Last Updated: 04/25/2025 1:39 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Kromek D3S-ID
Device
Published
10/28/2024 4:01 PM
247 - 257 - Prepare instrument; Remark:
Procedure Step
Published
04/11/2025 9:08 PM
257 - Prepare instrument
Instruction
Published
04/11/2025 9:07 PM
786 - 219 - Verify there are no sources of radiation near instrument; Remark:7
Procedure Step
Published
04/20/2025 6:37 PM
219 - Verify there are no sources of radiation near instrument
Instruction
Published
04/12/2025 4:20 PM
7 - No Radiation symbol
Remark
Published
07/12/2021 1:34 PM
4679 - 220 - Power on instrument; Remark:4572
Procedure Step
Published
06/23/2022 1:26 PM
220 - Power on instrument
Instruction
Published
04/06/2025 7:00 PM
4572 - Press and hold power button for about 1 second. The D3S indicator light will ...
Remark
Published
12/22/2022 12:14 PM
1496 - 1495 - Power on mobile device; Remark:
Procedure Step
Published
11/12/2020 4:40 PM
1495 - Power on mobile device
Instruction
Published
11/12/2020 4:20 PM
4577 - 4571 - Open D3S ID app; Remark:4825
Procedure Step
Published
06/15/2022 1:56 PM
4571 - Open D3S ID app
Instruction
Published
06/15/2022 1:55 PM
4825 - IO D3S App Icon Dark Background
Remark
Published
08/01/2022 12:33 PM
4702 - 4569 - Verify mobile device is communicating with D3S; Remark:4567
Procedure Step
Published
06/29/2022 11:56 PM
4569 - Verify mobile device is communicating with D3S
Instruction
Published
06/15/2022 1:58 PM
4567 - The indicator light will flash blue every 5 seconds when connected via Blueto...
Remark
Published
06/23/2022 12:03 PM
4568 - Additionally, the status icon will display a check mark when connected.
Remark
Published
06/15/2022 1:58 PM
4680 - 2295 - Verify sufficient battery charge; Remark:4587
Procedure Step
Published
06/23/2022 1:34 PM
2295 - Verify sufficient battery charge
Instruction
Published
04/06/2021 4:43 PM
4587 - Check the battery charge on the mobile device. Tap the status icon to verify ...
Remark
Published
06/23/2022 12:04 PM
4690 - 4662 - Verify instrument is in SEARCH mode; Remark:4586
Procedure Step
Published
06/23/2022 2:17 PM
4662 - Verify instrument is in SEARCH mode
Instruction
Published
06/30/2022 12:00 AM
4586 - SEARCH will be highlighted on the screen. If not, tap
Remark
Published
06/15/2022 2:03 PM
4691 - 4585 - Select alarm setting; Remark:4870
Procedure Step
Published
06/23/2022 2:32 PM
4585 - Select alarm setting
Instruction
Published
06/15/2022 2:04 PM
4870 - Tap menu icon.
Remark
Published
08/01/2022 12:34 PM
4689 - Tap speaker icon to enable both Sound and Vibrate.
Remark
Published
06/23/2022 2:20 PM
252 - 232 - Check instrument gamma response; Remark:
Procedure Step
Published
07/29/2021 11:10 AM
232 - Check instrument gamma response
Instruction
Published
07/23/2021 9:58 AM
4687 - 235 - Place instrument next to radioactive gamma source; Remark:4872
Procedure Step
Published
06/23/2022 2:23 PM
235 - Place instrument next to radioactive gamma source
Instruction
Published
07/09/2021 1:39 PM
4872 - IO D3S Move To Rad Source Trefoil
Remark
Published
07/29/2022 10:20 AM
4688 - 1439 - Verify instrument responds to radioactive source; Remark:4686
Procedure Step
Published
06/23/2022 2:24 PM
1439 - Verify instrument responds to radioactive source
Instruction
Published
06/29/2022 11:50 PM
4686 - The gamma radiation level will increase. The mobile device will vibrate and e...
Remark
Published
06/23/2022 2:24 PM
4658 - 4657 - Perform identification; Remark:
Procedure Step
Published
06/16/2022 4:40 PM
4657 - Perform identification
Instruction
Published
06/16/2022 4:40 PM
4683 - 4621 - Select confirmation mode; Remark:4620
Procedure Step
Published
06/23/2022 2:26 PM
4621 - Select confirmation mode
Instruction
Published
06/24/2022 9:01 AM
4620 - Tap CONFIRMATION or Move to Confirmation Mode
Remark
Published
06/15/2022 5:11 PM
4684 - 2218 - Place instrument close to measurement location; Remark:4619
Procedure Step
Published
06/23/2022 2:27 PM
2218 - Place instrument close to measurement location
Instruction
Published
04/06/2021 5:15 PM
4619 - Position the D3S close to RN material. For optimal measurement, the pointer w...
Remark
Published
06/15/2022 5:12 PM
4685 - 859 - Start identification measurement; Remark:4618
Procedure Step
Published
06/23/2022 2:29 PM
859 - Start identification measurement
Instruction
Published
10/26/2020 7:00 PM
4618 - Tap Start Scan. Stay in the same location for duration of me...
Remark
Published
06/15/2022 5:13 PM
