Last Updated: 04/25/2025 1:41 AM
The following table lists all the components that make up the document.
Replace gamma detector in Thermo Scientific FHT 1377 PackEye.
DET-HHD-THR-PAK-RM05, PackEye Gamma High Voltage Adjustment
DET-HHD-THR-PAK-RM08, PackEye Efficiency Test
Prior to arrival, notify operators concerning expected work, duration, and anticipated alarms and fault indications.
6676 - 6675 - Unpack Instrument; Remark:
6675 - Unpack Instrument
Unpack Instrument
3869 - 313 - Power off instrument; Remark:3850
313 - Power off instrument
Power off instrument
3850 - Press and hold the power button for approximately 10 seconds until indicator ...
Press and hold the power button for approximately 10 seconds until indicator lights are off.

4223 - 4222 - Place instrument on flat surface; Remark:4221
4222 - Place instrument on flat surface
Place instrument on flat surface
4221 - Shoulder straps should be facing down.
Shoulder straps should be facing down.
3926 - 3876 - Open top of backpack; Remark:3928
3876 - Open top of backpack
Open top of backpack
6217 - 6216 - Disconnect straps on sides of backpack; Remark:6215
6216 - Disconnect straps on sides of backpack
Disconnect straps on sides of backpack
6215 -

6709 - 6708 - Remove backpack contents; Remark:6707
6708 - Remove backpack contents
Remove backpack contents
6707 - Slowly slide foam assembly out of backpack. This process might take two peopl...
Slowly slide foam assembly out of backpack. This process might take two people; one to pull the backpack and the other to pull the foam assembly.

6710 - 6706 - Open protective foam; Remark:6705
6706 - Open protective foam
Open protective foam
6705 - Use utility knife or scissors. Cut tape on one side of foam assembly.
Use utility knife or scissors. Cut tape on one side of foam assembly.

6674 - 6673 - Remove Faulty Gamma Detector; Remark:
6673 - Remove Faulty Gamma Detector
Remove Faulty Gamma Detector
6225 - 6221 - Disconnect gamma detector; Remark:6220
6221 - Disconnect gamma detector
Disconnect gamma detector
6220 - The gamma detector is connected to the FHT 681 amplifier at PMT1
The gamma detector is connected to the FHT 681 amplifier at PMT1. Push toward connection, turn coupling nut counterclockwise, and pull away from connection.

6724 - 1673 - Remove gamma detector; Remark:6723
1673 - Remove gamma detector
Remove gamma detector
6725 - 6722 - Disconnect cable from gamma detector; Remark:6721
6722 - Disconnect cable from gamma detector
Disconnect cable from gamma detector
6721 - Keep the right-angle connector on the cable. Push toward connection, turn cou...
Keep the right-angle connector on the cable. Push toward connection, turn coupling nut counterclockwise, and pull away from connection.

6672 - 6671 - Install New Gamma Detector; Remark:
6671 - Install New Gamma Detector
Install New Gamma Detector
6719 - 6328 - Reseat gamma connectors; Remark:6718
6328 - Reseat gamma connectors
Reseat gamma connectors
6718 - Disconnect and reconnect both connections to the 90-degree elbow several time...
Disconnect and reconnect both connections to the 90-degree elbow several times to remove oxidation.

6720 - 6717 - Set gamma detector in place; Remark:6716
6717 - Set gamma detector in place
Set gamma detector in place
6716 - Insert the gamma detector into the foam. Route the cable through the foam to ...
Insert the gamma detector into the foam. Route the cable through the foam to the FHT681.
6332 - 6328 - Reseat gamma connectors; Remark:6326
6328 - Reseat gamma connectors
Reseat gamma connectors
6326 - The gamma detector is connected at the FHT 681 amplifier at the connection la...
The gamma detector is connected at the FHT 681 amplifier at the connection labeled PMT1. Disconnect and reconnect both connections to the 90-degree elbow several times to remove oxidation.
6327 -

303 - 302 - Reassemble instrument; Remark:
302 - Reassemble instrument
Reassemble instrument
6737 - 6736 - Secure protective foam together; Remark:6735
6736 - Secure protective foam together
Secure protective foam together
6735 - Re-wrap the foam assembly with electrical tape. Do not compress the foam....
Re-wrap the foam assembly with electrical tape. Do not compress the foam.

6333 - 6325 - Repack backpack; Remark:6323
6325 - Repack backpack
Repack backpack
6323 - Slide the protective foam assembly into backpack
Slide the protective foam assembly into backpack
6324 -

6342 - 6341 - Perform Gamma High Voltage Adjustment; Remark:6340
6341 - Perform Gamma High Voltage Adjustment
Perform Gamma High Voltage Adjustment
6340 - DET-HHD-THR-PAK-RM05, PackEye Gamma High Voltage Adjustment.
DET-HHD-THR-PAK-RM05, PackEye Gamma High Voltage Adjustment.
6435 - 2584 - Perform efficiency test; Remark:6434
2584 - Perform efficiency test
Perform efficiency test
6434 - DET-HHD-THR-RM08, PackEye Efficiency Test.
DET-HHD-THR-RM08, PackEye Efficiency Test.
5761 - 319 - Perform operational test; Remark:5760
319 - Perform operational test
Perform operational test
5760 - DET-HHD-THR-PAK-RM01, PackEye Operational Test
DET-HHD-THR-PAK-RM01, PackEye Operational Test
322 - 321 - Document maintenance actions; Remark:
321 - Document maintenance actions
Document maintenance actions
2869 - 2585 - Document maintenance performed; Remark:2866
2585 - Document maintenance performed
Document maintenance performed
2866 - Record observations, times, and results for the maintenance report.
Record observations, times, and results for the maintenance report.
4706 - 4705 - Submit report; Remark:4864
4705 - Submit report
Submit report
4864 - As specified by management or contractual obligations.
As specified by management or contractual obligations.