Search for and identify radioactive material with the FLIR identiFINDER 2 NG
Remove NGH references
3233 - 3232 - Prepare identiFINDER 2 for Operation; Remark:
3232 - Prepare identiFINDER 2 for Operation
Prepare identiFINDER 2 for Operation
3236 - 3235 - Disconnect identiFINDER 2 from external charger; Remark:3234
3229 - Caution : The identiFINDER 2 charger and the internal charging port can be damaged if t...
The identiFINDER 2 charger and the internal charging port can be damaged if the universal serial bus (USB) cable is not removed properly.
3235 - Disconnect identiFINDER 2 from external charger
Disconnect identiFINDER 2 from external charger
3234 - Remove the USB connector by pulling the connector toward the bottom of the in...
Remove the USB connector by pulling the connector toward the bottom of the instrument.

3239 - 3238 - Hold identiFINDER 2 by handle and secure safety strap to wrist; Remark:3237
3230 - Caution : Always use the safety strap to minimize potential damage to the instrument.
Always use the safety strap to minimize potential damage to the instrument.
3238 - Hold identiFINDER 2 by handle and secure safety strap to wrist
Hold identiFINDER 2 by handle and secure safety strap to wrist
3237 - Always use safety strap to minimize potential damage to the instrument.
Always use safety strap to minimize potential damage to the instrument.
3392 - 3391 - Verify no radioactive sources are near instrument; Remark:3240
3391 - Verify no radioactive sources are near instrument
Verify no radioactive sources are near instrument
3244 - 220 - Power on instrument; Remark:3243
3231 - Caution : Do not expose instrument to a sudden change in temperature of more than 30°C...
Do not expose instrument to a sudden change in temperature of more than 30°C.
220 - Power on instrument
Power on instrument
3243 - Hold down the power button until the start-up screen appears in the display.
Hold down the power button until the start-up screen appears in the display.

3247 - 2220 - Wait for measurement to complete; Remark:3245
2220 - Wait for measurement to complete
Wait for measurement to complete
3245 - Dose Rate screen indicates that the instrument is ready.
Dose Rate screen indicates that the instrument is ready.
3250 - 3249 - Verify sufficient battery power is available to conduct operation; Remark:3248
3249 - Verify sufficient battery power is available to conduct operation
Verify sufficient battery power is available to conduct operation
3248 - If only two bars are visible in the icon located in the upper left hand corne...
If only two bars are visible in the icon located in the upper left hand corner, remaining battery time is less than 60 minutes.
3254 - 1087 - Perform functional test; Remark:3252
1087 - Perform functional test
Perform functional test
3252 - Refer to RAD-HHD-IDF-RM08b, identiFINDER 2 Functional Test to verify equipmen...
Refer to RAD-HHD-IDF-RM08b, identiFINDER 2 Functional Test to verify equipment functionality before operation.
840 - 839 - Search Item/Area of Interest; Remark:
839 - Search Item/Area of Interest
Search Item/Area of Interest
3257 - 2209 - Verify instrument is in a suitable background radiation area; Remark:3255
2209 - Verify instrument is in a suitable background radiation area
Verify instrument is in a suitable background radiation area
3255 - The instrument should be located in an environment similar to the area being ...
The instrument should be located in an environment similar to the area being surveyed (e.g., outside, adjacent to inspection area), but far enough from item of interest to minimize any effects on the background measurement.
3315 - 3314 - Press L to enter Finder mode; Remark:3313
3314 - Press L to enter Finder mode
Press L to enter Finder mode
3319 - 2220 - Wait for measurement to complete; Remark:3316
2220 - Wait for measurement to complete
Wait for measurement to complete
3316 - Upon entering Finder mode, the instrument will collect a background for 10 se...
Upon entering Finder mode, the instrument will collect a background for 10 seconds.
3317 - Once the background is collected, the instrument is ready to search for radio...
Once the background is collected, the instrument is ready to search for radioactive sources.
3324 - 3321 - Search entire area of interest; Remark:3320
3322 - Note : If a search has already been performed using a survey instrument, the identiF...
If a search has already been performed using a survey instrument, the identiFINDER may be used to confirm search results.
3323 - Note : Slow, steady, methodical scans have the highest probability of detection....
Slow, steady, methodical scans have the highest probability of detection.
3321 - Search entire area of interest
Search entire area of interest
3320 - Move instrument slowly at approximately 20 cm per second. Maintain a distance...
Move instrument slowly at approximately 20 cm per second. Maintain a distance of less than 10 cm from survey area without making contact.

3328 - 849 - Observe instrument response for elevated radiation levels; Remark:3325
3327 - Note : Radiation may be evenly distributed and/or concentrated at a specific locatio...
Radiation may be evenly distributed and/or concentrated at a specific location.
849 - Observe instrument response for elevated radiation levels
Observe instrument response for elevated radiation levels
3325 - Listen for audible indicators, observe display screen (graph, dose rate, coun...
Listen for audible indicators, observe display screen (graph, dose rate, counts per second) and alarm lights to determine distribution of radiation.