4698 - 2220 - Wait for measurement to complete; Remark:4616
Procedure Step
Published
04/20/2025 1:01 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
4616 - Wait for at least 120 seconds. Tap STOP ID.
Remark
Published
06/23/2022 3:37 PM
4699 - 1442 - Verify identification results are correct; Remark:4826
Procedure Step
Published
06/23/2022 3:38 PM
1442 - Verify identification results are correct
Instruction
Published
09/02/2020 3:03 PM
4826 - IO D3S Gamma Identify Cs-137 4.43 uSv
Remark
Published
07/29/2022 9:40 AM
4700 - 241 - Move instrument away from radioactive gamma source; Remark:4871
Procedure Step
Published
06/23/2022 3:39 PM
241 - Move instrument away from radioactive gamma source
Instruction
Published
06/25/2021 12:22 PM
4871 - IO D3S Move Away Rad Source Trefoil
Remark
Published
07/29/2022 10:19 AM
4656 - 4480 - Check instrument neutron response (if available); Remark:
Procedure Step
Published
06/16/2022 4:42 PM
4480 - Check instrument neutron response (if available)
Instruction
Published
06/10/2022 7:19 AM
4693 - 3095 - Place instrument next to radioactive neutron source; Remark:4872
Procedure Step
Published
06/23/2022 3:40 PM
3095 - Place instrument next to radioactive neutron source
Instruction
Published
07/16/2021 10:49 AM
4694 - 1439 - Verify instrument responds to radioactive source; Remark:4692
Procedure Step
Published
06/23/2022 3:43 PM
4692 - The neutron radiation level will increase. The mobile device will vibrate and...
Remark
Published
06/23/2022 3:42 PM
4695 - 3098 - Move instrument away from radioactive neutron source; Remark:4871
Procedure Step
Published
06/23/2022 3:43 PM
3098 - Move instrument away from radioactive neutron source
Instruction
Published
07/16/2021 10:54 AM
4565 - 4564 - Tap SNOOZE; Remark:4566
Procedure Step
Published
06/15/2022 10:43 AM
4564 - Tap SNOOZE
Instruction
Published
06/15/2022 10:11 AM
4566 - IO D3S Snooze Neutron Red
Remark
Published
06/15/2022 10:43 AM
253 - 242 - Document and report any unresolved problems; Remark:
Procedure Step
Published
06/22/2021 10:23 AM
242 - Document and report any unresolved problems
Instruction
Published
04/21/2025 11:45 AM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
minutes
Duration
Published
05/26/2022 11:41 AM
Annually
Frequency
Published
04/01/2020 11:37 AM
Before each use
Frequency
Published
08/31/2020 5:00 PM
Monthly
Frequency
Published
04/01/2020 11:36 AM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
Power on/off D3S
Image
Published
07/26/2022 12:40 PM
D3S App Icon
Image
Published
07/26/2022 2:28 PM
KRO D3S Blue Indicator Light
Image
Published
07/11/2022 3:45 PM
KRO D3S Gamma 02uSv/h
Image
Published
07/12/2022 10:53 AM
KRO D3S Battery Status Icon
Image
Published
07/12/2022 10:56 AM
D3S Battery Status
Image
Published
07/26/2022 1:44 PM
D3S Battery Status Data
Image
Published
07/26/2022 1:44 PM
D3S Search mode
Image
Published
07/26/2022 1:44 PM
KRO D3S Menu Icon
Image
Published
07/12/2022 1:34 PM
D3S Sound Vibrate Icon
Image
Published
07/26/2022 2:24 PM
D3S Rad Source Closer
Image
Published
07/29/2022 10:12 AM
D3S Gamma Alarm
Image
Published
07/29/2022 9:07 AM
D3S Confirmation Mode
Image
Published
07/27/2022 11:23 AM
D3S Device distance
Image
Published
07/27/2022 11:23 AM
D3S Begin Scan
Image
Published
07/27/2022 11:23 AM
D3S Stop Scan ID
Image
Published
07/27/2022 11:23 AM
D3S Observe Results
Image
Published
07/27/2022 11:23 AM
D3S Rad Source Away
Image
Published
07/29/2022 10:14 AM
D3S Neutron Alarm
Image
Published
07/26/2022 2:24 PM
D3S Snooze High Neutron
Image
Published
06/15/2022 10:49 AM
DET-HHD-KRO-D3S-RM01

Verify instrument is operating correctly.

Equipment Hierarchy
Kromek D3S-ID
Frequency
Annually
Before each use
Monthly
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
minutes
Length of time
5
Materials Required
Radioactive gamma source of Cs-137 activity 37-370 kBq (1-10 μCi), or equivalent
Radioactive neutron source of Cf-252 activity 37-185 kBq (1-5 μCi), or equivalent
Originators
Demboski, Erica
Tucker, Brian
Originating Organization
PNNL
Approver
Simons, Brian
Safety Concerns
Radioactive Source
Final Documentation
None
Departmental Coordination

Prior to arrival, notify operators concerning expected work and duration

Date of Origin
Version
0.00
Step 1

4702 - 4569 - Verify mobile device is communicating with D3S; Remark:4567

Step 00
Step 2
Step 3
Document Type
Routine Maintenance Procedure
Percent Complete
100.00
Percent Complete W/Media