Last Updated: 04/25/2025 1:36 AM

The following table lists all the components that make up the document.

Title Type Status Last Updated
Atomtex AT6101C Backpack
Device
Published
06/27/2024 1:47 PM
1402 - 7927 - Establish Initial Conditions; Remark:
Procedure Step
Published
09/30/2022 1:47 PM
7927 - Establish Initial Conditions
Instruction
Published
07/12/2023 12:23 AM
1394 - 1393 - Verify no radioactive sources are near instrument; Remark:1392
Procedure Step
Published
11/24/2020 10:54 AM
1393 - Verify no radioactive sources are near instrument
Instruction
Published
05/25/2021 2:05 PM
1392 - Symbol - No radioactive material in immediate area
Remark
Published
09/01/2020 7:07 PM
1397 - 1396 - Verify check sample is in the side pocket; Remark:1395
Procedure Step
Published
11/24/2020 10:54 AM
1396 - Verify check sample is in the side pocket
Instruction
Published
11/24/2020 10:49 AM
1395 - Atomtex check sample in side pocket
Remark
Published
11/24/2020 9:41 AM
1400 - 1399 - Verify backpack is connected to mobile device; Remark:1398
Procedure Step
Published
11/24/2020 10:55 AM
1399 - Verify backpack is connected to mobile device
Instruction
Published
11/24/2020 10:48 AM
1398 - Scanning should appear in the status bar. Gamma and neutron values should be ...
Remark
Published
11/24/2020 9:40 AM
1404 - 1403 - Start Calibration; Remark:
Procedure Step
Published
11/24/2020 10:56 AM
1403 - Start Calibration
Instruction
Published
11/24/2020 10:47 AM
1407 - 1406 - Tap menu icon; Remark:1405
Procedure Step
Published
11/24/2020 10:56 AM
1406 - Tap menu icon
Instruction
Published
11/24/2020 10:46 AM
1405 - ATAS Menu Button
Remark
Published
11/17/2020 5:02 PM
1410 - 1409 - Tap Calibrate; Remark:1408
Procedure Step
Published
11/24/2020 10:56 AM
1409 - Tap Calibrate
Instruction
Published
11/24/2020 10:46 AM
1408 - ATAS_Calibrate_Menu_Option
Remark
Published
11/17/2020 5:01 PM
1413 - 1412 - Wait for calibration to complete; Remark:1411
Procedure Step
Published
11/24/2020 10:57 AM
1412 - Wait for calibration to complete
Instruction
Published
11/24/2020 10:45 AM
1411 - ATAS_Calibration_in_process
Remark
Published
11/17/2020 5:08 PM
1416 - 1415 - Tap OK; Remark:1414
Procedure Step
Published
11/24/2020 10:57 AM
1415 - Tap OK
Instruction
Published
11/24/2020 10:44 AM
1414 - Take note of the Gain code as it might be used for troubleshooting.
Remark
Published
11/24/2020 9:39 AM
1419 - 1418 - Enter scanning file name; Remark:1417
Procedure Step
Published
11/24/2020 10:58 AM
1418 - Enter scanning file name
Instruction
Published
09/29/2020 1:38 PM
1417 - ATAS_Scanning_File_Name
Remark
Published
11/17/2020 5:10 PM
1422 - 2220 - Wait for measurement to complete; Remark:1420
Procedure Step
Published
04/20/2025 12:51 PM
2220 - Wait for measurement to complete
Instruction
Published
10/22/2020 11:15 AM
1420 - ATAS_Background_Collection
Remark
Published
11/24/2020 9:42 AM
1425 - 1424 - Backpack radiation detector is ready for normal use; Remark:1423
Procedure Step
Published
11/24/2020 10:59 AM
1424 - Backpack radiation detector is ready for normal use
Instruction
Published
11/24/2020 10:41 AM
1423 - ATAS_Scanning_Screen
Remark
Published
11/17/2020 5:10 PM
Routine Maintenance Procedure
Document Type
Published
03/02/2020 3:31 PM
10 minutes
Duration
Published
04/01/2020 11:35 AM
As Needed
Frequency
Published
04/01/2020 11:36 AM
Technician 1
Skill Level
Published
05/14/2020 2:56 PM
None
Tools
Published
05/06/2020 9:50 PM
No Radiation Symbol
Image
Published
08/13/2024 2:56 PM
Atomtex_Backpack_Side_Pocket_Red_Arrow_0.png
Image
Published
09/04/2020 10:44 PM
AT6101C_Partial_pocket_side_KCL Potassium check sample_Puck_red_arrow_1.png
Image
Published
09/04/2020 10:26 PM
ATAS_Scanning.png
Image
Published
03/30/2022 2:22 AM
ATAS_Hamburger_Menu.png
Image
Published
03/30/2022 2:22 AM
ATAS_Calibrate_Menu_Option
Image
Published
03/30/2022 2:22 AM
ATAS_Calibration_in_process
Image
Published
03/30/2022 2:22 AM
ATAS_Calibration_Complete
Image
Published
03/30/2022 2:22 AM
ATAS_Scanning_File_Name
Image
Published
03/30/2022 2:22 AM
ATAS_Collecting_Background.png
Image
Published
03/30/2022 2:22 AM
ATAS_Scanning_Screen.png
Image
Published
03/30/2022 2:22 AM
DET-HHD-ATM-RM04

Perform an alignment (calibration) on the gamma channel

Equipment Hierarchy
Atomtex AT6101C Backpack
Frequency
As Needed
Technician Level
Technician Level
Technician 1
x 1
Duration
Time increment
10 minutes
Tools Required
None
Materials Required
Atomtex check sample
Originators
Small, Jack
Tucker, Brian
Originating Organization
PNNL
Additional Retest Requirements Information

None

Departmental Coordination

None

Date of Origin
Version
0.00
Percent Complete
100.00
Percent Complete W/Media
100.00