3331 - 3330 - Characterize radiation levels; Remark:3329
3330 - Characterize radiation levels
Characterize radiation levels
3329 - Use the “crosshair” technique to find the highest level of radiation....
Use the “crosshair” technique to find the highest level of radiation.
405 - 368 - Mark point(s) of interest; Remark:4277
368 - Mark point(s) of interest
Mark point(s) of interest
4277 - Place a marker (tape, pencil, etc.) on points exhibiting elevated radiation. ...
Place a marker (tape, pencil, etc.) on points exhibiting elevated radiation. There may be more than one point of interest.

3338 - 3337 - Document search results; Remark:3336
3337 - Document search results
Document search results
3336 - Document the distribution of radiation, location(s) of concentrated radiation...
Document the distribution of radiation, location(s) of concentrated radiation levels, and approximate radiation levels observed at all locations.
3340 - 3339 - Identify Radioactive Isotope(s); Remark:
3339 - Identify Radioactive Isotope(s)
Identify Radioactive Isotope(s)
3343 - 3342 - Position instrument at marked location; Remark:3341
3375 - Note : Incorrect identification results may occur due to large changes in temperatur...
Incorrect identification results may occur due to large changes in temperature (e.g., indoors vs. outdoors). The instrument should be calibrated after adjusting to ambient temperature.
3342 - Position instrument at marked location
Position instrument at marked location
3341 - Perform identification at each location previously marked for investigation.
Perform identification at each location previously marked for investigation.
Place the instrument close without making contact.
3346 - 859 - Start identification measurement; Remark:3344
859 - Start identification measurement
Start identification measurement
3344 - Press M to enter Identify mode from either Dose Rate or Find...
Press M to enter Identify mode from either Dose Rate or Finder mode.


3350 - 3349 - Follow instructions on display; Remark:3347
3349 - Follow instructions on display
Follow instructions on display
3347 - Optimal identification measurements are obtained when the signal bar is locat...
Optimal identification measurements are obtained when the signal bar is located between the right and left arrows on the screen.

3348 - If “Insufficient counts” message is displayed press M to...
If “Insufficient counts” message is displayed press M to continue measurement.
3353 - 2220 - Wait for measurement to complete; Remark:3351
2220 - Wait for measurement to complete
Wait for measurement to complete
3351 - Hold the instrument steady and as close as possible to the measurement locati...
Hold the instrument steady and as close as possible to the measurement location until the measurement is complete (typically 120 seconds).
3356 - 866 - Evaluate identification results; Remark:3354
866 - Evaluate identification results
Evaluate identification results
3354 - Once completed, the following information will be displayed: 1. Con...
Once completed, the following information will be displayed:
1. Confidence value: 4 (low) through 10 (high)
2. Isotope name (e.g. Co-57)
3. Isotope category
• IND – Industrial Isotope
• MED – Medical Isotope
• NUC – Special Nuclear Material
• NORM – Naturally Occurring Radioactive Material
• Annihilation radiation – produced by isotopes used in some medical scans

3359 - 3358 - Repeat identification if necessary; Remark:3357
3358 - Repeat identification if necessary
Repeat identification if necessary
3357 - If the result is unexpected, move away from the source and recalibrate by pow...
If the result is unexpected, move away from the source and recalibrate by powering the identiFINDER 2 off and on. Repeat identification.
If the identiFINDER 2 cannot find a match in the library, the instrument will display “Unk” (unknown).
3362 - 3361 - Make note of identification number; Remark:3360
3361 - Make note of identification number
Make note of identification number
3360 - The identiFINDER 2 will automatically save identification results. ...
The identiFINDER 2 will automatically save identification results.
The identification number is located the bottom of the screen.
3365 - 3364 - Press R to exit and return to Dose Rate mode; Remark:3363
3364 - Press R to exit and return to Dose Rate mode
Press R to exit and return to Dose Rate mode
3368 - 3367 - Document and report identification results; Remark:3366
3367 - Document and report identification results
Document and report identification results
3366 - Provide results from all identification locations.
Provide results from all identification locations.
3370 - 313 - Power off instrument; Remark:3369
313 - Power off instrument
Power off instrument
3369 - Press and hold power button, then press M to confirm shutdow...
Press and hold power button, then press M to confirm shutdown.

3374 - 3373 - Reconnect instrument to external power after use; Remark:3371
3373 - Reconnect instrument to external power after use
Reconnect instrument to external power after use
3371 - Push the USB connector toward the top (cap) of the instrument.
Push the USB connector toward the top (cap) of the instrument.
3372 - Verify the lights next to “Charging” and “Ext. Power” are lit when th...
Verify the lights next to “Charging” and “Ext. Power” are lit when the external power is first connected.