Last Updated: 04/25/2025 1:40 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Thermo FHT1377 PackEye
Device
Published
10/28/2024 3:37 PM
5767 - 5766 - Uninstall Faulty Power Supply Module; Remark:
Procedure Step
Published
11/03/2022 2:45 PM
5766 - Uninstall Faulty Power Supply Module
Instruction
Published
11/03/2022 2:45 PM
5660 - 313 - Power off instrument; Remark:4086
Procedure Step
Published
04/07/2025 4:34 PM
313 - Power off instrument
Instruction
Published
07/23/2021 2:41 PM
4086 - Press and hold the power button for approximately 10 seconds until indicator ...
Remark
Published
05/02/2022 6:04 PM
5666 - 4222 - Place instrument on flat surface; Remark:4221
Procedure Step
Published
11/03/2022 8:17 AM
4222 - Place instrument on flat surface
Instruction
Published
05/10/2022 1:54 PM
4221 - Shoulder straps should be facing down.
Remark
Published
05/10/2022 1:55 PM
5667 - 3876 - Open top of backpack; Remark:5665
Procedure Step
Published
11/03/2022 8:21 AM
3876 - Open top of backpack
Instruction
Published
04/04/2022 3:14 PM
5665 - IO PackEye Open Top for Battery
Remark
Published
11/09/2024 8:06 PM
5668 - 4219 - Remove distribution box; Remark:4218
Procedure Step
Published
11/03/2022 8:22 AM
4219 - Remove distribution box
Instruction
Published
05/10/2022 1:58 PM
4218 - The distribution box is usually located in the top of the backpack.
Remark
Published
05/10/2022 1:58 PM
5669 - 4217 - Disconnect battery cable; Remark:4216
Procedure Step
Published
11/03/2022 8:47 AM
5664 - Note : The other cord connected to the distribution box is the headphone plug.
Annotation
Published
11/03/2022 8:47 AM
4217 - Disconnect battery cable
Instruction
Published
05/10/2022 2:01 PM
4216 - Rotate the collar counterclockwise and pull straight out. The battery cable c...
Remark
Published
05/10/2022 2:03 PM
5673 - 5672 - Disconnect RS232 jumper; Remark:5670
Procedure Step
Published
11/03/2022 9:07 AM
5672 - Disconnect RS232 jumper
Instruction
Published
11/03/2022 9:03 AM
5670 - Use small, slotted screwdriver.
Remark
Published
11/03/2022 9:07 AM
5671 - IO PackEye RS232 Disconnect Cable
Remark
Published
11/09/2024 8:07 PM
5680 - 5679 - Disconnect Bluetooth adapter cable; Remark:5670
Procedure Step
Published
11/03/2022 9:15 AM
5679 - Disconnect Bluetooth adapter cable
Instruction
Published
11/03/2022 9:12 AM
5678 - IO PackEye Cable for BTA Disconnect
Remark
Published
11/09/2024 8:08 PM
5681 - 5677 - Remove wire ties; Remark:5676
Procedure Step
Published
11/03/2022 9:27 AM
5677 - Remove wire ties
Instruction
Published
11/03/2022 9:16 AM
5676 - Note location of wire ties for reassembly. Use wire cutter or scissors.
Remark
Published
11/03/2022 9:27 AM
5770 - IO PackEye Power Cable Ties Location
Remark
Published
11/09/2024 8:08 PM
5682 - 5675 - Remove power supply module; Remark:5674
Procedure Step
Published
11/03/2022 9:35 AM
5675 - Remove power supply module
Instruction
Published
11/03/2022 9:31 AM
5674 - There are Velcro strips securing the power supply module to the FHT 681 ampli...
Remark
Published
11/03/2022 9:35 AM
5765 - 5764 - Install New Power Supply Module; Remark:
Procedure Step
Published
11/03/2022 2:46 PM
5764 - Install New Power Supply Module
