Last Updated: 04/25/2025 1:37 AM
The following table lists all the components that make up the document.
Title | Type | Status | Last Updated |
---|---|---|---|
Sensor Tech Radiation Pager | Device
|
Published
|
10/28/2024 3:52 PM
|
2899 - 2871 - Access alarm threshold control dial; Remark: | Procedure Step
|
Published
|
06/18/2021 6:02 PM
|
2871 - Access alarm threshold control dial | Instruction
|
Published
|
06/17/2021 12:29 PM
|
2949 - 313 - Power off instrument; Remark:2948 | Procedure Step
|
Published
|
06/17/2021 12:59 PM
|
313 - Power off instrument | Instruction
|
Published
|
07/23/2021 2:41 PM
|
2948 - Pager - Power off instrument | Remark
|
Published
|
07/06/2021 10:35 AM
|
3014 - 2857 - Remove two (2) battery compartment screws; Remark:2965 | Procedure Step
|
Published
|
06/18/2021 6:03 PM
|
2857 - Remove two (2) battery compartment screws | Instruction
|
Published
|
06/17/2021 12:31 PM
|
2965 - Use a small Phillips (PH1) screwdriver. | Remark
|
Published
|
07/02/2021 12:27 PM
|
3015 - 2858 - Remove battery compartment cover; Remark:2953 | Procedure Step
|
Published
|
06/18/2021 6:04 PM
|
2858 - Remove battery compartment cover | Instruction
|
Published
|
06/17/2021 12:31 PM
|
2953 - Pager - Remove battery compartment cover | Remark
|
Published
|
06/22/2021 2:49 PM
|
2956 - 2859 - Remove batteries; Remark:2955 | Procedure Step
|
Published
|
06/18/2021 5:54 PM
|
2859 - Remove batteries | Instruction
|
Published
|
06/17/2021 12:32 PM
|
2955 - Pager - Remove batteries | Remark
|
Published
|
07/06/2021 10:36 AM
|
2900 - 2876 - Adjust threshold; Remark: | Procedure Step
|
Published
|
06/18/2021 6:05 PM
|
2876 - Adjust threshold | Instruction
|
Published
|
06/17/2021 12:33 PM
|
3017 - 2877 - Turn dial to change threshold; Remark:3016 | Procedure Step
|
Published
|
06/18/2021 6:06 PM
|
2886 - Caution : Adjusting the threshold too high will result in poor detection sensitivity. A... | Annotation
|
Published
|
06/16/2021 2:32 PM
|
2887 - Note : This is an iterative process which might require multiple trials to achieve d... | Annotation
|
Published
|
06/16/2021 2:33 PM
|
2877 - Turn dial to change threshold | Instruction
|
Published
|
06/17/2021 12:34 PM
|
3016 - Use a small slotted screwdriver. Turn clockwise to increase threshold (reduce... | Remark
|
Published
|
07/02/2021 12:27 PM
|
3019 - 2879 - Insert batteries; Remark:3018 | Procedure Step
|
Published
|
06/18/2021 6:08 PM
|
2879 - Insert batteries | Instruction
|
Published
|
06/17/2021 12:35 PM
|
3018 - Do not replace battery cover. | Remark
|
Published
|
07/29/2021 3:56 PM
|
3023 - 220 - Power on instrument; Remark:3022 | Procedure Step
|
Published
|
06/18/2021 6:09 PM
|
220 - Power on instrument | Instruction
|
Published
|
04/06/2025 7:00 PM
|
3022 - Pager - Power on instrument | Remark
|
Published
|
06/17/2021 12:45 PM
|
3025 - 2880 - Evaluate alarms; Remark:3024 | Procedure Step
|
Published
|
06/18/2021 6:09 PM
|
2880 - Evaluate alarms | Instruction
|
Published
|
06/17/2021 12:36 PM
|
3024 - Determine if instrument exhibits an unacceptable number of false alarms when ... | Remark
|
Published
|
06/23/2021 3:02 PM
|
2890 - 2882 - Repeat adjustment if necessary; Remark:2883 | Procedure Step
|
Published
|
06/18/2021 6:10 PM
|
2882 - Repeat adjustment if necessary | Instruction
|
Published
|
06/17/2021 12:37 PM
|
2883 - If too many false alarms, repeat steps to increase threshold. | Remark
|
Published
|
06/17/2021 12:47 PM
|
303 - 302 - Reassemble instrument; Remark: | Procedure Step
|
Published
|
10/07/2022 1:49 PM
