Search for and identify radioactive material.
247 - 257 - Prepare instrument; Remark:
257 - Prepare instrument
Prepare instrument
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
219 - Verify there are no sources of radiation near instrument
Verify there are no sources of radiation near instrument
7 - No Radiation symbol

2293 - 220 - Power on instrument; Remark:2231
2232 - Note : If an administrator password has not previously been set up, the instrument w...
If an administrator password has not previously been set up, the instrument will start up in Administrator mode. With a password, the system will start up in User mode. Instructions in this procedure are written for User mode.
2233 - Note : If the screen dims due to inactivity, press any button except ENTER to “wak...
If the screen dims due to inactivity, press any button except ENTER to “wake up.”
220 - Power on instrument
Power on instrument
2231 - Press and hold the BACK/POWER button for at least 1 second.
Press and hold the BACK/POWER button for at least 1 second.
2292 - 2220 - Wait for measurement to complete; Remark:2261
2234 - Caution : Do not attempt to take a new background until the start-up sequence is comple...
Do not attempt to take a new background until the start-up sequence is complete. This could adversely affect identification performance of the instrument.
2220 - Wait for measurement to complete
Wait for measurement to complete
2261 - This may take 1 to 3 minutes. When the start-up sequence is complete, the scr...
This may take 1 to 3 minutes. When the start-up sequence is complete, the screen will display “AutoCal complete!”

2291 - 817 - Verify sufficient battery level is available to conduct operation; Remark:2169
817 - Verify sufficient battery level is available to conduct operation
Verify sufficient battery level is available to conduct operation
2290 - 2206 - Connect headphones (optional); Remark:2262
2206 - Connect headphones (optional)
Connect headphones (optional)
2262 - The headphone jack is located on the left side of the instrument.
The headphone jack is located on the left side of the instrument.

2289 - 2207 - Connect shoulder strap (optional); Remark:2263
2207 - Connect shoulder strap (optional)
Connect shoulder strap (optional)
2263 - Hooks for the shoulder strap are located on the front and back of the instrum...
Hooks for the shoulder strap are located on the front and back of the instrument.

11052 - 3494 - Collect background measurement; Remark:
3494 - Collect background measurement
Collect background measurement
2288 - 2209 - Verify instrument is in a suitable background radiation area; Remark:2264
2209 - Verify instrument is in a suitable background radiation area
Verify instrument is in a suitable background radiation area
2264 - The instrument should be located near the secondary inspection area, in simil...
The instrument should be located near the secondary inspection area, in similar surroundings, but not directly adjacent to the conveyance, object, or person.
2287 - 2210 - Select Background; Remark:2265
2210 - Select Background
Select Background
2265 - Use the left and right buttons on the keypad to highlight Background on the s...
Use the left and right buttons on the keypad to highlight Background on the screen and press the ENTER button.
2170 - IO PG RIIDEye Background Enter


2286 - 2211 - Press ENTER button again to respond to on-screen reminder; Remark:2171
2211 - Press ENTER button again to respond to on-screen reminder
Press ENTER button again to respond to on-screen reminder
2285 - 2220 - Wait for measurement to complete; Remark:2266
2220 - Wait for measurement to complete
Wait for measurement to complete
2266 - This will take 60 seconds. You also have the option to add or subtract 30 sec...
This will take 60 seconds. You also have the option to add or subtract 30 seconds from the measurement time.

2252 - 2213 - Search for Radioactive Material; Remark:
2213 - Search for Radioactive Material
Search for Radioactive Material
2284 - 2214 - Select search screen (Dial or Finder); Remark:2270
2267 - Note : There are two (2) options for the Search screen. The default display is Dial....
There are two (2) options for the Search screen. The default display is Dial. The user must select Finder for the alternate display.
2214 - Select search screen (Dial or Finder)
Select search screen (Dial or Finder)
2270 - The default Dial screen displays count rate and dose rate on the left. The di...
The default Dial screen displays count rate and dose rate on the left. The dial displays the intensity of the radiation being measured.

2269 - The Finder screen displays count rate and dose rate in the upper right corner...
The Finder screen displays count rate and dose rate in the upper right corner. It also plots time and cps [counts per second] on a graph.