Instruction
Published
11/03/2022 2:45 PM
5843 - 5775 - Secure power supply module to FHT 681; Remark:5774
Procedure Step
Published
11/04/2022 10:56 AM
5775 - Secure power supply module to FHT 681
Instruction
Published
11/03/2022 2:46 PM
5774 - The module will snap into place.
Remark
Published
11/03/2022 2:50 PM
5844 - 5773 - Reconnect battery cable; Remark:5772
Procedure Step
Published
11/04/2022 10:57 AM
5773 - Reconnect battery cable
Instruction
Published
11/03/2022 2:51 PM
5772 - Align the connector and push in. Rotate collar clockwise and hand-tighten.
Remark
Published
11/03/2022 2:55 PM
5845 - 5771 - Install wire ties; Remark:5769
Procedure Step
Published
11/04/2022 10:59 AM
5771 - Install wire ties
Instruction
Published
11/03/2022 2:56 PM
5769 - If desired, install new wire ties and cut off excess.
Remark
Published
11/03/2022 3:05 PM
5846 - 4237 - Replace distribution box; Remark:4236
Procedure Step
Published
11/04/2022 11:03 AM
4237 - Replace distribution box
Instruction
Published
05/10/2022 2:15 PM
4236 - IO PackEye Distro Box Install
Remark
Published
10/24/2024 6:53 PM
5847 - 5788 - Reconnect Bluetooth adapter cable; Remark:5787
Procedure Step
Published
11/04/2022 11:05 AM
5788 - Reconnect Bluetooth adapter cable
Instruction
Published
11/03/2022 3:17 PM
5787 - IO PackEye Cable for BTA Connect
Remark
Published
11/09/2024 8:13 PM
5848 - 5786 - Reconnect RS232 jumper; Remark:5785
Procedure Step
Published
11/04/2022 11:06 AM
5786 - Reconnect RS232 jumper
Instruction
Published
11/03/2022 3:39 PM
5785 - Use small, slotted screwdriver to secure connector.
Remark
Published
11/16/2022 9:35 AM
5830 - IO PackEye RS232 Connect Cable
Remark
Published
11/09/2024 8:14 PM
5763 - 5762 - Check Battery Charging; Remark:
Procedure Step
Published
11/03/2022 2:48 PM
5762 - Check Battery Charging
Instruction
Published
11/03/2022 2:48 PM
5798 - 4718 - Connect charger; Remark:5797
Procedure Step
Published
11/03/2022 3:57 PM
4718 - Connect charger
Instruction
Published
06/29/2022 1:15 PM
5797 - IO PackEye Battery Charger Connection
Remark
Published
11/09/2024 8:15 PM
5807 - IO PackEye Battery Connect Cable
Remark
Published
11/09/2024 8:15 PM
5799 - 4716 - Verify instrument is charging; Remark:5752
Procedure Step
Published
11/03/2022 4:01 PM
4716 - Verify instrument is charging
Instruction
Published
06/29/2022 1:16 PM
5752 - CHARGE (yellow) and READY (green) LEDs will...
Remark
Published
11/10/2022 9:17 AM
5853 - 5796 - Charge Instrument; Remark:5795
Procedure Step
Published
11/04/2022 11:15 AM
5796 - Charge Instrument
Instruction
Published
11/03/2022 4:02 PM
5795 - It is recommended to fully charge the battery before first use. Only the gree...
Remark
Published
11/03/2022 4:04 PM
5761 - 319 - Perform operational test; Remark:5760
Procedure Step
Published
11/03/2022 2:58 PM
319 - Perform operational test
Instruction
Published
06/10/2021 8:31 PM
5760 - DET-HHD-THR-PAK-RM01, PackEye Operational Test
Remark
Published
11/03/2022 2:55 PM
322 - 321 - Document maintenance actions; Remark:
Procedure Step
Published
06/17/2021 9:49 AM
321 - Document maintenance actions
Instruction
Published
04/21/2025 11:19 AM
2869 - 2585 - Document maintenance performed; Remark:2866
