Last Updated: 04/25/2025 1:37 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Sensor Tech Radiation Pager
Device
Published
10/28/2024 3:52 PM
2899 - 2871 - Access alarm threshold control dial; Remark:
Procedure Step
Published
06/18/2021 6:02 PM
2871 - Access alarm threshold control dial
Instruction
Published
06/17/2021 12:29 PM
2949 - 313 - Power off instrument; Remark:2948
Procedure Step
Published
06/17/2021 12:59 PM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
2948 - Pager - Power off instrument
Remark
Published
07/06/2021 10:35 AM
3014 - 2857 - Remove two (2) battery compartment screws; Remark:2965
Procedure Step
Published
06/18/2021 6:03 PM
2857 - Remove two (2) battery compartment screws
Instruction
Published
06/17/2021 12:31 PM
2965 - Use a small Phillips (PH1) screwdriver.
Remark
Published
07/02/2021 12:27 PM
3015 - 2858 - Remove battery compartment cover; Remark:2953
Procedure Step
Published
06/18/2021 6:04 PM
2858 - Remove battery compartment cover
Instruction
Published
06/17/2021 12:31 PM
2953 - Pager - Remove battery compartment cover
Remark
Published
06/22/2021 2:49 PM
2956 - 2859 - Remove batteries; Remark:2955
Procedure Step
Published
06/18/2021 5:54 PM
2859 - Remove batteries
Instruction
Published
06/17/2021 12:32 PM
2955 - Pager - Remove batteries
Remark
Published
07/06/2021 10:36 AM
2900 - 2876 - Adjust threshold; Remark:
Procedure Step
Published
06/18/2021 6:05 PM
2876 - Adjust threshold
Instruction
Published
06/17/2021 12:33 PM
3017 - 2877 - Turn dial to change threshold; Remark:3016
Procedure Step
Published
06/18/2021 6:06 PM
2886 - Caution : Adjusting the threshold too high will result in poor detection sensitivity. A...
Annotation
Published
06/16/2021 2:32 PM
2887 - Note : This is an iterative process which might require multiple trials to achieve d...
Annotation
Published
06/16/2021 2:33 PM
2877 - Turn dial to change threshold
Instruction
Published
06/17/2021 12:34 PM
3016 - Use a small slotted screwdriver. Turn clockwise to increase threshold (reduce...
Remark
Published
07/02/2021 12:27 PM
3019 - 2879 - Insert batteries; Remark:3018
Procedure Step
Published
06/18/2021 6:08 PM
2879 - Insert batteries
Instruction
Published
06/17/2021 12:35 PM
3018 - Do not replace battery cover.
Remark
Published
07/29/2021 3:56 PM
3023 - 220 - Power on instrument; Remark:3022
Procedure Step
Published
06/18/2021 6:09 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
3022 - Pager - Power on instrument
Remark
Published
06/17/2021 12:45 PM
3025 - 2880 - Evaluate alarms; Remark:3024
Procedure Step
Published
06/18/2021 6:09 PM
2880 - Evaluate alarms
Instruction
Published
06/17/2021 12:36 PM
3024 - Determine if instrument exhibits an unacceptable number of false alarms when ...
Remark
Published
06/23/2021 3:02 PM
2890 - 2882 - Repeat adjustment if necessary; Remark:2883
Procedure Step
Published
06/18/2021 6:10 PM
2882 - Repeat adjustment if necessary
Instruction
Published
06/17/2021 12:37 PM
2883 - If too many false alarms, repeat steps to increase threshold.
Remark
Published
06/17/2021 12:47 PM
303 - 302 - Reassemble instrument; Remark:
Procedure Step
Published
10/07/2022 1:49 PM
302 - Reassemble instrument
Instruction
Published
06/16/2021 11:28 PM
2963 - 2862 - Replace battery compartment cover; Remark:2962
Procedure Step
Published
06/18/2021 5:56 PM
2862 - Replace battery compartment cover
Instruction
Published
06/17/2021 12:38 PM
2962 - Cover will snap into place.
Remark
Published
06/23/2021 2:58 PM
3031 - 2863 - Insert two (2) battery compartment screws; Remark:3030
Procedure Step
Published
06/18/2021 6:11 PM
569 - Caution : Do not overtighten screws.
Annotation
Published
06/16/2021 2:31 PM
2863 - Insert two (2) battery compartment screws
Instruction
Published
06/17/2021 12:38 PM
3030 - Pager - Battery compartment screws
Remark
Published
06/17/2021 12:49 PM
2868 - 2864 - Check operation; Remark:2865
Procedure Step
Published
06/18/2021 5:57 PM
2864 - Check operation
Instruction
Published
06/17/2021 12:39 PM
2865 - DET-HHD-STE-RPG-RM01, Radiation Pager Operational Test.
Remark
Published
08/02/2022 12:02 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
15 minutes
Duration
Published
04/01/2020 11:35 AM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Caution
Annotation Type
Published
03/02/2020 3:24 PM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 2
Skill Level
Published
04/09/2020 1:17 AM
Small Phillips (PH1) screwdriver
Tools
Published
05/14/2020 2:48 PM
Small slotted (flathead) screwdriver
Tools
Published
09/11/2020 11:23 AM
power_off_000564.png
Image
Published
03/26/2021 1:40 AM
Insert Screws into cover
Image
Published
03/27/2021 1:42 AM
Remove battery cover
Image
Published
03/27/2021 1:42 AM
Remove_Battery_000567.png
Image
Published
03/27/2021 1:42 AM
Change Threshold on Radiation Pager
Image
Published
03/27/2021 1:42 AM
stepager_di_batteryin_000563.png
Image
Published
03/24/2021 2:46 AM
Radiation Pager Power On
Image
Published
03/27/2021 1:42 AM
Radiation Pager evaluate alarm
Image
Published
03/27/2021 1:42 AM
Replace battery cover on radiation page
Image
Published
03/27/2021 1:42 AM
DET-HHD-STE-RPG-CM02

How to adjust the alarm threshold on the radiation pager. This is performed to minimize nuisance alarms in high radiation background environments.

Equipment Hierarchy
Sensor Tech Radiation Pager
Frequency
As Needed
Technician Level
Technician Level
Technician 2
x 1
Duration
Time increment
15 minutes
Length of time
15
Tools Required
Small Phillips (PH1) screwdriver
Small slotted (flathead) screwdriver
Materials Required
None
Retest Requirements
Final Documentation
Corrective Maintenance Report
Step 2

3017 - 2877 - Turn dial to change threshold; Remark:3016

Step 00
Step 4
Main Step
Document Type
Corrective Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00