|
302 - Reassemble instrument | Instruction
|
Published
|
06/16/2021 11:28 PM
|
2963 - 2862 - Replace battery compartment cover; Remark:2962 | Procedure Step
|
Published
|
06/18/2021 5:56 PM
|
2862 - Replace battery compartment cover | Instruction
|
Published
|
06/17/2021 12:38 PM
|
2962 - Cover will snap into place. | Remark
|
Published
|
06/23/2021 2:58 PM
|
3031 - 2863 - Insert two (2) battery compartment screws; Remark:3030 | Procedure Step
|
Published
|
06/18/2021 6:11 PM
|
569 - Caution : Do not overtighten screws. | Annotation
|
Published
|
06/16/2021 2:31 PM
|
2863 - Insert two (2) battery compartment screws | Instruction
|
Published
|
06/17/2021 12:38 PM
|
3030 - Pager - Battery compartment screws | Remark
|
Published
|
06/17/2021 12:49 PM
|
2868 - 2864 - Check operation; Remark:2865 | Procedure Step
|
Published
|
06/18/2021 5:57 PM
|
2864 - Check operation | Instruction
|
Published
|
06/17/2021 12:39 PM
|
2865 - DET-HHD-STE-RPG-RM01, Radiation Pager Operational Test. | Remark
|
Published
|
08/02/2022 12:02 PM
|
322 - 321 - Document maintenance actions; Remark: | Procedure Step
|
Published
|
06/17/2021 9:49 AM
|
321 - Document maintenance actions | Instruction
|
Published
|
04/21/2025 11:19 AM
|
2869 - 2585 - Document maintenance performed; Remark:2866 | Procedure Step
|
Published
|
06/18/2021 5:58 PM
|
2585 - Document maintenance performed | Instruction
|
Published
|
07/09/2021 12:15 PM
|
2866 - Record observations, times, and results for the maintenance report. | Remark
|
Published
|
04/21/2025 11:39 AM
|
4706 - 4705 - Submit report; Remark:4864 | Procedure Step
|
Published
|
06/29/2022 9:50 AM
|
4705 - Submit report | Instruction
|
Published
|
06/29/2022 9:50 AM
|
4864 - As specified by management or contractual obligations. | Remark
|
Published
|
07/28/2022 1:36 PM
|
Corrective Maintenance Procedure | Document Type
|
Published
|
03/02/2020 3:32 PM
|
15 minutes | Duration
|
Published
|
04/01/2020 11:35 AM
|
As Needed | Frequency
|
Published
|
04/01/2020 11:36 AM
|
Caution | Annotation Type
|
Published
|
03/02/2020 3:24 PM
|
Note | Annotation Type
|
Published
|
03/02/2020 3:25 PM
|
Technician 2 | Skill Level
|
Published
|
04/09/2020 1:17 AM
|
Small Phillips (PH1) screwdriver | Tools
|
Published
|
05/14/2020 2:48 PM
|
Small slotted (flathead) screwdriver | Tools
|
Published
|
09/11/2020 11:23 AM
|
power_off_000564.png | Image
|
Published
|
03/26/2021 1:40 AM
|
Insert Screws into cover | Image
|
Published
|
03/27/2021 1:42 AM
|
Remove battery cover | Image
|
Published
|
03/27/2021 1:42 AM
|
Remove_Battery_000567.png | Image
|
Published
|
03/27/2021 1:42 AM
|
Change Threshold on Radiation Pager | Image
|
Published
|
03/27/2021 1:42 AM
|
stepager_di_batteryin_000563.png | Image
|
Published
|
03/24/2021 2:46 AM
|
Radiation Pager Power On | Image
|
Published
|
03/27/2021 1:42 AM
|
Radiation Pager evaluate alarm | Image
|
Published
|
03/27/2021 1:42 AM
|
Replace battery cover on radiation page | Image
|
Published
|
03/27/2021 1:42 AM
|
How to adjust the alarm threshold on the radiation pager. This is performed to minimize nuisance alarms in high radiation background environments.
2899 - 2871 - Access alarm threshold control dial; Remark:
2871 - Access alarm threshold control dial
Access alarm threshold control dial
3014 - 2857 - Remove two (2) battery compartment screws; Remark:2965
2857 - Remove two (2) battery compartment screws
Remove two (2) battery compartment screws
2965 - Use a small Phillips (PH1) screwdriver.
Use a small Phillips (PH1) screwdriver.

