Last Updated: 05/01/2025 1:46 AM
The following table lists all the components that make up the document.
DET-HHD-RAP-PG01
Search for radioactive materials
Technician Level
Duration
Preparation
Step 1
Main Step
786 - 219 - Переконайтеся, що поряд з приладом немає радіоактивних джерел.; Remark:7
Step 00
219 - Verify there are no sources of radiation near instrument
Verify there are no sources of radiation near instrument
7 - No Radiation symbol
Image

Step 2
Step 3
Main Step
679 - 612 - ЗАЧЕКАЙТЕ , поки прилад завершить процес запуску.; Remark:
Step 00
612 - Wait for instrument to complete start-up sequence
Wait for instrument to complete start-up sequence
Step 4
Main Step
Step 6
Main Step
131 - 115 - SURVEY close to surface of item (5-10 centimeters); Remark:10467
Step 00
115 - SURVEY close to surface of item (5-10 centimeters)
SURVEY close to surface of item (5-10 centimeters)
10467 -
Image

Step 7
Main Step
10466 - 10465 - SURVEY SLOWLY over entire item (20 cm/s); Remark:
Step 00
10465 - SURVEY SLOWLY over entire item (20 cm/s)
SURVEY SLOWLY over entire item (20 cm/s)
Step 8
Main Step
10464 - 10463 - CHANGE to FIND MODE when alarm is constant; Remark:10462
Step 00
10463 - CHANGE to FIND MODE when alarm is constant
CHANGE to FIND MODE when alarm is constant
10462 -
Image

Step 9
Step 10
Main Step
133 - 368 - Позначити місце або місця, які потребують ...; Remark:132
Step 00
368 - Mark point(s) of interest
Mark point(s) of interest
Step 11
Main Step
136 - 117 - DOCUMENT results; Remark:
Step 00
117 - DOCUMENT results
DOCUMENT results
Step 12
Reminders
10458 - Reminder : It is recommended to perform an operational test daily before use. DET-HHD-RA...
Type
Reminder
It is recommended to perform an operational test daily before use. DET-HHD-RAP-RM01, PRM-470 Operational Test.
121 - Reminder : If instrument does not move for 60 seconds, it will switch to background mode...
Type
Reminder
If instrument does not move for 60 seconds, it will switch to background mode.
10459 - Reminder : Do not operate the instrument at temperatures lower than 0°C or higher than ...
Type
Reminder
Do not operate the instrument at temperatures lower than 0°C or higher than 38°C.
3076 - Reminder : Power off instrument prior to connecting to charger.
Type
Reminder
Power off instrument prior to connecting to charger.
Document Type
Pocket Guide
Percent Complete
100.00
Percent Complete W/Media
100.00