Last Updated: 05/02/2025 1:43 AM
The following table lists all the components that make up the document.
Use the Sensor Technology Engineering TA600 Radiation Pager in the interdiction and location of nuclear and other radioactive material.
247 - 214 - Pregătirea instrumentului pentru operare; Remark:
257 - Demontați instrumentul
Demontați instrumentul
786 - 219 - Verificați să nu existe surse de radiații în apropierea instrumentului; Remark:7
219 - Verificați să nu existe surse de radiații în apropierea instrumentului
Verificați să nu existe surse de radiații în apropierea instrumentului
7 - No Radiation symbol

7527 - 220 - Power on instrument; Remark:7511
220 - Power on instrument
Power on instrument
7511 - Select a mode for operation. (1) Audio only (2) Vibrate o...
Select a mode for operation.
(1) Audio only
(2) Vibrate only

7528 - 612 - Wait for instrument to complete start-up sequence; Remark:7512
612 - Wait for instrument to complete start-up sequence
Wait for instrument to complete start-up sequence
7512 - Red numerical LED display will flash while collecting background for 30 secon...
Red numerical LED display will flash while collecting background for 30 seconds.
Instrument is ready when alarm threshold number is displayed.


7529 - 319 - Perform operational test; Remark:7513
319 - Perform operational test
Perform operational test
7513 - It is recommended to perform an operational test daily before use. DET-HHD-ST...
It is recommended to perform an operational test daily before use. DET-HHD-STE-TA6-RM01, TA600 Radiation Pager Operational Test.
7530 - 7514 - Operate Instrument; Remark:
7514 - Operate Instrument
Operate Instrument
7531 - 4596 - Wear instrument during mission; Remark:7515
4596 - Wear instrument during mission
Wear instrument during mission
7515 - Carry instrument in holster attached to belt. Instrument remains powered on a...
Carry instrument in holster attached to belt. Instrument remains powered on at all times.

7532 - 1096 - Respond to alarms; Remark:7516
1096 - Respond to alarms
Respond to alarms
7516 - Momentarily press display button Radiation level is display...
Momentarily press display button
- Radiation level is displayed
- Yellow light flashes

7533 - 2905 - Follow standard operating procedures; Remark:7518
7517 - Note : An alarm is indicated by a numeric value (0-9) in the level display and a vib...
An alarm is indicated by a numeric value (0-9) in the level display and a vibration or audible tone. If the level display is 9, dose rate is 60 micro-Sieverts/hour (uSv/h) or greater. Operator should move to an area where levels are 8 or lower.
2905 - Follow standard operating procedures
Follow standard operating procedures
7518 - Standard operating procedures will define acceptable radiation levels from ra...
Standard operating procedures will define acceptable radiation levels from radioactive material.
7534 - 7519 - Locate radioactive material; Remark:
7519 - Locate radioactive material
Locate radioactive material
7535 - 7521 - Press and hold display button; Remark:7522
7520 - Note : It may take significant force to press and hold display button.
It may take significant force to press and hold display button.
7521 - Press and hold display button
Press and hold display button
7536 - 408 - Survey close to surface of item; Remark:7523
408 - Survey close to surface of item
Survey close to surface of item
7523 - Maintain a distance of approximately 5 to 10 centimeters (cm) from surface of...
Maintain a distance of approximately 5 to 10 centimeters (cm) from surface of item. Do not make contact.

7537 - 366 - Survey slowly over entire item; Remark:7524
366 - Survey slowly over entire item
Survey slowly over entire item
7524 - Move instrument approximately 20 cm per second. Thoroughly survey all accessi...
Move instrument approximately 20 cm per second. Thoroughly survey all accessible surfaces of item.

7538 - 367 - Observe instrument response; Remark:7525
367 - Observe instrument response
Observe instrument response
7525 - Numerical values and audible alerts or vibration rate will increase with radi...
Numerical values and audible alerts or vibration rate will increase with radiation levels. Use crosshair technique to localize.

405 - 368 - Marcați punctul/punctele de interes; Remark:390
368 - Marcați punctul/punctele de interes
Marcați punctul/punctele de interes
4277 - Plasați un marcaj (bandă adezivă, creion etc.) pe punctele care indică ni...
Plasați un marcaj (bandă adezivă, creion etc.) pe punctele care indică nivel de radiație crescut. Este posibil să existe mai mult de un punct de interes.

7540 - 313 - Power off instrument; Remark:7490
313 - Power off instrument
Power off instrument
7490 - Move mode switch to 0
Move mode switch to 0