3015 - 2858 - Remove battery compartment cover; Remark:2953
2858 - Remove battery compartment cover
Remove battery compartment cover
2900 - 2876 - Adjust threshold; Remark:
2876 - Adjust threshold
Adjust threshold
3017 - 2877 - Turn dial to change threshold; Remark:3016
2886 - Caution : Adjusting the threshold too high will result in poor detection sensitivity. A...
Adjusting the threshold too high will result in poor detection sensitivity. An alarm threshold higher than five (5) is not recommended.
2887 - Note : This is an iterative process which might require multiple trials to achieve d...
This is an iterative process which might require multiple trials to achieve desired results.
2877 - Turn dial to change threshold
Turn dial to change threshold
3016 - Use a small slotted screwdriver. Turn clockwise to increase threshold (reduce...
Use a small slotted screwdriver. Turn clockwise to increase threshold (reduce false alarms). Turn counterclockwise to decrease threshold (increase sensitivity). Dial can only be set to integer values.

3019 - 2879 - Insert batteries; Remark:3018
2879 - Insert batteries
Insert batteries
3018 - Do not replace battery cover.
Do not replace battery cover.

3025 - 2880 - Evaluate alarms; Remark:3024
2880 - Evaluate alarms
Evaluate alarms
3024 - Determine if instrument exhibits an unacceptable number of false alarms when ...
Determine if instrument exhibits an unacceptable number of false alarms when no sources of radiation are near.

2890 - 2882 - Repeat adjustment if necessary; Remark:2883
2882 - Repeat adjustment if necessary
Repeat adjustment if necessary
2883 - If too many false alarms, repeat steps to increase threshold.
If too many false alarms, repeat steps to increase threshold.
303 - 302 - Reassemble instrument; Remark:
302 - Reassemble instrument
Reassemble instrument
2963 - 2862 - Replace battery compartment cover; Remark:2962
2862 - Replace battery compartment cover
Replace battery compartment cover
2962 - Cover will snap into place.
Cover will snap into place.

3031 - 2863 - Insert two (2) battery compartment screws; Remark:3030
569 - Caution : Do not overtighten screws.
Do not overtighten screws.
2863 - Insert two (2) battery compartment screws
Insert two (2) battery compartment screws
2868 - 2864 - Check operation; Remark:2865
2864 - Check operation
Check operation
2865 - DET-HHD-STE-RPG-RM01, Radiation Pager Operational Test.
DET-HHD-STE-RPG-RM01, Radiation Pager Operational Test.
322 - 321 - Document maintenance actions; Remark:
321 - Document maintenance actions
Document maintenance actions
2869 - 2585 - Document maintenance performed; Remark:2866
2585 - Document maintenance performed
Document maintenance performed
2866 - Record observations, times, and results for the maintenance report.
Record observations, times, and results for the maintenance report.
4706 - 4705 - Submit report; Remark:4864
4705 - Submit report
Submit report
4864 - As specified by management or contractual obligations.
As specified by management or contractual obligations.