2268 - The screen can be changed by selecting either “Dial” or “Finder” and ...
The screen can be changed by selecting either “Dial” or “Finder” and pressing ENTER.

2283 - 2215 - Survey slowly over entire item; Remark:2271
2215 - Survey slowly over entire item
Survey slowly over entire item
2271 - Keep the instrument close to the entity being assessed, but not touching....
Keep the instrument close to the entity being assessed, but not touching.
2282 - 2216 - Observe display and listen to audible indicator (if enabled); Remark:2174
2216 - Observe display and listen to audible indicator (if enabled)
Observe display and listen to audible indicator (if enabled)
405 - 368 - Mark point(s) of interest; Remark:4277
368 - Mark point(s) of interest
Mark point(s) of interest
4277 - Place a marker (tape, pencil, etc.) on points exhibiting elevated radiation. ...
Place a marker (tape, pencil, etc.) on points exhibiting elevated radiation. There may be more than one point of interest.

2251 - 2217 - Perform Identification Measurement; Remark:
2217 - Perform Identification Measurement
Perform Identification Measurement
2280 - 2218 - Place instrument close to measurement location; Remark:2187
2218 - Place instrument close to measurement location
Place instrument close to measurement location
2279 - 2219 - Select Identify; Remark:2188
2219 - Select Identify
Select Identify
2273 - The screen will prompt the operator to move the instrument back or move it cl...
The screen will prompt the operator to move the instrument back or move it closer to the source of radiation as needed.

2278 - 2220 - Wait for measurement to complete; Remark:2275
2274 - Note : More identification time can be added by navigating to +30 and pressing the E...
More identification time can be added by navigating to +30 and pressing the ENTER button. You can also cancel the identification by navigating to Cancel and pressing the ENTER button.
2220 - Wait for measurement to complete
Wait for measurement to complete
2277 - 2221 - Observe identification results; Remark:2276
2221 - Observe identification results
Observe identification results
2276 - The screen displays any isotopes found along with a statistical confidence nu...
The screen displays any isotopes found along with a statistical confidence number.
2250 - 2222 - Record identification results; Remark:2238
2222 - Record identification results
Record identification results
2238 - Follow your standard operating procedure to record results. Spectra are autom...
Follow your standard operating procedure to record results. Spectra are automatically saved to the CF card with the event number in the filename.
2249 - 2223 - Press BACK/POWER button to return to search mode, if necessary; Remark:
2223 - Press BACK/POWER button to return to search mode, if necessary
Press BACK/POWER button to return to search mode, if necessary
2248 - 313 - Power off instrument; Remark:10741
313 - Power off instrument
Power off instrument
10741 - Select MENU
Select MENU
10743 - Highlight Shutdown
Highlight Shutdown
10669 - Choose SELECT
Choose SELECT
10746 - Select OK
Select OK
2247 - 313 - Power off instrument; Remark:10741
313 - Power off instrument
Power off instrument
10741 - Select MENU
Select MENU
10743 - Highlight Shutdown
Highlight Shutdown
10669 - Choose SELECT
Choose SELECT
10746 - Select OK
Select OK
2246 - 2224 - Connect instrument to charger; Remark:
2224 - Connect instrument to charger
Connect instrument to charger
2245 - 2225 - Offload Spectra onto Computer; Remark:
2225 - Offload Spectra onto Computer
Offload Spectra onto Computer
2244 - 2226 - Remove CF card from rear access panel; Remark:2190
2226 - Remove CF card from rear access panel
Remove CF card from rear access panel
2243 - 2227 - Connect CF card to USB CF card reader; Remark:2191
2227 - Connect CF card to USB CF card reader
Connect CF card to USB CF card reader
2242 - 2228 - Connect USB CF card reader to computer; Remark:
2228 - Connect USB CF card reader to computer
Connect USB CF card reader to computer
2241 - 2229 - Use Windows Explorer to navigate to Spectra folder; Remark:2192
2229 - Use Windows Explorer to navigate to Spectra folder
Use Windows Explorer to navigate to Spectra folder
2240 - 2230 - Copy *.N42 files onto computer; Remark:
2230 - Copy *.N42 files onto computer
Copy *.N42 files onto computer