Procedure Step
Published
06/18/2021 5:58 PM
2585 - Document maintenance performed
Instruction
Published
07/09/2021 12:15 PM
2866 - Record observations, times, and results for the maintenance report.
Remark
Published
04/21/2025 11:39 AM
4706 - 4705 - Submit report; Remark:4864
Procedure Step
Published
06/29/2022 9:50 AM
4705 - Submit report
Instruction
Published
06/29/2022 9:50 AM
4864 - As specified by management or contractual obligations.
Remark
Published
07/28/2022 1:36 PM
Corrective Maintenance Procedure
Document Type
Published
03/02/2020 3:32 PM
minutes
Duration
Published
05/26/2022 11:41 AM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Note
Annotation Type
Published
03/02/2020 3:25 PM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
Small slotted (flathead) screwdriver
Tools
Published
09/11/2020 11:23 AM
Wire cutter pliers or scissors
Tools
Published
11/03/2022 2:20 PM
Medium pliers
Tools
Published
11/03/2022 2:20 PM
Power off Packeye
Image
Published
02/19/2022 1:48 AM
packeye_di_opentop_redarrow_00758.jpg
Image
Published
10/20/2023 12:55 PM
packeye_di_distrobox_remove.jpg
Image
Published
10/20/2023 4:00 PM
packeye_di_batterycable_remove.jpg
Image
Published
10/20/2023 3:54 PM
packeye_di_batterycable_2prong.jpg
Image
Published
10/20/2023 3:51 PM
packeye_di_rs232cable_remove.jpg
Image
Published
10/23/2023 9:49 AM
packeye_di_rs232btadisconnect.jpg
Image
Published
10/23/2023 1:46 PM
packeye_di_powercable_ties_locate.jpg
Image
Published
10/20/2023 4:02 PM
packeye_di_powermodule_remove.jpg
Image
Published
10/20/2023 4:04 PM
packeye_di_powermodule_install.jpg
Image
Published
10/20/2023 4:03 PM
packeye_di_batterycable_install.jpg
Image
Published
10/20/2023 3:53 PM
packeye_di_distrobox_install.jpg
Image
Published
10/20/2023 3:59 PM
packeye_di_rs232btaconnectinsert.jpg
Image
Published
10/23/2023 1:44 PM
packeye_di_rs232cable_insert.jpg
Image
Published
10/23/2023 9:48 AM
packeye_di_battery_chargerconnector.jpg
Image
Published
10/23/2023 2:17 PM
packeye_di_charger_connect_redarrow_00783.jpg
Image
Published
10/20/2023 12:53 PM
packeye_di_batteryCHARGING_0.jpg
Image
Published
10/23/2023 1:43 PM
packeye_di_batteryREADY.jpg
Image
Published
10/23/2023 9:34 AM
DET-HHD-THR-PAK-CM03

Replace power supply module in Thermo Scientific FHT 1377 PackEye.

Equipment Hierarchy
Thermo FHT1377 PackEye
Frequency
As Needed
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
minutes
Length of time
10
Tools Required
Small slotted (flathead) screwdriver
Wire cutter pliers or scissors
Medium pliers
Materials Required
PackEye power supply module (Power supply FHT 1377 Thermo Part # S25505110)
Wire ties
Originators
Kilgannon, Erica
Tucker, Brian
Originating Organization
PNNL
Approver
Cody Hostick
Safety Concerns
None
Retest Requirements
Final Documentation
Corrective Maintenance Report
Departmental Coordination

Prior to arrival, notify operators concerning expected work, duration, and anticipated alarms and fault indications.

Date of Origin
Version
0.00
Step 1

5669 - 4217 - Disconnect battery cable; Remark:4216

Step 00
Step 2
Step 4
Main Step
Document Type
Corrective Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media
100.00