-260
8861 - Did this fix the Issue?