Last Updated: 05/02/2025 1:38 AM
The following table lists all the components that make up the document.
Remove and replace identiFINDER battery pack
3192 - 3174 - Prior to beginning, brief system operators on expected work, duration, and te...; Remark:
3174 - Prior to beginning, brief system operators on expected work, duration, and te...
Prior to beginning, brief system operators on expected work, duration, and temporary lack of instrument availability
3193 - 3173 - Loosen screw at end of handle; Remark:3181
3189 - Note : Small flat-blade screwdriver if using the powerPACK, Phillips PH1 screwdriver...
Small flat-blade screwdriver if using the powerPACK, Phillips PH1 screwdriver if using the powerPACK 2
3173 - Loosen screw at end of handle
Loosen screw at end of handle
3181 - Do not remove the screw completely; just loosen until it freely slides in its...
Do not remove the screw completely; just loosen until it freely slides in its bearing.

3194 - 3175 - Slide handle towards back of instrument until it slips free; Remark:3182
3175 - Slide handle towards back of instrument until it slips free
Slide handle towards back of instrument until it slips free
3195 - 3176 - Remove existing battery pack from handle; Remark:3183
3176 - Remove existing battery pack from handle
Remove existing battery pack from handle
3196 - 3177 - Insert replacement battery pack into handle; Remark:3184
3190 - Note : The powerPACK and powerPACK 2 are containers made to insert four (4) AA batte...
The powerPACK and powerPACK 2 are containers made to insert four (4) AA batteries (Alkaline or NiMH) and use as a backup for the rechargeable battery packs. Both of these containers have internal circuitry that prevents them from charging while they are installed in the identiFINDER.
3191 - Note : The powerPACK plus, ultra, and ultra 2 are rechargeable self-contained batter...
The powerPACK plus, ultra, and ultra 2 are rechargeable self-contained battery packs. Over time, these packs will lose their ability to maintain a charge and will need to be replaced.
3177 - Insert replacement battery pack into handle
Insert replacement battery pack into handle
3184 - Observe the orientation notch to find the correct orientation.
Observe the orientation notch to find the correct orientation.

3185 - Top of the battery pack must fall in line with the top of the handle; if it s...
Top of the battery pack must fall in line with the top of the handle; if it sticks out, the orientation is wrong.

3197 - 3178 - Align handle with groove on side and move handle toward front of instrument; Remark:3186
3178 - Align handle with groove on side and move handle toward front of instrument
Align handle with groove on side and move handle toward front of instrument
3199 - 3179 - Fasten screw; Remark:3187
3179 - Fasten screw
Fasten screw
3187 - Take care not to over-tighten compartment screw.
Take care not to over-tighten compartment screw.

3198 - 3180 - A rechargeable battery pack may need to be charged; Remark:3188
3180 - A rechargeable battery pack may need to be charged
A rechargeable battery pack may need to be charged
3188 - Connect instrument to charger for 3 to 4 hours.
Connect instrument to charger for 3 to 4 hours.